Fluid Check Valves and Shuttle Valves for Reagent and Pneumatic Lines
Product Specifications
Product Overview
A check valve fails in two directions. It can leak backwards, which is the failure everyone expects, or it can refuse to open, which is the failure that stops a run. The second one is more common on a reagent line, because the pressure available is small. A bottle standing 500 mm above the valve gives about 0.7 psi of head, and a valve with a 3 psi spring will never open on that head no matter how long the pump runs. The cracking pressure has to sit below the pressure the line can actually produce at the flow the method needs.
Cracking pressure is the forward differential at which the valve just begins to pass, and it is not the only number that matters. The reseal pressure is where it closes again, and on a diaphragm valve that is close to the cracking figure because the diaphragm is free floating and follows the slightest pressure change. On a spring valve the spring sets both, and the spring itself carries a tolerance. A 3 psi nominal spring with a 30 percent tolerance may open at 2.1 psi or at 3.9 psi, so a line that only makes 2.5 psi is a coin toss rather than a specification.
The seat and the diaphragm are what the reagent touches. A spring valve in natural nylon with Buna-N seals is the general purpose choice for water based reagents, and it is the wrong choice for a solvent or a strong base. Viton covers most solvents and acids, EPDM covers caustic and hot water, and silicone is chosen where the line is disposable and the seal has to meet a biocompatibility class. A valve whose body is fine and whose seal swells will leak at the seat, and the leak shows up as a slow drift in a reagent blank rather than as a puddle.
A shuttle valve is a different part doing a different job: two inlets, one outlet, and the higher of the two pressures wins. It is common on pneumatic lines where a manual push button and an automatic signal both have to drive the same cylinder, and it is rarely used on a liquid reagent line, because two check valves feeding a tee give the same result with smaller dead volume. The shuttle parts on this page are listed for air and inert gas duty, and their cracking figures are around 1 psi or less so that either input can take over without a pressure spike.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Valve types | diaphragm, umbrella, duckbill and spring check valves, plus poppet shuttle valves | the type decides the cracking range more than the body does |
| Cracking pressure | 0.11 psi on a Luer diaphragm valve, 0.25 psi on a barbed diaphragm valve, 0.5 psi on a molded diaphragm valve, 0.5 to 10 psi on spring valves | 0.11 psi is about 77 mm of water, 3 psi is about 2.1 m |
| Spring tolerance | plus or minus 30 percent on the nominal cracking pressure | a 3 psi spring may open between 2.1 and 3.9 psi |
| Reseal pressure | 1.9 psi on the molded diaphragm family, close to cracking on a free floating diaphragm | reseal sets how much reverse flow slips past before the seat closes |
| Maximum working pressure | 50 psi on the molded diaphragm family, 25 or 50 psi on the barbed spring family, 250 psig back pressure on the duckbill body | the barbed duckbill version is limited to 100 psig by the tube |
| Body materials | natural nylon, polypropylene, PVDF, polycarbonate, polysulfone, ABS on the Luer family | PVDF for aggressive reagents, polypropylene for cost |
| Seal materials | silicone, EPDM, Viton, Buna-N, neoprene, and a silicone umbrella or diaphragm | the seal is the part the reagent attacks first |
| Connections | hose barbs from 1/32 to 3/4 inch, 1/8 MNPT, male and female Luer, 10-32 and 1/8 NPT on the shuttle family | the barb has to match the tube the method already uses |
| Luer standard | ISO 80369-7 on the Luer diaphragm family | the small bore standard for medical fluid paths |
| Cleanliness | USP Class VI, ISO 10993-1 and FDA 21 CFR 181.32 on the barbed diaphragm family | what a disposable or a patient contact line is asked for |
| Mounting attitude | any position on the diaphragm family, because a free floating diaphragm has no spring to fight | a spring valve adds its offset in every attitude |
| Shuttle cracking | under 1.0 psi on the 1/8 NPT shuttle family, about 1 psig on the poppet type | either inlet can take over without a pressure spike |
| Shuttle pressure | 0 to 125 psi on the pneumatic shuttle family, 200 psi on the inline type, 300 psig on the poppet type | air and inert gas duty |
| Reverse leakage | under 0.06 standard cubic inches per minute at 1 psi on the umbrella family | a leak figure, not a drop tight claim |
Included Models and Order Codes
The table covers the four checking principles that turn up on analyser and instrument lines. Diaphragm valves are the low pressure end, from 0.11 psi at a Luer port to 0.5 psi on a molded barbed body, and they are used where a pump makes only a little pressure. Umbrella valves sit at 0.25 to 0.75 psi in a plastic body with no spring at all. The duckbill body is the smallest of the group and needs more than 5 psig of back pressure to seat reliably, which is a real limit rather than a footnote. Spring valves start at 0.5 psi and go to 10 psi, and are the ones to use where the line pressure is high enough to pay for the offset and where the reverse pressure has to hold a pump or a bottle.
