Quartz Flow Cells and Cuvettes for Photometric Instruments
Product Specifications
Product Overview
A flow cell is the point where the optics and the fluidics meet, and the two have opposite wishes. The optics want a long path to absorb more light, which lowers the detection limit. The fluidics want a small volume so that a peak does not smear and a wash step does not take minutes. A 10 mm path with a 3 mm aperture holds 80 microlitres, while the same path in a 12.5 mm square body with a 11 by 6.5 mm aperture holds 750 microlitres, and the difference in peak shape between those two is what a method development run is looking at.
The material matters before the geometry does. Quartz glass carries from 200 to 2500 nm with more than 82 percent transmission on an empty cell, the extended range quartz goes to 3500 nm for near infrared work, and optical glass starts at 320 nm. A method that reads at 340 nm on a glass cell is reading the cell, not the sample. For work below 240 nm there is no substitute for quartz.
Path length tolerance is the quiet specification. A 10 mm cell held to plus or minus 0.01 mm gives an absorbance that can be compared between cells, which is what a batch of measurements needs. The same cell in a demountable body, taken apart for cleaning and put back together, has to be re checked, because the two windows are no longer at a fixed distance. That is why demountable cells are used for viscous samples and not for reference work.
The connections are the last decision and the one that leaks. Flow cells in this range use M6 by 1 screw connectors with FEP tubing, or a small bore tube pushed onto a barb, and the tube bore has to match the flow. A 0.5 mm path cell with a 1 mm tube and a peristaltic pump is a cell that never clears, because the swept volume of the tubing is larger than the cell itself.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Path lengths | 0.1, 0.2, 0.5, 1, 2, 5, 10, 50 mm in the flow cell range | 0.1 to 2 mm for small volumes, 10 mm for absorbance work |
| Volumes | 6 to 3200 microlitres depending on path and aperture | the 3 by 3 mm cell at 100 microlitres is the common micro choice |
| Window materials | quartz glass 200 to 2500 nm, extended range quartz 200 to 3500 nm, special optical glass 320 to 2500 nm | a glass cell cannot be used below 320 nm |
| Transmission | more than 82 percent on an empty quartz cell across 200 to 2500 nm | more than 80 percent for the extended range and the optical glass |
| Path length tolerance | plus or minus 0.003 mm at 0.01 to 0.05 mm, plus or minus 0.005 mm at 0.1 to 0.2 mm, plus or minus 0.01 mm at 0.5 to 20 mm, plus or minus 0.02 mm at 30 to 100 mm | the tolerance is what makes two cells comparable |
| Centre height | 8.5 mm or 15 mm on the standard flow cells | the height has to match the instrument holder |
| Outside dimensions | 35 x 12.5 x 12.5 mm on the micro cells, 45 x 12.5 x 12.5 mm on the 10 mm cells | the holder is built around these two sizes |
| Aperture | round 1.5, 2 or 3 mm, or rectangular 11 by 6.5 mm and 17.5 by 6.5 mm | the aperture sets how much of the beam reaches the sample |
| Connections | M6 by 1 screw connectors with 500 mm of FEP tubing, or a barb for small bore tube | the connector is part of the cell, not an accessory |
| Flow direction | either direction on the straight cells, one direction on the cells with a bypass | a bypass cell has an inlet and an outlet that cannot be swapped |
| Temperature use | limited by the connector and the tubing, not by the quartz | above 60 C the FEP tube softens at the fitting |
| Cleaning | demountable bodies allow the windows to be separated, fixed bodies do not | a fixed body is cleaned by flushing, not by taking it apart |
Included Models and Order Codes
The flow cells are grouped by what the method is trying to do. The micro cells at the top keep the volume small so a peak stays sharp. The 10 mm cells in the middle are the absorbance workhorses, and the two path cell at the end allows a dilute and a concentrated reading without changing the plumbing. The demountable body is last because it is the one for samples that will not flush out.
| Part | Type | Path | Volume | Aperture | Outside size | Note |
|---|---|---|---|---|---|---|
| 176.751-QS | flow cell, two path lengths in one body | 3 x 3 mm | 100 microlitres | 3 mm | 35 x 12.5 x 12.5 mm | centre height 15 mm, M6 by 1 connectors |
| 176.760-QS | flow cell, two path lengths in one body | 5 mm and 10 mm | two volumes in one body | as ordered | 35 x 12.5 x 12.5 mm | turn the cell through 90 degrees to change path, tubes stay connected |
| 178.712-QS | flow cell, micro | 10 mm | 18 microlitres | 1.5 mm round | 35 x 12.5 x 12.5 mm | smallest aperture in the 10 mm group |
| 178.710-QS | flow cell, micro | 10 mm | 80 microlitres | 3 mm round | 35 x 12.5 x 12.5 mm | centre height 15 mm or 8.5 mm |
| 178.710-QS, 50 mm | flow cell, long path | 50 mm | 370 microlitres | 3 mm round | 35 x 12.5 x 52.5 mm | for low concentration work |
| 178.711-OS | flow cell, optical glass | 10 mm | 30 microlitres | 2 mm round | 35 x 12.5 x 12.5 mm | visible range only, 320 to 2500 nm |
| 175.000-QS | flow cell, standard | 10 mm | 750 microlitres | 11 x 6.5 mm | 45 x 12.5 x 12.5 mm | centre height 15 mm, the absorbance workhorse |
| 137-QS | flow cell, standard | 10 mm | 2600 microlitres | 20 x 9 mm | 45 x 12.5 x 12.5 mm | larger aperture for a wide beam |
| 130-QS | flow cell, standard | 10 mm | 3200 microlitres | 33 x 9.5 mm | 45 x 12.5 x 12.5 mm | largest volume in the range |
| 170.700-QS | flow cell, ultra micro with bypass | 1 mm | 62 microlitres | 17.5 x 3.5 mm | 35 x 12.5 x 12.5 mm | bypass improves flow through a thin channel |
| 170.700-QS, 0.1 mm | flow cell, ultra micro with bypass | 0.1 mm | 6.2 microlitres | 17.5 x 3.5 mm | 35 x 12.5 x 12.5 mm | smallest volume on the page |
| 124-QS, 0.5 mm | demountable cell with tubes | 0.5 mm | 85 microlitres | 15 mm inside diameter | 22 mm outside diameter | takes apart for cleaning, path has to be rechecked |
| 124-QS, 0.1 mm | demountable cell with tubes | 0.1 mm | 18 microlitres | 15 mm inside diameter | 22 mm outside diameter | viscous samples, separate holder required |
Choosing Between Path Length and Volume
The two numbers move in opposite directions, and the method decides which one wins. The table is the short version of that choice.
