Syringe Filters and Filter Paper for Sample Prep
Product Specifications
Product Overview
Sample preparation is the step where an assay is usually lost rather than won. A filter that binds the analyte lowers recovery, a filter with a large hold-up volume dilutes a 200 microlitre sample, and a filter run above its rated pressure bursts at the housing seam rather than at the membrane. The three numbers that decide the outcome are the membrane chemistry, the pore size and the hold-up volume, in that order, with the pressure rating as the hard limit.
Membrane chemistry follows the solvent. PVDF is the low protein binding choice for aqueous and mild organic samples and is the default for HPLC and LC-MS injection. PES gives the highest flow per unit area and is used where a biological sample has to pass quickly at low pressure. SFCA, surfactant-free cellulose acetate, suits aqueous solutions in the pH 4 to 8 range and is the common choice for media and buffers. Hydrophobic PTFE is used for venting and for gases, and for solvents that attack the other membranes. Nylon is used where a solvent such as acetonitrile has to pass without extractables.
Pore size sets what is removed. A 0.22 micrometre membrane removes bacteria and is used where the sample has to be sterile before injection. A 0.45 micrometre membrane is the standard clarification step for HPLC samples and removes particles that would block a guard column. A 1.2 or 5 micrometre membrane is used for viscous samples, for slurries and for prefiltration ahead of a 0.22 micrometre device, which is cheaper than replacing a blocked fine filter. For a sample that is already clear, going finer than 0.45 micrometres buys nothing and costs flow.
Hold-up volume is the number that quietly changes results. A 4 mm device holds less than 10 microlitres after an air purge, a 13 mm device holds 15 to 50 microlitres depending on the family, a 25 mm device holds about 0.1 to 0.2 millilitres and a 28 to 33 mm device holds 0.1 to 0.15 millilitres. On a 1 mL sample a 0.2 mL hold-up is a 20 percent loss, so the device is chosen to match the sample volume, not the other way round. Sartorius publishes the volume bands directly, with the 4 mm family for less than 1 mL, the 15 mm family for 1 to 10 mL and the 25 mm family for 10 to 100 mL.
Pressure and temperature limits are set by the housing, not by the membrane. The small 4 mm PVDF unit carries 14 bar because its housing is a solid barrel, while a 13 mm unit is limited to 5.2 bar and a 25 mm vented unit to 5.1 bar. A 33 mm PES unit is rated 10.3 bar because its housing is a different moulding. On the Sartorius MBS housing the limit is 6.0 bar for the high flow PES version and 4.5 bar for the SFCA and PTFE versions, with a burst pressure of at least 7 bar and a maximum temperature of 60 C. Filter paper follows a different rule: the limit is mechanical strength when wet, and a soft grade tears before it blinds.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Device diameters | 4, 13, 15, 25, 28 and 33 mm | device size is chosen from the sample volume, not the syringe |
| Effective filtration area | 0.07 to 6.2 cm2 | 0.1 cm2 at 4 mm, 0.65 to 0.8 cm2 at 13 mm, 4 cm2 at 25 mm, 4.52 cm2 at 33 mm, 6.2 cm2 at 28 mm |
| Membrane options | PVDF, PES, SFCA, cellulose acetate, nylon, hydrophobic PTFE | PVDF for low protein binding, PES for flow, SFCA for aqueous media |
| Pore sizes | 0.1, 0.22, 0.45, 0.65, 0.8, 1.2 and 5 um | 0.22 um for sterile filtration, 0.45 um for routine clarification |
| Hold-up volume, 4 mm | under 10 uL after air purge | the smallest sample band, for volumes under 1 mL |
| Hold-up volume, 13 mm | 15 uL on the non-sterile PVDF sheet, up to 50 uL on the medical sheet | confirm the sheet before a low volume method is transferred |
| Hold-up volume, 25 to 33 mm | 0.1 to 0.2 mL after air purge | used for 10 to 100 mL sample volumes |
