Inline Fluid Heaters and Peltier Chillers for Analysers
Product Specifications
Product Overview
An inline heater adds energy to a moving stream, so its useful output is not the wattage on the label but the temperature rise it can hold at the flow the method needs. The relation is simple: watts equal flow in litres per second times 4.18 times the rise in kelvin, times about 1000 for water. A 1000 W heater raises water flowing at 10 mL per second by about 24 K, and the same heater on a 30 mL per second line raises it by 8 K. That is why a heater is chosen from the flow first and from the wattage second.
Watt density is the second number, and it is the one that decides whether the heater survives. Power has to cross the wall into the liquid, and if the liquid is not moving fast enough the wall runs hotter than the fluid. A layered heater on a thin stainless substrate reaches 450 W per square inch on water, which is 70 W per square centimetre, and that figure only holds while the tube is full and flowing. Run the same heater dry and the wall temperature runs away in seconds, which is why the water rating and the air rating differ by a factor of three.
The standard parts carry their own high limit thermocouple in the heated section, and that sensor is not a substitute for a flow switch. The control loop reads the wall, the flow switch proves the liquid is there, and the two together are what keep a dry start from turning into a repair. On liquid duty the heater also has to be mounted vertically with the inlet at the bottom, so that a bubble leaves the tube instead of sitting on the element.
Cooling below ambient is a different problem, because there is nothing to burn and everything to gain from stability. A thermoelectric chiller moves heat with a Peltier stack instead of a compressor, so it holds a setpoint without cycling and it does not need refrigerant. The trade is capacity: a 150 to 400 W unit covers a detector, a sample loop or a small laser, and anything above that is a compressor job. The control range of the units below is minus 12 to plus 40 C at the liquid outlet, and stability is quoted at the setpoint rather than over the whole band.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Heater power range | 250 W to 4000 W in the standard offering, in 120 V and 240 V versions | the 240 V part is the same length as the 120 V part with twice the power |
| Water side watt density | up to 450 W per square inch, which is 70 W per square centimetre | only while the tube is full and flowing |
| Air side watt density | up to 150 W per square inch, which is 23 W per square centimetre | the water figure does not transfer to gas duty |
| Heated length | 3.00 to 7.75 inch outer tube, 76 to 197 mm | length sets the surface area and therefore the density |
| Heating zones | one or two circuits per heater | two zones allow a distributed profile along the tube |
| Voltage and current | up to 240 V, up to 15 A per zone | resistance tolerance plus 10 and minus 5 percent |
| Maximum pressure | 150 psi, which is 10.2 bar | the fitting, not the tube, is usually the limit |
| Maximum temperature | 100 C on water, 350 C on gas, as read by the internal thermocouple | liquid duty is limited by boiling at the wall |
| Wall materials | 444 stainless substrate, 316L stainless fittings and baffle | the baffle promotes turbulent flow and raises the transfer rate |
| Mounting | vertical with the inlet at the bottom on liquid duty, any attitude on gas duty | a trapped bubble leaves the element dry |
| Heat needed for a rise | watts equal flow in litres per second times 4.18 times the rise in kelvin times 1000 for water | 10 mL per second and 1000 W gives about 24 K |
| Chiller capacity | 150 W and 300 W in the MRC family, 400 W in the benchtop unit, 140 W in the liquid to air assembly | capacity is quoted at a stated temperature difference |
| Chiller control range | minus 12 to plus 40 C at the liquid outlet | below ambient the loop needs a glycol mix |
| Chiller stability | plus or minus 0.05 C at steady state on the benchtop unit | stability is a setpoint figure, not a range figure |
| Chiller interface | RS232 or RS485 on the recirculating units, 115 or 230 VAC supply | the controller has to speak the same protocol |
Included Models and Order Codes
The heater table is the maker standard offering, so every line is a part that exists at that voltage, wattage and length. Read the density column with the length, because two heaters of the same wattage at different lengths do not behave the same way in the same flow. The chiller table follows, with the recirculating thermoelectric units first and the liquid to air assembly last, since that one cools a loop into room air rather than holding a setpoint on its own.
