Retention time moves with column temperature. A bench that swings two degrees over a shift will drift peaks enough to fail a system suitability check, and the usual fix is not a new pump but a column oven or a heater block that actually holds set point. This page covers both: Peltier chambers that heat and chill a whole compartment, and aluminium blocks that clamp a single column and take it to 150 C.
The two families solve different problems. A Peltier oven is chosen when the method needs sub ambient control, because it can hold 4.0 C to 70.0 C and bring a column back down between runs. A block heater is chosen when the method runs hot, up to 150 C, and when the column is long: the block bodies listed here take columns up to 50 cm and can carry two Rheodyne valves in the heated zone, which removes a cold transfer line.
What decides a replacement is chamber size and control stability, not the outside dimensions. A unit that holds four 30 cm by 1 inch columns cannot be replaced by a single column oven without changing the method, and a block rated to 0.1 C stability cannot be swapped for one that only holds 1 C if the method is sensitive to retention time.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Peltier oven range | 4.0 C to 70.0 C | Heating and chilling |
| Heating only oven range | 20.0 C to 90.0 C | Single column chamber |
| Stability | +/- 0.1 C to +/- 0.2 C | After warm up |
| Accuracy | +/- 0.2 C | Both oven families |
| Chamber uniformity | +/- 0.5 C | Across the chamber |
| Column capacity | four 30 cm x 1 in columns | High capacity Peltier oven |
| Block heater range | Ambient to 150 C | Aluminium block bodies |
| Block element power | 125 W, 300 W, 500 W | One, four and long column bodies |
| Block sensor | 1000 ohm RTD | Platinum resistance sensor |
| Control | PID with RS232 on the ovens | Ramp and soak available |
| Over temperature cut out | 180 C +/- 10 C | Independent protection |
Included Models and Order Codes
The table mixes ovens and blocks because they are ordered for the same slot on the bench. Capacity is given as columns by length, since a unit that holds four 30 cm columns is a different replacement from one that holds a single column, even when the temperature range is identical.
| Part Number | Type | Temperature Range | Capacity | Control | Note |
|---|---|---|---|---|---|
| CO20 | Peltier column oven | 20.0 C to 90.0 C | One column, 30 cm, 1/4 or 3/8 in | PID, RTD sensor | 40 W heater |
| CO30 | Peltier column oven | 4.0 C to 70.0 C | Four columns, 30 cm x 1 in | PID, RS232 | 100 W Peltier, 230 W max |
| CH-30 | Heater block body | Ambient to 150 C | One column up to 1/2 in OD, 30 cm | External controller | 125 W element |
| CH-430 | Heater block body | Ambient to 150 C | Four columns, 30 cm | External controller | 300 W, two valve positions |
| CH-460 | Heater block body | Ambient to 150 C | Four columns, up to 50 cm | External controller | 500 W element |
| CH-500 | Heater and controller | Ambient to 150 C | Two 1/4 in and two 1/8 in blocks | Built in controller | 120 V and 240 V versions |
| 500879 | Replacement block | Ambient to 150 C | 4 in blocks, pack of two | Fits CH-500 | Spare part |
| 500895 | Replacement block | Ambient to 150 C | 1 in blocks, pack of two | Fits CH-500 | Spare part |
| WAT038039 | Temperature control system | Ambient to 150 C | One heater module plus controller | Programmable | Controls three modules |
| WAT038040 | Column heater module | Ambient to 150 C | Four columns to 30 cm | From WAT038039 | Additional module |
Key Features
Sub ambient control – the Peltier chamber holds 4.0 C, which a block heater cannot reach without a chiller
0.1 C stability – retention time stays inside a system suitability window over a full shift
Four column capacity – a high capacity chamber holds four 30 cm columns at one set point
Blocks to 150 C – aluminium bodies take a column well above the range of a Peltier oven
Valve in the heated zone – the four column block has provision for two Rheodyne valves, which removes a cold transfer line
Long column support – one body takes columns up to 50 cm, so a long method does not need a second oven
Replacement blocks – block halves are sold as spares, so a damaged channel does not scrap the controller
Over temperature protection – an independent cut out at 180 C protects the column if a sensor fails
Applications
The choice between a chamber and a block follows the method, not the bench space.
Pharmaceutical method development – sub ambient control holds a method at 4.0 C to 70.0 C through a sequence of runs
Food and flavour analysis – block heaters take a column to 150 C where a Peltier chamber cannot reach
Multi column systems – a four column chamber keeps every column in a method at the same set point
Column temperature control is supplied with SUPERB PCB and SUPERB PCBA work when the analyser build also needs its controller board.
Sourcing and Cross-Reference
A column oven is replaced on capacity and range, not on outside size. Four 30 cm columns in one chamber is a different requirement from one column in a block, and a unit that meets the temperature range but not the capacity will change the method even though it fits the bench.
Block heaters are simpler: the column outside diameter, the column length and the block length decide the part, and the controller is separate. That separation is useful, because the block and its element wear while the controller usually does not, and replacement blocks are available on their own.
Where a method is already validated, the set point stability matters more than the headline range. A unit that holds 0.1 C will reproduce a retention time; one that drifts a degree will not, whatever its maximum temperature says. We quote the stability figure with the range rather than after it.
Controller boards for ovens and blocks run through SUPERB Assembly before the chamber is calibrated.
Q: Can you supply a replacement heater block without a new controller?
A: Yes. SUPERB Tech supplies heater blocks as separate parts so a damaged block does not require a new controller.
Why Choose SUPERB for Column Temperature Control
- Capacity before size: the replacement is matched on columns held and length, which is what changes a method
- Stability stated: the control stability is quoted with the range, because that is what holds a retention time
- Blocks sold apart: a worn block is replaced without buying a controller again
- Valve heating available: a heated valve position removes the cold transfer line that causes peak broadening
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Rotary and Injection Valves, Sample Loops and Injection Loops, Inline Fluid Heaters and Peltier Chillers, Needle and Metering Valves, Tube Fittings and Ferrules, Bubble Traps and Debubblers.