Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen

A water quality analyser is a probe plus a small computer, and the computer is the part that decides whether the reading holds its accuracy. The circuits below take pH, conductivity, dissolved oxygen, redox and platinum RTD probes and return a finished value over UART or I2C. Each one is calibrated in one, two or three points and each carries its own ISO reference for the quantity it reads. They run from 3.3 V to 5 V on a 14 by 20 mm board, which is small enough to sit inside a probe head rather than in a cabinet.
quote now

Product Specifications

Product Overview

A pH electrode is a very high impedance source, a conductivity cell is a very low impedance one, and a dissolved oxygen probe is a small current source that depends on temperature, salinity and pressure. Those three signals have almost nothing in common, which is why an analyser with four parameters used to mean four separate front ends, four calibrations and four sets of drift. The circuits on this page put each front end on its own board with its own converter and its own calibration store, and expose one command set, so the host talks to four devices that behave the same way.

The second reason these parts are used is isolation. A pH electrode in a grounded metal tank, a conductivity cell in the same tank and a dosing pump on the same rail form a loop that puts pump noise straight into the reading, and the usual symptom is a pH value that moves when the pump starts. The EC circuit is sensitive enough to read microvolt levels bled into the water, so the fix is to isolate the front end or to isolate the interface, and to keep the probe cable away from the pump cable. The measurement itself is not improved by a better probe if the loop is not broken.

The third point is calibration practice. A pH circuit is calibrated in two or three buffers, an ORP circuit in a single reference solution, a conductivity circuit in a standard solution at a known temperature and a dissolved oxygen circuit in air at a known pressure. The calibration is stored in the circuit, so a probe can be replaced without recalibrating the host, and a circuit can be replaced without recalibrating the probe. That split is what makes a field service visit a probe swap rather than a full revalidation.

The fourth point is the number of parameters that come out of one measurement. The conductivity circuit also reports total dissolved solids, salinity and specific gravity, all derived from the same reading. The dissolved oxygen circuit applies temperature, salinity and pressure compensation from values the host sends. The RTD circuit reads the platinum probe that the other three use for their own temperature compensation, which is why one PT-1000 sensor in the flow cell can serve the whole analyser.

Key Specifications

ParameterValueNote
Quantities coveredpH, conductivity with TDS, salinity and specific gravity, dissolved oxygen, redox potential, temperatureone circuit per quantity, one command set for all of them
pH range and accuracy0.001 to 14.000, resolution 0.001, accuracy plus or minus 0.002calibrated in one, two or three buffers
Conductivity range and accuracy0.07 to 500,000 uS/cm, accuracy plus or minus 2 percentreading time 600 ms, TDS and salinity derived from the same value
Dissolved oxygen range and accuracy0.00 to 100 mg/L, 0 to 350 percent saturation, accuracy plus or minus 0.05 mg/Ltemperature, salinity and pressure compensation sent by the host
Redox range and accuracy-1020 to 1020 mV, accuracy plus or minus 1 mVsingle point calibration against a reference solution
Temperature range and accuracy-126.000 to 1254 C, resolution 0.001, accuracy plus or minus (0.1 + 0.0017 x C)accepts PT-100 and PT-1000, two, three or four wire
Supply voltage3.3 to 5 V on the pH, EC, DO and ORP circuits, 3.3 to 5.5 V on the RTD circuit5 V tolerant 3.3 V logic on the serial lines
InterfaceUART or I2C, ASCII command setaddresses 99, 100, 97, 98 and 102 in decimal
Board size13.97 by 20.16 mm, 1.76 to 1.86 gfits a probe head, a cable gland housing or a DIN carrier
Standards referencedISO 10523 for pH, ISO 7888 for conductivity, ISO 11271 for redox potentialthe circuits are specified against the method, not only against a bench meter
Supported probesany pH or ORP electrode, any conductivity probe from K 0.01 to K 10.2, any galvanic dissolved oxygen probe, PT-100 or PT-1000probe brand is not restricted by the circuit
Calibration storageheld in the circuit, not in the hosta probe swap does not need a host recalibration
Data logging50 readings held on the RTD circuituseful where a temperature excursion has to be shown after the fact
Probe connectionBNC or bare wire depending on probe familykeep the probe cable clear of pump and valve cables