| Part | Type | Cracking | Connections | Body and seal | Note |
|---|---|---|---|---|---|
| CV7 series | diaphragm, free floating | 0.11 psi | male or female Luer, or Luer to hose barb | ABS body, polysulfone insert, silicone disk | ISO 80369-7, smallest cracking figure on the page |
| CVJT series | diaphragm | 0.25 psi or less | 1/32, 1/16 and 1/8 inch barbs | USP Class VI materials | USP Class VI, ISO 10993-1 and FDA 21 CFR 181.32 |
| CVN, CVP, CVK series | molded diaphragm | 0.5 psi | 1/8 to 3/8 inch barb, 1/8 MNPT | nylon, polycarbonate, PVDF or polypropylene body | seal 1.9 psi, maximum 50 psi, silicone, Viton or neoprene diaphragm |
| CVNUCB-58-BU-1/2# | spring | 1/2 psi | 5/8 inch barb, equal | natural nylon, Buna-N | spring tolerance plus or minus 30 percent |
| CVNUCB-58-BU-1# | spring | 1 psi | 5/8 inch barb, equal | natural nylon, Buna-N | general purpose water line |
| CVNUCB-58-EPDM-1# | spring | 1 psi | 5/8 inch barb, equal | natural nylon, EPDM | caustic and hot water duty |
| CVNUCB-58-BU-3# | spring | 3 psi | 5/8 inch barb, equal | natural nylon, Buna-N | needs 2.1 m of water head to open |
| CVNUCB-516-V-3# | spring | 3 psi | 5/16 inch barb, equal | natural nylon, Viton | solvent and acid duty |
| CVNUCB-516-V-10# | spring | 10 psi | 5/16 inch barb, equal | natural nylon, Viton | for a line that holds pressure against a pump |
| F2804410075A8NV | umbrella | 0.75 psi | 1/4 inch barb | nylon body, Viton umbrella | 1.80 scfm forward at 5 psi, reverse under 0.06 SCIM at 1 psi |
| F2804410025I7PPV | umbrella | 0.25 psi | 3/16 inch barb | polypropylene body, Viton umbrella | lowest cracking figure in a barbed plastic body |
| CKV-M3 | duckbill | under 0.5 psig | 5 mm hex body, barbed or threaded options | brass or 303 stainless, about 2 g | 250 psig back pressure, orifice 0.035 to 0.040 inch, needs over 5 psig back pressure to check |
| SV11-125-44 | poppet shuttle | under 1.0 psi | 1/8 NPT female output, two inputs | pneumatic body | Cv 0.48, 0 to 125 psi, 15.0 scfm at 50 psi and 33 scfm at 125 psi |
| MJSV-1 | poppet shuttle | about 1 psig to shift | 1/8 NPT | pneumatic body | 300 psig maximum, 30 scfm at 50 psig, not a pressure selector |
| R431003347 | diaphragm shuttle, inline | 1 psi differential | 1/8 NPT | pneumatic body | Cv 1.25, 200 psi maximum, minus 40 to 160 F |
Matching a Check Valve to the Line
The same line can be protected four ways, and the choice follows from the pressure the pump makes and the pressure the valve has to hold back. The table is the short version of that decision.
| Situation | Valve to use | Why | What to check |
|---|---|---|---|
| Gravity fed reagent bottle, no pump | diaphragm at 0.11 to 0.25 psi | bottle head is under 1 psi, so a spring valve would never open | the head height and the tube bore |
| Peristaltic pump outlet against a closed valve | spring at 3 to 10 psi | the pump makes pressure, and the spring holds it back when the pump stops | the pump pressure at the working flow |
| Solvent line through a pump | Viton or PVDF wetted parts | Buna-N and nylon swell in solvents | the seal list against the reagent list |
| Disposable patient line | diaphragm in USP Class VI materials | the line is thrown away and has to meet a biocompatibility class | the class and the Luer standard |
| Two pneumatic inputs, one cylinder | shuttle valve | the higher pressure wins without a control signal | the pressure difference needed to shift, about 1 psi |
| Two liquid sources into one line | two check valves into a tee | a shuttle valve is an air part, and its dead volume is larger | the cracking pressure of both valves |
Key Features
Cracking pressure per part – from 0.11 psi at a Luer port to 10 psi on a barbed spring valve, written next to the part instead of a range for the family.