| Part | Voltage | Power | Heated length | Zones | Watt density | Note |
|---|---|---|---|---|---|---|
| FLUENT 240 V, 500 W | 240 V | 500 W | 3.00 inch, 76 mm | 1 | 210 W per square inch | shortest body in the standard range |
| FLUENT 120 V, 250 W | 120 V | 250 W | 4.25 inch, 108 mm | 1 | 57 W per square inch | lowest density, for a slow flow |
| FLUENT 240 V, 1000 W | 240 V | 1000 W | 4.25 inch, 108 mm | 1 | 228 W per square inch | about 24 K rise at 10 mL per second |
| FLUENT 120 V, 375 W | 120 V | 375 W | 5.25 inch, 133 mm | 1 | 62 W per square inch | long body, low density |
| FLUENT 240 V, 1500 W | 240 V | 1500 W | 5.25 inch, 133 mm | 1 | 247 W per square inch | mid range single zone |
| FLUENT 120 V, 500 W | 120 V | 500 W | 6.50 inch, 165 mm | 1 | 63 W per square inch | long low density body |
| FLUENT 240 V, 2000 W | 240 V | 2000 W | 6.50 inch, 165 mm | 1 | 250 W per square inch | single zone, high power |
| FLUENT 120 V, 750 W | 120 V | 750 W | 6.50 inch, 165 mm | 2 | 94 W per square inch | two zones on a 120 V supply |
| FLUENT 240 V, 3000 W | 240 V | 3000 W | 6.50 inch, 165 mm | 2 | 375 W per square inch | highest density at this length |
| FLUENT 120 V, 1000 W | 120 V | 1000 W | 7.75 inch, 197 mm | 2 | 100 W per square inch | longest 120 V body |
| FLUENT 240 V, 4000 W | 240 V | 4000 W | 7.75 inch, 197 mm | 2 | 400 W per square inch | top of the standard range |
| MRC150-DH2-DV | 115 or 230 VAC | 150 W cooling | closed loop | 1 | not applicable | RS232, cooling only |
| MRC150-DH2-HT-DV | 115 or 230 VAC | 150 W cooling | closed loop | 1 | not applicable | RS232, cooling and heating |
| MRC300-DH2-DV | 115 or 230 VAC | 300 W cooling | closed loop | 1 | not applicable | RS232, cooling only |
| MRC300-DH4-HT-DV | 115 or 230 VAC | 300 W cooling | closed loop | 1 | not applicable | RS485, cooling and heating |
| NRC400 | benchtop, mains | 400 W cooling | closed loop | 1 | not applicable | plus or minus 0.05 C at steady state, about 21 kg |
| SuperCool liquid to air | 24 VDC | 140 W at zero temperature difference | liquid heat exchanger | 1 | not applicable | 230 x 150 x 109 mm, cools a loop into room air |
Sizing a Heater or a Chiller from the Flow
The two decisions are made from the same two numbers, the flow and the temperature change wanted, and the table is the short form of both.
