Sample Loops and Injection Loops for Valve Injection
Product Specifications
Product Overview
A sample loop is a length of tubing with two fittings, and the volume is set by the bore and the length. The loop is filled by a syringe, either completely so that the loop volume determines the injection, or partially so that the syringe volume determines it. In complete filling the loop is overfilled and the excess goes to waste; in partial filling the sample is metered by the syringe and the loop is only a conduit. The two methods need different loop sizes for the same injection, which is the most common source of a wrong volume.
The numbers that matter are the loop volume, the bore and the tubing outside diameter. A 20 microlitre loop in a 0.30 mm bore is short and narrow, so the sample plug stays compact. A 20 microlitre loop in a 0.51 mm bore is shorter but wider, and the wider bore gives a slightly different peak shape at the same volume. The outside diameter is a mechanical constraint: a 1/16 inch loop fits an analytical valve, and a 1/8 inch loop is used on a preparative valve where the port is larger.
Pressure and temperature decide the material. A 316 stainless loop with a Vespel ferrule carries 48 MPa (about 7,000 psi) and 80 C, which covers an analytical system. A PEEK loop carries about 34 MPa (5,000 psi) and 50 C and gives a metal free path, which is used for proteins and for anything that would pick up iron. A preparative valve is rated lower, 28 MPa on one model with a 21 MPa shipping setting, because the port and the rotor are larger.
Loading volume is the last number. In complete filling the load volume is about 95 percent of the loop volume for a full mass injection, and in partial filling the injected volume is limited to half the loop volume for a reproducible result. That means a 100 microlitre loop is used for a 50 microlitre partial injection, and a 20 microlitre loop is used for a 20 microlitre complete injection. A loop chosen without that rule is either twice as large as it needs to be, which broadens the peak, or too small, which truncates the injection.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Volume range | 2 microlitre to 20 mL | 5 microlitre to 5 mL is the analytical range |
| Materials | 316 stainless steel with a Vespel ferrule, PEEK | PEEK for a metal free path |
| Bore range | 0.13, 0.18, 0.20, 0.25, 0.30, 0.51, 0.76, 1.0 and 2.0 mm | the bore and the length set the volume |
| Tubing outside diameter | 1/16 in analytical, 1/8 in preparative, 0.020 in micro | the OD has to match the valve port |
| Pressure, analytical stainless | 48 MPa, about 7,000 psi | 7725 and 9725 family |
| Pressure, PEEK | 34 MPa, about 5,000 psi | 9725 and 9010 family, 50 C |
| Pressure, preparative | 28 MPa, shipping setting 21 MPa | 3725 and 3725i |
| Temperature, stainless | 80 C | the ferrule sets the limit |
| Temperature, PEEK | 50 C | above this the PEEK creeps |
| Complete filling | loop volume determines the injection, load to 95 percent of the loop | the excess goes to waste |
| Partial filling | syringe determines the injection, limited to half the loop volume | the loop is a conduit |
| Valve families | 7125 and 7010 use the 7020 series, 7725 and 7725i use 7755, 9725 and 9010 use 9055, 3725 uses 3055 and 3065 | a loop is not interchangeable between families |
| Ferrule | Vespel on the stainless loops, PEEK on the PEEK loops | a stainless ferrule on a PEEK port damages the port |
| Marking | volume and part number on the tag | two loops of the same length look identical |
Included Models and Order Codes
The table is grouped by valve family, because that is the constraint that cannot be changed. Within a family the rows run from the smallest analytical volume to the largest preparative volume, with the bore and the tubing outside diameter, so a loop is ordered by volume and by fit in one step.
| Part | Volume | Bore | Tubing OD | Material | Valve family |
|---|---|---|---|---|---|
| 7755-020 | 5 microlitre | 0.18 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7755-021 | 10 microlitre | 0.30 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7755-022 | 20 microlitre | 0.30 mm | 1/16 in | 316 stainless steel | 7725 and 7725i, the standard loop |
| 7755-023 | 50 microlitre | 0.51 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7755-024 | 100 microlitre | 0.51 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7755-025 | 200 microlitre | 0.76 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7755-026 | 500 microlitre | 0.76 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7755-027 | 1 mL | 0.76 mm | 1/16 in | 316 stainless steel | 7725 and 7725i |
| 7020 | 5 microlitre | 0.18 mm | 1/16 in | 316 stainless steel | 7125 and 7010, not for 7725 |
| 7022 | 20 microlitre | 0.51 mm | 1/16 in | 316 stainless steel | 7125 and 7010 |
| 7024 | 100 microlitre | 0.51 mm | 1/16 in | 316 stainless steel | 7125 and 7010 |
| 7026 | 500 microlitre | 0.76 mm | 1/16 in | 316 stainless steel | 7125 and 7010 |
| 7027 | 1 mL | 0.76 mm | 1/16 in | 316 stainless steel | 7125 and 7010 |
| 7028 | 2 mL | 1.0 mm | 1/16 in | 316 stainless steel | 7125 and 7010 |
| 7029 | 5 mL | 1.0 mm | 1/16 in | 316 stainless steel | 7125 and 7010 |
| 9055-022 | 20 microlitre | 0.25 mm | 1/16 in | PEEK | 9725 and 9010 |
| 9055-024 | 100 microlitre | 0.51 mm | 1/16 in | PEEK | 9725 and 9010 |
| 9055-026 | 500 microlitre | 0.76 mm | 1/16 in | PEEK | 9725 and 9010 |
| 3055-018 | 2 mL | 1.6 mm | 1/8 in | PEEK | 3725 and 3725i |
| 3055-023 | 10 mL | 2.0 mm | 1/8 in | PEEK | 3725 and 3725i |
| 3055-025 | 20 mL | 2.0 mm | 1/8 in | PEEK | 3725 and 3725i |
| 3065-018 | 2 mL | 2.0 mm | 1/8 in | 316 stainless steel | 3725-038 and 3725i-038 |
| 3065-023 | 10 mL | 2.0 mm | 1/8 in | 316 stainless steel | 3725-038 and 3725i-038 |
| 3065-025 | 20 mL | 2.0 mm | 1/8 in | 316 stainless steel | 3725-038 and 3725i-038 |
Loop Volume and Injection Volume
The loop volume is not the injection volume in every method. This table gives the usable injection for a loop under the two loading methods, which is the calculation that decides whether a loop is twice the size it needs to be.
