UV and Visible LED Sources for Photometric Detection
Product Specifications
Product Overview
A photometric channel is an LED, a sample, a filter and a detector, and the LED is the part that is most often chosen on the wrong number. Luminous flux in lumens is weighted to the human eye and falls to nothing below 400 nm, so it says nothing useful about a 275 nm source. What the measurement needs is radiant power at the wavelength, the width of the emission band and the stability of that band with current and temperature. The parts on this page are listed with those figures rather than with a brightness claim.
The deep UV end starts at 255 nm, where a TO-39 part delivers around 0.2 milliwatts at 20 mA with a forward voltage between 6.5 and 8 volts. At 275 nm the same package delivers around 0.6 milliwatts. Those are small numbers, and they set the rest of the channel: a short optical path, a detector with enough gain, and a reference channel if the reading has to hold over a shift. The 340 nm part is a different class, an SMD emitter at 350 mA that delivers 40 milliwatts with a 10 nm band, and it is the one to use when the assay reads NADH and the sample absorbs strongly.
The visible end is served by direct colour emitters rather than by phosphor converted parts, because a phosphor broadens the spectrum and a photometric channel needs a narrow band. The blue part is binned from 460 to 485 nm with a 20 nm half width, the green from 520 to 540 nm with 30 nm, and the red from 620 to 645 nm with 20 nm. That binning is the reason two LEDs with the same part number can read differently: a 25 nm spread across the blue bin is wider than the absorption band of some assays, so the bin has to be fixed or the channel has to carry a filter.
Two rules keep these parts alive. The first is drive: a deep UV LED has a forward voltage that varies by more than a volt between parts, so a series resistor from a fixed rail gives a current that varies just as much, and the maker states that a voltage source will destroy the part. The second is thermal: the 340 nm SMD has a thermal resistance of 8.5 C/W and a junction limit of 95 C, so the board it sits on is part of the optical specification. A shift in junction temperature moves the wavelength by 0.05 nm per degree on the visible parts and changes the output of the UV parts by a comparable amount.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Wavelength range | 255 nm to 627 nm across the family | 255 and 275 nm for deep UV, 340 nm for near UV, 470, 530 and 627 nm for visible |
| Peak tolerance | plus or minus 3 nm on the UV parts, binned to 10 or 20 nm bands on the visible parts | the bin is part of the order code on the visible parts |
| Spectral width | 10 nm full width at half maximum on the 340 nm SMD, 11 to 15 nm on the TO-39 UV parts, 20 to 30 nm on the visible parts | a narrow band is what makes a fixed wavelength channel work |
| Radiant power, deep UV | 0.2 mW typical at 255 nm and 20 mA, 0.6 mW typical at 275 nm and 20 mA | microwatt class, so the path is short and the detector has gain |
| Radiant power, near UV | 40 mW at 340 nm and 350 mA | the strongest UV part on the page |
| Flux, visible | 41 lm at 470 nm, 102 lm at 530 nm, 46 lm at 627 nm, all at 350 mA | lumens are quoted for the visible parts only |
| Forward voltage | 6.5 to 8.0 V at 255 nm, 6.0 to 7.2 V at 275 nm, 4.1 V at 340 nm, 2.3 to 3.5 V on the visible parts | a current source is mandatory on the UV parts |
| Test current | 20 mA on the TO-39 UV parts, 350 mA on the 340 nm SMD and the visible parts | the drive current sets the junction temperature and the output |
| Maximum current | 30 mA DC and 200 mA pulsed at 1 percent duty on the TO-39 UV parts, 350 mA DC on the 340 nm SMD | pulsing raises the peak power without raising the average |
| Power dissipation | 180 mW on the TO-39 UV parts, 1.68 W on the 340 nm SMD | the board is the heatsink on the SMD part |
| Thermal resistance | 50 C/W at 255 nm, 8.5 C/W at 340 nm, 10 to 12 C/W on the visible parts | junction temperature moves both output and wavelength |
| Viewing angle | 115 to 125 degrees, hemispherical lens on the HL parts, flat window on the FW parts | the lens changes the coupling into a fibre or a flow cell |
| Operating temperature | minus 30 to plus 55 C on the TO-39 UV parts, minus 40 to plus 135 C on the visible parts | the UV parts are the narrower of the two |
| Package | TO-39 metal glass hermetic with a UV window, or SMD for the 340 nm and visible parts | a hermetic package is what keeps a deep UV part alive in humid air |
Included Models and Order Codes
Six parts cover the bands that a fixed wavelength photometer or a small spectrometer uses. The two TO-39 parts are the deep UV end, 255 nm and 275 nm, and they are the ones with the widest spread of forward voltage, so they are driven from a current source and not from a resistor on a rail. The 340 nm SMD part is the workhorse for an enzymatic assay at 350 mA, and it is the only UV part here that delivers tens of milliwatts. The three visible parts are direct colour emitters with a half width of 20 to 30 nm, which is narrow enough for a dye endpoint and wide enough that a filter is still worth having on a strong absorber.