Included Models and Order Codes

The five circuits are ordered by the quantity they read, and the part number is the same shape in each family: EZO-pH, EZO-EC, EZO-DO, EZO-ORP and EZO-RTD. They share the board outline, the supply range and the command set, so a four parameter analyser can carry four of them on one bus with four different addresses and one firmware path. The pH circuit reads 0.001 to 14.000 with a resolution of 0.001 and an accuracy of plus or minus 0.002, and it accepts any electrode, which is why it is usually paired with a lab grade or an industrial probe rather than a fixed one. The EC circuit scales its resolution with the reading, from 0.01 uS/cm at the low end to 100 uS/cm above 100 mS/cm, so the K value of the probe has to match the range that matters. The DO circuit reports both mg/L and percent saturation and takes the salinity and pressure figures from the host. The ORP circuit is a single point calibration and reads to plus or minus 1 mV. The RTD circuit auto detects PT-100 and PT-1000 and holds 50 readings, which covers the temperature channel of the whole analyser.

Part numberReadsRangeAccuracyInterfaceNote
EZO-pHpH0.001 to 14.000plus or minus 0.002UART or I2C, address 99one, two or three point calibration, ISO 10523
EZO-ECconductivity, TDS, salinity, specific gravity0.07 to 500,000 uS/cmplus or minus 2 percentUART or I2C, address 100600 ms per reading, probes from K 0.01 to K 10.2, ISO 7888
EZO-DOdissolved oxygen0.00 to 100 mg/L and 0 to 350 percent saturationplus or minus 0.05 mg/LUART or I2C, address 97temperature, salinity and pressure compensation sent by the host
EZO-ORPredox potential-1020 to 1020 mVplus or minus 1 mVUART or I2C, address 98single point calibration, ISO 11271
EZO-RTDtemperature-126.000 to 1254 Cplus or minus (0.1 + 0.0017 x C)UART or I2C, address 102PT-100 or PT-1000 auto detect, 50 reading logger

Matching a Probe to the Reading Range

The circuit is only half of the measurement. The probe decides the range the reading is accurate over, and the conductivity family is where this matters most: the same circuit gives its best resolution at a different band for each K value. The table below lists the bands published with the EC circuit and the probe types used with the other four.

MeasurementProbe requirementAccurate reading bandNote
Conductivity, K 0.1 probetwo electrode cell, K 0.10.07 to 50,000 uS/cmultra pure water and low conductivity loops
Conductivity, K 1.0 probetwo electrode cell, K 1.05 to 200,000 uS/cmgeneral water and process loops
Conductivity, K 10 probetwo electrode cell, K 1010 uS/cm to 1 S/cmbrine, sea water and concentrated chemical lines
pHany glass electrode with a BNC lead0.001 to 14.000 pHtwo or three buffers bracket the working range
Redoxany platinum or gold ORP electrode-1020 to 1020 mVsingle reference solution sets the offset
Dissolved oxygengalvanic probe with a PTFE membrane0.00 to 100 mg/L0 mV at zero oxygen is the one constant across probes
TemperaturePT-100 or PT-1000, two, three or four wire-126.000 to 1254 Cthe circuit detects the type and reads either class

Key Features

Finished value on the bus – the converter, the calibration and the command parser sit on the board, so the host reads a number rather than a raw electrode voltage.

One command set for five parameters – pH, conductivity, oxygen, redox and temperature answer to the same ASCII protocol, which keeps the analyser firmware to one driver.

