Syringe Filters and Filter Paper for Sample Prep

A syringe filter is the last thing a sample touches before the column, so the membrane, the pore size and the hold-up volume decide both the recovery and the blank. This page covers 4 to 33 mm devices in PVDF, PES, SFCA and hydrophobic PTFE, with the pressure, temperature and hold-up figures for each housing size. Filter paper grades are listed with the same two numbers that matter in practice, particle retention and Herzberg filtration time. Glass microfiber discs are included for prefiltration and for gravimetric work where a cellulose grade sheds too much fibre.
quote now

Product Specifications

Product Overview

Sample preparation is the step where an assay is usually lost rather than won. A filter that binds the analyte lowers recovery, a filter with a large hold-up volume dilutes a 200 microlitre sample, and a filter run above its rated pressure bursts at the housing seam rather than at the membrane. The three numbers that decide the outcome are the membrane chemistry, the pore size and the hold-up volume, in that order, with the pressure rating as the hard limit.

Membrane chemistry follows the solvent. PVDF is the low protein binding choice for aqueous and mild organic samples and is the default for HPLC and LC-MS injection. PES gives the highest flow per unit area and is used where a biological sample has to pass quickly at low pressure. SFCA, surfactant-free cellulose acetate, suits aqueous solutions in the pH 4 to 8 range and is the common choice for media and buffers. Hydrophobic PTFE is used for venting and for gases, and for solvents that attack the other membranes. Nylon is used where a solvent such as acetonitrile has to pass without extractables.

Pore size sets what is removed. A 0.22 micrometre membrane removes bacteria and is used where the sample has to be sterile before injection. A 0.45 micrometre membrane is the standard clarification step for HPLC samples and removes particles that would block a guard column. A 1.2 or 5 micrometre membrane is used for viscous samples, for slurries and for prefiltration ahead of a 0.22 micrometre device, which is cheaper than replacing a blocked fine filter. For a sample that is already clear, going finer than 0.45 micrometres buys nothing and costs flow.

Hold-up volume is the number that quietly changes results. A 4 mm device holds less than 10 microlitres after an air purge, a 13 mm device holds 15 to 50 microlitres depending on the family, a 25 mm device holds about 0.1 to 0.2 millilitres and a 28 to 33 mm device holds 0.1 to 0.15 millilitres. On a 1 mL sample a 0.2 mL hold-up is a 20 percent loss, so the device is chosen to match the sample volume, not the other way round. Sartorius publishes the volume bands directly, with the 4 mm family for less than 1 mL, the 15 mm family for 1 to 10 mL and the 25 mm family for 10 to 100 mL.

Pressure and temperature limits are set by the housing, not by the membrane. The small 4 mm PVDF unit carries 14 bar because its housing is a solid barrel, while a 13 mm unit is limited to 5.2 bar and a 25 mm vented unit to 5.1 bar. A 33 mm PES unit is rated 10.3 bar because its housing is a different moulding. On the Sartorius MBS housing the limit is 6.0 bar for the high flow PES version and 4.5 bar for the SFCA and PTFE versions, with a burst pressure of at least 7 bar and a maximum temperature of 60 C. Filter paper follows a different rule: the limit is mechanical strength when wet, and a soft grade tears before it blinds.

Key Specifications

ParameterValueNote
Device diameters4, 13, 15, 25, 28 and 33 mmdevice size is chosen from the sample volume, not the syringe
Effective filtration area0.07 to 6.2 cm20.1 cm2 at 4 mm, 0.65 to 0.8 cm2 at 13 mm, 4 cm2 at 25 mm, 4.52 cm2 at 33 mm, 6.2 cm2 at 28 mm
Membrane optionsPVDF, PES, SFCA, cellulose acetate, nylon, hydrophobic PTFEPVDF for low protein binding, PES for flow, SFCA for aqueous media
Pore sizes0.1, 0.22, 0.45, 0.65, 0.8, 1.2 and 5 um0.22 um for sterile filtration, 0.45 um for routine clarification
Hold-up volume, 4 mmunder 10 uL after air purgethe smallest sample band, for volumes under 1 mL
Hold-up volume, 13 mm15 uL on the non-sterile PVDF sheet, up to 50 uL on the medical sheetconfirm the sheet before a low volume method is transferred
Hold-up volume, 25 to 33 mm0.1 to 0.2 mL after air purgeused for 10 to 100 mL sample volumes
Hold-up volume, 28 mm MBS housing150 uL or lessSartorius quotes this for the High Flow, NML and HY families
Maximum inlet pressure, 4 mm PVDF14 bar, which is 200 psisolid barrel housing, the highest rating in the group
Maximum inlet pressure, 13 mm PVDF5.2 bar, which is 75 psihousing limited, not membrane limited
Maximum inlet pressure, 25 mm vented PVDF5.1 bar, which is 75 psiflow at 2.1 bar is 40 mL/min or more
Maximum inlet pressure, 33 mm PES10.3 bar, which is 150 psiflow 150 mL/min or more, process volume 200 mL
Maximum pressure, 28 mm PES high flow6.0 bar, which is 87 psiSFCA and PTFE versions of the same housing stop at 4.5 bar
Housing burst pressure7 bar or more on the MBS housingthe burst figure is a safety margin, not an operating point
Maximum temperature45 C on the PVDF units, 60 C on the MBS unitsabove this the housing creeps and the seal leaks
Bubble point, water3.4 bar at PES 0.22 um, 2.0 bar at PES 0.45 um3.2 bar at SFCA 0.2 um, 2.0 bar at SFCA 0.45 um, 5.0 bar at PES 0.1 um
Filter paper retention0.7 to 25 umGrade 42 and GF/F at the fine end, Grade 4 and Grade 41 at the fast end
Filter paper ash0.06 percent for qualitative grades, 0.007 percent for ashless gradesashless grades are the ones used before ashing or gravimetry
Filter paper thickness150 to 675 umGF/B and GF/D are the thick loading grades at 675 um