Spring tolerance stated – plus or minus 30 percent, so a 3 psi nominal valve is understood as 2.1 to 3.9 psi before it is ordered.
Head height in plain terms – 0.11 psi is about 77 mm of water and 3 psi is about 2.1 m, which is the comparison that decides a gravity fed line.
Body and seal listed separately – nylon, polypropylene, PVDF and polysulfone bodies with silicone, EPDM, Viton or Buna-N seals, so the reagent check is made on the right column.
Four checking principles – diaphragm, umbrella, duckbill and spring, because the cracking range follows the principle more than the body size.
Duckbill limit stated – the CKV-M3 needs more than 5 psig of back pressure to check reliably, and that limit is on the page rather than in a footnote.
Shuttle valves kept separate – cracking under 1 psi and Cv 0.48 to 1.25 on air and inert gas, with a note that two check valves into a tee is the liquid answer.
Cleanliness classes – USP Class VI, ISO 10993-1, FDA 21 CFR 181.32 and ISO 80369-7 listed on the families that carry them.
Applications
Four places where the cracking pressure decides the part rather than the port size.
Reagent bottle anti siphon – a diaphragm valve at 0.11 psi under the bottle cap stops the line draining back into the bottle when the pump stops, and it opens on the small head the bottle provides.
Pump outlet protection – a spring valve at 3 psi on the outlet of a diaphragm pump keeps the reagent from being pulled back through the pump head when the inlet valve closes.
Wash line isolation – a Viton seated valve separates the wash bottle from the sample path, so a wash step does not dilute the sample left in the needle.
Pneumatic cylinder selection – a shuttle valve lets a manual button and a controller output drive the same cylinder without a logic valve in between.
On a fluidic panel the valves sit below the electronics, so the SUPERB PCBA and the SUPERB PCB stay clear of the drip line and a weeping seat does not reach a connector.
Sourcing and Cross-Reference
The first question on a check valve order is the pressure budget, not the port. A line that makes 1 psi cannot use a 3 psi spring, and the mistake is usually found after the valve is fitted, when the pump runs and nothing moves. Send the head height, the pump pressure at the working flow and the back pressure the valve has to hold, and the cracking pressure follows from those three numbers.
The second question is the reagent. A nylon body with a Buna-N seal is the default and covers water based reagents; a solvent needs Viton or PVDF, and a caustic wash needs EPDM or polypropylene. A valve that passes a water test can still swell on a cleaning solution, and the swelling shows as a seat leak a week later rather than at the bench.
The third question is the tube, because a barb is sized to the tube the method already uses and an adapter adds a joint that can weep. For a disposable or a patient contact line, the Class VI and ISO 10993 declarations have to come with the part. See medical device components for the documentation side, and BOM shortage recovery where one valve is holding up a build.
The SUPERB Assembly team pressure tests each valve seat after it is fitted, because a check valve that leaks backwards is invisible until a blank run comes out high.
Q: Can you supply a 0.5 psi cracking check valve in PVDF with a 1/8 inch barb, and quote the same body with a Viton seal for a solvent line?
A: Yes. Send the head height, the pump pressure, the back pressure and the reagent, and the SUPERB Tech team returns the cracking pressure, the body and the seal that fit.
Why Instrument Builders Source Check Valves Through SUPERB
- Cracking pressure first: the quotation starts from the pressure the line can make, so a valve that cannot open is not offered.
- Seal against the reagent: body and seal are separate columns, and the wetted material is confirmed against the reagent before the order.
- Matched barb sizes: 1/32 to 3/4 inch barbs, Luer and NPT are listed per part, so the valve fits the tube rather than the other way round.
- Air and liquid kept apart: shuttle valves are quoted for pneumatic duty and liquid selection is done with two check valves, which keeps the dead volume small.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Solenoid and Pinch Valves for Fluid Handling, Peristaltic and Diaphragm Micro Pumps, Liquid Flow Sensors for Medical and Analytical Instruments, Quick Disconnect Couplings and Luer Fittings, Air Bubble Detectors for Infusion and Reagent Lines, Differential Pressure Sensors for Ventilators, Medical Device Components, BOM Shortage Recovery.