| Hold-up volume, 28 mm MBS housing | 150 uL or less | Sartorius quotes this for the High Flow, NML and HY families |
| Maximum inlet pressure, 4 mm PVDF | 14 bar, which is 200 psi | solid barrel housing, the highest rating in the group |
| Maximum inlet pressure, 13 mm PVDF | 5.2 bar, which is 75 psi | housing limited, not membrane limited |
| Maximum inlet pressure, 25 mm vented PVDF | 5.1 bar, which is 75 psi | flow at 2.1 bar is 40 mL/min or more |
| Maximum inlet pressure, 33 mm PES | 10.3 bar, which is 150 psi | flow 150 mL/min or more, process volume 200 mL |
| Maximum pressure, 28 mm PES high flow | 6.0 bar, which is 87 psi | SFCA and PTFE versions of the same housing stop at 4.5 bar |
| Housing burst pressure | 7 bar or more on the MBS housing | the burst figure is a safety margin, not an operating point |
| Maximum temperature | 45 C on the PVDF units, 60 C on the MBS units | above this the housing creeps and the seal leaks |
| Bubble point, water | 3.4 bar at PES 0.22 um, 2.0 bar at PES 0.45 um | 3.2 bar at SFCA 0.2 um, 2.0 bar at SFCA 0.45 um, 5.0 bar at PES 0.1 um |
| Filter paper retention | 0.7 to 25 um | Grade 42 and GF/F at the fine end, Grade 4 and Grade 41 at the fast end |
| Filter paper ash | 0.06 percent for qualitative grades, 0.007 percent for ashless grades | ashless grades are the ones used before ashing or gravimetry |
| Filter paper thickness | 150 to 675 um | GF/B and GF/D are the thick loading grades at 675 um |
Included Models and Order Codes
Part numbers below are the maker codes for single use devices. The suffixes carry meaning: on the Sartorius range the letter after the dash is the pack and the outlet style, with K for a sterile pack of 50 and Q for a non-sterile pack of 500, so 16541-K and 16541-Q are the same filter in different packaging. On the Merck range the last two characters mark sterility and packaging, so a change from R to X is a change from sterile to non-sterile rather than a change of membrane.
| Part | Membrane and pore | Diameter and area | Hold-up volume | Maximum pressure | Format |
|---|---|---|---|---|---|
| SLGV004SL | PVDF, hydrophilic, 0.22 um | 4 mm, 0.1 cm2 | under 10 uL | 14 bar, 200 psi | sterile, ethylene oxide, 45 C limit |
| SLGVR13SL | PVDF, hydrophilic, 0.22 um | 13 mm, 0.65 cm2 | 50 uL or less on the medical sheet | 5.2 bar, 75 psi | sterile, ethylene oxide, housing 14.7 mm |
| SLGVX13 | PVDF, hydrophilic, 0.22 um | 13 mm, 0.8 cm2 | 15 uL or less after air purge | 5.2 bar family limit | non-sterile, yellow band, bubble point 3.4 bar |
| SLGVVR255F | PVDF, hydrophilic, 0.22 um, vented | 25 mm, 4 cm2 | 0.1 mL or less after air purge | 5.1 bar, 75 psi | sterile, vented dome, 40 mL/min at 2.1 bar |
| SLGP033RS | PES, hydrophilic, 0.22 um | 33 mm, 4.52 cm2 | under 0.1 mL | 10.3 bar, 150 psi | sterile, 150 mL/min, 200 mL process volume |
| 16541-K | PES, 0.22 um | 28 mm, 6.2 cm2 | 150 uL or less | 6.0 bar, 87 psi | sterile, male luer slip, pack of 50 |
| 16541-Q | PES, 0.22 um | 28 mm, 6.2 cm2 | 150 uL or less | 6.0 bar, 87 psi | non-sterile, male luer slip, pack of 500 |
| 16537-K | PES, 0.45 um | 28 mm, 6.2 cm2 | 150 uL or less | 6.0 bar, 87 psi | sterile, male luer lock, amber code |
| 16534-K | SFCA, 0.2 um | 28 mm, 6.2 cm2 | 150 uL or less | 4.5 bar, 65 psi | sterile, male luer lock, blue code |
| 16555-K | SFCA, 0.45 um | 28 mm, 6.2 cm2 | 150 uL or less | 4.5 bar, 65 psi | sterile, male luer lock, yellow code |
Filter Paper and Glass Microfiber Discs, Grade by Grade
Filter paper is chosen on two numbers that pull against each other. Particle retention is the finest particle the grade holds at 98 percent efficiency in liquid, and the Herzberg time is how long 100 mL of water takes to pass 100 cm2 under gravity, so a lower time is a faster paper. Basis weight and thickness tell how much solids the disc holds before it blinds. The ash figure decides whether the grade can be used before ignition or gravimetric weighing.
