Quick Disconnect Couplings and Luer Fittings for Reagent Lines
Product Specifications
Product Overview
A quick disconnect does two jobs at once: it holds a tube joint that has to come apart, and it stops the liquid when it does. On a reagent bottle the second job matters more than the first, because a drip at every bottle change ends up on the instrument floor and in the service log. A valved coupling closes both halves as they separate, and a non spill coupling is shaped so that the small volume left in each half stays inside instead of running down the outside.
The first figure to compare is the flow coefficient against the tube bore. A coupling with a Cv of 0.6 on a quarter inch line is a restriction, and on a gravity fed reagent line that restriction becomes a slow fill rather than a fast one. The second is the pressure and temperature pair, because the same body is rated differently for water and for a warm cleaning solution, and a coupling chosen on pressure alone can be outside its temperature rating at the cleaning step.
The third figure is the wetted material. Acetal bodies with Buna-N seals are the general purpose choice for mild reagents, and they are not the choice for a solvent or for a strong base. A luer fitting in nylon is fine for a small bore sample line and is not fine for a line that is autoclaved. The material list on the coupling has to match the reagent list on the bottle, and that check is the one that gets skipped when a coupling is bought as an accessory.
The fourth point is the tubing interface. A hose barb takes a push fit and a clamp, a threaded port takes a fitting, and a panel mount puts the coupling through the wall of the instrument so the tube does not have to be reached inside. Getting that wrong costs an adapter, and every adapter is another joint that can leak. The parts on this page are listed with their termination so the panel layout can be made from the same table.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Coupling types | valved and non valved quick disconnects, plus luer fittings | a valved coupling closes both halves on separation |
| Working pressure | vacuum to 120 psi, 8.3 bar on both the quarter inch and the non spill families | the body rating, not the tube rating |
| Temperature | 0 to 71 C on the non spill family, minus 40 to plus 82 C on the quarter inch valved family | a warm cleaning step has to be inside this range |
| Flow coefficient | Cv 0.6 on the non spill coupling, 1.40 maximum on the quarter inch family, 0.82 and 0.51 on the reduced bore versions | a coupling is a restriction and the Cv says how much |
| Spill volume | under 0.10 cc per disconnect at all rated pressures on the non spill family | the figure that decides whether a bottle change needs a wipe |
| Air inclusion | 0.26 cc per connect on the non spill family | air in the line is read as a bubble by a flow sensor |
| Body materials | acetal on the quarter inch family, with 316 stainless internal springs and stainless external latch | mild reagents only |
| Seals | Buna-N as standard, other materials available | the seal is the part the reagent attacks |
| Lubricant | Krytox PFPE on the non spill family | inert, so it does not react with the reagent |
| Tubing range | 1/4 to 3/8 inch, 6.4 to 9.5 mm on the quarter inch family | tube size is chosen before the coupling |
| Luer fittings | female luer to 1/16 inch hose barb in nylon | small bore lines inside a manifold |
| Terminations | hose barb, pipe thread in NPT and BSPT, panel mount, in line | a panel mount keeps the tube inside the instrument |
| Mating rule | the quarter inch family mates with the metal and polypropylene versions of the same series | different series with the same bore do not mate |
| Cleaning | the non spill family is quoted for cleaning in place at its rated temperature | check the seal before a caustic wash |
Included Models and Order Codes
The table lists the parts that cover a reagent bottle, a panel pass through and a manifold joint. The non spill insert and body pair is the one to use where a bottle is changed with liquid in the line, because the spill figure is under 0.10 cc per disconnect. The quarter inch valved family covers the same duty at a lower cost and a higher flow coefficient, with a valved insert quoted as PLCD 240-04 and an unvalved insert as PLC 240-04, both on 1/4 inch NPT. The panel mount valved insert PLCD42004 puts the joint through the instrument wall on a 1/4 inch hose barb, and the luer fitting LF2131 joins a 1/16 inch tube inside a manifold.
| Part | Type | Termination | Flow | Note |
|---|---|---|---|---|
| CPC NS4 insert and body | non spill, valved both halves | hose barb and thread | Cv about 0.6 | under 0.10 cc spill per disconnect, 8.3 bar, 0 to 71 C |
| CPC PLCD 240-04 | quick disconnect, valved insert | 1/4 inch NPT male | Cv 1.40 maximum | acetal body, Buna-N seal, 8.3 bar, minus 40 to plus 82 C |
| CPC PLC 240-04 | quick disconnect, unvalved insert | 1/4 inch NPT male | Cv 1.40 maximum | straight through, for a line that is already shut off |
| CPC PLCD 240-040 | quick disconnect, valved insert | 1/4 inch BSPT male | Cv 1.40 maximum | the BSPT thread for a European panel |
| CPC PLCD 240-060 | quick disconnect, valved insert | 3/8 inch BSPT male | Cv 1.40 maximum | larger thread on the same body family |
| CPC PLCD42004 | quick disconnect, valved panel mount insert | 1/4 inch hose barb | Cv 1.40 maximum | panel mount, so the tube stays inside the instrument |
| CPC LF2131 | luer fitting | female luer to 1/16 inch hose barb | small bore | nylon body, black, for manifold joints |
Choosing a Coupling for a Reagent Bottle
The same bottle can be connected four ways, and the choice follows from how often it is changed and what is in it. The table is the short version of that decision.