Included Models and Order Codes

Part numbers below are the maker codes for single use devices. The suffixes carry meaning: on the Sartorius range the letter after the dash is the pack and the outlet style, with K for a sterile pack of 50 and Q for a non-sterile pack of 500, so 16541-K and 16541-Q are the same filter in different packaging. On the Merck range the last two characters mark sterility and packaging, so a change from R to X is a change from sterile to non-sterile rather than a change of membrane.

PartMembrane and poreDiameter and areaHold-up volumeMaximum pressureFormat
SLGV004SLPVDF, hydrophilic, 0.22 um4 mm, 0.1 cm2under 10 uL14 bar, 200 psisterile, ethylene oxide, 45 C limit
SLGVR13SLPVDF, hydrophilic, 0.22 um13 mm, 0.65 cm250 uL or less on the medical sheet5.2 bar, 75 psisterile, ethylene oxide, housing 14.7 mm
SLGVX13PVDF, hydrophilic, 0.22 um13 mm, 0.8 cm215 uL or less after air purge5.2 bar family limitnon-sterile, yellow band, bubble point 3.4 bar
SLGVVR255FPVDF, hydrophilic, 0.22 um, vented25 mm, 4 cm20.1 mL or less after air purge5.1 bar, 75 psisterile, vented dome, 40 mL/min at 2.1 bar
SLGP033RSPES, hydrophilic, 0.22 um33 mm, 4.52 cm2under 0.1 mL10.3 bar, 150 psisterile, 150 mL/min, 200 mL process volume
16541-KPES, 0.22 um28 mm, 6.2 cm2150 uL or less6.0 bar, 87 psisterile, male luer slip, pack of 50
16541-QPES, 0.22 um28 mm, 6.2 cm2150 uL or less6.0 bar, 87 psinon-sterile, male luer slip, pack of 500
16537-KPES, 0.45 um28 mm, 6.2 cm2150 uL or less6.0 bar, 87 psisterile, male luer lock, amber code
16534-KSFCA, 0.2 um28 mm, 6.2 cm2150 uL or less4.5 bar, 65 psisterile, male luer lock, blue code
16555-KSFCA, 0.45 um28 mm, 6.2 cm2150 uL or less4.5 bar, 65 psisterile, male luer lock, yellow code

Filter Paper and Glass Microfiber Discs, Grade by Grade

Filter paper is chosen on two numbers that pull against each other. Particle retention is the finest particle the grade holds at 98 percent efficiency in liquid, and the Herzberg time is how long 100 mL of water takes to pass 100 cm2 under gravity, so a lower time is a faster paper. Basis weight and thickness tell how much solids the disc holds before it blinds. The ash figure decides whether the grade can be used before ignition or gravimetric weighing.

CodeGradeRetentionHerzberg timeThicknessBasis weightAsh
1001-047Grade 1, qualitative, 47 mm circle11 um150 s180 um88 g/m20.06 percent
1001-090Grade 1, qualitative, 90 mm circle11 um150 s180 um88 g/m20.06 percent
1001-125Grade 1, qualitative, 125 mm circle11 um150 s180 um88 g/m20.06 percent
1002-090Grade 2, qualitative, 90 mm circle8 um240 s190 um103 g/m20.06 percent
1004-090Grade 4, qualitative, 90 mm circle20 to 25 um37 s205 um96 g/m20.06 percent
1005-090Grade 5, qualitative, 90 mm circle2.5 um1420 s200 um98 g/m20.06 percent
1442-110Grade 42, ashless quantitative, 110 mm2.5 um1870 s200 um100 g/m20.007 percent
1820-047GF/A, glass microfiber, 47 mm1.6 umair flow 4.3 s260 um53 g/m2not applicable
1821-047GF/B, glass microfiber, 47 mm1.0 umair flow 12 s675 um143 g/m2not applicable
1823-047GF/D, glass microfiber, 47 mm2.7 umair flow 2.2 s675 um121 g/m2not applicable
1825-047GF/F, glass microfiber, 47 mm0.7 umair flow 19 s420 um75 g/m2not applicable

Key Features

Six device sizes from one range – 4, 13, 15, 25, 28 and 33 mm bodies cover a 100 microlitre injection and a 200 mL buffer prep with the same handling routine.

