Quartz Flow Cells and Cuvettes for Photometric Instruments

In a flow cell the optical path and the volume pull against each other, and the method decides which one gives way. The parts below cover all quartz flow cells from 0.1 to 50 mm path with volumes from 6 to 3200 microlitres, plus cuvettes for batch work with a held path length tolerance. Each entry gives path length, centre height, outside dimensions, aperture and volume, so the cell is matched to the light beam and to the tubing that feeds it. Window material is listed with the wavelength band it transmits, because a quartz cell and an optical glass cell are not interchangeable at 340 nm.
quote now

Product Specifications

Product Overview

A flow cell is the point where the optics and the fluidics meet, and the two have opposite wishes. The optics want a long path to absorb more light, which lowers the detection limit. The fluidics want a small volume so that a peak does not smear and a wash step does not take minutes. A 10 mm path with a 3 mm aperture holds 80 microlitres, while the same path in a 12.5 mm square body with a 11 by 6.5 mm aperture holds 750 microlitres, and the difference in peak shape between those two is what a method development run is looking at.

The material matters before the geometry does. Quartz glass carries from 200 to 2500 nm with more than 82 percent transmission on an empty cell, the extended range quartz goes to 3500 nm for near infrared work, and optical glass starts at 320 nm. A method that reads at 340 nm on a glass cell is reading the cell, not the sample. For work below 240 nm there is no substitute for quartz.

Path length tolerance is the quiet specification. A 10 mm cell held to plus or minus 0.01 mm gives an absorbance that can be compared between cells, which is what a batch of measurements needs. The same cell in a demountable body, taken apart for cleaning and put back together, has to be re checked, because the two windows are no longer at a fixed distance. That is why demountable cells are used for viscous samples and not for reference work.

The connections are the last decision and the one that leaks. Flow cells in this range use M6 by 1 screw connectors with FEP tubing, or a small bore tube pushed onto a barb, and the tube bore has to match the flow. A 0.5 mm path cell with a 1 mm tube and a peristaltic pump is a cell that never clears, because the swept volume of the tubing is larger than the cell itself.

Key Specifications

ParameterValueNote
Path lengths0.1, 0.2, 0.5, 1, 2, 5, 10, 50 mm in the flow cell range0.1 to 2 mm for small volumes, 10 mm for absorbance work
Volumes6 to 3200 microlitres depending on path and aperturethe 3 by 3 mm cell at 100 microlitres is the common micro choice
Window materialsquartz glass 200 to 2500 nm, extended range quartz 200 to 3500 nm, special optical glass 320 to 2500 nma glass cell cannot be used below 320 nm
Transmissionmore than 82 percent on an empty quartz cell across 200 to 2500 nmmore than 80 percent for the extended range and the optical glass
Path length toleranceplus or minus 0.003 mm at 0.01 to 0.05 mm, plus or minus 0.005 mm at 0.1 to 0.2 mm, plus or minus 0.01 mm at 0.5 to 20 mm, plus or minus 0.02 mm at 30 to 100 mmthe tolerance is what makes two cells comparable
Centre height8.5 mm or 15 mm on the standard flow cellsthe height has to match the instrument holder
Outside dimensions35 x 12.5 x 12.5 mm on the micro cells, 45 x 12.5 x 12.5 mm on the 10 mm cellsthe holder is built around these two sizes
Apertureround 1.5, 2 or 3 mm, or rectangular 11 by 6.5 mm and 17.5 by 6.5 mmthe aperture sets how much of the beam reaches the sample
ConnectionsM6 by 1 screw connectors with 500 mm of FEP tubing, or a barb for small bore tubethe connector is part of the cell, not an accessory
Flow directioneither direction on the straight cells, one direction on the cells with a bypassa bypass cell has an inlet and an outlet that cannot be swapped
Temperature uselimited by the connector and the tubing, not by the quartzabove 60 C the FEP tube softens at the fitting
Cleaningdemountable bodies allow the windows to be separated, fixed bodies do nota fixed body is cleaned by flushing, not by taking it apart

Included Models and Order Codes

The flow cells are grouped by what the method is trying to do. The micro cells at the top keep the volume small so a peak stays sharp. The 10 mm cells in the middle are the absorbance workhorses, and the two path cell at the end allows a dilute and a concentrated reading without changing the plumbing. The demountable body is last because it is the one for samples that will not flush out.

PartTypePathVolumeApertureOutside sizeNote
176.751-QSflow cell, two path lengths in one body3 x 3 mm100 microlitres3 mm35 x 12.5 x 12.5 mmcentre height 15 mm, M6 by 1 connectors
176.760-QSflow cell, two path lengths in one body5 mm and 10 mmtwo volumes in one bodyas ordered35 x 12.5 x 12.5 mmturn the cell through 90 degrees to change path, tubes stay connected
178.712-QSflow cell, micro10 mm18 microlitres1.5 mm round35 x 12.5 x 12.5 mmsmallest aperture in the 10 mm group
178.710-QSflow cell, micro10 mm80 microlitres3 mm round35 x 12.5 x 12.5 mmcentre height 15 mm or 8.5 mm
178.710-QS, 50 mmflow cell, long path50 mm370 microlitres3 mm round35 x 12.5 x 52.5 mmfor low concentration work
178.711-OSflow cell, optical glass10 mm30 microlitres2 mm round35 x 12.5 x 12.5 mmvisible range only, 320 to 2500 nm
175.000-QSflow cell, standard10 mm750 microlitres11 x 6.5 mm45 x 12.5 x 12.5 mmcentre height 15 mm, the absorbance workhorse
137-QSflow cell, standard10 mm2600 microlitres20 x 9 mm45 x 12.5 x 12.5 mmlarger aperture for a wide beam
130-QSflow cell, standard10 mm3200 microlitres33 x 9.5 mm45 x 12.5 x 12.5 mmlargest volume in the range
170.700-QSflow cell, ultra micro with bypass1 mm62 microlitres17.5 x 3.5 mm35 x 12.5 x 12.5 mmbypass improves flow through a thin channel
170.700-QS, 0.1 mmflow cell, ultra micro with bypass0.1 mm6.2 microlitres17.5 x 3.5 mm35 x 12.5 x 12.5 mmsmallest volume on the page
124-QS, 0.5 mmdemountable cell with tubes0.5 mm85 microlitres15 mm inside diameter22 mm outside diametertakes apart for cleaning, path has to be rechecked
124-QS, 0.1 mmdemountable cell with tubes0.1 mm18 microlitres15 mm inside diameter22 mm outside diameterviscous samples, separate holder required

