Solenoid and Pinch Valves for Fluid Handling in Analysers
Product Specifications
Product Overview
An analyser valve has to do three things at once: close against the system pressure without leaking, open fast enough for the dosing step, and survive the reagent. A direct acting solenoid valve does all three with a plunger that lifts off a seat, and because the coil acts on the plunger directly it works from zero pressure, which is what a gravity fed reagent line needs. A servo assisted valve needs a pressure difference to stay open or closed and is the wrong choice on a line that runs dry.
The pinch valve is the answer to a different problem. When the liquid is a blood sample, a reagent that attacks every elastomer except one, or a slurry that would jam a seat, the valve must not have a seat at all. A pinch valve squeezes a sleeve shut with air pressure, so the flow path is a straight tube and the only wetted part is the sleeve. That is why sample lines and waste lines in clinical analysers are often pinch valves while the reagent lines are solenoid valves.
The second figure is the orifice. Flow through a valve at a given pressure difference scales with the orifice area, and a 1.2 mm orifice gives a fifth of the flow of a 2.4 mm one. The valve is also a restriction in the middle of the fluid path, so a valve sized for the dose still has to pass the flush volume without stalling the pump. On a 24 VDC coil the opening time is in the millisecond range and the closing time is longer, which the firmware has to know if the dose is short.
The third figure is the seal. Nitrile, FKM, EPDM and PTFE are the common choices, and the reagent list decides between them: a dilute acid is fine on FKM, a solvent is not, and a strong base attacks many of the cheaper seals. The body material matters too, because a brass body with a stainless seat is not a stainless valve. The last line of any valve quotation should therefore be the wetted material list, not the coil voltage.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Valve families | direct acting 2/2 solenoid, and air operated pinch | solenoid for reagent lines, pinch for sample and waste lines |
| Solenoid orifice range | 1.2, 1.6 and 2.0 mm on the 6011 body, 1 to 4 mm on the VDW family | flow through the valve scales with orifice area |
| Solenoid pressure range | 0 to 21 bar depending on orifice and coil | a smaller orifice holds more pressure, which is why the range moves with the size |
| Flow coefficient | kv 0.045 to 0.11 on the 6011, Cv 0.04 to 0.44 on the VDW family | the valve is a restriction in the middle of the fluid path |
| Circuit function | 2/2 normally closed as standard | normally open and three port versions exist in the same families |
| Coil voltage | 24 VDC and 230 VAC 50 Hz on the 6011, 24 VDC, 12 VDC and 100 to 220 VAC on the VDW family | power is 2.5 to 3 W on the compact VDW sizes |
| Response time | millisecond opening, longer closing | a short dose is limited by the closing time, not the opening time |
| Pinch valve closing | air or fluid pressure with a minimum 2 bar differential | the sleeve closes fully at that differential |
| Pinch valve line pressure | 2 to 6 bar depending on the valve diameter | the sleeve is the pressure limit, not the body |
| Pinch valve sizes | DN 10 to DN 100 | body in stainless steel 1.4404, wetted surface Ra 0.5 |
| Wetted material, solenoid | brass, polyamide, stainless steel 1.4305, with NBR, FKM or EPDM seals | the seal is the part the reagent attacks first |
| Wetted material, pinch | the sleeve only, in the elastomer chosen for the reagent | the flow path is a straight tube with no seat |
| Leak rate, solenoid | permitted leakage 10 to the minus 4 mbar litre per second | the figure a leak test is written against |
| Protection | IP65 with a cable plug on the 6011 | without the plug the coil is not sealed |
Included Models and Order Codes
The table lists the sizes that cover a small analyser fluid path. The 6011 is a direct acting plunger valve with a welded guide tube, offered in 1.2, 1.6 and 2.0 mm orifice, and the item numbers below are 24 VDC and 230 VAC versions of the same body. The maker has marked the 6011 range as replaced by the 7011, so stock on the listed numbers is a last time buy question rather than a catalogue question, and that is exactly the case where a stock check is worth more than a data sheet. The VDW family is the compact alternative, a direct acting two port valve from 1 mm to 4 mm orifice with a resin, brass or stainless body. The pinch valve closes on a sleeve with 2 bar of air and takes line pressure up to 6 bar depending on size.
| Part | Type | Orifice | Pressure range | Flow | Note |
|---|---|---|---|---|---|
| Bürkert 134084 | 6011, 2/2 direct acting | 1.2 mm, M5 port | 0 to 12 bar | kv 0.045 | 24 VDC coil, brass body |
| Bürkert 134085 | 6011, 2/2 direct acting | 1.2 mm, M5 port | 0 to 21 bar | kv 0.045 | 24 V 50 Hz coil, highest pressure in the family |
| Bürkert 137794 | 6011, 2/2 direct acting | 1.6 mm, G1/8 port | 0 to 6 bar | kv 0.06 | 24 VDC coil, brass or stainless body |
| Bürkert 137796 | 6011, 2/2 direct acting | 2.0 mm, G1/8 port | 0 to 4.5 bar | kv 0.11 | 24 VDC coil, highest flow of the three sizes |
| SMC VDW10 | 2 port direct acting | 1.0 and 1.6 mm, M5 | 0 to 1.0 MPa | Cv 0.04 to 0.07 | 2.5 W, resin, brass or stainless body |
| SMC VDW20 | 2 port direct acting | 2.3 and 3.2 mm, 1/8 | 0 to 1.0 MPa | Cv 0.18 to 0.30 | 3 W, water and medium vacuum duty |
| SMC VDW30 | 2 port direct acting | 2 to 4 mm, 1/4 | 0 to 1.0 MPa | Cv 0.16 to 0.44 | highest flow of the family |
| AKO VMC | air operated pinch | DN 10 to DN 100 | line 2 to 6 bar | straight through bore | closes at 2 bar air differential, body 1.4404 |
Sizing a Valve for a Dosing Line
The dose is set by the open time and the flow through the valve, so the valve has to pass the dose at the pressure the line actually holds. The table is the working version of that sum for the sizes above.
