Differential Pressure Sensors for Ventilators and Process Lines
Product Specifications
Product Overview
Most flow measurement in a ventilator or an analyser is a pressure measurement in disguise: the gas passes a restriction and the drop across it follows the square of the flow. That makes the sensor a low pressure device, because the drop at a normal breath is often below 100 Pa. A sensor with a 500 Pa range and a 3 percent span accuracy resolves about 15 Pa at full scale, which is why the zero point accuracy of 0.1 Pa matters more than the span figure for the low end of the breath.
The second requirement is speed. A breathing cycle is slow compared with electronics, but the valve that the loop controls is not, and a sensor that averages over a hundred milliseconds turns a fast inspiratory phase into a lag. The thermal principle used in these bodies responds in under 3 ms and samples at 2 kHz, so the signal is ahead of the valve rather than behind it. The same property helps in a leak test, where the pressure step itself is the measurement.
The third decision is the body. A manifold mount sensor screws directly onto a machined block with an O ring, which keeps the internal volume small and the response fast. A tube connection body sits on a bracket and reaches the port through a short hose, which is easier to route in a crowded enclosure. A reflow solderable body goes on the board itself, where the pneumatic path is part of the manifold and the board area is the limit. All three use the same sensing principle, so the choice is mechanical.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Measuring principle | thermal flow through a micro channel | measures differential pressure of air and non aggressive gases |
| Pressure ranges | plus or minus 125 Pa, plus or minus 500 Pa, plus or minus 2 kPa | bi-directional, zero at mid scale |
| Zero point accuracy | 0.08 Pa on the 125 Pa bodies, 0.1 Pa on the 500 Pa bodies | the figure that decides the low end of the range |
| Span accuracy | 3 percent of reading | plus the zero point error, not instead of it |
| Offset stability | below 0.05 Pa per year on the SDP800 bodies, below 0.01 Pa per year on the SDP31 | no field zeroing expected |
| Response time | under 3 ms to 63 percent of a step | measurement time 40 to 50 ms in triggered mode |
| Resolution | 16 bit | pressure and temperature share the same interface |
| Interface | I2C at 400 to 1000 kHz, or ratiometric analog 0.5 to 4.5 V | digital bodies sample at 2 kHz |
| Supply | 2.7 to 5.5 V on the SDP800 bodies, 3 to 3.6 V on the SDP31 | 3.3 V nominal |
| Supply current | 3.8 mA measuring, 1.1 mA idle, 1 uA in sleep | on the SDP800 bodies |
| Media | air, nitrogen, oxygen, non condensing | calibrated for air and nitrogen |
| Overpressure | 1 bar allowable, 5 bar burst on the SDP800 bodies | 3 bar burst on the SDP31 |
| Temperature | minus 40 to plus 85 C, accuracy 2 C from minus 10 to plus 60 C | the reading is the silicon temperature, not the gas |
| Dimensions | 29 x 18 x 27.05 mm on the tube body, 5 x 8 x 5 mm on the SDP31 | weight 6 g and 0.2 g respectively |
| Wetted materials | PBT, glass, silicon, gold, FR4, silicone seal, epoxy, copper alloy | lead free solder, REACH and RoHS compliant |
| Mounting | manifold with O ring, tube connection, or reflow on board | per body, same sensing principle |
Included Models and Order Codes
The body decides the mounting and the range decides the resolution. SDP810-500Pa takes 4 mm tubing and covers plus or minus 500 Pa with a 0.1 Pa zero, which is the general purpose part on a bench analyser. SDP810-125Pa covers plus or minus 125 Pa on the same body, which is the range for a ventilator flow restriction where the drop is small. SDP800-500Pa is the same sensor with a manifold connection, sealed on an O ring, so there is no hose between the port and the sensor. SDP31 is the reflow solderable body at 5 by 8 by 5 mm, covering plus or minus 500 Pa, and SDP32 is the same body at 125 Pa. The analog parts are the option where the controller has no spare I2C channel: MPXV7002DP covers plus or minus 2 kPa and gives a 0.5 to 4.5 V ratiometric output from a 5 V rail, which is the range for a filter or a duct line rather than for a breath.
| Model | Range | Output and connection | Zero accuracy | Notes |
|---|---|---|---|---|
| SDP810-500Pa | plus or minus 500 Pa | I2C, 4 mm tube ports | 0.1 Pa | order code 1-101532-01, I2C address 0x25 |
| SDP810-125Pa | plus or minus 125 Pa | I2C, 4 mm tube ports | 0.08 Pa | same body, smaller range |
| SDP800-500Pa | plus or minus 500 Pa | I2C, manifold with O ring | 0.1 Pa | order code 1-101551-01 |
| SDP801-500Pa | plus or minus 500 Pa | I2C, manifold, address 0x26 | 0.1 Pa | two sensors on one bus |
| SDP31 | plus or minus 500 Pa | I2C, reflow solderable, 5 x 8 x 5 mm | 0.1 Pa | 0.2 g, offset stability below 0.01 Pa per year |
| SDP32 | plus or minus 125 Pa | I2C, reflow solderable, 5 x 8 x 5 mm | 0.08 Pa | same footprint as SDP31 |
| MPXV7002DP | plus or minus 2 kPa | analog 0.5 to 4.5 V, dual port | per maker sheet | 5 V supply, ratiometric output |
Range Conversion for the Units on Customer Drawings
A pneumatic drawing is as likely to be in inches of water as in pascal, and a filter specification is often in millibar. The table converts the ranges in this family into the four units that appear in customer documents, using 249.09 Pa per inch of water and 100 Pa per millibar. It shows why the 125 Pa body is the one for a breathing circuit: 125 Pa is half an inch of water, which is the order of magnitude of the drop across a flow restriction at a normal breath.