| Part | Wavelength | Power or flux | Test current | Forward voltage | Package |
|---|---|---|---|---|---|
| UVTOP255-HL-TO39 | 255 nm, plus or minus 5 nm | 0.12 to 0.20 mW | 20 mA | 6.5 to 8.0 V | TO-39 hermetic, hemispherical lens |
| UVTOP250-FW-TO39 | 255 nm typical, 250 to 260 nm | 0.2 to 0.4 mW | 20 mA | 6.0 to 7.5 V | TO-39 hermetic, flat window |
| UVTOP275-HL-TO39 | 275 nm, 270 to 280 nm | 0.36 to 0.60 mW | 20 mA | 6.0 to 7.2 V | TO-39 hermetic, hemispherical lens |
| UVTOP340H-FW-SMD | 345 nm typical, 340 to 350 nm | 40 mW | 350 mA | 4.1 V | SMD, flat window, 115 degrees |
| LXML-PB01-0040 | 470 nm typical, 460 to 485 nm | 41 lm | 350 mA | 2.55 to 3.51 V | SMD, 125 degrees viewing |
| LXML-PM01-0100 | 530 nm typical, 520 to 540 nm | 102 lm | 350 mA | 2.55 to 3.51 V | SMD, 125 degrees viewing |
| LXML-PD01-0040 | 627 nm typical, 620 to 645 nm | 46 lm | 350 mA | 2.31 to 3.51 V | SMD, 125 degrees viewing |
Which Band a Photometric Channel Uses
The wavelength follows from the chemistry, and the part follows from the power that the chemistry needs. The table is the short version of that mapping for the bands on this page.
| Band | What it is chosen for | Part to start with | What to check |
|---|---|---|---|
| 255 and 275 nm | direct UV absorbance and the reference channel of a UV photometer | UVTOP255-HL-TO39 or UVTOP275-HL-TO39 | quartz optics, sub milliwatt power, detector gain |
| 340 nm | NADH and NADPH linked enzymatic assays, where the sample absorbs strongly | UVTOP340H-FW-SMD | junction temperature and the board as heatsink |
| 470 nm | blue absorbing dye endpoints and colourimetric tests read between 450 and 480 nm | LXML-PB01-0040 | the wavelength bin, 460 to 485 nm is a wide spread |
| 530 nm | green absorbing dye endpoints, including DPD chlorine which reads between 515 and 540 nm | LXML-PM01-0100 | the 30 nm half width against the assay band |
| 627 nm | red absorbing endpoints and a second reference wavelength | LXML-PD01-0040 | the lower forward voltage of the red part |
Key Features
Radiant power instead of lumens – the deep UV and near UV parts are quoted in milliwatts, because lumens do not describe a 275 nm source.
Spectral width per part – 10 nm on the 340 nm SMD, 11 to 15 nm on the TO-39 parts, 20 to 30 nm on the visible parts, so the filter decision is made on a number.
