Inline Filters and Frits for Reagent Lines
Product Specifications
Product Overview
A reagent line has three places where a filter earns its place, and they are not interchangeable. The first is the reservoir inlet, where a cup on the end of the supply tube keeps fibres, septum fragments and dust out of the pump head. The second is between the pump and the injection valve, where the filter catches particles shed by the pump seal and the valve rotor before they reach the column. The third is between the valve and the column, where a precolumn body catches precipitate from the sample itself. An inlet cup is coarse, 10 or 20 micron, because it has to pass the whole flow without cavitating the pump. An inline or precolumn body is fine, 0.5 or 2 micron, because the flow through it is small and the particle load is low.
Pore size sets the pressure drop as well as the cut point. A frit is not a sieve of straight holes; it is a sintered or moulded disc, and the fraction of its volume that is open to flow falls as the pores get smaller. Published figures for one family run 18 percent open at 0.2 micron, 20 percent at 0.5, 24 percent at 2, 26 percent at 5, 28 percent at 10 and 30 percent at 20 micron. At the same flow, a 0.5 micron frit therefore costs noticeably more pressure than a 2 micron frit of the same diameter, and a line that already runs near its pump limit will show that difference as a pressure alarm rather than as a smoother baseline.
Swept volume is the second number that decides whether a filter can be used at all. A precolumn body with a 1.3 to 1.4 microlitre swept volume sits comfortably ahead of a 4.6 mm column, but the same body ahead of a 1 mm column is a mixing chamber. The semi-prep bodies in this family run 89 to 103 microlitre, which is fine on a 10 mm column and wrong on anything analytical. At the other end, the encapsulated micro bodies reach 10 to 85 nL, and those are the ones that go into a capillary line where every extra microlitre shows as a broader peak. The rule is simple: put the filter where the volume is not sensitive, and if the volume is sensitive, use the smallest body that still holds the frit.
Materials and pressure rating close the choice. A 316 stainless body with a stainless frit covers most reversed phase work and carries 6,000 psi. A PEEK body with a PEEK frit gives a fully biocompatible wetted path at 5,000 psi, which matters for proteins and for anything that would pick up iron. Titanium frits are the third option where stainless is not acceptable. Replacement policy differs by family as well: the precolumn bodies take a new frit in minutes, while the biocompatible inline bodies are exchanged as a complete end fitting, which is faster on a production line but costs more per change. Both policies are listed on the model table, because a line that changes frits weekly and a line that changes them yearly do not want the same body.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Pore size range | 0.2 to 20 micron, with 0.5 and 2 micron the common analytical ratings | inlet cups are 10 or 20 micron, precolumn and inline bodies 0.5 or 2 micron |
| Frit open fraction | 18 percent at 0.2 micron, 20 at 0.5, 24 at 2, 26 at 5, 28 at 10, 30 at 20 micron | the open fraction is what turns pore size into back pressure |
| Swept volume | 10 nL to 103 microlitre across the families | precolumn 1.3 to 1.4 microlitre, micro bodies 10 to 85 nL |
| Pressure rating | 4,000 to 6,000 psi, which is 276 to 414 bar | the PEEK precolumn bodies are rated 5,000 psi |
| Body materials | 316 stainless steel, PEEK, with titanium frit options | PEEK bodies give a fully biocompatible wetted path |
| Frit materials | 316 stainless steel, PEEK, titanium, UHMWPE cup on the inlet filters | PEEK frits hold a more uniform porosity from batch to batch |
| Tubing sizes | 1/16 in OD, 1/8 in OD, 1/32 in OD, 360 micron OD on the micro bodies | the micro bodies need the matching sleeve or ferrule set |
| Port threads | 10-32 coned, 1/4-28 flat bottom, M6, 5/16-24 | a flat bottom port needs a ferrule with a flat seat |
| Maximum flow, inlet cups | 10 mL/min on the 2 micron cup, 40 mL/min on the 10 micron cups | the finer cup limits flow before it limits particles |
| Maximum flow, inline bodies | 10 mL/min on the narrow body, 25 mL/min on the wider biocompatible body | semi-prep bodies take the higher flow of a 10 mm column |
| Thru hole on micro filters | 0.006 in, which is 150 micron | the hole is the bore, the frit is what filters |
| Inlet cup reach | draws to within 2 mm of the bottle bottom, 1/8 in on the bottom of bottle body | leaves less dead volume in the reservoir |
| Frit change policy | precolumn bodies take a replacement frit; biocompatible inline bodies take a new end fitting | cost per change differs by an order of magnitude |
| Biocompatibility | PEEK and titanium wetted parts, no stainless in the flow path | required where a metal ion would change the assay |
Included Models and Order Codes
The table is grouped by position on the line rather than by part number, because that is how a line is specified: pick the inlet cup first, then the body that sits ahead of the column, then the frit that goes in it. Voltages and flows are the maker figures for that exact part, not a family average. Where a body is sold with a frit included, the included frit is named, so a replacement can be ordered without guessing.
