Peristaltic and Diaphragm Micro Pumps for Analysers
Product Specifications
Product Overview
An analyser moves small volumes on a schedule, and the pump is the part that has to repeat the same volume a hundred thousand times. Two families do that work. A peristaltic head squeezes a tube with rollers, so the liquid only ever touches the tube, and a tube change is the whole maintenance procedure. A diaphragm head pulls liquid through two check valves with a membrane, so it holds pressure better and can run dry without destroying itself, but the valves and the membrane are wetted parts that have to be compatible with the reagent.
The first specification to read is not the flow. It is the pressure the head holds against the back pressure of the fluid path: a filter, a narrow bore, a check valve and a length of small tubing all add up, and a pump quoted at its free flow will deliver far less once the path is connected. The second is the suction height, because an analyser that draws from a bottle below the instrument is asking the pump to lift liquid, and a head that is quoted only on pressure will not do it. The third is the volume per revolution, which is what the firmware actually counts when it doses.
Peristaltic heads have a property that suits reagents: the liquid is confined to the tube, so a change of reagent is a change of tube and nothing else. That is why they are used for sample lines, waste lines and colour reagents. The cost is tube life, which falls as the rollers turn and as the back pressure rises, and pulsation, which is smoothed by a longer tube run or a smaller bore. Diaphragm heads run faster, hold more pressure in a smaller body and tolerate a dry inlet, which is why they are used for filter purge, degassing and pressure-driven sample injection.
The last choice is the motor. A stepper gives a known volume per step and needs no encoder, which is what a dosing channel wants. A brushless DC motor gives speed and life in a smaller body, which is what a continuous circulation channel wants. Either way the pump head and the motor are usually ordered together, and the figure that matters at the quotation stage is the volume per revolution at the bore that the method uses, not the maximum flow on the front of the catalogue.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Pump families | Peristaltic tube heads and diaphragm heads with check valves | tube side is wetted on a peristaltic head, valve and membrane on a diaphragm head |
| Diaphragm flow range | 0.018 to 0.8 l/min across the small OEM series | free flow at the head, before the fluid path is connected |
| Diaphragm pressure head | 1 to 6 bar relative, depending on the series | the small FMM 20 is a 1 bar head, the larger OEM heads reach 6 bar |
| Peristaltic flow range | up to 300 ml/min on the 114 head with 4.8 mm bore, up to 3,000 ml/min intermittent on the 300 series | flow is a function of bore and speed, not of the head alone |
| Peristaltic tube sizes | 0.5, 0.8, 1.6, 2.4, 3.2, 4.0 and 4.8 mm bore on the 114 head | Pumpsil tubing is offered to 4.8 mm bore, other tubing to 3.2 mm |
| Suction height | 3 to 4.5 m of water column on the OEM diaphragm heads | quoted for water at 20 C, falls with viscosity |
| Pressure head, diaphragm | 6 bar on FD 1.200, FF 1.12, FMS-FC 1.25 and FP 1.70, 1 bar on FMM 20 | back pressure from a filter has to be counted against this figure |
| Motor options | stepper for dosing, brushless DC for circulation | volume per step or per revolution is the figure the firmware counts |
| Wetted materials | PP, PVDF, PTFE and stainless steel on diaphragm heads, silicone, Pumpsil and Norprene on tube heads | the reagent list decides the material, not the flow |
| Dry running | permitted on diaphragm heads, not on a peristaltic tube under pressure | a dry tube heats and takes a set |
| Tube life | falls with back pressure and roller speed | a tube change is the whole maintenance step on a peristaltic line |
| Mounting | OEM panel mount with the head on the instrument chassis | motor and head are normally ordered as one unit |
Included Models and Order Codes
Five diaphragm heads and two tube heads cover the flow range that most analysers need. The diaphragm heads run from the FMM 20 at 18 ml per minute up to the FP 1.70 at 800 ml per minute, and all but the smallest hold 6 bar, which is the figure that matters when a filter sits downstream. The tube heads are quoted by bore and speed: the 114 head takes seven bores from 0.5 mm to 4.8 mm and reaches 300 ml per minute, and the 300 series reaches 3,000 ml per minute intermittent, which is the range for waste removal and for circulating a bath. Order codes follow the head plus the motor, and the tube is ordered separately because it is a consumable.
| Pump | Type | Max flow | Max pressure | Suction height | Note |
|---|---|---|---|---|---|
| KNF FMM 20 | diaphragm, OEM | 0.018 l/min | 1 bar | 3 m H2O | smallest head in the range, for micro dosing |
| KNF FD 1.200 | diaphragm, OEM | 0.04 l/min | 6 bar | 4.5 m H2O | highest suction of the small heads |
| KNF FF 1.12 | diaphragm, OEM | 0.13 l/min | 6 bar | 3 m H2O | general reagent transfer |
| KNF FMS-FC 1.25 | diaphragm, OEM | 0.22 l/min | 6 bar | 3 m H2O | filtration and purge duty |
| KNF FP 1.70 | diaphragm, OEM | 0.8 l/min | 6 bar | 3 m H2O | largest of the OEM heads listed |
| Watson-Marlow 114 | peristaltic head | 300 ml/min | pressure set by tube and bore | lift by tube and speed | seven bores 0.5 to 4.8 mm |
| Watson-Marlow 300 series | peristaltic head | 3,000 ml/min intermittent, 2,400 ml/min continuous | pressure set by tube and bore | lift by tube and speed | 400 rpm maximum, cased or panel version |
Choosing Between a Tube Head and a Diaphragm Head
The two families overlap in flow, so the choice is usually made on the reagent and the maintenance interval rather than on the specification sheet. The table below is the short version of that decision.
