Air Bubble Detectors for Infusion and Reagent Lines
Product Specifications
Product Overview
A bubble detector is a go or no go sensor with a number behind it, and the number is the smallest bubble it will catch reliably. On an infusion line that figure is what decides whether a 4 microlitre bubble reaches the patient or trips the alarm, and on a reagent line it decides whether a segment of air spoils a calibration. The detector does not measure volume, it measures the change in transmission across the tube, so the same sensor reports a different bubble size on a different tube. That is why every part on this page is quoted with the tube it was characterised on rather than with a single sensitivity figure.
The ultrasonic type sends a burst through the tube wall and the liquid and reads what arrives on the other side. Air attenuates the signal strongly, so a bubble shows as a drop in amplitude, and the measurement cycle is 200 microseconds with a response between 0.2 and 2 milliseconds. The advantage is that the tube can be opaque, which matters on a loaded PVC or a silicone line where an optical sensor would see nothing. The limit is that the sound has to couple through the wall, so the tube hardness and wall thickness have to match the sensor channel. A wall of 10 to 20 percent of the outer diameter is the range these sensors are set up for.
The optical type puts an infrared LED on one side and a phototransistor on the other, and reads the change in transmitted power when the liquid is replaced by air. It is faster, 20 to 30 microseconds on the parts listed here, and it detects a fixed air gap rather than a bubble volume, so a 0.8 mm gap is caught on a 2 mm, 3 mm or 4 mm tube. The condition is a transparent tube with a clean wall, because a residue on the inside of the tube changes the reading as much as a bubble does. On a reagent line that carries a dye or a precipitate, that residue is the normal case rather than the exception.
Two behaviours catch people out. Foam reads as air on both principles, because the sound and the light see a mixture with a different index, so a line that foams at the top of a bottle will alarm on foam rather than on a bubble. And a bubble that is smaller than 30 percent of the inner diameter can pass an ultrasonic sensor without a reliable trip, even though its volume is well above the sensitivity figure, because a small bubble that does not span the channel does not change the path enough. The tables below give both figures, the diameter and the volume, so the two are not confused.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Measuring principle | ultrasonic transmission, or optical transmission with an infrared LED and a phototransistor | ultrasound works through opaque tube, optics need a clear one |
| Bubble size, ultrasonic | greater than 0.3 microlitre on the ABD05 family, greater than about one third of the inner diameter on the ABD06 family | a bubble that does not span the channel is not caught |
| Bubble size, optical | an air gap of 0.8 mm or more on the Panasonic BE-A family | a fixed gap, not a volume, so tube bore changes the volume |
| Response time | 0.2 to 2 ms on the ultrasonic family, 20 to 30 microseconds on the optical family for a bubble | the pump stop distance follows from the response and the flow |
| Measuring cycle | 200 microseconds on the ultrasonic family | the output updates faster than a peristaltic pulse |
| Supply | 5 V DC plus or minus 0.2 V on the ABD05 family, 5 to 24 V DC on the optical family | the ABD05 output is 5 V logic |
| Output | TTL switching, serial or PWM on the ABD05 family, NPN or PNP open collector at 50 mA on the optical family | two outputs on the optical family, liquid absent on and liquid present on |
| Channel width | 1.8 mm to 15.5 mm across the ultrasonic family | the channel is chosen by the tube outer diameter |
| Tube range | 2.4 mm to 22 mm outer diameter on the ultrasonic parts, 2 to 4 mm on the optical parts | wall thickness 10 to 20 percent of the outer diameter |
| Ingress protection | IP67 on the ABD05 family with potted electronics, IP40 on the optical family | a washdown panel needs the potted version |
| Operating temperature | plus 5 to plus 45 C on ABD05, plus 5 to plus 60 C on ABD06, minus 25 to plus 55 C on the optical family | a cold room line is outside the ultrasonic range |
| Materials in the optical path | POM housing and PMMA measuring cell on ABD05, PET housing with a POM cover on ABD06 | resistant to disinfection solutions, not to all solvents |
| Foam | detected as air on both principles | a foaming reagent will alarm before a bubble arrives |
| Mounting | clamp on the tube for ABD05, two M4 threaded holes on the rear for ABD06 | the sensor is a fixed part of the machine, not a probe |
Included Models and Order Codes
The ultrasonic family is split by channel width, and the channel is chosen from the tube rather than from the sensitivity. ABD05.56 takes a 2.4 mm outer diameter tube with a 0.8 mm bore, which is the smallest line on the page and the one that catches the smallest bubble, 0.5 microlitre. The 3.4 mm and 5.2 mm channels cover the usual infusion and reagent tube, 4.2 mm and 6.8 mm outer diameter, and the response is the same 1 millisecond typical across the family. The ABD06 parts take harder tube and higher pressure, from 6 mm up to 22 mm outer diameter, and they detect a bubble above about one third of the bore rather than a fixed volume. The optical parts are the fast option on a clear 2 to 4 mm tube, and they are listed with the air gap they catch rather than with a volume.
