Solenoid and Pinch Valves for Fluid Handling in Analysers

A valve in an analyser does one of two jobs: it isolates a reagent from a cell, or it lets a sample flow through a tube that must not touch metal. The parts below split along that line. A direct acting solenoid valve gives a fast, small, repeatable seat with a metal or plastic body. An air operated pinch valve closes by squeezing a sleeve, so the liquid only meets the sleeve and the flow path is straight through. Both are quoted with orifice, pressure range and wetted material, because those three decide whether a valve can sit in the fluid path at all.
quote now

Product Specifications

Product Overview

An analyser valve has to do three things at once: close against the system pressure without leaking, open fast enough for the dosing step, and survive the reagent. A direct acting solenoid valve does all three with a plunger that lifts off a seat, and because the coil acts on the plunger directly it works from zero pressure, which is what a gravity fed reagent line needs. A servo assisted valve needs a pressure difference to stay open or closed and is the wrong choice on a line that runs dry.

The pinch valve is the answer to a different problem. When the liquid is a blood sample, a reagent that attacks every elastomer except one, or a slurry that would jam a seat, the valve must not have a seat at all. A pinch valve squeezes a sleeve shut with air pressure, so the flow path is a straight tube and the only wetted part is the sleeve. That is why sample lines and waste lines in clinical analysers are often pinch valves while the reagent lines are solenoid valves.

The second figure is the orifice. Flow through a valve at a given pressure difference scales with the orifice area, and a 1.2 mm orifice gives a fifth of the flow of a 2.4 mm one. The valve is also a restriction in the middle of the fluid path, so a valve sized for the dose still has to pass the flush volume without stalling the pump. On a 24 VDC coil the opening time is in the millisecond range and the closing time is longer, which the firmware has to know if the dose is short.

The third figure is the seal. Nitrile, FKM, EPDM and PTFE are the common choices, and the reagent list decides between them: a dilute acid is fine on FKM, a solvent is not, and a strong base attacks many of the cheaper seals. The body material matters too, because a brass body with a stainless seat is not a stainless valve. The last line of any valve quotation should therefore be the wetted material list, not the coil voltage.

Key Specifications

ParameterValueNote
Valve familiesdirect acting 2/2 solenoid, and air operated pinchsolenoid for reagent lines, pinch for sample and waste lines
Solenoid orifice range1.2, 1.6 and 2.0 mm on the 6011 body, 1 to 4 mm on the VDW familyflow through the valve scales with orifice area
Solenoid pressure range0 to 21 bar depending on orifice and coila smaller orifice holds more pressure, which is why the range moves with the size
Flow coefficientkv 0.045 to 0.11 on the 6011, Cv 0.04 to 0.44 on the VDW familythe valve is a restriction in the middle of the fluid path
Circuit function2/2 normally closed as standardnormally open and three port versions exist in the same families
Coil voltage24 VDC and 230 VAC 50 Hz on the 6011, 24 VDC, 12 VDC and 100 to 220 VAC on the VDW familypower is 2.5 to 3 W on the compact VDW sizes
Response timemillisecond opening, longer closinga short dose is limited by the closing time, not the opening time
Pinch valve closingair or fluid pressure with a minimum 2 bar differentialthe sleeve closes fully at that differential
Pinch valve line pressure2 to 6 bar depending on the valve diameterthe sleeve is the pressure limit, not the body
Pinch valve sizesDN 10 to DN 100body in stainless steel 1.4404, wetted surface Ra 0.5
Wetted material, solenoidbrass, polyamide, stainless steel 1.4305, with NBR, FKM or EPDM sealsthe seal is the part the reagent attacks first
Wetted material, pinchthe sleeve only, in the elastomer chosen for the reagentthe flow path is a straight tube with no seat
Leak rate, solenoidpermitted leakage 10 to the minus 4 mbar litre per secondthe figure a leak test is written against
ProtectionIP65 with a cable plug on the 6011without the plug the coil is not sealed

Included Models and Order Codes

The table lists the sizes that cover a small analyser fluid path. The 6011 is a direct acting plunger valve with a welded guide tube, offered in 1.2, 1.6 and 2.0 mm orifice, and the item numbers below are 24 VDC and 230 VAC versions of the same body. The maker has marked the 6011 range as replaced by the 7011, so stock on the listed numbers is a last time buy question rather than a catalogue question, and that is exactly the case where a stock check is worth more than a data sheet. The VDW family is the compact alternative, a direct acting two port valve from 1 mm to 4 mm orifice with a resin, brass or stainless body. The pinch valve closes on a sleeve with 2 bar of air and takes line pressure up to 6 bar depending on size.