| Requirement | Tube head | Diaphragm head | Reason |
|---|---|---|---|
| Reagent changes often | preferred | possible | the liquid stays inside a tube that is replaced with the reagent |
| Back pressure above 2 bar | not preferred | preferred | a diaphragm head holds 6 bar, a tube deforms and loses flow |
| Runs dry | avoid | preferred | a dry tube takes a set, a dry membrane does not |
| Air bubbles in the line | tolerated | pulled through the valves | bubbles alter the dose on a diaphragm head |
| Dose below 100 microlitre | preferred | possible | a small bore tube gives a small volume per revolution |
| Lift from a bottle below the bench | depends on bore | 4.5 m on the FD 1.200 | suction height is quoted per head |
| Waste removal from a cell | preferred | possible | waste is the line that clogs first, and a tube is quick to change |
Key Features
Volume per revolution stated – each head is quoted with its bore and speed, so the dose follows from the count the firmware already keeps.
Suction height published – lifting from a bottle under the bench is a normal analyser layout, and the 4.5 m figure on the FD 1.200 is the one that decides it.
Pressure head against a filter – the 6 bar heads hold their flow when a filter and a narrow bore sit downstream.
Tube is the only wetted part – on a tube head a reagent change is a tube change, which removes a cleaning step from the service schedule.
Dry running on the diaphragm side – a dry inlet during prime or purge does not damage a diaphragm head.
OEM mounting – heads and motors are panel mounted on the instrument chassis, so the pump sits where the fluid path already runs.
Two motor classes – a stepper for dosing channels and a brushless DC motor for circulation channels, from the same head range.
Seven bores on one head – the 114 head takes 0.5 mm to 4.8 mm tubing, so one spare head covers the whole flow range of a small analyser.
Applications
Three analyser layouts where the pump choice follows from the reagent and the maintenance interval rather than from the flow figure.
Clinical chemistry reagent lines – a tube head doses reagent into a cuvette and a diaphragm head pushes the waste out, so the reagent only meets a tube and the waste line is the one that gets changed.
Water and wastewater sample conditioning – a diaphragm head draws sample through a filter and holds pressure against the filter as it loads, which is the duty a tube head cannot take.
Gas and liquid degassing – a diaphragm head pulls a vacuum on a membrane contactor while a tube head meters the liquid that has to be degassed, and the two duties need different heads.
A pump channel is a small board with a driver and a current limit, so the SUPERB PCBA and the SUPERB PCB are laid out with the motor return kept away from the sensor front end in the same enclosure.
Sourcing and Cross-Reference
The first question on a pump build is the fluid path, not the flow. A reagent that attacks silicone needs a different tube, and a reagent that crystallises needs a head that can be flushed without dismantling it. On the diaphragm side the membrane and the valve seats are the parts that see the liquid, so the material list has to be checked against the reagent before a head is quoted.
The second question is the duty cycle. A head that runs a few seconds per sample lasts far longer than one that circulates a bath all day, and the tube on a peristaltic line is a consumable that has to be in the spares list. Where a tube head is used for waste, the tube is the part that clogs, so the service kit is a tube and nothing else.
The third question is what happens when the line is empty. A diaphragm head primes from dry, a tube head does not, and a method that starts with an empty bottle will behave differently on the two. See lead time and availability for a specific head and BOM shortage recovery where a build is held up by one pump.
The SUPERB Assembly team keeps the tube run short and the head vertical, because a trapped bubble in a horizontal tube reads as a short dose.
Q: Can you supply the pump head and the tubing together, with the bore already matched to the dose the method needs?
A: Yes. Send the reagent list, the dose, the back pressure and the lift, and the SUPERB Tech team returns the head and the bore that fit.
Why Analyser Builders Source Micro Pumps Through SUPERB
- Head figures first: flow at free delivery is quoted next to pressure head and suction height, so a pump is not chosen on a number it cannot hold in the fluid path.
- Tube as a consumable: the tube is listed as a separate line, because on a peristaltic line it is the part that wears and the part that has to be in the spares kit.
- Two families quoted together: a dose channel and a waste channel are usually different pumps, and both are quoted against the same reagent list.
- Dry running stated: whether a head can prime from empty is written down, since that decides how the method starts.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Liquid Flow Sensors for Medical and Analytical Instruments, Solenoid and Pinch Valves for Fluid Handling, Water Quality Sensor Circuits for pH and Conductivity, Differential Pressure Sensors for Ventilators, Level Sensors and Float Switches for Tanks, BOM Shortage Recovery.