| Situation | Coupling to use | Why | What to check |
|---|---|---|---|
| Bottle changed daily with liquid in the line | non spill valved pair | under 0.10 cc spill per disconnect | the seal material against the reagent |
| Bottle changed weekly, line emptied first | valved insert on a quarter inch body | the valve already stops the flow | the tube bore against the Cv |
| Panel pass through | panel mount insert | the joint sits in the wall, not inside | the panel cut out and the thread |
| Manifold joint inside the instrument | luer fitting | small bore, low dead volume | the barb size against the tube |
| Line that is clamped off anyway | unvalved insert | no valve to clean | the clamp is the shutoff, so it must be rated |
| Cleaning in place with a warm solution | non spill pair at its rated temperature | the body and the seal are rated for it | the temperature and the caustic concentration |
Key Features
Spill volume stated – under 0.10 cc per disconnect is the figure that decides whether a bottle change needs a wipe down.
Air inclusion stated – 0.26 cc per connect, because air in a reagent line is read as a bubble by the flow sensor downstream.
Flow coefficient per part – Cv from 0.6 to 1.40, so the restriction is known before the line is built.
Pressure and temperature together – 8.3 bar and 71 C on one family, 8.3 bar and 82 C on the other, both written next to the part.
Panel mount option – the joint can sit in the instrument wall, which keeps the tube out of the way of the door.
Mating rule stated – the quarter inch family mates with the metal and polypropylene versions of the same series, and same bore parts from other series do not.
Inert lubricant – Krytox PFPE on the non spill family, so the lubricant does not react with the reagent.
Luer for manifold joints – a female luer to 1/16 inch barb joins small bore tubing with a low dead volume.
Applications
Three joints where the spill volume and the termination decide the coupling rather than the flow.
Reagent bottle connection – a non spill pair on the bottle cap keeps the bench dry when a bottle is swapped with liquid still in the line.
Instrument panel pass through – a panel mount insert puts the joint in the wall, so the tube does not have to be reached past the door.
Manifold and probe joints – a luer fitting joins 1/16 inch tubing inside the fluidic manifold, where the dead volume matters more than the flow.
A fluidic panel is a set of joints on a plate, so the SUPERB PCBA and the SUPERB PCB are mounted above the drip line and the couplings are kept below it.
Sourcing and Cross-Reference
The first question on a coupling build is the reagent list. Acetal and Buna-N cover mild reagents and are the wrong choice for a solvent, and a coupling that passes a water test can swell on a washing solution. The second question is the tube, because the barb size and the Cv have to match the bore that the method already uses.
The second question is what happens at a disconnect. A non spill pair keeps the spill under 0.10 cc, which is the difference between a clean change and a service call. A valved insert on a cheaper body stops the flow but leaves a film on the outside, and that film is what attracts dust and crystals.
The third question is the panel. A coupling inside the instrument has to be reached with the door open, while a panel mount is reached from outside, and that changes the service time on every bottle change. See lead time and availability for a specific coupling and BOM shortage recovery where a build is held up by one part.
The SUPERB Assembly team wet tests every joint before the panel is closed, because a coupling that weeps shows up as a dried crystal a week later.
Q: Can you supply the coupling as a matched pair, with the insert and the body quoted together so the two halves mate?
A: Yes. Send the reagent, the tube size, the pressure and the panel cut out, and the SUPERB Tech team returns the mating pairs that fit.
Why Instrument Builders Source Couplings Through SUPERB
- Matched pairs: the insert and the body are quoted together, because two halves from different series with the same bore do not mate.
- Spill figure stated: the volume lost per disconnect is written down, which is what a clean room bottle change is judged on.
- Seal against the reagent: the wetted material list is quoted next to the part, so a solvent line is not built with a water coupling.
- Termination listed: barb, NPT, BSPT and panel mount are separate lines, so the panel layout comes from the same table.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Solenoid and Pinch Valves for Fluid Handling, Peristaltic and Diaphragm Micro Pumps, Liquid Flow Sensors for Medical and Analytical Instruments, Water Quality Sensor Circuits for pH and Conductivity, Wire to Board Crimp Housings and Headers, BOM Shortage Recovery.