A reagent line fails quietly. The pump seal wears, the injection valve rotor sheds a few particles, a buffer precipitates at the first cold spot, and the damage shows up as a rising back pressure or a split peak rather than as a leak. The filters below cover the three places that matter: an inlet cup in the reservoir, an inline body between the pump and the valve, and a precolumn body just ahead of the column. Pore size runs from 0.2 to 20 micron, swept volume from 10 nL to 103 microlitre, and the wetted path is 316 stainless steel, PEEK or titanium, so the same family covers an HPLC line and a clinical reagent path. Each line gives the pore size, the swept volume and the pressure rating together, because a filter chosen on pore size alone can add more dispersion than the column can absorb.
quote now

Product Specifications

Product Overview

A reagent line has three places where a filter earns its place, and they are not interchangeable. The first is the reservoir inlet, where a cup on the end of the supply tube keeps fibres, septum fragments and dust out of the pump head. The second is between the pump and the injection valve, where the filter catches particles shed by the pump seal and the valve rotor before they reach the column. The third is between the valve and the column, where a precolumn body catches precipitate from the sample itself. An inlet cup is coarse, 10 or 20 micron, because it has to pass the whole flow without cavitating the pump. An inline or precolumn body is fine, 0.5 or 2 micron, because the flow through it is small and the particle load is low.

Pore size sets the pressure drop as well as the cut point. A frit is not a sieve of straight holes; it is a sintered or moulded disc, and the fraction of its volume that is open to flow falls as the pores get smaller. Published figures for one family run 18 percent open at 0.2 micron, 20 percent at 0.5, 24 percent at 2, 26 percent at 5, 28 percent at 10 and 30 percent at 20 micron. At the same flow, a 0.5 micron frit therefore costs noticeably more pressure than a 2 micron frit of the same diameter, and a line that already runs near its pump limit will show that difference as a pressure alarm rather than as a smoother baseline.

Swept volume is the second number that decides whether a filter can be used at all. A precolumn body with a 1.3 to 1.4 microlitre swept volume sits comfortably ahead of a 4.6 mm column, but the same body ahead of a 1 mm column is a mixing chamber. The semi-prep bodies in this family run 89 to 103 microlitre, which is fine on a 10 mm column and wrong on anything analytical. At the other end, the encapsulated micro bodies reach 10 to 85 nL, and those are the ones that go into a capillary line where every extra microlitre shows as a broader peak. The rule is simple: put the filter where the volume is not sensitive, and if the volume is sensitive, use the smallest body that still holds the frit.

Materials and pressure rating close the choice. A 316 stainless body with a stainless frit covers most reversed phase work and carries 6,000 psi. A PEEK body with a PEEK frit gives a fully biocompatible wetted path at 5,000 psi, which matters for proteins and for anything that would pick up iron. Titanium frits are the third option where stainless is not acceptable. Replacement policy differs by family as well: the precolumn bodies take a new frit in minutes, while the biocompatible inline bodies are exchanged as a complete end fitting, which is faster on a production line but costs more per change. Both policies are listed on the model table, because a line that changes frits weekly and a line that changes them yearly do not want the same body.

Key Specifications

ParameterValueNote
Pore size range0.2 to 20 micron, with 0.5 and 2 micron the common analytical ratingsinlet cups are 10 or 20 micron, precolumn and inline bodies 0.5 or 2 micron
Frit open fraction18 percent at 0.2 micron, 20 at 0.5, 24 at 2, 26 at 5, 28 at 10, 30 at 20 micronthe open fraction is what turns pore size into back pressure
Swept volume10 nL to 103 microlitre across the familiesprecolumn 1.3 to 1.4 microlitre, micro bodies 10 to 85 nL
Pressure rating4,000 to 6,000 psi, which is 276 to 414 barthe PEEK precolumn bodies are rated 5,000 psi
Body materials316 stainless steel, PEEK, with titanium frit optionsPEEK bodies give a fully biocompatible wetted path
Frit materials316 stainless steel, PEEK, titanium, UHMWPE cup on the inlet filtersPEEK frits hold a more uniform porosity from batch to batch
Tubing sizes1/16 in OD, 1/8 in OD, 1/32 in OD, 360 micron OD on the micro bodiesthe micro bodies need the matching sleeve or ferrule set
Port threads10-32 coned, 1/4-28 flat bottom, M6, 5/16-24a flat bottom port needs a ferrule with a flat seat
Maximum flow, inlet cups10 mL/min on the 2 micron cup, 40 mL/min on the 10 micron cupsthe finer cup limits flow before it limits particles
Maximum flow, inline bodies10 mL/min on the narrow body, 25 mL/min on the wider biocompatible bodysemi-prep bodies take the higher flow of a 10 mm column
Thru hole on micro filters0.006 in, which is 150 micronthe hole is the bore, the frit is what filters
Inlet cup reachdraws to within 2 mm of the bottle bottom, 1/8 in on the bottom of bottle bodyleaves less dead volume in the reservoir
Frit change policyprecolumn bodies take a replacement frit; biocompatible inline bodies take a new end fittingcost per change differs by an order of magnitude
BiocompatibilityPEEK and titanium wetted parts, no stainless in the flow pathrequired where a metal ion would change the assay

