Turbidity and Photometric Detector Modules for Water Analysis

Turbidity is a scattered light measurement, and the number a plant reports depends on the wavelength, the angle and the standard the sensor was checked against. The modules below use an 860 nm source at 90 degrees, which is the geometry ISO 7027 asks for, and they report in NTU or FNU. Ranges run from 0 to 10 NTU on a drinking water point to 0 to 4,000 NTU on a sludge line, and each one is quoted with its accuracy at that range. Each entry also lists the flow or immersion requirement, because a turbidity sensor that sits in a dead pocket reads the pocket and not the water.
quote now

Product Specifications

Product Overview

A turbidity sensor does not measure dirt. It measures light scattered by particles, and the reading is only comparable to another instrument if the wavelength, the angle and the calibration standard match. ISO 7027 fixes the source at 860 nm in the near infrared and the detector at 90 degrees to the beam, which is why almost every process sensor is built that way. A sensor built at 650 nm reads differently on the same water, and a sensor that reads back scatter instead of side scatter reads differently again.

The first figure to check is the range against the point being monitored. A filter outlet that normally sits at 0.1 NTU needs a range that starts at 0 and a resolution in hundredths, while the same instrument on a return sludge line needs 0 to 4,000 NTU and cares about fouling rather than resolution. Quoting one sensor for both points is how a plant ends up with a reading that is either always at zero or always at full scale.

The second figure is fouling. A turbidity sensor sits in water that deposits a film on the window, and a film scatters light exactly like the particles it is meant to measure. The answer is either a wiper or a cleaning brush on the sensor, or a flow cell that keeps the window washed, and both of those are listed on the part rather than added afterwards. A dry secondary standard is the other half of the answer, because it lets the zero be checked without formazin.

The third figure is the output the plant can use. A 4 to 20 mA loop goes into an existing controller, and Modbus RTU goes into a plant network, and many sensors offer both at once. The supply follows from that: a two wire loop powered sensor draws its power from the loop, while a four wire sensor with a wiper needs a separate supply, and the wiper draws more current than the measurement circuit.

Key Specifications

ParameterValueNote
Measurement principle90 degree scattered light with an 860 nm sourcethe geometry ISO 7027 asks for
Ranges available0 to 10, 0 to 100, 0 to 1,000 and 0 to 4,000 NTU depending on the modelrange is chosen against the point being monitored, not the plant
Accuracyplus or minus 1 to 5 percent of full scale, or 2 percent of reading on the low range sensorsread the figure with the range it belongs to
Resolution0.01 NTU below 10 NTU, 0.1 NTU abovethe resolution a filter outlet needs is in hundredths
Wavelength860 nm, tolerance plus or minus 30 nm on the laboratory referencea 650 nm sensor reads differently on the same water
UnitsNTU, FNU and FTU, where 1 FTU equals 1 NTU equals 1 FNUthe three names describe the same scale
Supply7 to 28 VDC on the inline probes, 12 VDC on the flow cell sensorsa wiper adds current on top of the measurement circuit
Output4 to 20 mA and RS485 Modbus RTU, often at the same timea loop and a network can both be fed from one sensor
Pressure and depthup to 0.8 MPa and 60 m immersion on the submersible probesthe immersion version is not a flow cell
Temperature range0 to 50 C on the optical probes, minus 5 to 60 C on the brush cleaned versiona sample above 50 C needs a cooler
Fouling controlwiper, cleaning brush or a flow cell that washes the windowa film on the window scatters light like the particles
Calibrationformazin standard, or a dry secondary standard where one is supplieda dry standard lets the zero be checked without chemicals

Included Models and Order Codes

Four modules cover the plant from a filter outlet to a sludge line. The FIEDLER S461 comes in a low range version at 0 to 10 and 0 to 100 NTU with 0.01 NTU resolution, and a high range version at 0 to 1,000 and 0 to 4,000 NTU, both with Modbus RTU and IP68 bodies. The TurbiTechw2 LR is a flow cell sensor with a 0 to 30 FTU range and a wiper, quoted as 225125, and a 0 to 500 FTU version quoted as 227386. The brush cleaned inline probe covers 0 to 10 up to 0 to 4,000 NTU in one body with both a 4 to 20 mA output and Modbus. The laboratory reference at 0 to 1,000 NTU is listed because a plant sensor is normally verified against one.

ModuleRangeAccuracyOutput and supplyNote
FIEDLER S461/LT0 to 10 and 0 to 100 NTUbetter than 5 percent of rangeModbus RTU, 12 to 24 VDC, under 3 Wimmersion probe, IP68, 4 bar, 42 mm by 231 mm
FIEDLER S461/TN0 to 1,000 and 0 to 4,000 NTUbetter than 5 percent of rangeModbus RTU, 12 to 24 VDCsame body, for return sludge and raw water
TurbiTechw2 LR 2251250 to 30 FTUplus or minus 2 percent of reading or 0.02 NTURS485, 12 VDC from the monitorflow cell with wiper, 1 bar, 0.5 to 5 l/min, IP65
TurbiTechw2 LR 2273860 to 500 FTUplus or minus 2 percent of reading or 0.1 FTURS485, 12 VDC from the monitorsame flow cell, high range
Inline probe, brush cleaned0 to 10, 0 to 100, 0 to 1,000 and 0 to 4,000 NTUplus or minus 1 percent of full scale, 2 percent on the brush version4 to 20 mA and RS485 Modbus RTU, 7 to 28 VDC0.01 NTU resolution below 10 NTU, IP68, 0.8 MPa
TL2310 laboratory reference0 to 1,000 NTUplus or minus 2 percent of reading plus 0.01 NTUmains powered bench unit860 nm source, 0.001 NTU resolution, used to verify plant sensors