| Loop volume | Complete filling | Partial filling | Best for | Note |
|---|---|---|---|---|
| 5 microlitre | 5 microlitre, load to 4.8 | 2.5 microlitre maximum | a narrow bore column | the smallest standard loop |
| 10 microlitre | 10 microlitre, load to 9.5 | 5 microlitre maximum | a 2.1 mm column | the common small injection |
| 20 microlitre | 20 microlitre, load to 19 | 10 microlitre maximum | a 4.6 mm column | the standard analytical loop |
| 100 microlitre | 100 microlitre, load to 95 | 50 microlitre maximum | a method that needs a larger injection | the loop is twice the partial volume |
| 500 microlitre | 500 microlitre, load to 475 | 250 microlitre maximum | trace analysis | the plug is long, check the peak shape |
| 2 mL | 2 mL, load to 1.9 | 1 mL maximum | preparative | 1/8 in tubing |
| 10 mL | 10 mL, load to 9.5 | 5 mL maximum | preparative | 2.0 mm bore |
| 20 mL | 20 mL, load to 19 | 10 mL maximum | process scale | the largest standard loop |
Key Features
5 microlitre to 20 mL – one list covers a narrow bore analytical injection and a preparative load
Bore given with the volume – the bore and the length set the volume, and the bore also sets the peak shape
Valve family on every row – a 7755 loop does not fit a 7125 injector, so the family is named rather than implied
Stainless and PEEK – 48 MPa and 80 C in stainless, 34 MPa and 50 C in PEEK, with a metal free path
Complete and partial filling explained – the same loop gives a different injection depending on the loading method
1/16 in and 1/8 in tubing – analytical and preparative ports are covered without an adapter
Vespel and PEEK ferrules – the ferrule is matched to the port material, which is what avoids a damaged port
Volume and part number on the tag – two loops of the same length are otherwise indistinguishable
Applications
A loop is replaced to change the injection volume, to move to a different sample, or because the old loop is worn at the ferrule. The three cases below set the volume and the material.
Analytical HPLC – a 4.6 mm column at 1 mL/min is injected with 20 microlitres. A 7755-022 loop in a 0.30 mm bore keeps the plug compact, and the 48 MPa rating covers a system that runs at 20 MPa. A 100 microlitre loop would be used only if the method needs a larger injection, and then the peak shape is checked.
Protein and ion chromatography – a PEEK 9055 loop with a PEEK ferrule keeps iron out of the flow path, which matters for a protein that binds metal. The 34 MPa rating is enough for a low pressure column, and the 50 C limit is checked before a method that runs warm.
Preparative and process scale – a 3725 injector loads 10 or 20 millilitres through a 2.0 mm bore loop in 1/8 inch tubing. The pressure rating is 28 MPa with a 21 MPa shipping setting, and the loop is changed as a complete assembly rather than re ferruled on the skid.
The valve and pump control boards in these systems are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the injection timing to the detector interface.
Sourcing and Cross-Reference
A loop is ordered by volume, bore and valve family, and the family is the field that is missed. The 7020 series and the 7755 series look similar and do not fit the same injector, and the 7725 port is a 30 degree seat while the 7125 is not. We confirm the valve model and the existing loop number before naming a part.
Where a volume is no longer supplied, the substitute is chosen on the same volume and the same bore. A 20 microlitre loop in a 0.51 mm bore and a 20 microlitre loop in a 0.30 mm bore hold the same volume and give a different plug length, which shows up as a small change in peak width. Where the method is sensitive, the bore is matched as well as the volume.
Ferrules and fittings are part of the order. A stainless ferrule on a PEEK port damages the port, and a PEEK ferrule on a stainless port may not hold pressure. We ship the loop with the ferrule that matches the port and note the torque, because a loop that leaks at the ferrule is usually over tightened rather than under tightened.
Where a valve is assembled with its loop and fittings, the set is picked together, which is how SUPERB Assembly ships an injection module.
Q: Can you supply the same loop volume for a valve whose original loop is no longer listed?
A: SUPERB Tech can confirm the volume, the bore and the valve family against the old loop before the order is placed.
Why Buy Sample Loops from SUPERB
- Valve family named: the loop is matched to the injector, not only to the volume, because the seat angle differs
- Bore and volume together: the same volume in two bores gives two plug lengths, so both are quoted
- Complete and partial filling shown: the usable injection is given for each loading method
- Stainless and PEEK on one list: 48 MPa and 80 C, or 34 MPa and 50 C with a metal free path
- Ferrule matched to the port: the fitting is supplied with the loop and the torque is stated
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Rotary and Injection Valves for Analyser Sample Paths, Syringe Pumps and Precision Dispensers for Analysers, Inline Static Mixers for Reagent Streams, Inline Filters and Frits for Reagent Lines, Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Component Lead Time Solutions.