| Part | Principle | Channel or tube | Bubble limit | Output | Note |
|---|---|---|---|---|---|
| SONOCHECK ABD05.56, 200 02 0057 | ultrasonic | 1.8 mm channel, 2.4 mm tube | reliably above 1.0 mm, 0.5 microlitre | TTL, serial or PWM | smallest bore on the page, PVC tube 0.8 mm inner diameter |
| SONOCHECK ABD05.54, 200 02 0054 | ultrasonic | 2.2 mm channel, 3.0 mm tube | reliably above 1.5 mm, 1.5 microlitre | TTL, serial or PWM | rectangular body, 2.0 mm bore |
| SONOCHECK ABD05.55, 200 02 0056 | ultrasonic | 2.2 mm channel, 3.0 mm tube | reliably above 1.5 mm, 1.5 microlitre | TTL, serial or PWM | circular body, same channel as the 0054 |
| SONOCHECK ABD05.50, 200 02 0050 | ultrasonic | 3.4 mm channel, 4.2 mm tube | reliably above 2.0 mm, 4 microlitre | TTL, serial or PWM | circular body, the common infusion tube |
| SONOCHECK ABD05.51, 200 02 0051 | ultrasonic | 3.4 mm channel, 4.2 mm tube | reliably above 2.0 mm, 4 microlitre | TTL, serial or PWM | rectangular body, same channel as the 0050 |
| SONOCHECK ABD05.57, 200 02 0107 | ultrasonic | 4.4 mm channel, 5.5 mm tube | reliably above 2.9 mm, 10 microlitre | TTL, serial or PWM | circular body, 3.5 mm bore |
| SONOCHECK ABD05.52, 200 02 0052 | ultrasonic | 5.2 mm channel, 6.8 mm tube | reliably above 3.5 mm, 20 microlitre | TTL, serial or PWM | circular body, reagent line size |
| SONOCHECK ABD05.53, 200 02 0053 | ultrasonic | 5.2 mm channel, 6.8 mm tube | reliably above 3.5 mm, 20 microlitre | TTL, serial or PWM | rectangular body, same channel as the 0052 |
| SONOCHECK ABD06.115 | ultrasonic | 5.6 mm channel, 6.0 to 8.0 mm tube | above about one third of the bore | switching or analogue | hard tube and higher pressure, PET housing |
| SONOCHECK ABD06.114 | ultrasonic | 8.0 mm channel, 8.9 to 11.4 mm tube | above about one third of the bore | switching or analogue | 10.0 mm bore maximum on this channel |
| SONOCHECK ABD06.119 | ultrasonic | 11 mm channel, 12.0 to 15.5 mm tube | above about one third of the bore | switching or analogue | largest channel, 13.5 mm bore |
| SONOCHECK ABD06.120 | ultrasonic | 7.2 mm channel, 8.0 to 10.0 mm tube | above about one third of the bore | switching or analogue | hinged cover, tube changed without tools |
| SONOCHECK ABD06.121 | ultrasonic | 15.5 mm channel, 17.0 to 22.0 mm tube | above about one third of the bore | switching or analogue | hinged cover, 20.0 mm bore |
| Panasonic BE-A201 and BE-A201P | optical, 855 nm | 2 mm outer, 1 mm inner, clear PFA type | 0.8 mm air gap | NPN or PNP open collector | 30 microseconds, IP40, 15 g |
| Panasonic BE-A301 and BE-A301P | optical, 855 nm | 3 mm outer, 2 mm inner, clear PFA type | 0.8 mm air gap | NPN or PNP open collector | 20 microseconds, IP40, 15 g |
| Panasonic BE-A401 and BE-A401P | optical, 855 nm | 4 mm outer, 2.4 mm inner, clear PVC type | 0.8 mm air gap | NPN or PNP open collector | 20 microseconds, IP40, 15 g |
| Elveflow MBD | optical, 890 nm | 1/16 or 1/8 inch outer diameter | liquid to air transition, no volume figure | analogue, true or false | detection module 68 by 29 by 33 mm, separate amplifier |
Bubble Size That Is Actually Detected
The volume figure on an ultrasonic sensor is the volume of a sphere at the detected diameter, not the length of the bubble in the tube, and the difference is worth a look. The table is the characterised data for the ABD05 family on the tube it was measured with.
| Sensor | Tube, outer and inner | Mostly detected | Reliably detected |
|---|---|---|---|
| ABD05.50 and ABD05.51 | PVC, 4.0 mm and 1.6 mm | above 1.5 mm, 2 microlitre | above 2.0 mm, 4 microlitre |
| ABD05.52 and ABD05.53 | PVC, 6.8 mm and 4.3 mm | above 2.0 mm, 4 microlitre | above 3.5 mm, 20 microlitre |
| ABD05.54 and ABD05.55 | PVC 2.8 mm and 2.0 mm, silicone 3.0 mm and 1.0 mm | above 1.2 mm, 1.0 microlitre | above 1.5 mm, 1.5 microlitre |
| ABD05.56 | PVC, 2.4 mm and 0.8 mm | above 0.8 mm on the bench | above 1.0 mm, 0.5 microlitre |
| ABD05.57 | PVC, 5.5 mm and 3.5 mm | above 2.0 mm, 4 microlitre | above 2.9 mm, 10 microlitre |
Key Features
Bubble limit per tube – every part is quoted with the diameter and the volume it catches on a named tube, instead of a single sensitivity claim for the family.