Reading a Turbidity Specification

The same water can be reported at three different numbers by three sensors, and the specification explains most of the difference. The table is the short list of what to compare before two instruments are called equivalent.

Specification lineWhy it mattersWhat to compare
Wavelengthscattering depends on wavelength860 nm for ISO 7027, 650 nm is a different reading
Angleside scatter and back scatter see different particle sizes90 degrees is the standard geometry
Range against the pointa 4,000 NTU sensor cannot resolve 0.1 NTUread the resolution at the low end of the range
Accuracy basispercent of reading and percent of full scale differ by the rangeboth figures with the range they belong to
Standard usedformazin, kaolin and latex give different numbersthe standard named on the certificate
Fouling controla film scatters light like particleswiper, brush or flow cell, and the cleaning interval
Flow at the windowa dead pocket reads the pocket0.5 to 5 l/min on a flow cell, or a wiper on immersion

Key Features

ISO 7027 geometry – an 860 nm source with a 90 degree detector is the geometry the standard describes, so readings are comparable between plants.

Range chosen by point – 0 to 10 NTU for a filter outlet and 0 to 4,000 NTU for a sludge line are separate parts, not one part at two settings.

Resolution at the low end – 0.01 NTU below 10 NTU is what makes a small rise visible before it reaches a filter run.

Fouling control on the part – a wiper, a brush or a flow cell is listed with the sensor, because a film on the window reads as turbidity.

Dry secondary standard – the zero can be checked without mixing formazin, which turns a daily check into a two minute task.

Loop and network together – 4 to 20 mA into an existing controller and Modbus RTU into a plant network from the same sensor.

Immersion and flow cell bodies – a submersible probe for a channel and a flow cell for a pipe, both at 860 nm, so the same calibration practice applies.

Pressure and depth stated – 0.8 MPa and 60 m immersion are written down, which is what a borehole or a pressurised line needs to know.

Applications

Three water points where the range and the fouling control are decided by the process rather than by the sensor.

Drinking water filter outlet – a 0 to 10 NTU sensor with 0.01 NTU resolution watches the outlet, and the dry standard lets the zero be verified before each run.

Wastewater return sludge – a 0 to 4,000 NTU probe with a brush reads the return line, where the window would otherwise be coated within a shift.

Industrial process water – a flow cell at 0.5 to 5 l/min keeps the window washed on a line that would otherwise sit stagnant between batches.

A turbidity head is an LED, a photodiode and a transimpedance stage, so the SUPERB PCBA and the SUPERB PCB keep the LED return away from the photodiode input on the same small board.

Sourcing and Cross-Reference

The first question on a turbidity build is the range and the point. A sensor bought for the plant range is usually wrong for the outlet, and a sensor bought for the outlet is blinded on the sludge line. The second question is the standard the certificate names, because formazin and kaolin give different numbers on the same water and a validation file has to say which one was used.

The second question is the cleaning interval. A wiper or a brush adds a moving part and a current draw, and a flow cell adds a pump, and each of those is a maintenance item that has to be in the plan. A sensor with a dry standard is worth more than a sensor with a better accuracy figure, because the zero check is what actually detects fouling.

The third question is the output. A 4 to 20 mA loop is what an existing controller reads, and Modbus is what a plant historian reads, and a sensor that offers both avoids a converter. See lead time and availability for a specific module and BOM shortage recovery where a build is held up by one probe.

The SUPERB Assembly team blackens the inside of the optical housing before the window is bonded, because a reflection off the wall reads as turbidity at zero.

Q: Can you supply the turbidity module with the range matched to our outlet and the dry standard included for the zero check?

A: Yes. Send the range, the point being monitored, the output and the cleaning method, and the SUPERB Tech team returns the modules that fit.

Why Water Analyser Builders Source Turbidity Modules Through SUPERB

  1. Range matched to the point: the outlet and the sludge line are quoted as different parts, because one sensor cannot resolve both.
  2. Cleaning method on the part: wiper, brush and flow cell are listed with the sensor, so the maintenance interval is known before the order.
  3. Standard named: the calibration standard behind the accuracy figure is stated, which is the line a validation file quotes.
  4. Both outputs: a loop and a network output on the same sensor, so no converter has to be added to the panel.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Water Quality Sensor Circuits for pH and Conductivity, Liquid Flow Sensors for Medical and Analytical Instruments, Peristaltic and Diaphragm Micro Pumps, Level Sensors and Float Switches for Tanks, PIN and Avalanche Photodiodes for Optical Receivers, Component Lead Time Solutions.