Included Models and Order Codes

The table is grouped by position on the line rather than by part number, because that is how a line is specified: pick the inlet cup first, then the body that sits ahead of the column, then the frit that goes in it. Voltages and flows are the maker figures for that exact part, not a family average. Where a body is sold with a frit included, the included frit is named, so a replacement can be ordered without guessing.

PartType and positionPorosityWetted pathSwept volumeMaximum flowNote
A-240Inlet solvent filter, reservoir10 micron316 stainless steelnot applicable40 mL/mincoarse cup for the main supply line
A-242Inlet solvent filter, reservoir2 micron316 stainless steelnot applicable10 mL/minfiner cup, lower flow, for a small bore line
A-302Inlet solvent filter with stem10 micron316 stainless steelnot applicable40 mL/minstem fits 1/16 in ID tubing
A-302AInlet solvent filter with flangeless fittings10 micronPCTFE and 316 stainless steelnot applicable40 mL/minfor 1/8 in OD tubing, ships with one XP-315
A-309Inlet solvent filter, bottom of the bottle10 micron316 stainless steelnot applicable40 mL/minside port draws within 1/8 in of the bottom
A-315Precolumn filter assembly2 micron316 stainless steel1.4 microlitrenot stated for the body6,000 psi, ships with one A-101 frit, 10-32 coned
A-355Precolumn filter, biocompatible2 micronPEEK1.4 microlitrenot stated for the body5,000 psi, ships with one A-700 frit
A-356Precolumn filter, biocompatible0.5 micronPEEK1.3 microlitrenot stated for the body5,000 psi, ships with one A-701 frit
A-410Semi-prep inline filter, biocompatible2 micronPEEK wetted, stainless body89 microlitrenot stated for the body6,000 psi, ships with one OC-802 frit
A-411Semi-prep inline filter, biocompatible10 micronPEEK wetted, stainless body103 microlitrenot stated for the body6,000 psi, ships with one OC-803 frit
A-430Inline filter, biocompatible0.5 or 2 micronPEEK end fittings, stainless bodynot stated for the body25 mL/minfrit is replaced with the end fitting
A-431Inline filter, biocompatible, narrow bore0.5 or 2 micronPEEK end fittings, stainless bodynot stated for the body10 mL/minfor a low flow line with a small bore tube
M-520Inline micro filter0.5 micron PEEK fritPEEKnot stated for the bodynot stated for the body150 micron thru hole, connects with MicroTight sleeves
M-525Inline micro filter, direct mount0.5 micron PEEK fritPEEKnot stated for the bodynot stated for the bodytakes 1/32 in OD tubing without a sleeve
M-530Mini micro filter, encapsulated2 micron stainless screenPEEK capsule85 nLnot stated for the body4,000 psi, capsule is replaced as a unit
M-537Mini micro filter, nano1 micron stainless fritPEEK capsule10 nLnot stated for the body4,000 psi, for 360 micron OD tubing
A-101Replacement frit2 micron316 stainless steel0.74 microlitrenot applicablefits the A-315 body and the standard inline bodies
A-102Replacement frit0.5 micron316 stainless steel0.61 microlitrenot applicablesame seat as A-101, finer cut
A-700Replacement frit, PEEK2 micronPEEK0.74 microlitrenot applicablebiocompatible replacement for A-101
A-701Replacement frit, PEEK0.5 micronPEEK0.61 microlitrenot applicablelowest swept volume in the precolumn family
C-425xSmall diameter frit, 10 pack0.5 micron316 stainless steel0.1 microlitrenot applicable0.192 in ring, for a tight volume budget
C-414Semi-prep frit2 micron316 stainless steel26.9 microlitrenot applicable0.375 in disc for the larger bodies
C-415Semi-prep frit2 micron316 stainless steel47.9 microlitrenot applicable0.500 in disc, highest flow of the three

Where the Filter Sits and What It Protects

The same pore size in two positions does two different jobs. This table is the short form used when a line is mapped out, and it is also the one to check when a back pressure problem is blamed on the column.