| Dose per step | Valve to use | What limits the dose | What to check |
|---|---|---|---|
| Under 50 microlitre | 1.2 mm orifice, direct acting | closing time, not flow | coil suppression and the tube volume after the valve |
| 50 to 500 microlitre | 1.6 mm orifice, direct acting | flow at the line pressure | line pressure stable during the step |
| 0.5 to 5 ml | 2.0 to 3.2 mm orifice | flow at the line pressure | the pump has to keep up with the open valve |
| Flush volume above 5 ml | 3 to 4 mm orifice, or a pinch valve | bore of the whole path | every fitting between valve and cell adds volume |
| Sample containing solids | pinch valve | sleeve bore | sleeve material against the sample |
| Line that runs dry | direct acting solenoid | zero pressure operation | a servo assisted valve will not open |
Key Features
Works from zero pressure – a direct acting valve opens with no pressure difference, which is what a gravity fed reagent bottle needs.
Orifice and flow coefficient together – the kv figure is listed next to each orifice, so the dose can be worked out without a bench test.
Pressure range per size – a 1.2 mm valve holds 21 bar while a 2.0 mm one holds 4.5 bar, and both numbers are on the page rather than in a footnote.
Millisecond opening – the coil acts on the plunger directly, so the dosing step is limited by the closing time and not by the valve opening.
Wetted materials listed – brass, polyamide and stainless bodies with NBR, FKM or EPDM seals, so a reagent list can be checked against the valve.
Pinch valve with a straight bore – the sample only meets the sleeve, which is what a blood or slurry line needs.
Air operated closing – 2 bar of air closes the sleeve fully, and the line pressure limit is the sleeve rather than the body.
Leak rate stated – the permitted leakage figure is written down, so a leak test has a number to test against.
Applications
Three fluid paths where the choice between a seat and a sleeve follows from what is in the liquid.
Clinical chemistry reagent lines – a 1.2 mm direct acting valve doses reagent into a cuvette, and the closing time sets the smallest step the method can use.
Sample and waste lines – a pinch valve closes on the outside of the tube, so a sample that would jam a seat passes straight through and the sleeve is the only part to change.
Flush and purge circuits – a 2.0 mm valve passes the flush volume fast enough that the wash step does not set the cycle time of the analyser.
A valve driver is a small board with a flyback path, so the SUPERB PCBA and the SUPERB PCB keep the coil return short and the analogue input away from the switching node.
Sourcing and Cross-Reference
The first question on a valve build is the reagent list, because the seal is the part that fails. A dilute acid, a solvent and a strong base each need a different elastomer, and a valve quoted with the wrong seal will pass a bench test and fail in the field. The second question is whether the line ever runs dry, because a servo assisted valve needs pressure to switch and a direct acting one does not.
The second question is the dead volume between the valve and the cell. A valve that is right for the flow can still hold too much liquid in the fittings after it, and that volume shows up as carry over between samples. The fix is to place the valve close to the cell and to keep the bore of the path constant, not to look for a valve with a smaller orifice.
The third question is availability. Several of the small direct acting families have been replaced by successors, and the listed numbers are then a last time buy rather than a catalogue line. See obsolete part cross reference for a successor check and BOM shortage recovery where a build is held up by one valve.
The SUPERB Assembly team mounts the pinch valve so the sleeve hangs vertically, because a sleeve that sits in a puddle of reagent ages faster than one that drains.
Q: Can you confirm which of these valve sizes are still in production and quote the successor where the maker has replaced one?
A: Yes. Send the reagent list, the dose, the line pressure and the voltage, and the SUPERB Tech team returns the valves and the seals that fit.
Why Analyser Builders Source Fluid Valves Through SUPERB
- Seal checked against the reagent: the wetted material list is quoted with the part, because the seal is what fails first in a reagent line.
- Zero pressure operation stated: whether a valve opens with no pressure difference is written down, which decides how the bottle is mounted.
- Successor checked: where a family has been replaced by the maker, the successor is looked up rather than quoted from an old catalogue.
- Pinch option for dirty lines: a sleeve valve is quoted next to the seat valves, because the sample line is where a seat clogs.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Peristaltic and Diaphragm Micro Pumps, Liquid Flow Sensors for Medical and Analytical Instruments, Water Quality Sensor Circuits for pH and Conductivity, Level Sensors and Float Switches for Tanks, Quick Disconnect Couplings and Luer Fittings, Obsolete Part Cross Reference.