| Part | Type and position | Porosity | Wetted path | Swept volume | Maximum flow | Note |
|---|---|---|---|---|---|---|
| A-240 | Inlet solvent filter, reservoir | 10 micron | 316 stainless steel | not applicable | 40 mL/min | coarse cup for the main supply line |
| A-242 | Inlet solvent filter, reservoir | 2 micron | 316 stainless steel | not applicable | 10 mL/min | finer cup, lower flow, for a small bore line |
| A-302 | Inlet solvent filter with stem | 10 micron | 316 stainless steel | not applicable | 40 mL/min | stem fits 1/16 in ID tubing |
| A-302A | Inlet solvent filter with flangeless fittings | 10 micron | PCTFE and 316 stainless steel | not applicable | 40 mL/min | for 1/8 in OD tubing, ships with one XP-315 |
| A-309 | Inlet solvent filter, bottom of the bottle | 10 micron | 316 stainless steel | not applicable | 40 mL/min | side port draws within 1/8 in of the bottom |
| A-315 | Precolumn filter assembly | 2 micron | 316 stainless steel | 1.4 microlitre | not stated for the body | 6,000 psi, ships with one A-101 frit, 10-32 coned |
| A-355 | Precolumn filter, biocompatible | 2 micron | PEEK | 1.4 microlitre | not stated for the body | 5,000 psi, ships with one A-700 frit |
| A-356 | Precolumn filter, biocompatible | 0.5 micron | PEEK | 1.3 microlitre | not stated for the body | 5,000 psi, ships with one A-701 frit |
| A-410 | Semi-prep inline filter, biocompatible | 2 micron | PEEK wetted, stainless body | 89 microlitre | not stated for the body | 6,000 psi, ships with one OC-802 frit |
| A-411 | Semi-prep inline filter, biocompatible | 10 micron | PEEK wetted, stainless body | 103 microlitre | not stated for the body | 6,000 psi, ships with one OC-803 frit |
| A-430 | Inline filter, biocompatible | 0.5 or 2 micron | PEEK end fittings, stainless body | not stated for the body | 25 mL/min | frit is replaced with the end fitting |
| A-431 | Inline filter, biocompatible, narrow bore | 0.5 or 2 micron | PEEK end fittings, stainless body | not stated for the body | 10 mL/min | for a low flow line with a small bore tube |
| M-520 | Inline micro filter | 0.5 micron PEEK frit | PEEK | not stated for the body | not stated for the body | 150 micron thru hole, connects with MicroTight sleeves |
| M-525 | Inline micro filter, direct mount | 0.5 micron PEEK frit | PEEK | not stated for the body | not stated for the body | takes 1/32 in OD tubing without a sleeve |
| M-530 | Mini micro filter, encapsulated | 2 micron stainless screen | PEEK capsule | 85 nL | not stated for the body | 4,000 psi, capsule is replaced as a unit |
| M-537 | Mini micro filter, nano | 1 micron stainless frit | PEEK capsule | 10 nL | not stated for the body | 4,000 psi, for 360 micron OD tubing |
| A-101 | Replacement frit | 2 micron | 316 stainless steel | 0.74 microlitre | not applicable | fits the A-315 body and the standard inline bodies |
| A-102 | Replacement frit | 0.5 micron | 316 stainless steel | 0.61 microlitre | not applicable | same seat as A-101, finer cut |
| A-700 | Replacement frit, PEEK | 2 micron | PEEK | 0.74 microlitre | not applicable | biocompatible replacement for A-101 |
| A-701 | Replacement frit, PEEK | 0.5 micron | PEEK | 0.61 microlitre | not applicable | lowest swept volume in the precolumn family |
| C-425x | Small diameter frit, 10 pack | 0.5 micron | 316 stainless steel | 0.1 microlitre | not applicable | 0.192 in ring, for a tight volume budget |
| C-414 | Semi-prep frit | 2 micron | 316 stainless steel | 26.9 microlitre | not applicable | 0.375 in disc for the larger bodies |
| C-415 | Semi-prep frit | 2 micron | 316 stainless steel | 47.9 microlitre | not applicable | 0.500 in disc, highest flow of the three |
Where the Filter Sits and What It Protects
The same pore size in two positions does two different jobs. This table is the short form used when a line is mapped out, and it is also the one to check when a back pressure problem is blamed on the column.