Choosing Between Path Length and Volume

The two numbers move in opposite directions, and the method decides which one wins. The table is the short version of that choice.

SituationCell to useWhyWhat to watch
Peak has to stay sharp on a fast chromatogram3 by 3 mm cell at 100 microlitres or the 18 microlitre micro cellthe swept volume is small, so the peak does not smeartube bore and connector dead volume
Trace absorbance at 340 nm10 mm path quartz cellpath length buys absorbance, and quartz is needed below 320 nmstray light and cell matching
Sample is viscous or leaves a filmdemountable 124 series cellthe windows can be separated and cleanedpath length has to be rechecked after reassembly
Both a dilute and a concentrated sample in one run176.760-QS two path celltwo path lengths in one body, tubes stay connectedturning the cell changes the optical axis
Very small sample, 20 microlitres or less170.700-QS at 0.1 mm or 178.712-QS at 1.5 mm aperturethe channel is thin and the aperture is narrowparticles block a thin channel, so filter first

Key Features

Quartz from 200 nm – the windows are synthetic quartz, so a reading at 214 or 254 nm is the sample and not the cell

Extended range option – the extended range quartz goes to 3500 nm for near infrared methods on the same instrument

Held path length tolerance – 0.5 to 20 mm paths are held to plus or minus 0.01 mm, which is what makes two cells comparable

Two path lengths in one body – turn the cell through 90 degrees and the tubes stay screwed in, so a method can read dilute and concentrated samples

Bypass on the ultra micro cell – the bypass keeps a thin channel flowing, which is the only way a 6 microlitre cell clears properly

M6 by 1 screw connectors – the tubing is screwed to the cell rather than pushed on, so a 2 bar line does not leak at the fitting

Demountable body for difficult samples – the windows separate for cleaning when a flush will not remove the residue

Standard outside dimensions – 35 and 45 mm bodies fit the holders that instruments are already built around

Applications

Flow cells and cuvettes sit inside photometric instruments, where the optical bench and the fluid path are fixed in the instrument. The three cases below are where the cell specification is decided.

Clinical chemistry and immunoassay analysers – absorbance read at a fixed wavelength after a reagent step. A 10 mm cell with a 750 microlitre volume is the common answer, and a two path cell is used where the same channel reads a blank and a high concentration.

High performance liquid chromatography detectors – a micro cell with a small swept volume keeps a peak from broadening. The 3 by 3 mm body at 100 microlitres and the 18 microlitre micro cell are the two used most, and the choice follows the column bore.

Water and environmental photometry – colour, turbidity and nutrient methods read in a flow cell or a cuvette. Quartz is required for the ultraviolet methods, and a demountable body is used where the sample leaves a film that a flush will not clear.

The detector boards that read these cells are built in the same programmes as SUPERB PCB and SUPERB PCBA work, from the photodiode input to the absorbance calculation.

Sourcing and Cross-Reference

A flow cell is ordered by part number and then by connector, and the connector is where a substitute goes wrong. M6 by 1 threads, barbed inlets and 1/4-28 fittings all appear on cells of the same external size, and the tubing in the instrument only fits one of them.

Path length tolerance is the property that cannot be seen. Two cells of the same dimensions but different tolerance give two different absorbances for the same sample, so where a method compares against a stored calibration, the tolerance has to match the original.

Cuvettes and cells are also the parts that break in handling rather than in service. We supply them in the maker packaging with the material code marked on the body, because a cell with no marking cannot be identified once it is out of the box.

A cell with its tubing and fittings is shipped as one set when an instrument is being rebuilt, which is how SUPERB Assembly handles it.

Q: Can you supply a flow cell with the same path length, volume and connector as the one in the instrument?

A: SUPERB Tech can match path length, volume and connector against the cell already in the instrument before quoting.

Why Buy Flow Cells and Cuvettes from SUPERB

  1. Path, volume and aperture together: the three numbers that decide whether a method transfers are quoted on every line
  2. Window material stated with its band: quartz, extended range quartz and optical glass are named with the wavelength range they carry
  3. Tolerance, not just dimensions: a cell is only comparable to another cell when the path tolerance is the same
  4. Connector confirmed against the instrument: M6 by 1, barbed and threaded options are checked before the order goes out
  5. Matched cells on request: cells can be supplied as a set so that a batch of measurements uses one path length

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: UV and Visible LED Sources for Photometric Detection, Turbidity and Photometric Detector Modules for Water Analysis, Bandpass and Interference Filters for Photometric Channels, Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen, Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Passive Components, Obsolete Part Cross Reference.