| Code | Grade | Retention | Herzberg time | Thickness | Basis weight | Ash |
|---|---|---|---|---|---|---|
| 1001-047 | Grade 1, qualitative, 47 mm circle | 11 um | 150 s | 180 um | 88 g/m2 | 0.06 percent |
| 1001-090 | Grade 1, qualitative, 90 mm circle | 11 um | 150 s | 180 um | 88 g/m2 | 0.06 percent |
| 1001-125 | Grade 1, qualitative, 125 mm circle | 11 um | 150 s | 180 um | 88 g/m2 | 0.06 percent |
| 1002-090 | Grade 2, qualitative, 90 mm circle | 8 um | 240 s | 190 um | 103 g/m2 | 0.06 percent |
| 1004-090 | Grade 4, qualitative, 90 mm circle | 20 to 25 um | 37 s | 205 um | 96 g/m2 | 0.06 percent |
| 1005-090 | Grade 5, qualitative, 90 mm circle | 2.5 um | 1420 s | 200 um | 98 g/m2 | 0.06 percent |
| 1442-110 | Grade 42, ashless quantitative, 110 mm | 2.5 um | 1870 s | 200 um | 100 g/m2 | 0.007 percent |
| 1820-047 | GF/A, glass microfiber, 47 mm | 1.6 um | air flow 4.3 s | 260 um | 53 g/m2 | not applicable |
| 1821-047 | GF/B, glass microfiber, 47 mm | 1.0 um | air flow 12 s | 675 um | 143 g/m2 | not applicable |
| 1823-047 | GF/D, glass microfiber, 47 mm | 2.7 um | air flow 2.2 s | 675 um | 121 g/m2 | not applicable |
| 1825-047 | GF/F, glass microfiber, 47 mm | 0.7 um | air flow 19 s | 420 um | 75 g/m2 | not applicable |
Key Features
Six device sizes from one range – 4, 13, 15, 25, 28 and 33 mm bodies cover a 100 microlitre injection and a 200 mL buffer prep with the same handling routine.
Membrane choice by solvent – PVDF for low protein binding, PES for the highest flow, SFCA for aqueous media at pH 4 to 8, hydrophobic PTFE for venting and gases.
Hold-up volume stated per size – Under 10 microlitres at 4 mm, 15 to 50 microlitres at 13 mm, 0.1 to 0.2 millilitres at 25 mm and above, so the dilution step can be calculated.
Pressure rating set by housing – 14 bar on the 4 mm barrel, 5.2 bar on 13 mm, 5.1 bar on the vented 25 mm, 10.3 bar on the 33 mm PES unit and 6.0 bar on the 28 mm PES unit.
Sterile and non-sterile in one family – The same membrane is offered sterile by ethylene oxide or gamma, and non-sterile in bulk packs, with the code difference limited to the suffix.
Filter paper graded by retention and speed – Grade 1 at 11 micrometres and 150 seconds, Grade 4 at 20 to 25 micrometres and 37 seconds, Grade 42 at 2.5 micrometres and 1870 seconds.
Ashless grades for gravimetry – Grade 40, 41 and 42 carry 0.007 percent ash, which is the figure that matters before ignition or before a weighing step.