PartTypeOrificePressure rangeFlowNote
Bürkert 1340846011, 2/2 direct acting1.2 mm, M5 port0 to 12 barkv 0.04524 VDC coil, brass body
Bürkert 1340856011, 2/2 direct acting1.2 mm, M5 port0 to 21 barkv 0.04524 V 50 Hz coil, highest pressure in the family
Bürkert 1377946011, 2/2 direct acting1.6 mm, G1/8 port0 to 6 barkv 0.0624 VDC coil, brass or stainless body
Bürkert 1377966011, 2/2 direct acting2.0 mm, G1/8 port0 to 4.5 barkv 0.1124 VDC coil, highest flow of the three sizes
SMC VDW102 port direct acting1.0 and 1.6 mm, M50 to 1.0 MPaCv 0.04 to 0.072.5 W, resin, brass or stainless body
SMC VDW202 port direct acting2.3 and 3.2 mm, 1/80 to 1.0 MPaCv 0.18 to 0.303 W, water and medium vacuum duty
SMC VDW302 port direct acting2 to 4 mm, 1/40 to 1.0 MPaCv 0.16 to 0.44highest flow of the family
AKO VMCair operated pinchDN 10 to DN 100line 2 to 6 barstraight through borecloses at 2 bar air differential, body 1.4404

Sizing a Valve for a Dosing Line

The dose is set by the open time and the flow through the valve, so the valve has to pass the dose at the pressure the line actually holds. The table is the working version of that sum for the sizes above.

Dose per stepValve to useWhat limits the doseWhat to check
Under 50 microlitre1.2 mm orifice, direct actingclosing time, not flowcoil suppression and the tube volume after the valve
50 to 500 microlitre1.6 mm orifice, direct actingflow at the line pressureline pressure stable during the step
0.5 to 5 ml2.0 to 3.2 mm orificeflow at the line pressurethe pump has to keep up with the open valve
Flush volume above 5 ml3 to 4 mm orifice, or a pinch valvebore of the whole pathevery fitting between valve and cell adds volume
Sample containing solidspinch valvesleeve boresleeve material against the sample
Line that runs drydirect acting solenoidzero pressure operationa servo assisted valve will not open

Key Features

Works from zero pressure – a direct acting valve opens with no pressure difference, which is what a gravity fed reagent bottle needs.

Orifice and flow coefficient together – the kv figure is listed next to each orifice, so the dose can be worked out without a bench test.

Pressure range per size – a 1.2 mm valve holds 21 bar while a 2.0 mm one holds 4.5 bar, and both numbers are on the page rather than in a footnote.

Millisecond opening – the coil acts on the plunger directly, so the dosing step is limited by the closing time and not by the valve opening.

Wetted materials listed – brass, polyamide and stainless bodies with NBR, FKM or EPDM seals, so a reagent list can be checked against the valve.

Pinch valve with a straight bore – the sample only meets the sleeve, which is what a blood or slurry line needs.

Air operated closing – 2 bar of air closes the sleeve fully, and the line pressure limit is the sleeve rather than the body.

Leak rate stated – the permitted leakage figure is written down, so a leak test has a number to test against.

Applications

Three fluid paths where the choice between a seat and a sleeve follows from what is in the liquid.

Clinical chemistry reagent lines – a 1.2 mm direct acting valve doses reagent into a cuvette, and the closing time sets the smallest step the method can use.

Sample and waste lines – a pinch valve closes on the outside of the tube, so a sample that would jam a seat passes straight through and the sleeve is the only part to change.

Flush and purge circuits – a 2.0 mm valve passes the flush volume fast enough that the wash step does not set the cycle time of the analyser.

A valve driver is a small board with a flyback path, so the SUPERB PCBA and the SUPERB PCB keep the coil return short and the analogue input away from the switching node.

Sourcing and Cross-Reference

The first question on a valve build is the reagent list, because the seal is the part that fails. A dilute acid, a solvent and a strong base each need a different elastomer, and a valve quoted with the wrong seal will pass a bench test and fail in the field. The second question is whether the line ever runs dry, because a servo assisted valve needs pressure to switch and a direct acting one does not.

The second question is the dead volume between the valve and the cell. A valve that is right for the flow can still hold too much liquid in the fittings after it, and that volume shows up as carry over between samples. The fix is to place the valve close to the cell and to keep the bore of the path constant, not to look for a valve with a smaller orifice.

The third question is availability. Several of the small direct acting families have been replaced by successors, and the listed numbers are then a last time buy rather than a catalogue line. See obsolete part cross reference for a successor check and BOM shortage recovery where a build is held up by one valve.

The SUPERB Assembly team mounts the pinch valve so the sleeve hangs vertically, because a sleeve that sits in a puddle of reagent ages faster than one that drains.

Q: Can you confirm which of these valve sizes are still in production and quote the successor where the maker has replaced one?

A: Yes. Send the reagent list, the dose, the line pressure and the voltage, and the SUPERB Tech team returns the valves and the seals that fit.

Why Analyser Builders Source Fluid Valves Through SUPERB

  1. Seal checked against the reagent: the wetted material list is quoted with the part, because the seal is what fails first in a reagent line.
  2. Zero pressure operation stated: whether a valve opens with no pressure difference is written down, which decides how the bottle is mounted.
  3. Successor checked: where a family has been replaced by the maker, the successor is looked up rather than quoted from an old catalogue.
  4. Pinch option for dirty lines: a sleeve valve is quoted next to the seat valves, because the sample line is where a seat clogs.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Peristaltic and Diaphragm Micro Pumps, Liquid Flow Sensors for Medical and Analytical Instruments, Water Quality Sensor Circuits for pH and Conductivity, Level Sensors and Float Switches for Tanks, Quick Disconnect Couplings and Luer Fittings, Obsolete Part Cross Reference.