Syringe Pumps and Precision Dispensers for Analysers

A syringe pump delivers volume by moving a plunger, so the number that matters is not the flow rate but the resolution of that movement. The parts below cover 60 mm stroke OEM pumps from 25 microlitres to 50 millilitres per stroke, with step counts from 6000 up to 384000 on a smooth flow version. Each entry gives resolution, speed range, precision, accuracy and the syringe sizes the frame accepts, so the pump is matched to the volume before it is matched to the protocol. The slowest speed is quoted as a stroke time rather than a flow, because a nanolitre per minute flow is easier to read as hours per stroke.
quote now

Product Specifications

Product Overview

A syringe pump is a lead screw, a stepper motor and a plunger, and the whole specification follows from how finely the screw can be moved. A 60 mm stroke divided into 6000 steps gives 10 micrometres per step, which in a 500 microlitre syringe is about 80 nanolitres. Divide the same stroke into 48000 microsteps and the same syringe delivers about 10 nanolitres per step. The step count, not the motor, is the resolution.

Precision and accuracy are separate promises. Accuracy is how close the delivered volume is to the volume asked for, and it is quoted at plus or minus 1 percent of full stroke on the pumps below. Precision is how close repeated deliveries are to each other, and it is quoted at 0.05 percent or better at full stroke, which is where the lead screw is most consistent. Near the bottom of the stroke the same pump is less precise, because the same mechanical play is a larger fraction of a short move.

The valve on top of the pump is the other half of the system. A three port valve gives aspirate, dispense and a reagent port, and a nine port distribution valve turns the same pump into a selector that can draw from any of several bottles. The turn time matters as much as the flow, because a 250 millisecond port change is part of the cycle time when a pump is dispensing into a plate.

The practical limit of a syringe pump is the syringe, not the drive. A 50 millilitre syringe at atmospheric pressure needs a much larger force than a 50 microlitre one, which is why the same frame is offered with different force ratings, and why the seal material is chosen from the reagent rather than from the volume. PTFE seals cover most aqueous reagents and dilute acids, and UHMWPE seals are used where the plunger has to be softer.

Key Specifications

ParameterValueNote
Stroke60 mm on both pump familiesthe stroke and the syringe bore together set the volume
Resolution6000 steps standard, 48000 microsteps, 384000 steps on the smooth flow version10 micrometres per step at 6000, which is about 80 nanolitres in a 500 microlitre syringe
Syringe volume25 microlitres to 50 millilitresa 50 millilitre syringe needs a reinforced frame on one family
Accuracyplus or minus 1 percent at full strokequoted at full stroke, not over the whole stroke
Precision0.05 percent or better at full stroke on a 500 microlitre syringe and larger0.1 percent on the 50 and 100 microlitre syringes
Speed2.0 s to 100 minutes per stroke on the standard drive, 28.2 s to 16 hours on the smooth flow drivethe slow end is what makes nanolitre per minute flows possible
Flow range, smooth flow26 nanolitres per minute to 53 millilitres per minutethe range belongs to the drive and the syringe together
Valve3 port, 9 port distribution and other configurationsport change time is part of the cycle
Valve turn timeunder 250 ms between adjacent portsa stepper with an optical encoder positions the port
InterfaceRS-232, RS-485 or CANup to 15 pumps addressed on one bus
Supply24 VDC, 2.0 A peak on one family, 850 mA average and 1.5 A maximum on the otherthe peak current is the motor, not the idle draw
Life4 million cycles on the maker life testthe seal and the valve are the parts that wear first
Seal materialsPTFE or UHMWPE plunger tips, borosilicate glass barrelsPTFE for reagents, UHMWPE where a softer tip is wanted
Environment15 to 40 C, 20 to 80 percent relative humiditycondensation inside the frame is the usual failure in a cold room

Included Models and Order Codes

The two families below are both 60 mm stroke OEM pumps and they are not interchangeable, because one is built around the valve and the other around the drive. The XLP parts are listed by syringe volume, since the same drive takes the whole range and the syringe is the part that is ordered. The PSD parts are listed by variant, because the difference between them is resolution and force rather than volume.

PartFamilyResolutionVolumeSpeedNote
Cavro XLP 6000, 50 microlitreXLP 6000 drive with 26796116000 steps, 48000 microsteps50 microlitres1.2 s to 160 min per strokePTFE seal, 1/4-28 UNF front fitting
Cavro XLP 6000, 250 microlitreXLP 6000 drive with 26798116000 steps, 48000 microsteps250 microlitres1.2 s to 160 min per strokethe common analytical volume
Cavro XLP 6000, 500 microlitreXLP 6000 drive with 26799116000 steps, 48000 microsteps500 microlitres1.2 s to 160 min per stroke0.05 percent precision at full stroke
Cavro XLP 6000, 1 mLXLP 6000 drive with 26790116000 steps, 48000 microsteps1 millilitre1.2 s to 160 min per strokeborosilicate barrel, PTFE tip
Cavro XLP 6000, 5 mLXLP 6000 drive with 26790516000 steps, 48000 microsteps5 millilitres1.2 s to 160 min per strokeHDPE seal version is 2679055
Cavro XLP 6000, 25 mLXLP 6000 drive with 26790916000 steps, 48000 microsteps25 millilitres1.2 s to 160 min per strokehigh volume syringe, slower top speed
Cavro XLP 6000, 50 mLXLP 6000 drive with 26791016000 steps, 48000 microsteps50 millilitres1.2 s to 160 min per strokeneeds the reinforced frame
Hamilton PSD/6 standard 63133-01PSD/648000 microsteps25 microlitres to 50 millilitres2.0 s to 100 min per stroke20 lbf linear force, 25 in oz valve torque
Hamilton PSD/6 smooth flow 99199-01PSD/6384000 steps25 microlitres to 50 millilitres28.2 s to 16 hours per stroke26 nanolitres per minute to 53 millilitres per minute
Hamilton PSD/6 high lift 99196-01PSD/6109714 steps25 microlitres to 50 millilitres8.1 s to 228 min per stroke40 lbf linear force for a large syringe
Hamilton PSD/6 high valve torque 99197-01PSD/648000 microsteps25 microlitres to 50 millilitres2.0 s to 100 min per stroke75 in oz valve torque for a stiff valve
Hamilton PSD/6 high torque and lift 99198-01PSD/6109714 steps25 microlitres to 50 millilitres8.1 s to 228 min per stroke40 lbf and 75 in oz together

