AD5933YRSZ
Product Specifications
AD5933YRSZ
Analog Devices — Analog & Mixed-Signal ICs
Product Overview
The AD5933YRSZ from Analog Devices is a high-performance analog / mixed-signal IC providing precision signal acquisition, conversion, and conditioning. Analog Devices (ADI) is a world leader in high-performance analog, mixed-signal, and DSP ICs for instrumentation, healthcare, automotive, and communications.
Superb Automation partners with leading brands including Qorvo, TI, ADI, XILINX, NVIDIA, Intel, ST, MACOM, Renesas, and Coilcraft to provide complete BOM and supply-chain solutions.
Key Features
- High-resolution data conversion with excellent SNR and THD
- Wide dynamic range and high linearity across operating conditions
- Integrated voltage reference with low temperature drift
- Flexible digital interface — SPI, I2C, or parallel options
- Industrial/automotive temperature range with guaranteed specs
- Low power consumption with power-down modes
Key Specifications
| Resolution | bits |
| Channels | — |
| Sample Rate | kSPS |
| Interface | — |
| Supply Voltage | V |
| Operating Temp | °C |
| Package | — |
| INL | LSB |
| SNR | dB |
| THD | dB |
Applications
Request Quote — In Stock, 100% New & Original, Full Warranty
AD5933YRSZ: A 27 Bit Tuning Word on a 100 kHz Sweep Is 0.75 mHz of Step
The AD5933 is an impedance converter that combines a frequency generator, a 12 bit 1 MSPS converter and a discrete Fourier transform engine on one die, communicating over an I2C interface. The excitation frequency is set by a 27 bit tuning word, which on a 16.776 MHz master clock gives a frequency resolution below 0.1 Hz across a sweep up to 100 kHz, and the impedance measurement range is 1 kohm to 10 Mohm with a system accuracy of 0.5 percent. It runs from 2.7 V to 5.5 V and includes a temperature sensor with 2 degrees Celsius accuracy.
What the DFT engine returns and what it does not
At each frequency the engine returns a real and an imaginary data word, which is the response of the impedance at that frequency expressed as two numbers. The magnitude and the phase are computed off chip from those two words, and the conversion from code to impedance requires a calibration with a known resistor at each frequency point. That calibration step is not optional: the gain of the receive path varies with frequency and with the value of the feedback resistor, so a sweep without per point calibration reads a curve that includes the instrument's own response.
Where the 12 bits go
The 12 bit converter sets the dynamic range of a single measurement, which is 72 dB before averaging. The impedance range of 1 kohm to 10 Mohm spans four decades, so a single setting of the feedback resistor cannot cover it: the data sheet specifies a range of about four to one in impedance per setting, and the 100 ohm to 1 kohm decade needs additional circuitry. A sweep across several decades therefore needs either a switchable feedback resistor or several passes.
| Parameter | Value | What it decides |
|---|---|---|
| Excitation frequency | Up to 100 kHz | Electrochemical and bioimpedance range |
| Frequency resolution | 27 bit, below 0.1 Hz | 0.75 mHz steps near 100 kHz |
| Impedance range | 1 kohm to 10 Mohm | Four decades with a switched feedback resistor |
| System accuracy | 0.5 percent | After calibration at each frequency |
| Converter | 12 bit, 1 MSPS | 72 dB of single measurement range |
| Output type | Real and imaginary words | Magnitude and phase computed off chip |
| Excitation amplitude | Programmable peak to peak | Trade between signal and cell disturbance |
| Temperature sensor | Plus or minus 2 degrees C | Correction for a temperature dependent cell |
| Interface | I2C, up to 400 kHz | Two wires to a host processor |
| Supply | 2.7 V to 5.5 V | Runs from a 3.3 V rail |
| Temperature range | -40 C to +125 C | Automotive qualified |
| Clock | Internal 16.776 MHz or external | A shared system clock is an option |
| Package | 16-lead SSOP | Small board area for a complete analyser |
The feedback resistor sets the range
The receive path measures the current through the impedance under test by passing it through a feedback resistor, so the output code is proportional to that resistor and inversely proportional to the impedance. Choosing a feedback resistor close to the impedance under test puts the signal in the middle of the converter's range, and choosing one a decade away loses three bits. In a sweep over a decade of impedance the feedback resistor should be switched at the boundaries, which is why a practical instrument has several resistors and a multiplexer rather than one fixed value.
Calibration and the two point check
A calibration with a known resistor at each frequency removes the gain and the offset of the receive path. A second point, either a short or a known capacitor, removes the phase error. Without the second point the phase reading carries a fixed offset that looks like a property of the cell. A cell that reads a constant 3 degree phase at every frequency is more likely to be an uncalibrated instrument than a physical effect.
Buying an impedance front end
The comparison across candidates is the frequency range, the impedance range per setting and the accuracy after calibration, and the order code fixes the package. A part with the same function and a lower maximum frequency cannot serve an electrochemical application that sweeps to 100 kHz. Availability is tracked through ADC DAC cross reference, and the wider question of a sensor bill of materials is set out in medical device components. Where a short part forces a change, the checks are in how to verify a drop in replacement.