PositionPore sizeVolume that mattersPart that is protectedFailure it prevents
Reservoir inlet10 to 20 micronnone, this is the supply sidepump head and check valvescavitation and seal wear from fibres and dust
Pump outlet, before the valve0.5 to 2 micron1.3 to 1.4 microlitrevalve rotor sealseal particles reaching the column
Between valve and column0.5 to 2 micron1.3 to 1.4 microlitrecolumn bed and fritsample precipitate blocking the column inlet
Capillary or nano line0.5 micron10 to 85 nLcapillary and unionplugging at the first union after the valve
Semi-prep or prep line2 to 10 micron89 to 103 microlitreprep columnflow restriction at high sample load
Sample preparation, off line0.2 to 2 micronnot applicableautosampler needle and loopfibres drawn from a vial septum

Key Features

Pore size from 0.2 to 20 micron – one family covers the inlet cup, the inline body and the precolumn body, so the pore size is chosen against the particle load rather than against the catalogue section

Open fraction stated per pore size – 18 percent at 0.2 micron rising to 30 percent at 20 micron, which is what makes a pressure estimate possible before a part is ordered

Swept volume from 10 nL – the encapsulated micro bodies are small enough for a capillary line, and the precolumn bodies stay at 1.3 to 1.4 microlitre

PEEK wetted path on the biocompatible range – no stainless steel in the flow path, which keeps proteins and metal sensitive assays out of trouble

Replaceable frit in the precolumn bodies – a new frit goes in with the body still on the line, so the cost per change is the frit rather than the assembly

Pressure rating to 6,000 psi – 414 bar covers a UHPLC line, and the PEEK precolumn bodies are rated at 5,000 psi for the same duty

150 micron thru hole on the micro filters – the bore stays open while the frit does the filtering, which keeps the dead volume at the union low

Inlet cups that reach the bottle bottom – the bottom of bottle body draws within 1/8 in of the base, which reduces the volume left behind when a batch is changed

Applications

These parts sit on the wet side of an instrument, in the path between the reagent bottle and the detector. The three cases below are the ones that set the pore size and the body volume.

HPLC and UHPLC mobile phase – a 10 micron inlet cup in the reservoir and a 0.5 micron precolumn body ahead of the column. On a sub 2 micron column the precolumn body is the part that saves the column, and its 1.3 microlitre volume is small enough not to spoil the peak shape.

Clinical chemistry reagent lines – a reagent pack sits for days and grows particles at the pierce point. A 2 micron inline body between the pump and the valve keeps those particles off the rotor seal, and a PEEK wetted path keeps metal ions out of the reaction.

Process analyser sample lines – sample arrives from a sample probe with fibres and scale. A 10 micron cup at the sample inlet and a 2 micron body ahead of the injection valve give two stages, so the coarse filter can be cleaned while the fine one stays in place.

The instrument boards behind these lines are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the pump drive to the detector front end.

Sourcing and Cross-Reference

Filters are ordered by pore size, body volume and connection, and the connection is the part that goes wrong on a replacement. A 10-32 coned port and a 1/4-28 flat bottom port look similar in a drawing and do not seal into each other, so the port on the existing body is confirmed before a part is named.

Where a line is being copied from an older instrument, the frit seat is the constraint rather than the body. A 0.250 in seat takes the standard frit and a 0.192 in seat does not, and a body that accepts the wrong disc will leak at the seal rather than at the thread. We check the seat diameter and the ring material together.

For a discontinued body, the replacement is chosen on swept volume first and pore size second, because a body with the right pore size and twice the volume changes the chromatography. Where no equivalent exists at the same volume, we say so before shipping a part that fits the thread but shifts the peak.

When a line is down, the body, the frit and the fittings are picked as one set, which is how SUPERB Assembly ships a filter change.

Q: Can you supply a precolumn filter with the same swept volume when the original body is discontinued?

A: SUPERB Tech can confirm the port thread, the frit seat and the swept volume against the old body before the order is placed.

Why Buy Filters and Frits from SUPERB

  1. Position asked before the part: the reservoir cup, the inline body and the precolumn body are specified separately, because the same pore size does different work in each place
  2. Volume and pore size quoted together: a body is never offered on pore size alone, since swept volume is what changes the chromatogram
  3. Connection and frit seat confirmed: the port thread and the disc seat are checked against the existing body before an order is released
  4. PEEK and titanium options on the same list: a biocompatible path is a line item rather than a special request
  5. Maker part number on the paperwork: the frit and the body are both named in the delivery file, which is what a method transfer needs

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Peristaltic and Diaphragm Micro Pumps for Analysers, Inline Degassing Modules and Vacuum Degassers for Liquid Lines, Fluid Check Valves and Shuttle Valves for Reagent and Pneumatic Lines, Syringe Pumps and Precision Dispensers for Analysers, Component Lead Time Solutions.