From Step Count to Delivered Volume

The step count is the only number that turns into a delivered volume, and it turns into a different volume in every syringe. The table is the short form of that arithmetic.

SyringeVolume per step at 6000 stepsVolume per step at 48000 stepsWhat it means
50 microlitresabout 8 nanolitresabout 1 nanolitremicrostep mode is the only way to dispense single microlitres
250 microlitresabout 42 nanolitresabout 5 nanolitresstandard mode is enough for 2 microlitre steps
500 microlitresabout 83 nanolitresabout 10 nanolitresthe volume where 0.05 percent precision is quoted
5 millilitresabout 0.83 microlitresabout 0.1 microlitreslarge volume, so the step is coarse by comparison
50 millilitresabout 8.3 microlitresabout 1 microlitredispensing below 10 microlitres is not meaningful here

Key Features

6000 and 48000 step drives – the same frame covers coarse and fine work, and the microstep mode is what makes a 50 microlitre syringe useful

Smooth flow drive at 384000 steps – the slow end reaches 16 hours per stroke, which is how a nanolitre per minute flow is held steady

Valve with a 250 ms port change – a distribution valve turns one pump into a selector, and the port change is part of the cycle time

Optical encoder for step loss – the drive reports lost steps, so a jammed syringe shows up as an error instead of as a short delivery

Force variants for large syringes – a 40 lbf version carries a 50 millilitre syringe that a 20 lbf frame would stall on

Seals chosen from the reagent – PTFE and UHMWPE tips are both available across the volume range

Serial bus with 15 addresses – several pumps share one link, which is how a multi channel analyser is built

Daisy chain to 16 units – the other family addresses 16 pumps on one chain, with the same command set

Applications

These pumps sit in analysers that have to move a known volume without a continuous flow. The three cases below decide the resolution and the force.

Clinical chemistry and immunoassay dispensers – reagent and sample are aspirated and dispensed into a cuvette. A 250 or 500 microlitre syringe on a 48000 step drive gives a few nanolitres per step, which is what a 2 microlitre dispense needs.

Sample preparation and liquid handling – serial dilution and plate filling. A 5 or 25 millilitre syringe with a distribution valve draws from one bottle and delivers to several wells, and the valve turn time sets the throughput.

Chromatography and mass spectrometry loading – a sample loop is filled slowly to avoid cavitation. The smooth flow drive holds a flow of a few microlitres per minute, and the slow stroke is the reason for the 384000 step version.

The driver boards that move these pumps are the kind of assembly SUPERB PCB and SUPERB PCBA programmes build, from the stepper driver to the encoder input.

Sourcing and Cross-Reference

A pump is ordered by stroke, resolution and valve, and the syringe is ordered beside it. A pump that arrives with the right drive and the wrong syringe seal is a pump that has to be taken apart before it can run, and the seal is the part the reagent attacks first.

Where an OEM pump is no longer listed, the command set is the constraint. Two pumps with the same stroke and volume can speak different protocols, and a replacement that needs new firmware is a different project from a drop in part.

Valves and syringes are consumables on this family, and they wear at different rates. We keep the valve, the syringe and the seal as separate lines so a maintenance kit can be ordered without a new pump.

A pump with its syringe, valve and seal is shipped as one kit when an analyser is being rebuilt, which is how SUPERB Assembly handles it.

Q: Can you supply an OEM syringe pump with the same stroke, volume and command set as the one in the instrument?

A: SUPERB Tech can match stroke, resolution and interface against the pump already in the instrument before quoting.

Why Buy Syringe Pumps from SUPERB

  1. Resolution quoted with the syringe: the step count is turned into a volume per step, so the resolution is read in microlitres rather than in steps
  2. Accuracy and precision separated: the two figures are quoted at full stroke, which is where both are measured
  3. Force and valve torque stated: a large syringe needs more force, and the variant is chosen from the syringe rather than from the volume range
  4. Consumables quoted with the pump: valve, syringe and seal are separate lines, so a maintenance order does not become a pump order
  5. Command set checked on a substitute: the interface is confirmed against the controller before a replacement is named

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Rotary and Injection Valves for Analyser Sample Paths, Peristaltic and Diaphragm Micro Pumps for Analysers, Inline Degassing Modules and Vacuum Degassers, Solenoid and Pinch Valves for Fluid Handling in Analysers, Air Bubble Detectors for Infusion and Reagent Lines, Sensors and Detectors, Component Lead Time Solutions.