Sample Loops and Injection Loops for Valve Injection

A sample loop is the volume that reaches the column, and it is the cheapest way to change an injection volume without changing the valve. The loop also sets the delay and the peak shape, because a loop that is too large for the injection is a mixing chamber. This page covers loops from 5 microlitre to 20 millilitre in 316 stainless steel and PEEK, with the bore, the tubing outside diameter and the valve family for each part. Complete filling and partial filling are both covered, because the same loop gives a different injection volume depending on how it is loaded. The valve family is named on every row, because a loop that fits a 7725 injector does not fit a 7125 injector even when the volume is the same.
quote now

Product Specifications

Product Overview

A sample loop is a length of tubing with two fittings, and the volume is set by the bore and the length. The loop is filled by a syringe, either completely so that the loop volume determines the injection, or partially so that the syringe volume determines it. In complete filling the loop is overfilled and the excess goes to waste; in partial filling the sample is metered by the syringe and the loop is only a conduit. The two methods need different loop sizes for the same injection, which is the most common source of a wrong volume.

The numbers that matter are the loop volume, the bore and the tubing outside diameter. A 20 microlitre loop in a 0.30 mm bore is short and narrow, so the sample plug stays compact. A 20 microlitre loop in a 0.51 mm bore is shorter but wider, and the wider bore gives a slightly different peak shape at the same volume. The outside diameter is a mechanical constraint: a 1/16 inch loop fits an analytical valve, and a 1/8 inch loop is used on a preparative valve where the port is larger.

Pressure and temperature decide the material. A 316 stainless loop with a Vespel ferrule carries 48 MPa (about 7,000 psi) and 80 C, which covers an analytical system. A PEEK loop carries about 34 MPa (5,000 psi) and 50 C and gives a metal free path, which is used for proteins and for anything that would pick up iron. A preparative valve is rated lower, 28 MPa on one model with a 21 MPa shipping setting, because the port and the rotor are larger.

Loading volume is the last number. In complete filling the load volume is about 95 percent of the loop volume for a full mass injection, and in partial filling the injected volume is limited to half the loop volume for a reproducible result. That means a 100 microlitre loop is used for a 50 microlitre partial injection, and a 20 microlitre loop is used for a 20 microlitre complete injection. A loop chosen without that rule is either twice as large as it needs to be, which broadens the peak, or too small, which truncates the injection.

Key Specifications

ParameterValueNote
Volume range2 microlitre to 20 mL5 microlitre to 5 mL is the analytical range
Materials316 stainless steel with a Vespel ferrule, PEEKPEEK for a metal free path
Bore range0.13, 0.18, 0.20, 0.25, 0.30, 0.51, 0.76, 1.0 and 2.0 mmthe bore and the length set the volume
Tubing outside diameter1/16 in analytical, 1/8 in preparative, 0.020 in microthe OD has to match the valve port
Pressure, analytical stainless48 MPa, about 7,000 psi7725 and 9725 family
Pressure, PEEK34 MPa, about 5,000 psi9725 and 9010 family, 50 C
Pressure, preparative28 MPa, shipping setting 21 MPa3725 and 3725i
Temperature, stainless80 Cthe ferrule sets the limit
Temperature, PEEK50 Cabove this the PEEK creeps
Complete fillingloop volume determines the injection, load to 95 percent of the loopthe excess goes to waste
Partial fillingsyringe determines the injection, limited to half the loop volumethe loop is a conduit
Valve families7125 and 7010 use the 7020 series, 7725 and 7725i use 7755, 9725 and 9010 use 9055, 3725 uses 3055 and 3065a loop is not interchangeable between families
FerruleVespel on the stainless loops, PEEK on the PEEK loopsa stainless ferrule on a PEEK port damages the port
Markingvolume and part number on the tagtwo loops of the same length look identical

Included Models and Order Codes

The table is grouped by valve family, because that is the constraint that cannot be changed. Within a family the rows run from the smallest analytical volume to the largest preparative volume, with the bore and the tubing outside diameter, so a loop is ordered by volume and by fit in one step.

PartVolumeBoreTubing ODMaterialValve family
7755-0205 microlitre0.18 mm1/16 in316 stainless steel7725 and 7725i
7755-02110 microlitre0.30 mm1/16 in316 stainless steel7725 and 7725i
7755-02220 microlitre0.30 mm1/16 in316 stainless steel7725 and 7725i, the standard loop
7755-02350 microlitre0.51 mm1/16 in316 stainless steel7725 and 7725i
7755-024100 microlitre0.51 mm1/16 in316 stainless steel7725 and 7725i
7755-025200 microlitre0.76 mm1/16 in316 stainless steel7725 and 7725i
7755-026500 microlitre0.76 mm1/16 in316 stainless steel7725 and 7725i
7755-0271 mL0.76 mm1/16 in316 stainless steel7725 and 7725i
70205 microlitre0.18 mm1/16 in316 stainless steel7125 and 7010, not for 7725
702220 microlitre0.51 mm1/16 in316 stainless steel7125 and 7010
7024100 microlitre0.51 mm1/16 in316 stainless steel7125 and 7010
7026500 microlitre0.76 mm1/16 in316 stainless steel7125 and 7010
70271 mL0.76 mm1/16 in316 stainless steel7125 and 7010
70282 mL1.0 mm1/16 in316 stainless steel7125 and 7010
70295 mL1.0 mm1/16 in316 stainless steel7125 and 7010
9055-02220 microlitre0.25 mm1/16 inPEEK9725 and 9010
9055-024100 microlitre0.51 mm1/16 inPEEK9725 and 9010
9055-026500 microlitre0.76 mm1/16 inPEEK9725 and 9010
3055-0182 mL1.6 mm1/8 inPEEK3725 and 3725i
3055-02310 mL2.0 mm1/8 inPEEK3725 and 3725i
3055-02520 mL2.0 mm1/8 inPEEK3725 and 3725i
3065-0182 mL2.0 mm1/8 in316 stainless steel3725-038 and 3725i-038
3065-02310 mL2.0 mm1/8 in316 stainless steel3725-038 and 3725i-038
3065-02520 mL2.0 mm1/8 in316 stainless steel3725-038 and 3725i-038

