  • Differential Pressure Sensors for Ventilators and Process Lines

Differential Pressure Sensors for Ventilators and Process Lines

A ventilator measures flow by measuring the pressure drop across a fixed restriction, so the sensor has to resolve a few pascal. The 125 Pa and 500 Pa bodies hold a zero point accuracy of 0.08 and 0.1 Pa, with no measurable drift. Response is under 3 ms and the digital versions sample at 2 kHz, which is faster than the breathing cycle needs. Ranges run from 125 Pa through 2 kPa, in manifold mount, tube connection and reflow solderable bodies.
quote now

Product Specifications

Product Overview

Most flow measurement in a ventilator or an analyser is a pressure measurement in disguise: the gas passes a restriction and the drop across it follows the square of the flow. That makes the sensor a low pressure device, because the drop at a normal breath is often below 100 Pa. A sensor with a 500 Pa range and a 3 percent span accuracy resolves about 15 Pa at full scale, which is why the zero point accuracy of 0.1 Pa matters more than the span figure for the low end of the breath.

The second requirement is speed. A breathing cycle is slow compared with electronics, but the valve that the loop controls is not, and a sensor that averages over a hundred milliseconds turns a fast inspiratory phase into a lag. The thermal principle used in these bodies responds in under 3 ms and samples at 2 kHz, so the signal is ahead of the valve rather than behind it. The same property helps in a leak test, where the pressure step itself is the measurement.

The third decision is the body. A manifold mount sensor screws directly onto a machined block with an O ring, which keeps the internal volume small and the response fast. A tube connection body sits on a bracket and reaches the port through a short hose, which is easier to route in a crowded enclosure. A reflow solderable body goes on the board itself, where the pneumatic path is part of the manifold and the board area is the limit. All three use the same sensing principle, so the choice is mechanical.

Key Specifications

ParameterValueNote
Measuring principlethermal flow through a micro channelmeasures differential pressure of air and non aggressive gases
Pressure rangesplus or minus 125 Pa, plus or minus 500 Pa, plus or minus 2 kPabi-directional, zero at mid scale
Zero point accuracy0.08 Pa on the 125 Pa bodies, 0.1 Pa on the 500 Pa bodiesthe figure that decides the low end of the range
Span accuracy3 percent of readingplus the zero point error, not instead of it
Offset stabilitybelow 0.05 Pa per year on the SDP800 bodies, below 0.01 Pa per year on the SDP31no field zeroing expected
Response timeunder 3 ms to 63 percent of a stepmeasurement time 40 to 50 ms in triggered mode
Resolution16 bitpressure and temperature share the same interface
InterfaceI2C at 400 to 1000 kHz, or ratiometric analog 0.5 to 4.5 Vdigital bodies sample at 2 kHz
Supply2.7 to 5.5 V on the SDP800 bodies, 3 to 3.6 V on the SDP313.3 V nominal
Supply current3.8 mA measuring, 1.1 mA idle, 1 uA in sleepon the SDP800 bodies
Mediaair, nitrogen, oxygen, non condensingcalibrated for air and nitrogen
Overpressure1 bar allowable, 5 bar burst on the SDP800 bodies3 bar burst on the SDP31
Temperatureminus 40 to plus 85 C, accuracy 2 C from minus 10 to plus 60 Cthe reading is the silicon temperature, not the gas
Dimensions29 x 18 x 27.05 mm on the tube body, 5 x 8 x 5 mm on the SDP31weight 6 g and 0.2 g respectively
Wetted materialsPBT, glass, silicon, gold, FR4, silicone seal, epoxy, copper alloylead free solder, REACH and RoHS compliant
Mountingmanifold with O ring, tube connection, or reflow on boardper body, same sensing principle

Included Models and Order Codes

The body decides the mounting and the range decides the resolution. SDP810-500Pa takes 4 mm tubing and covers plus or minus 500 Pa with a 0.1 Pa zero, which is the general purpose part on a bench analyser. SDP810-125Pa covers plus or minus 125 Pa on the same body, which is the range for a ventilator flow restriction where the drop is small. SDP800-500Pa is the same sensor with a manifold connection, sealed on an O ring, so there is no hose between the port and the sensor. SDP31 is the reflow solderable body at 5 by 8 by 5 mm, covering plus or minus 500 Pa, and SDP32 is the same body at 125 Pa. The analog parts are the option where the controller has no spare I2C channel: MPXV7002DP covers plus or minus 2 kPa and gives a 0.5 to 4.5 V ratiometric output from a 5 V rail, which is the range for a filter or a duct line rather than for a breath.