| Duty | First number to fix | Then check | What fails otherwise |
|---|---|---|---|
| Heat water 10 mL per second by 30 K | about 1250 W of useful power | watt density at the length chosen, so the wall stays wet | a short body at high density trips its high limit |
| Heat a slow flow of 1 mL per second by 50 K | about 210 W of useful power | a low density long body, because the flow cannot carry heat away | a 3 inch 500 W body boils the film at the wall |
| Heat a gas stream | the mass flow and the specific heat of the gas | the air side density of 23 W per square centimetre and the baffle | a water rated density destroys the element |
| Hold a detector at 15 C in a 25 C room | the heat load and 10 K of lift | chiller capacity at that lift and the pump pressure drop | the loop settles above setpoint and never reaches it |
| Hold a sample loop at 4 C | the heat load plus the lift from ambient | glycol mix, insulation and condensation control | the setpoint is reachable on paper and not in the room |
Key Features
Layered heater on a thin wall – the circuit is sprayed onto a 444 stainless substrate, so the mass is low and the response is measured in seconds rather than minutes
Density quoted for water and for air – 450 W per square inch on water and 150 on air, which stops a water figure being used on a gas line
One or two heating zones – a distributed profile along the tube, useful where the inlet is cold and the outlet must not overshoot
High limit thermocouple inside – the control loop reads the wall temperature inside the heated section, not a sensor bolted to the outside
Vertical mounting on liquid duty – inlet at the bottom so a bubble leaves the tube instead of sitting on the element
Thermoelectric cooling without refrigerant – a Peltier stack holds a setpoint without compressor cycling and without a refrigerant charge
Control range from minus 12 to plus 40 C – the recirculating units cover chilled sample loops and warm rinse lines with one body
Serial interface on the chiller – RS232 or RS485 lets the instrument controller read and set the loop instead of running it open loop
Applications
These parts sit where a liquid has to arrive at a temperature, either hot for a reaction or cold for a detector. The three cases below are the ones that fix the specification.
Reagent and sample preheating – a colorimetric or enzymatic method runs at 37 C and the sample arrives at room temperature. A 250 to 1000 W heater on a low flow line gives the rise without a hot spot, and the wall thermocouple is what the controller regulates.
Detector and laser temperature control – a photometric detector or a diode laser drifts with case temperature. A thermoelectric chiller holds the loop within a fraction of a degree, and the serial interface lets the instrument log the setpoint alongside the reading.
Rinse and wash stations – wash water has to arrive hot enough to dissolve a residue and cool enough not to distort a cell. A two zone heater allows the inlet section to take the bulk of the power and the outlet section to trim it.
The control boards that drive these heaters and chillers are the sort of assembly SUPERB PCB and SUPERB PCBA programmes build, from the thermocouple input to the solid state output.
Sourcing and Cross-Reference
A heater is ordered by voltage, wattage, length and fitting, and the fitting is the part that goes wrong most often on a replacement. We confirm the inlet and outlet fitting and the lead length before naming a part, because a 1/4 inch tube fitting and a 1/4 inch NPT fitting are not the same connection.
For a chiller, the loop is the deciding factor. The coolant, the pump pressure drop, the hose bore and the heat load all change the capacity that reaches the setpoint, and a unit chosen from its nameplate alone usually settles a few degrees high.
Where the original heater or chiller is discontinued, the electrical rating is the constraint. Resistance tolerance and voltage have to match the controller that drives it, and we state plainly when a substitute needs a different control output rather than quietly shipping a part that fits the thread.
A heater with its fitting, lead and thermocouple is shipped as one set when a line is down, which is how SUPERB Assembly handles a replacement.
Q: Can you match a heater by voltage, wattage and fitting when the original part number is no longer listed?
A: SUPERB Tech can confirm voltage, wattage and fitting against the old part before a substitute is quoted.
Why Buy Heaters and Chillers from SUPERB
- Duty read before the part: flow, temperature rise and the liquid are asked for first, because wattage alone does not size a heater
- Density and length quoted together: the same wattage at two lengths behaves differently, so both numbers appear on every line
- Fittings and leads confirmed: tube fittings, NPT fittings and lead length are checked against the assembly before an order is released
- Chiller capacity at the working lift: capacity is quoted at a stated temperature difference, and the loop pressure drop is part of the question
- Documentation with the shipment: maker data sheet and the electrical rating travel with the part, which is what a validation file needs
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Thermoelectric Coolers and Peltier Modules for Precision Cooling, Medical Isolation Transformers for Diagnostic and Laboratory Equipment, Turbidity and Photometric Detector Modules for Water Analysis, Peristaltic and Diaphragm Micro Pumps for Analysers, Gas Mass Flow Sensors for Ventilators and Analysers, Passive Components, BOM Shortage Recovery.