Immersion Assemblies and Sample Probes for Online Analysers

A pH or conductivity sensor is only as good as the point it sits in. An immersion assembly holds the probe at a depth where the reading is representative, seals the vessel, and lets a technician reach the sensor without draining the tank. The bodies below run from a polypropylene dip tube for an open channel up to a stainless retractable body rated 10 bar at 140 C, with hygienic versions that take 30 minutes of saturated steam at 120 C. Immersion depth runs from 48 mm on a short retractable body to 3,000 mm on a dip tube, and the process connection is a DN 100 or DN 80 flange, an ANSI 3 in flange, or a G 1 1/4 thread. Each line gives depth, pressure, temperature and wetted material together, because the depth that fits a tank and the pressure that fits a pipe rarely come from the same body.
quote now

Product Specifications

Product Overview

An online analyser reads what the sensor sees, and the sensor sees what the assembly puts in front of it. That is the whole reason an immersion assembly exists: the probe has to sit in flowing, representative medium, deep enough that a foam layer or a sludge blanket does not cover it, and far enough from a wall that the reading is not taken in a stagnant pocket. A dip tube hung over the side of a channel and a retractable body bolted into a pressurized pipe solve the same problem in two different ways, and the choice is made by the vessel rather than by the sensor.

The parts of the assembly are always the same, even when the shapes differ. There is a tube or a housing that sets the depth, a holder that grips the 12 mm sensor or the 120 mm electrode, a seal that keeps the process out of the junction, a process connection that mates with the vessel, and a head that takes the cable and the potential matching pin. The pin matters more than it looks: a pH loop in a plastic tank needs a reference path through the medium, and a body that offers a pin in Alloy C4, tantalum or 316 gives the loop that path. Without it the reading wanders and no amount of calibration fixes it.

Depth is a compromise between representativeness and mechanics. A 1,000 mm dip tube reaches below the foam in an aeration basin and is short enough not to bend in a current. A 3,000 mm tube reaches the bottom of a storage tank but needs a flange stiff enough to carry its own weight and the drag of the medium. A retractable body in a pipe is the other extreme: 48 to 235 mm of immersion, because the pipe bore decides the depth and a long body would block the flow. The tables below list depth per body, and the selection table turns vessel type into a starting depth.

Pressure and temperature narrow the list quickly. A polypropylene dip tube is fine in an open channel but is limited to 5 bar absolute in its pressurized version and 80 C in all versions. A PVDF body reaches 6 bar and 150 C, and a stainless body reaches 10 bar and 150 C with the same sensor. The retractable bodies add a second limit: 4 bar when the stroke is moved by hand and 10 bar when it is moved by air, because the pressure cylinder is what holds the seal while the sensor is withdrawn. A line that runs at 6 bar and expects manual actuation is therefore a line that needs a different body, not a stronger gland.

Retraction is what makes maintenance possible without a shutdown. The sensor is pulled out of the process into a rinse chamber, washed, calibrated and pushed back, all while the pipe stays full. On a hygienic line the same motion is what allows the body to be steamed in place: 30 minutes of saturated steam at 120 C is the documented cycle for the hygienic body in this family. What the retraction does not do is remove the need for a rinse supply and a drain, so a retractable body is specified together with its water and air connections rather than on its own.

Key Specifications

ParameterValueNote
Immersion depth48 mm on the shortest retractable body to 3,000 mm on a dip tubethe vessel decides the depth, not the sensor
Process pressure4 bar with manual actuation, 10 bar with pneumatic actuation on the retractable bodies; 5 bar absolute on the pressurized plastic dip tubethe pressure cylinder holds the seal during a stroke
Process temperatureminus 10 to plus 80 C on PP, up to 150 C on PVDF and stainless140 C for 1 hour at 3 bar on the hygienic body
Sensor sizes12 mm glass or ISFET sensors, and 120 mm or 225 mm electrodes in the retractable bodiesthe holder is specific to the length, not just the diameter
Wetted materialsPP, PP-GF20 head, PVDF, stainless steel 1.4404the head is outside the process and can differ from the tube
Seal materialsEPDM, FKM, Chemraz, Fluoraz, Kalrez on the retractable bodiesseal choice follows the medium and the steam cycle
Process connectionsDN 100 flange, DN 80 PN16, ANSI 3 in 150 lb, JIS 10K 80A, G 1 1/4 threadthe flange standard is often the reason a body cannot be reused
Potential matchingpin in Alloy C4, tantalum or 316 stainless steelrequired for a stable pH reading in a plastic or coated vessel
Cable entries1 x Pg 13.5 and 2 x Pg 16 on the dip tube headsthe second entry takes a cleaning line or a spare gland
Pneumatic supply4 to 6 bar, filtered to 40 micron, Oil and water freeno continuous air consumption in the measure position
Rinse supply2 to 6 bar of clean water at the rinse port of a retractable bodya rinse supply is part of the installation, not an accessory
Sterilisation30 minutes of saturated steam at 120 C on the hygienic bodydocumented as in line steam sterilizable including the seals
Weight2.5 to 3.0 kg on a PVDF body, 8.0 to 12.0 kg on a stainless body, about 4.0 kg on the PP dip tubeweight is carried by the flange, so it sets the nozzle stiffness
Mounting attitudevertical preferred, flange square to the nozzlea tilted dip tube traps air under the holder