Glass microfiber without fibre shed – GF/A, GF/B, GF/D and GF/F hold 0.7 to 2.7 micrometres and are used ahead of a fine membrane where a cellulose disc would blind.
Applications
The same parts appear in three places in an analyser: the sample path, the mobile phase path and the vent path.
HPLC and LC-MS sample prep – A 13 mm PVDF 0.22 micrometre device on a 1 to 10 mL sample removes particles ahead of the column, and the 15 microlitre hold-up keeps the first injection representative.
Media, buffer and reagent filtration – A 28 or 33 mm PES unit filters 100 to 200 mL at 150 mL/min or more, which is the step where a sterile 0.22 micrometre device pays for itself in time saved.
Venting and gas paths – A hydrophobic PTFE device passes air and stops liquid, so a reagent bottle breathes without drawing room air into the liquid.
Gravimetric and ash work – Grade 42 at 2.5 micrometres and 0.007 percent ash holds barium sulfate and calcium carbonate without adding weight to the crucible.
The same devices are held for SUPERB PCB and SUPERB PCBA builds of analyser boards, where the filter sits between the sample probe and the injection valve.
Sourcing and Cross-Reference
Filters are consumables with a short shelf life and a long qualification tail, so the sourcing question is usually continuity rather than price. Merck quotes a three year shelf life on the PVDF range, and a method that was validated on one lot expects the next lot to give the same bubble point and the same hold-up. When a pack size changes, the device does not: 16541-K and 16541-Q differ in packaging and sterility only, which is the difference that has to be documented before a switch.
Where a membrane has been discontinued or a pack size is out of stock, the cross reference is made on four fields together, membrane chemistry, pore size, effective area and hold-up volume. A 0.22 micrometre PES device with 4.5 cm2 and 0.1 mL hold-up is interchangeable with another 0.22 micrometre PES device of the same area; a 0.45 micrometre device is not, even when the housing looks the same. The syringe fitting is the last check, because a luer slip outlet will not take the same downstream connector as a luer lock.
Filter paper and glass microfiber are sourced by grade and by diameter rather than by brand alone, since the grade number is the specification. A request for Grade 1 is a request for 11 micrometres at 150 seconds and 0.06 percent ash, and a substitute at 8 micrometres is Grade 2 and a different method. For discs that are punched in house, the sheet format is available in 460 by 570 mm and the yield per sheet is calculated from the disc diameter before the order is placed.
Kits can be assembled and labelled for SUPERB Assembly lines, with the device, the syringe and the label in one bag.
Q: Can you supply the same membrane in a different diameter without requalifying the method?
A: SUPERB Tech confirms the membrane, the pore size and the housing before quoting, because the same diameter is sold with three different membranes.
Why Buy Sample Prep Consumables Through SUPERB
- One part number, one lot trail: Each shipment is booked against the maker code and the lot, so a method transfer can name the exact device that was used.
- Membrane and paper in one order: Syringe devices, membrane discs, qualitative paper and glass microfiber are consolidated, which removes a second purchase order and a second freight charge.
- Consumable plus hardware: The filter order can be combined with the pump tubing, the inline filter and the flow cell that sit on the same skid.
- Short lead time on the fast movers: 0.22 and 0.45 micrometre devices in 13, 25 and 28 mm, together with Grade 1 and GF/A discs, are the lines held for repeat orders.
- Cross reference on discontinued items: Where a device has been replaced, the nearest equivalent is proposed on membrane, pore, area and hold-up, with the differences stated before the order.
- Documentation for audits: Maker code, lot and certificate reference are kept with the shipment record, which is what an audit asks for on a consumable.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Inline Filters and Frits for Reagent Lines, Quartz Flow Cells and Cuvettes, Calibration Buffers and Conductivity Standards, Syringe Pumps and Precision Dispensers, Turbidity and Photometric Detector Modules, Sample Loops and Injection Loops, BOM Shortage Recovery.