ModelRangeOutput and connectionZero accuracyNotes
SDP810-500Paplus or minus 500 PaI2C, 4 mm tube ports0.1 Paorder code 1-101532-01, I2C address 0x25
SDP810-125Paplus or minus 125 PaI2C, 4 mm tube ports0.08 Pasame body, smaller range
SDP800-500Paplus or minus 500 PaI2C, manifold with O ring0.1 Paorder code 1-101551-01
SDP801-500Paplus or minus 500 PaI2C, manifold, address 0x260.1 Patwo sensors on one bus
SDP31plus or minus 500 PaI2C, reflow solderable, 5 x 8 x 5 mm0.1 Pa0.2 g, offset stability below 0.01 Pa per year
SDP32plus or minus 125 PaI2C, reflow solderable, 5 x 8 x 5 mm0.08 Pasame footprint as SDP31
MPXV7002DPplus or minus 2 kPaanalog 0.5 to 4.5 V, dual portper maker sheet5 V supply, ratiometric output

Range Conversion for the Units on Customer Drawings

A pneumatic drawing is as likely to be in inches of water as in pascal, and a filter specification is often in millibar. The table converts the ranges in this family into the four units that appear in customer documents, using 249.09 Pa per inch of water and 100 Pa per millibar. It shows why the 125 Pa body is the one for a breathing circuit: 125 Pa is half an inch of water, which is the order of magnitude of the drop across a flow restriction at a normal breath.

PascalInches of waterMillibarKilopascal
100 Pa0.40 inH2O1.0 mbar0.10 kPa
125 Pa0.50 inH2O1.25 mbar0.125 kPa
250 Pa1.00 inH2O2.5 mbar0.25 kPa
500 Pa2.01 inH2O5.0 mbar0.50 kPa
1000 Pa4.01 inH2O10.0 mbar1.0 kPa
2000 Pa8.03 inH2O20.0 mbar2.0 kPa

Key Features

Zero point in fractions of a pascal – 0.08 and 0.1 Pa on the 125 Pa and 500 Pa bodies is what lets a breath be resolved at the start and the end of inspiration.

No offset drift to compensate – offset stability below 0.05 Pa per year, and below 0.01 Pa per year on the reflow body, so a field zeroing routine is not needed.

Response under 3 ms – the signal is ahead of the valve it controls, which is what a closed loop dosing or ventilation cycle needs.

2 kHz sampling – the digital bodies update at 2 kHz in continuous mode, so a leak test sees the step rather than an average of it.

Bi-directional range – the range is plus and minus around zero, which covers both inspiration and expiration through one restriction.

Three mounting styles – manifold with an O ring, tube connection and reflow solderable, so the pneumatic path is a mechanical decision rather than a new board layout.

Media compatibility stated – air, nitrogen and oxygen, non condensing, calibrated for air and nitrogen, which covers a ventilator gas path.

Sleep mode at 1 uA – the digital bodies drop to a microamp between measurements, which matters in a battery instrument that samples in bursts.

Applications

Three measurements where a low pressure sensor is the whole instrument, not an accessory to it.

Ventilator flow and airway pressure – the drop across a fixed restriction gives flow, and the same sensor reads the airway pressure during the expiratory phase, with the 125 Pa body covering a normal breath.

Filter and duct monitoring – a 500 Pa body across a filter shows the loading before the airflow falls, and the 2 kPa analog part covers a duct line with a higher drop.

Leak and pressure decay test – the 2 kHz sampling and 3 ms response catch the pressure step of a small leak, which a slower sensor averages into the noise.

A sensor on a pneumatic block is usually part of a small controller board, so the SUPERB PCBA and the SUPERB PCB are checked against the same port layout and the hose routing is reviewed once for both.

Sourcing and Cross-Reference

The first question is the range, and it is set by the restriction, not by the supply pressure. A 500 Pa body on a circuit that only develops 40 Pa leaves the reading in the bottom tenth of the range, where the zero point accuracy is doing all the work. The second question is the body: a manifold part needs the machined block and the O ring, and a tube part needs a short, equal length pair of hoses or the reading becomes a flow measurement of the hoses.

The third question is the interface. Two digital sensors on one bus need different I2C addresses, which is why the 0x26 variant exists. A controller without a spare I2C channel takes the analog body, at the cost of an ADC and a ratiometric reference.

Where a body has been superseded, the usual route is the same range in another mounting style, since the pneumatic port and the calibration stay. See lead time and availability for a specific model, and how to verify a drop in replacement where a sensor sits inside an existing manifold.

The SUPERB Assembly team seals the manifold ports with the maker gasket and keeps the two hoses equal in length, because an unequal pair reads as a pressure offset rather than as a flow.

Q: Can you confirm the pressure range and the mounting style, and whether the two ports are manifold or tube, because our pneumatic block is already machined for one of them?

A: Yes. Send the pressure range, the mounting style, the interface the controller uses, the gas and the ambient temperature, and the SUPERB Tech team returns the body and the order code that fit.

Why Instrument Builders Source Differential Pressure Sensors Through SUPERB

  1. Range chosen against the restriction: each body is quoted with its zero point accuracy and its span figure, so the low end of the reading is not left to the span percentage.
  2. Mounting stated per code: manifold, tube and reflow bodies are kept apart with their order codes, because the pneumatic block is machined before the sensor is ordered.
  3. Drift figure included: offset stability per year is part of the offer, which is what decides whether a field zeroing step is needed in the service manual.
  4. Media and temperature limits: gas compatibility and the temperature range are listed, so the wetted material check does not have to be reconstructed from a drawing.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Liquid Flow Sensors for Medical and Analytical Instruments, Industrial Pressure Sensors and Transducers, Temperature and Humidity Sensors, Level Sensors and Float Switches for Tanks and Sumps, IEC 60601 Power Supplies for Medical Diagnostic Equipment, How to Verify a Drop in Replacement.