Response in microseconds – 20 to 30 microseconds on the optical parts and 1 millisecond typical on the ultrasonic parts, which sets how far the bubble travels before the pump stops.
Opaque tube option – the ultrasonic parts read through loaded PVC and silicone, where an optical sensor sees nothing.
Wall thickness rule – 10 to 20 percent of the outer diameter is the range the ultrasonic channel is set for, and a thin wall tube couples differently.
Foam declared – both principles read foam as air, so a foaming reagent is an alarm condition rather than a surprise.
Potted IP67 body – the ABD05 electronics are potted and rated IP67, which is what a washdown panel next to a reagent bottle needs.
Serial and PWM output – bubble size and alarms can be coded on a serial frame or a PWM period, not just a switching edge.
Optical air gap – 0.8 mm on the Panasonic parts, stated as a gap so the same figure holds on a 2 mm, 3 mm or 4 mm tube.
Applications
Four places where the bubble figure and the tube decide the sensor rather than the output type.
Infusion pump air in line – an ABD05 on a 4.2 mm tube trips at a 4 microlitre bubble, which is the figure an infusion safety review asks for rather than a percentage of the line.
Reagent line on an analyser – an ABD05.52 or ABD05.53 on a 6.8 mm tube catches a 20 microlitre segment before it reaches the flow cell, where a bubble would read as absorbance.
Sample probe and wash station – an optical BE-A on a clear 3 mm tube responds in 20 microseconds, so the probe can be stopped before the bubble reaches the syringe.
Microfluidic chip feed – an Elveflow MBD on 1/16 inch tubing reports a liquid to air transition without touching the fluid, which keeps the chip path disposable.
On an analyser the detector sits on the tube run beside the SUPERB PCBA and the SUPERB PCB, close enough that the cable is short and the alarm reaches the controller without a long unshielded run.
Sourcing and Cross-Reference
The first question on a bubble detector is the tube, because the sensor is characterised on one and the figures do not carry across. Send the outer diameter, the inner diameter and the wall material, and the channel follows. A sensor bought for a 4.2 mm tube will not clamp a 6.8 mm line, and a shim to make it fit changes the coupling and the sensitivity with it.
The second question is the smallest bubble that has to be caught, in millimetres rather than microlitres on a small bore line. On a 0.8 mm bore, a 0.5 microlitre bubble is about 1 mm long, so a sensor that catches 1 mm is the one that meets the specification. On a 4.3 mm bore the same 1 mm bubble is 14 microlitres and the sensor figure moves with it.
The third question is what the line carries. A dye or a precipitate coats the inside of the tube and an optical sensor reads the coating as a change in transmission, while an ultrasonic sensor is less affected by the film but needs the tube wall to stay coupled. For a patient contact line the sensor has no wetted part at all, which is why it can sit on a disposable tube. See medical device components for the documentation, and last time buy and LTB planning where a sensor family is being narrowed.
The SUPERB Assembly team runs a wet and dry test on every detector it fits, because a channel that is clamped too tight reads a clear tube as air.
Q: Can you supply the detector with the tube sample matched, so the channel is set for our 4.2 mm PVC line rather than a generic 4 mm bore?
A: Yes. Send a length of the tube with the outer diameter, the inner diameter and the wall material, and the SUPERB Tech team returns the channel and the bubble figure that go with it.
Why Instrument Builders Source Bubble Detectors Through SUPERB
- Tube matched before order: the channel is chosen from the outer diameter and wall thickness that the method already uses.
- Bubble figure in diameter and volume: both numbers are quoted, because a volume on a large bore is a much shorter bubble than the same volume on a small one.
- Opaque and clear lines covered: ultrasonic and optical parts are on the same page, so the choice follows the tube rather than the catalogue.
- No wetted part: the sensor clamps the outside of the tube, so a disposable line stays disposable.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Solenoid and Pinch Valves for Fluid Handling, Peristaltic and Diaphragm Micro Pumps, Liquid Flow Sensors for Medical and Analytical Instruments, Fluid Check Valves and Shuttle Valves, Ultrasonic Transducers and Distance Sensors, Turbidity and Photometric Detector Modules, Quick Disconnect Couplings and Luer Fittings, Medical Device Components.