RTD and Thermocouple Probes for Analyser Blocks
Product Specifications
Product Overview
The temperature at a reaction block, a column oven or a sample cell is usually measured through a small port, into a metal body that is hotter than the air around it. That geometry decides more about the reading than the sensor inside: a 3 mm sheath responds in seconds, a 6 mm sheath lasts longer, and a probe that touches the bottom of a blind port reads the block while a probe hanging in the gas space reads nothing useful.
Two sensor families cover the job. A Pt100 element changes resistance by roughly 0.385 ohms per ohm per degree and is linear enough to be read without correction across minus 100 to 400 C, which is why it is the default for block temperature. A type K thermocouple produces about 41 uV per degree and reaches 1149 C in an Inconel sheath, which is the choice for heaters, hot gas lines and anything a Pt100 cannot survive.
The lead connection is where most installations go wrong. A two wire Pt100 adds the lead resistance to the reading, which shows up as a fixed offset of a degree or more. Three wire compensation removes most of it and four wire removes all of it, so the probe should be chosen with the input the controller actually has, not the other way round.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| RTD element | Pt100, alpha 0.00385, class A or 1/3 DIN | Pt1000 and 2 wire versions available |
| Wiring | 2, 3 or 4 wire | 3 wire compensates lead resistance |
| Thermocouple types | K, J, T, N and E | Class 1 tolerance as standard |
| Sheath material | 304 stainless, 316 stainless, Inconel 600 | Inconel for high temperature |
| Sheath diameter | 1.5, 3, 4.5 and 6 mm | 3 and 6 mm are the stocked sizes |
| Probe length | 50 to 500 mm | Longer and lagged versions to order |
| RTD temperature range | Minus 100 to 400 C | Class A or better |
| Thermocouple range | Up to 1149 C in Inconel, 899 C in 304 | Per calibration and sheath |
| Pressure at the sheath | 20 bar at 3 mm, 40 bar at 6 mm | Probe body rating |
| Response, T66 in liquid | 3 s for a Pt100, 4 s for a thermocouple | Closed end sheath |
| Threads | M8 x 1, M10 x 1, G1/8, G1/4, G1/2, 1/4 NPT | Match the block port |
| Lead insulation | PVC, silicone, FEP, fibreglass | Sets the lead temperature limit |
Included Models and Order Codes
The thermocouple block is grouped by calibration and sheath: SS for 304 stainless, IN for Inconel 600, XL for the compacted mineral sheath, with the junction letter and the length in inches inside the part number. The RTD rows are the Pt100 assemblies used on analyser blocks, with the element class and the wiring stated. The last row is an ordering code rather than a stocked number, which is how a Pt100 with a specific thread and lead length is specified.
| Part Number | Type | Sheath | Diameter | Length | Range | Use |
|---|---|---|---|---|---|---|
| TJ36-CASS-18U-6 | Type K, ungrounded | 304 stainless | 3 mm | 150 mm | 899 C | General block temperature |
| TJ36-CASS-116U-6 | Type K, ungrounded | 304 stainless | 1.5 mm | 150 mm | 899 C | Fast response in small ports |
| TJ36-CAIN-18U-6 | Type K, ungrounded | Inconel 600 | 3 mm | 150 mm | 1071 C | Heaters and hot gas |
| TJ36-CAIN-18U-12 | Type K, ungrounded | Inconel 600 | 3 mm | 300 mm | 1071 C | Deeper immersion |
| TJ36-ICSS-18U-6 | Type J, ungrounded | 304 stainless | 3 mm | 150 mm | 521 C | Reducing atmospheres |
| TJ36-ICIN-116G-12 | Type J, grounded | Inconel 600 | 1.5 mm | 300 mm | 441 C | Grounded tip on metal blocks |
| TJ36-CAXL-18U-6 | Type K, ungrounded | Mineral XL | 3 mm | 150 mm | 1149 C | Long life at high temperature |
| TJ36-NNIN-18U-6 | Type N, ungrounded | Inconel 600 | 3 mm | 150 mm | 1200 C nominal | Drift resistant at temperature |
| P3A-S316-125-PX-6-PFXX-80-STWL | Pt100, 3 wire, class A | 316 stainless | 3.2 mm | 150 mm | 260 C | Wet sample lines |
| P-A-6-100-M8-XM-2 | Pt100 ordering code, class A | 304 stainless | 6 mm | 100 mm | 400 C | Thread and lead made to order |
Key Features
Element class stated – class A, 1/3 DIN and 1/10 DIN Pt100 options, so the tolerance is known before the loop is tuned
Wiring matched to the input – 2, 3 and 4 wire builds, with 3 wire as the usual compromise for lead resistance
Junction choice on thermocouples – grounded for fast response on metal, ungrounded for isolation, exposed for gas
Sheath by temperature – 304 stainless to 899 C, Inconel 600 to 1071 C, mineral XL to 1149 C
Thread to the port – M8, M10, G1/8, G1/4, G1/2 and 1/4 NPT, with adjustable compression fittings where the depth has to move
Lead insulation by environment – PVC to 100 C, silicone to 200 C, FEP to 250 C, fibreglass to 400 C
Applications
These probes are fitted to a metal body with a port, so the application is usually described by the block rather than by the sensor.
Analyser reaction and sample blocks – a 3 mm Pt100 in a 1/4 NPT port for block temperature, with a type K alongside as an over temperature check
Column ovens and heater blocks – a Pt100 in a G1/4 port, 3 wire to the controller, with the sheath length set by the block depth
Hot gas and exhaust lines – a type K in an Inconel sheath, ungrounded, where a Pt100 would not survive the sheath temperature
Temperature probes and the boards that read them are quoted together, so SUPERB PCB and SUPERB PCBA orders can include the probe and its front end.
Sourcing and Cross-Reference
A probe is a mechanical part as much as a sensor. The sheath length has to reach the middle of the flow, the thread has to match the port without an adaptor, and the lead has to leave the block without touching it. When a probe is quoted from a drawing we ask for three dimensions: port thread, insertion depth and the space available for the lead to bend.
Interchangeability is the second point. A class A Pt100 from one maker and a class B from another will read several tenths of a degree apart at 200 C, which matters when a method names a block temperature. We quote the element class and the calibration basis together so two batches of probes read the same.
Thermocouple ageing is worth a note in the maintenance file. Type K drifts in the first tens of hours at high temperature and then settles, so a probe that reads 3 C high after a week in a 900 C oven is often behaving normally. For long term use above 800 C, type N in an Inconel sheath holds its calibration better and is stocked for that reason.
Probe, gland and connector are kitted through SUPERB Assembly, with the lead length cut to the enclosure drawing.
Q: Can you supply a Pt100 probe with a 1/4 NPT thread, a 100 mm sheath and a 3 wire FEP lead for a sample block?
A: Yes. SUPERB Tech quotes Pt100 and thermocouple probes against the port thread, the insertion depth and the controller input, and states the element class.
Why Choose SUPERB for RTD and Thermocouple Probes
- Port dimensions first: thread, insertion depth and lead clearance confirmed before the probe is quoted
- Class and calibration stated: element class and tolerance named, so two batches read the same
- Sheath by temperature: 304, 316, Inconel and mineral XL options with the working limit given
- Lead insulation matched: PVC, silicone, FEP and fibreglass options tied to the temperature at the gland
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Thermocouple and RTD Front Ends, NTC Thermistors and Temperature Probes, Column Ovens and Heater Blocks, Inline Fluid Heaters and Peltier Chillers, Immersion Assemblies and Sample Probes, Calibration Buffers and Conductivity Standards.