Loop Volume and Injection Volume

The loop volume is not the injection volume in every method. This table gives the usable injection for a loop under the two loading methods, which is the calculation that decides whether a loop is twice the size it needs to be.

Loop volumeComplete fillingPartial fillingBest forNote
5 microlitre5 microlitre, load to 4.82.5 microlitre maximuma narrow bore columnthe smallest standard loop
10 microlitre10 microlitre, load to 9.55 microlitre maximuma 2.1 mm columnthe common small injection
20 microlitre20 microlitre, load to 1910 microlitre maximuma 4.6 mm columnthe standard analytical loop
100 microlitre100 microlitre, load to 9550 microlitre maximuma method that needs a larger injectionthe loop is twice the partial volume
500 microlitre500 microlitre, load to 475250 microlitre maximumtrace analysisthe plug is long, check the peak shape
2 mL2 mL, load to 1.91 mL maximumpreparative1/8 in tubing
10 mL10 mL, load to 9.55 mL maximumpreparative2.0 mm bore
20 mL20 mL, load to 1910 mL maximumprocess scalethe largest standard loop

Key Features

5 microlitre to 20 mL – one list covers a narrow bore analytical injection and a preparative load

Bore given with the volume – the bore and the length set the volume, and the bore also sets the peak shape

Valve family on every row – a 7755 loop does not fit a 7125 injector, so the family is named rather than implied

Stainless and PEEK – 48 MPa and 80 C in stainless, 34 MPa and 50 C in PEEK, with a metal free path

Complete and partial filling explained – the same loop gives a different injection depending on the loading method

1/16 in and 1/8 in tubing – analytical and preparative ports are covered without an adapter

Vespel and PEEK ferrules – the ferrule is matched to the port material, which is what avoids a damaged port

Volume and part number on the tag – two loops of the same length are otherwise indistinguishable

Applications

A loop is replaced to change the injection volume, to move to a different sample, or because the old loop is worn at the ferrule. The three cases below set the volume and the material.

Analytical HPLC – a 4.6 mm column at 1 mL/min is injected with 20 microlitres. A 7755-022 loop in a 0.30 mm bore keeps the plug compact, and the 48 MPa rating covers a system that runs at 20 MPa. A 100 microlitre loop would be used only if the method needs a larger injection, and then the peak shape is checked.

Protein and ion chromatography – a PEEK 9055 loop with a PEEK ferrule keeps iron out of the flow path, which matters for a protein that binds metal. The 34 MPa rating is enough for a low pressure column, and the 50 C limit is checked before a method that runs warm.

Preparative and process scale – a 3725 injector loads 10 or 20 millilitres through a 2.0 mm bore loop in 1/8 inch tubing. The pressure rating is 28 MPa with a 21 MPa shipping setting, and the loop is changed as a complete assembly rather than re ferruled on the skid.

The valve and pump control boards in these systems are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the injection timing to the detector interface.

Sourcing and Cross-Reference

A loop is ordered by volume, bore and valve family, and the family is the field that is missed. The 7020 series and the 7755 series look similar and do not fit the same injector, and the 7725 port is a 30 degree seat while the 7125 is not. We confirm the valve model and the existing loop number before naming a part.

Where a volume is no longer supplied, the substitute is chosen on the same volume and the same bore. A 20 microlitre loop in a 0.51 mm bore and a 20 microlitre loop in a 0.30 mm bore hold the same volume and give a different plug length, which shows up as a small change in peak width. Where the method is sensitive, the bore is matched as well as the volume.

Ferrules and fittings are part of the order. A stainless ferrule on a PEEK port damages the port, and a PEEK ferrule on a stainless port may not hold pressure. We ship the loop with the ferrule that matches the port and note the torque, because a loop that leaks at the ferrule is usually over tightened rather than under tightened.

Where a valve is assembled with its loop and fittings, the set is picked together, which is how SUPERB Assembly ships an injection module.

Q: Can you supply the same loop volume for a valve whose original loop is no longer listed?

A: SUPERB Tech can confirm the volume, the bore and the valve family against the old loop before the order is placed.

Why Buy Sample Loops from SUPERB

  1. Valve family named: the loop is matched to the injector, not only to the volume, because the seat angle differs
  2. Bore and volume together: the same volume in two bores gives two plug lengths, so both are quoted
  3. Complete and partial filling shown: the usable injection is given for each loading method
  4. Stainless and PEEK on one list: 48 MPa and 80 C, or 34 MPa and 50 C with a metal free path
  5. Ferrule matched to the port: the fitting is supplied with the loop and the torque is stated

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Rotary and Injection Valves for Analyser Sample Paths, Syringe Pumps and Precision Dispensers for Analysers, Inline Static Mixers for Reagent Streams, Inline Filters and Frits for Reagent Lines, Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Component Lead Time Solutions.