| Position | Pore size | Volume that matters | Part that is protected | Failure it prevents |
|---|---|---|---|---|
| Reservoir inlet | 10 to 20 micron | none, this is the supply side | pump head and check valves | cavitation and seal wear from fibres and dust |
| Pump outlet, before the valve | 0.5 to 2 micron | 1.3 to 1.4 microlitre | valve rotor seal | seal particles reaching the column |
| Between valve and column | 0.5 to 2 micron | 1.3 to 1.4 microlitre | column bed and frit | sample precipitate blocking the column inlet |
| Capillary or nano line | 0.5 micron | 10 to 85 nL | capillary and union | plugging at the first union after the valve |
| Semi-prep or prep line | 2 to 10 micron | 89 to 103 microlitre | prep column | flow restriction at high sample load |
| Sample preparation, off line | 0.2 to 2 micron | not applicable | autosampler needle and loop | fibres drawn from a vial septum |
Key Features
Pore size from 0.2 to 20 micron – one family covers the inlet cup, the inline body and the precolumn body, so the pore size is chosen against the particle load rather than against the catalogue section
Open fraction stated per pore size – 18 percent at 0.2 micron rising to 30 percent at 20 micron, which is what makes a pressure estimate possible before a part is ordered
Swept volume from 10 nL – the encapsulated micro bodies are small enough for a capillary line, and the precolumn bodies stay at 1.3 to 1.4 microlitre
PEEK wetted path on the biocompatible range – no stainless steel in the flow path, which keeps proteins and metal sensitive assays out of trouble
Replaceable frit in the precolumn bodies – a new frit goes in with the body still on the line, so the cost per change is the frit rather than the assembly
Pressure rating to 6,000 psi – 414 bar covers a UHPLC line, and the PEEK precolumn bodies are rated at 5,000 psi for the same duty
150 micron thru hole on the micro filters – the bore stays open while the frit does the filtering, which keeps the dead volume at the union low
Inlet cups that reach the bottle bottom – the bottom of bottle body draws within 1/8 in of the base, which reduces the volume left behind when a batch is changed
Applications
These parts sit on the wet side of an instrument, in the path between the reagent bottle and the detector. The three cases below are the ones that set the pore size and the body volume.
HPLC and UHPLC mobile phase – a 10 micron inlet cup in the reservoir and a 0.5 micron precolumn body ahead of the column. On a sub 2 micron column the precolumn body is the part that saves the column, and its 1.3 microlitre volume is small enough not to spoil the peak shape.
Clinical chemistry reagent lines – a reagent pack sits for days and grows particles at the pierce point. A 2 micron inline body between the pump and the valve keeps those particles off the rotor seal, and a PEEK wetted path keeps metal ions out of the reaction.
Process analyser sample lines – sample arrives from a sample probe with fibres and scale. A 10 micron cup at the sample inlet and a 2 micron body ahead of the injection valve give two stages, so the coarse filter can be cleaned while the fine one stays in place.
The instrument boards behind these lines are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the pump drive to the detector front end.
Sourcing and Cross-Reference
Filters are ordered by pore size, body volume and connection, and the connection is the part that goes wrong on a replacement. A 10-32 coned port and a 1/4-28 flat bottom port look similar in a drawing and do not seal into each other, so the port on the existing body is confirmed before a part is named.
Where a line is being copied from an older instrument, the frit seat is the constraint rather than the body. A 0.250 in seat takes the standard frit and a 0.192 in seat does not, and a body that accepts the wrong disc will leak at the seal rather than at the thread. We check the seat diameter and the ring material together.
For a discontinued body, the replacement is chosen on swept volume first and pore size second, because a body with the right pore size and twice the volume changes the chromatography. Where no equivalent exists at the same volume, we say so before shipping a part that fits the thread but shifts the peak.
When a line is down, the body, the frit and the fittings are picked as one set, which is how SUPERB Assembly ships a filter change.
Q: Can you supply a precolumn filter with the same swept volume when the original body is discontinued?
A: SUPERB Tech can confirm the port thread, the frit seat and the swept volume against the old body before the order is placed.
Why Buy Filters and Frits from SUPERB
- Position asked before the part: the reservoir cup, the inline body and the precolumn body are specified separately, because the same pore size does different work in each place
- Volume and pore size quoted together: a body is never offered on pore size alone, since swept volume is what changes the chromatogram
- Connection and frit seat confirmed: the port thread and the disc seat are checked against the existing body before an order is released
- PEEK and titanium options on the same list: a biocompatible path is a line item rather than a special request
- Maker part number on the paperwork: the frit and the body are both named in the delivery file, which is what a method transfer needs
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Peristaltic and Diaphragm Micro Pumps for Analysers, Inline Degassing Modules and Vacuum Degassers for Liquid Lines, Fluid Check Valves and Shuttle Valves for Reagent and Pneumatic Lines, Syringe Pumps and Precision Dispensers for Analysers, Component Lead Time Solutions.