Included Models and Order Codes

The table is ordered by how the body meets the process: first the dip tubes that hang into an open or closed vessel, then the retractable bodies that stroke through a pipe wall. Every line is a body that exists at that depth and that pressure, and the note names the part that usually has to be ordered with it. Where a temperature is published only for a shorter cycle, the cycle is written next to the number rather than dropped.

BodyWetted materialImmersion depthPressureTemperatureConnectionNote
Dipfit CPA111-APP tube, PP-GF20 holder1,000 or 2,000 mmunpressurizedminus 10 to 80 CDN 100 flangeopen tank or channel, bayonet holder
Dipfit CPA111-BPP tube, PP-GF20 holder1,000 or 2,000 mmunpressurizedminus 10 to 80 Cadjustable DN 100 flangedepth set on site with two half shells
Dipfit CPA111-CPP tube, PP-GF20 holder1,000 or 2,000 mm5 bar absolute at 20 Cminus 10 to 80 Cpressurized DN 100 flangeclosed vessel, unpressurized above 80 C
Dipfit CPA111-DPP tube, stainless bracket1,000 mmunpressurizedminus 10 to 80 Csuspension brackethangs over a channel, no nozzle needed
Dipfit CPA140, PVDFPVDF tube and holder500 to 2,500 mm6 barup to 150 CDN 80, ANSI 3 in, JIS 10K 80Athree sensor slots plus a cleaning head
Dipfit CPA140, stainless1.4404 tube and holder500 to 2,500 mm10 barup to 150 CDN 80, ANSI 3 in, JIS 10K 80Aaggressive media, 8 to 12 kg depending on length
Cleanfit CPA471, short1.440495 mm4 bar manual, 10 bar air0 to 140 CG 1 1/4120 mm gel or 225 mm liquid KCl electrode
Cleanfit CPA471, long1.4404201 mm4 bar manual, 10 bar air0 to 140 CG 1 1/4225 mm electrode, longer stroke into the pipe
Cleanfit CPA473, PA cylinder1.4404 tube, PA cylinder61 to 235 mm6 barup to 80 CG 1 1/4scraper seals, manual actuation only
Cleanfit CPA473, stainless cylinder1.4404 tube and cylinder61 to 235 mm10 bar continuous at 6 bar and 100 Cup to 100 CG 1 1/4sticky or fibrous media, sleeve seals
Cleanfit CPA475, short1.440448 mm4 bar manual, 10 bar air0 to 100 C, 140 C for 1 h at 3 barhygienic clamp23 mm stroke, in line steam sterilizable
Cleanfit CPA475, long1.4404101 mm4 bar manual, 10 bar air0 to 100 C, 140 C for 1 h at 3 barhygienic clamp65 mm stroke, 120 C steam for 30 minutes

Matching the Body to the Vessel

The vessel is the first fact in the choice, and it settles the connection, the depth and the actuation before the sensor is discussed. This table is the short form of that decision.

VesselConnectionBody to start withDepth ruleWhat goes wrong otherwise
Open channel or sumpsuspension bracketCPA111-Dsensor at least 50 mm below the lowest water linethe sensor dries out between batches and the reading drifts
Tank with a DN 100 nozzleDN 100 flangeCPA111-A or CPA111-Ctube ends 100 mm above the tank floorsludge covers the junction and the reading falls away
Pressurized pipe up to 10 barG 1 1/4CPA471 or CPA473short body unless the bore is largea long body blocks the bore and the pump sees the restriction
Hygienic or aseptic linehygienic clampCPA475short stroke, depth set by the seal seatproduct is trapped in the rinse chamber and cannot be cleaned
Sticky or fibrous mediumG 1 1/4CPA473scraper body, depth to clear the wall crustthe seal fouls and the sensor reads the crust, not the medium
Deep tank of 3 mDN 80 flangeCPA1402,500 mm tube plus extensionthe tube resonates in the current and cracks at the flange
Plastic or coated vesselanybody with a potential matching pinpin must be in the medium at all timesthe pH reading wanders and calibration cannot fix it