Current source called out – the TO-39 parts have a forward voltage spread of more than a volt, and the maker states that a voltage source destroys them.
Thermal resistance stated – 8.5 C/W at 340 nm with a 95 C junction limit, which makes the board part of the optical specification.
Hermetic TO-39 UV parts – a metal glass package with a UV window keeps a deep UV emitter alive in humid air.
Lens and window options – hemispherical lens for coupling into a flow cell, flat window for a collimated path.
Binned visible parts – 460 to 485 nm on the blue, 520 to 540 nm on the green and 620 to 645 nm on the red, with the bin in the order code.
Wide operating range on visible parts – minus 40 to plus 135 C, which covers an instrument that runs warm inside a cabinet.
Applications
Four channels where the band and the radiant power decide the part rather than the package.
Enzymatic analyser at 340 nm – the SMD part delivers 40 mW at 350 mA, which is enough to pass a 10 mm cuvette and still leave signal on the detector.
Water chlorine measurement – a 530 nm emitter with a 30 nm half width covers the DPD absorbance band without a narrow filter in front of the detector.
UV absorbance reference – a 255 nm TO-39 part gives the short wavelength end of a dual beam channel, with the second beam on a reference detector.
Handheld colorimeter – a 470 nm part and a matched photodiode make a compact channel for a dye test that reads in the blue.
In line turbidity and colour – a 627 nm part works as the reference wavelength where the sample scatters more than it absorbs.
The LED sits on the same optical board as the SUPERB PCBA and the SUPERB PCB, so the thermal pad under the emitter is part of the board rather than a separate heatsink.
Sourcing and Cross-Reference
The first question on a photometric LED is the band, and the band comes from the chemistry rather than from the catalogue. Once the band is fixed, the second question is the power that has to reach the detector through the sample, because a 0.2 milliwatt deep UV part and a 40 milliwatt near UV part need different optics and different detectors. Send the wavelength, the path length and the expected absorbance range, and the part follows.
The second question is the bin. A blue emitter binned from 460 to 485 nm can sit 25 nm away from the wavelength the assay was set up on, and a photometer calibrated on one bin will read differently on another. The same applies to the UV parts, where the peak is quoted with a tolerance of 3 nm and the band moves with junction temperature. Fixing the bin and the test current is what makes two instruments built a month apart read the same.
The third question is the drive and the board. The TO-39 parts are current driven and will not survive a voltage source, and the 340 nm SMD needs a copper area that keeps the junction below 95 C at 350 mA. A part that is correct on paper and hot on the board loses output and shifts wavelength at the same time. See lead time and availability for a specific bin, and BOM shortage recovery where a bin change would need a recalibration.
The SUPERB Assembly team keeps the emitter off the board until the reflow profile for the optical side is fixed, because a deep UV window does not survive a second pass.
Q: Can you supply the 470 nm and 530 nm emitters in a fixed wavelength bin, so a replacement part does not need a recalibration?
A: Yes. Send the wavelength, the bin, the test current and the detector, and the SUPERB Tech team returns the part and the drive condition that hold the reading.
Why Instrument Builders Source Photometric LEDs Through SUPERB
- Band first: the wavelength is chosen from the chemistry, and the part is chosen from the power that has to reach the detector.
- Bin held: the visible parts are ordered on a fixed bin, so a replacement does not move the calibration.
- Drive stated: the current source requirement and the forward voltage spread are on the page, which is where a UV channel usually fails.
- Thermal path: thermal resistance and junction limit are quoted, so the board can be sized before the first prototype.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Turbidity and Photometric Detector Modules, PIN Photodiodes and Avalanche Photodiodes, Phototransistors for Optical Sensing, Water Quality Sensor Circuits for pH and Conductivity, High Power LED Emitters for Illumination and Infrared, Laser Diodes and VCSEL Emitters, Component Lead Time Solutions, BOM Shortage Recovery.