Membrane choice by solvent – PVDF for low protein binding, PES for the highest flow, SFCA for aqueous media at pH 4 to 8, hydrophobic PTFE for venting and gases.

Hold-up volume stated per size – Under 10 microlitres at 4 mm, 15 to 50 microlitres at 13 mm, 0.1 to 0.2 millilitres at 25 mm and above, so the dilution step can be calculated.

Pressure rating set by housing – 14 bar on the 4 mm barrel, 5.2 bar on 13 mm, 5.1 bar on the vented 25 mm, 10.3 bar on the 33 mm PES unit and 6.0 bar on the 28 mm PES unit.

Sterile and non-sterile in one family – The same membrane is offered sterile by ethylene oxide or gamma, and non-sterile in bulk packs, with the code difference limited to the suffix.

Filter paper graded by retention and speed – Grade 1 at 11 micrometres and 150 seconds, Grade 4 at 20 to 25 micrometres and 37 seconds, Grade 42 at 2.5 micrometres and 1870 seconds.

Ashless grades for gravimetry – Grade 40, 41 and 42 carry 0.007 percent ash, which is the figure that matters before ignition or before a weighing step.

Glass microfiber without fibre shed – GF/A, GF/B, GF/D and GF/F hold 0.7 to 2.7 micrometres and are used ahead of a fine membrane where a cellulose disc would blind.

Applications

The same parts appear in three places in an analyser: the sample path, the mobile phase path and the vent path.

HPLC and LC-MS sample prep – A 13 mm PVDF 0.22 micrometre device on a 1 to 10 mL sample removes particles ahead of the column, and the 15 microlitre hold-up keeps the first injection representative.

Media, buffer and reagent filtration – A 28 or 33 mm PES unit filters 100 to 200 mL at 150 mL/min or more, which is the step where a sterile 0.22 micrometre device pays for itself in time saved.

Venting and gas paths – A hydrophobic PTFE device passes air and stops liquid, so a reagent bottle breathes without drawing room air into the liquid.

Gravimetric and ash work – Grade 42 at 2.5 micrometres and 0.007 percent ash holds barium sulfate and calcium carbonate without adding weight to the crucible.

The same devices are held for SUPERB PCB and SUPERB PCBA builds of analyser boards, where the filter sits between the sample probe and the injection valve.

Sourcing and Cross-Reference

Filters are consumables with a short shelf life and a long qualification tail, so the sourcing question is usually continuity rather than price. Merck quotes a three year shelf life on the PVDF range, and a method that was validated on one lot expects the next lot to give the same bubble point and the same hold-up. When a pack size changes, the device does not: 16541-K and 16541-Q differ in packaging and sterility only, which is the difference that has to be documented before a switch.

Where a membrane has been discontinued or a pack size is out of stock, the cross reference is made on four fields together, membrane chemistry, pore size, effective area and hold-up volume. A 0.22 micrometre PES device with 4.5 cm2 and 0.1 mL hold-up is interchangeable with another 0.22 micrometre PES device of the same area; a 0.45 micrometre device is not, even when the housing looks the same. The syringe fitting is the last check, because a luer slip outlet will not take the same downstream connector as a luer lock.

Filter paper and glass microfiber are sourced by grade and by diameter rather than by brand alone, since the grade number is the specification. A request for Grade 1 is a request for 11 micrometres at 150 seconds and 0.06 percent ash, and a substitute at 8 micrometres is Grade 2 and a different method. For discs that are punched in house, the sheet format is available in 460 by 570 mm and the yield per sheet is calculated from the disc diameter before the order is placed.

Kits can be assembled and labelled for SUPERB Assembly lines, with the device, the syringe and the label in one bag.

Q: Can you supply the same membrane in a different diameter without requalifying the method?

A: SUPERB Tech confirms the membrane, the pore size and the housing before quoting, because the same diameter is sold with three different membranes.

Why Buy Sample Prep Consumables Through SUPERB

  1. One part number, one lot trail: Each shipment is booked against the maker code and the lot, so a method transfer can name the exact device that was used.
  2. Membrane and paper in one order: Syringe devices, membrane discs, qualitative paper and glass microfiber are consolidated, which removes a second purchase order and a second freight charge.
  3. Consumable plus hardware: The filter order can be combined with the pump tubing, the inline filter and the flow cell that sit on the same skid.
  4. Short lead time on the fast movers: 0.22 and 0.45 micrometre devices in 13, 25 and 28 mm, together with Grade 1 and GF/A discs, are the lines held for repeat orders.
  5. Cross reference on discontinued items: Where a device has been replaced, the nearest equivalent is proposed on membrane, pore, area and hold-up, with the differences stated before the order.
  6. Documentation for audits: Maker code, lot and certificate reference are kept with the shipment record, which is what an audit asks for on a consumable.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Inline Filters and Frits for Reagent Lines, Quartz Flow Cells and Cuvettes, Calibration Buffers and Conductivity Standards, Syringe Pumps and Precision Dispensers, Turbidity and Photometric Detector Modules, Sample Loops and Injection Loops, BOM Shortage Recovery.