Key Features

Depth from 48 mm to 3,000 mm – a short retractable body in a pipe and a long dip tube in a tank come from the same range, so a plant can standardise on one sensor type

Manual and pneumatic actuation – 4 bar by hand and 10 bar by air on the same body, which matters where the line pressure is above what a hand stroke can hold

Five body materials – PP, PP-GF20, PVDF, 1.4404 and a PA pressure cylinder, chosen against the medium rather than against the price

Potential matching pin in three materials – Alloy C4, tantalum and 316, so a plastic tank or a coated wall still gives a stable pH loop

Hygienic body with a documented steam cycle – 30 minutes of saturated steam at 120 C, seals included, which is the line that a validation file asks for

Three sensor slots on the heavy duty body – pH, ORP and temperature at one nozzle, or two sensors plus a cleaning head

Rinse and air ports as part of the body – 2 x G 1/4 rinse and 2 x G 1/8 air on the plastic retractable body, so the cleaning circuit is not improvised on site

Bayonet sensor holder – the holder comes out without tools, which turns a calibration into a bench job rather than a line stoppage

Applications

These bodies sit at the point where a sensor meets the process, which is the point where most analyser faults start. The three cases below are the ones that set the depth, the material and the actuation.

Municipal and industrial wastewater – a dip tube in an aeration basin reaches below the foam, and a retractable body in the return line allows a pH electrode to be cleaned without draining the pipe. The plastic bodies are the usual choice here because the medium is aggressive to metal and the pressure is low.

Chemical process and tank farms – a PVDF or stainless body with three sensor slots puts pH, ORP and temperature at one nozzle, and the 10 bar rating covers a pressurized vessel. The potential matching pin is what keeps the loop stable in a lined tank.

Pharmaceutical and food water systems – a hygienic retractable body is steamed in place and stroked for calibration without breaking the aseptic boundary. The 120 C steam cycle and the 140 C short excursion are the two numbers the validation file needs.

The analyser electronics behind these probes are the kind of build SUPERB PCB and SUPERB PCBA programmes support, from the high impedance input to the isolated current loop.

Sourcing and Cross-Reference

An assembly is ordered by the vessel connection first, then by depth, then by material, and the connection is the part that decides whether an existing nozzle can be reused. A DN 100 flange, a DN 80 flange and an ANSI 3 in flange are three different bodies with the same sensor, so the nozzle drawing is checked before a part is named.

Depth is confirmed against the vessel rather than against the old body. A tube that ended 100 mm above the floor in a clean tank can be the wrong length once a sludge blanket forms, and the assembly that replaced it is then blamed for a reading that is actually a siting problem. We ask for the low water line and the floor distance before quoting a length.

Where a body is obsolete, the sensor and the nozzle are the fixed points and the body is what changes. A short body can sometimes be replaced by a long one with a spacer, but a manual body cannot be replaced by a pneumatic one without an air supply, and we say that before shipping rather than after the flange is bolted.

A body, a sensor and a cable gland are ordered as one set when a line is down, which is how SUPERB Assembly ships a replacement.

Q: Can you match an immersion assembly to an existing DN 100 nozzle and keep the same 120 mm electrode?

A: SUPERB Tech can check the nozzle standard, the electrode length and the seal material against the old body before an order is released.

Why Buy Immersion Assemblies from SUPERB

  1. Vessel asked before the body: connection, depth and medium are checked first, because a body that fits the sensor and not the nozzle is a return
  2. Depth and pressure on every line: the two numbers that rule out a body appear together, which stops a 6 bar pipe being fitted with a 4 bar body
  3. Pin and seal materials named: the potential matching pin and the O ring material are line items, not assumptions
  4. Rinse and air ports specified with the body: a retractable body is quoted with the connections it needs, so the cleaning circuit is not discovered at commissioning
  5. Maker document with the shipment: the technical data sheet travels with the body, which is what a validation or an audit file expects

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: pH and ORP Electrodes for Process Analysers, Conductivity Cells and Toroidal Sensors for Water Analysers, Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen, Turbidity and Photometric Detector Modules for Water Analysis, Inline Filters and Frits for Reagent Lines, Obsolete Part Cross Reference Hub.