| Pascal | Inches of water | Millibar | Kilopascal |
|---|---|---|---|
| 100 Pa | 0.40 inH2O | 1.0 mbar | 0.10 kPa |
| 125 Pa | 0.50 inH2O | 1.25 mbar | 0.125 kPa |
| 250 Pa | 1.00 inH2O | 2.5 mbar | 0.25 kPa |
| 500 Pa | 2.01 inH2O | 5.0 mbar | 0.50 kPa |
| 1000 Pa | 4.01 inH2O | 10.0 mbar | 1.0 kPa |
| 2000 Pa | 8.03 inH2O | 20.0 mbar | 2.0 kPa |
Key Features
Zero point in fractions of a pascal – 0.08 and 0.1 Pa on the 125 Pa and 500 Pa bodies is what lets a breath be resolved at the start and the end of inspiration.
No offset drift to compensate – offset stability below 0.05 Pa per year, and below 0.01 Pa per year on the reflow body, so a field zeroing routine is not needed.
Response under 3 ms – the signal is ahead of the valve it controls, which is what a closed loop dosing or ventilation cycle needs.
2 kHz sampling – the digital bodies update at 2 kHz in continuous mode, so a leak test sees the step rather than an average of it.
Bi-directional range – the range is plus and minus around zero, which covers both inspiration and expiration through one restriction.
Three mounting styles – manifold with an O ring, tube connection and reflow solderable, so the pneumatic path is a mechanical decision rather than a new board layout.
Media compatibility stated – air, nitrogen and oxygen, non condensing, calibrated for air and nitrogen, which covers a ventilator gas path.
Sleep mode at 1 uA – the digital bodies drop to a microamp between measurements, which matters in a battery instrument that samples in bursts.
Applications
Three measurements where a low pressure sensor is the whole instrument, not an accessory to it.
Ventilator flow and airway pressure – the drop across a fixed restriction gives flow, and the same sensor reads the airway pressure during the expiratory phase, with the 125 Pa body covering a normal breath.
Filter and duct monitoring – a 500 Pa body across a filter shows the loading before the airflow falls, and the 2 kPa analog part covers a duct line with a higher drop.
Leak and pressure decay test – the 2 kHz sampling and 3 ms response catch the pressure step of a small leak, which a slower sensor averages into the noise.
A sensor on a pneumatic block is usually part of a small controller board, so the SUPERB PCBA and the SUPERB PCB are checked against the same port layout and the hose routing is reviewed once for both.
Sourcing and Cross-Reference
The first question is the range, and it is set by the restriction, not by the supply pressure. A 500 Pa body on a circuit that only develops 40 Pa leaves the reading in the bottom tenth of the range, where the zero point accuracy is doing all the work. The second question is the body: a manifold part needs the machined block and the O ring, and a tube part needs a short, equal length pair of hoses or the reading becomes a flow measurement of the hoses.
The third question is the interface. Two digital sensors on one bus need different I2C addresses, which is why the 0x26 variant exists. A controller without a spare I2C channel takes the analog body, at the cost of an ADC and a ratiometric reference.
Where a body has been superseded, the usual route is the same range in another mounting style, since the pneumatic port and the calibration stay. See lead time and availability for a specific model, and how to verify a drop in replacement where a sensor sits inside an existing manifold.
The SUPERB Assembly team seals the manifold ports with the maker gasket and keeps the two hoses equal in length, because an unequal pair reads as a pressure offset rather than as a flow.
Q: Can you confirm the pressure range and the mounting style, and whether the two ports are manifold or tube, because our pneumatic block is already machined for one of them?
A: Yes. Send the pressure range, the mounting style, the interface the controller uses, the gas and the ambient temperature, and the SUPERB Tech team returns the body and the order code that fit.
Why Instrument Builders Source Differential Pressure Sensors Through SUPERB
- Range chosen against the restriction: each body is quoted with its zero point accuracy and its span figure, so the low end of the reading is not left to the span percentage.
- Mounting stated per code: manifold, tube and reflow bodies are kept apart with their order codes, because the pneumatic block is machined before the sensor is ordered.
- Drift figure included: offset stability per year is part of the offer, which is what decides whether a field zeroing step is needed in the service manual.
- Media and temperature limits: gas compatibility and the temperature range are listed, so the wetted material check does not have to be reconstructed from a drawing.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Liquid Flow Sensors for Medical and Analytical Instruments, Industrial Pressure Sensors and Transducers, Temperature and Humidity Sensors, Level Sensors and Float Switches for Tanks and Sumps, IEC 60601 Power Supplies for Medical Diagnostic Equipment, How to Verify a Drop in Replacement.