| Situation | Cell to use | Why | What to watch |
|---|---|---|---|
| Peak has to stay sharp on a fast chromatogram | 3 by 3 mm cell at 100 microlitres or the 18 microlitre micro cell | the swept volume is small, so the peak does not smear | tube bore and connector dead volume |
| Trace absorbance at 340 nm | 10 mm path quartz cell | path length buys absorbance, and quartz is needed below 320 nm | stray light and cell matching |
| Sample is viscous or leaves a film | demountable 124 series cell | the windows can be separated and cleaned | path length has to be rechecked after reassembly |
| Both a dilute and a concentrated sample in one run | 176.760-QS two path cell | two path lengths in one body, tubes stay connected | turning the cell changes the optical axis |
| Very small sample, 20 microlitres or less | 170.700-QS at 0.1 mm or 178.712-QS at 1.5 mm aperture | the channel is thin and the aperture is narrow | particles block a thin channel, so filter first |
Key Features
Quartz from 200 nm – the windows are synthetic quartz, so a reading at 214 or 254 nm is the sample and not the cell
Extended range option – the extended range quartz goes to 3500 nm for near infrared methods on the same instrument
Held path length tolerance – 0.5 to 20 mm paths are held to plus or minus 0.01 mm, which is what makes two cells comparable
Two path lengths in one body – turn the cell through 90 degrees and the tubes stay screwed in, so a method can read dilute and concentrated samples
Bypass on the ultra micro cell – the bypass keeps a thin channel flowing, which is the only way a 6 microlitre cell clears properly
M6 by 1 screw connectors – the tubing is screwed to the cell rather than pushed on, so a 2 bar line does not leak at the fitting
Demountable body for difficult samples – the windows separate for cleaning when a flush will not remove the residue
Standard outside dimensions – 35 and 45 mm bodies fit the holders that instruments are already built around
Applications
Flow cells and cuvettes sit inside photometric instruments, where the optical bench and the fluid path are fixed in the instrument. The three cases below are where the cell specification is decided.
Clinical chemistry and immunoassay analysers – absorbance read at a fixed wavelength after a reagent step. A 10 mm cell with a 750 microlitre volume is the common answer, and a two path cell is used where the same channel reads a blank and a high concentration.
High performance liquid chromatography detectors – a micro cell with a small swept volume keeps a peak from broadening. The 3 by 3 mm body at 100 microlitres and the 18 microlitre micro cell are the two used most, and the choice follows the column bore.
Water and environmental photometry – colour, turbidity and nutrient methods read in a flow cell or a cuvette. Quartz is required for the ultraviolet methods, and a demountable body is used where the sample leaves a film that a flush will not clear.
The detector boards that read these cells are built in the same programmes as SUPERB PCB and SUPERB PCBA work, from the photodiode input to the absorbance calculation.
Sourcing and Cross-Reference
A flow cell is ordered by part number and then by connector, and the connector is where a substitute goes wrong. M6 by 1 threads, barbed inlets and 1/4-28 fittings all appear on cells of the same external size, and the tubing in the instrument only fits one of them.
Path length tolerance is the property that cannot be seen. Two cells of the same dimensions but different tolerance give two different absorbances for the same sample, so where a method compares against a stored calibration, the tolerance has to match the original.
Cuvettes and cells are also the parts that break in handling rather than in service. We supply them in the maker packaging with the material code marked on the body, because a cell with no marking cannot be identified once it is out of the box.
A cell with its tubing and fittings is shipped as one set when an instrument is being rebuilt, which is how SUPERB Assembly handles it.
Q: Can you supply a flow cell with the same path length, volume and connector as the one in the instrument?
A: SUPERB Tech can match path length, volume and connector against the cell already in the instrument before quoting.
Why Buy Flow Cells and Cuvettes from SUPERB
- Path, volume and aperture together: the three numbers that decide whether a method transfers are quoted on every line
- Window material stated with its band: quartz, extended range quartz and optical glass are named with the wavelength range they carry
- Tolerance, not just dimensions: a cell is only comparable to another cell when the path tolerance is the same
- Connector confirmed against the instrument: M6 by 1, barbed and threaded options are checked before the order goes out
- Matched cells on request: cells can be supplied as a set so that a batch of measurements uses one path length
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: UV and Visible LED Sources for Photometric Detection, Turbidity and Photometric Detector Modules for Water Analysis, Bandpass and Interference Filters for Photometric Channels, Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen, Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Passive Components, Obsolete Part Cross Reference.