Calibration held on the board – a probe can be replaced in the field without touching the host calibration, and a board can be replaced without recalibrating the probe.

Resolution that follows the range – the conductivity circuit reports 0.01 uS/cm at the low end and 100 uS/cm above 100 mS/cm, so the reading keeps four valid digits across the span.

Derived values included – total dissolved solids, salinity and specific gravity come from the conductivity reading, which removes three separate instruments from a skid.

Standard referenced per quantity – pH is specified against ISO 10523, conductivity against ISO 7888 and redox against ISO 11271, which is the language a validation file needs.

Small enough for a probe head – 14 by 20 mm on 3.3 to 5 V means the electronics can travel with the probe instead of sitting in a cabinet at the end of a long cable.

Address set per board – five default addresses let all five circuits share one bus without a multiplexer, which is what a compact four parameter analyser needs.

Applications

Three water circuits where the choice between a lab grade probe, an industrial probe and a flow cell is decided by the maintenance interval rather than by the electronics.

Dialysis water loop – conductivity and pH are watched at the reverse osmosis outlet and at the loop return, and the 0.07 uS/cm floor of the K 0.1 probe is what makes a small rise visible before it reaches a patient.

In vitro diagnostics analyser – a pH and a redox circuit sit inside the instrument next to the reagent lines, and the board size lets them live in the reagent compartment rather than in a separate box.

Boiler and cooling water – conductivity drives the blowdown valve while a PT-1000 on the same bus supplies the temperature compensation, so the two readings are taken at the same moment in the same pipe.

An analyser front end is a small board with sensitive inputs, so the SUPERB PCBA and the SUPERB PCB are laid out with the pH input kept away from the pump rail and the conductivity drive.

Sourcing and Cross-Reference

The first question on a water quality build is the probe, not the circuit. A pH electrode with a blocked junction gives a stable reading in buffer and a drifting one in process, and no circuit can correct that. The conductivity probe has to match the band that matters, because a K 10 cell in pure water reads noise and a K 0.1 cell in brine reads a short. The dissolved oxygen probe has to be a galvanic type with a serviceable membrane, since the electrolyte and the membrane are the parts that age.

The second question is where the calibration lives. With the calibration stored on the circuit, a replacement probe needs one calibration in the flow cell, and a replacement circuit takes the stored values with the old board. That is the reason a service kit usually carries a circuit and a probe together rather than a probe alone.

The third question is the cable and the earth. A pH electrode is a high impedance source, so a damp connector reads as a drifting value, and a conductivity cell is sensitive to noise coupled in from pumps and dosing valves. See lead time and availability for a specific board and BOM shortage recovery where a build is held up by one of the five.

The SUPERB Assembly team keeps the pH connector dry and the probe leads short, because a damp BNC reads as a drifting value that looks like a failed electrode.

Q: Can you supply the pH and conductivity circuits with the probe K value already matched, so the analyser does not need a new calibration procedure?

A: Yes. Send the parameter list, the probe type and K value, the supply and the bus, and the SUPERB Tech team returns the circuits and the addresses that fit.

Why Analyser Builders Source Water Quality Circuits Through SUPERB

  1. Range and probe band together: the K value that gives the accurate band is listed next to the circuit, so a pure water loop and a brine line are not quoted the same probe.
  2. Calibration location stated: the calibration is held on the board, which is what makes a field probe swap a single step instead of a host recalibration.
  3. Standards named: each quantity carries its ISO reference, which is the line a validation file has to quote.
  4. One bus for five quantities: the default addresses let pH, conductivity, oxygen, redox and temperature share a bus, which keeps the analyser wiring to four wires plus power.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Liquid Flow Sensors for Medical and Analytical Instruments, Level Sensors and Float Switches for Tanks, Differential Pressure Sensors for Ventilators, Gas Sensors for Air Quality and Leak Detection, IEC 60601 Power Supplies for Medical Diagnostic Equipment, BOM Shortage Recovery.