Inline Fluid Heaters and Peltier Chillers for Analysers

Heating a flowing liquid is a race between the power going in and the heat the tube takes out, and the number that settles it is watt density, not wattage. The parts below cover layered inline heaters from 250 W to 4 kW with water side densities up to 70 W per square centimetre, and thermoelectric chillers that hold a closed loop between minus 12 and plus 40 C. Each entry gives voltage, wattage, heated length, the number of zones and the density at that length, so a heater is chosen from the flow and the temperature rise rather than from the label. Chillers are listed with cooling capacity, control range and the interface, because a chiller that cannot be read by the instrument controller is a chiller that runs open loop.
quote now

Product Specifications

Product Overview

An inline heater adds energy to a moving stream, so its useful output is not the wattage on the label but the temperature rise it can hold at the flow the method needs. The relation is simple: watts equal flow in litres per second times 4.18 times the rise in kelvin, times about 1000 for water. A 1000 W heater raises water flowing at 10 mL per second by about 24 K, and the same heater on a 30 mL per second line raises it by 8 K. That is why a heater is chosen from the flow first and from the wattage second.

Watt density is the second number, and it is the one that decides whether the heater survives. Power has to cross the wall into the liquid, and if the liquid is not moving fast enough the wall runs hotter than the fluid. A layered heater on a thin stainless substrate reaches 450 W per square inch on water, which is 70 W per square centimetre, and that figure only holds while the tube is full and flowing. Run the same heater dry and the wall temperature runs away in seconds, which is why the water rating and the air rating differ by a factor of three.

The standard parts carry their own high limit thermocouple in the heated section, and that sensor is not a substitute for a flow switch. The control loop reads the wall, the flow switch proves the liquid is there, and the two together are what keep a dry start from turning into a repair. On liquid duty the heater also has to be mounted vertically with the inlet at the bottom, so that a bubble leaves the tube instead of sitting on the element.

Cooling below ambient is a different problem, because there is nothing to burn and everything to gain from stability. A thermoelectric chiller moves heat with a Peltier stack instead of a compressor, so it holds a setpoint without cycling and it does not need refrigerant. The trade is capacity: a 150 to 400 W unit covers a detector, a sample loop or a small laser, and anything above that is a compressor job. The control range of the units below is minus 12 to plus 40 C at the liquid outlet, and stability is quoted at the setpoint rather than over the whole band.

Key Specifications

ParameterValueNote
Heater power range250 W to 4000 W in the standard offering, in 120 V and 240 V versionsthe 240 V part is the same length as the 120 V part with twice the power
Water side watt densityup to 450 W per square inch, which is 70 W per square centimetreonly while the tube is full and flowing
Air side watt densityup to 150 W per square inch, which is 23 W per square centimetrethe water figure does not transfer to gas duty
Heated length3.00 to 7.75 inch outer tube, 76 to 197 mmlength sets the surface area and therefore the density
Heating zonesone or two circuits per heatertwo zones allow a distributed profile along the tube
Voltage and currentup to 240 V, up to 15 A per zoneresistance tolerance plus 10 and minus 5 percent
Maximum pressure150 psi, which is 10.2 barthe fitting, not the tube, is usually the limit
Maximum temperature100 C on water, 350 C on gas, as read by the internal thermocoupleliquid duty is limited by boiling at the wall
Wall materials444 stainless substrate, 316L stainless fittings and bafflethe baffle promotes turbulent flow and raises the transfer rate
Mountingvertical with the inlet at the bottom on liquid duty, any attitude on gas dutya trapped bubble leaves the element dry
Heat needed for a risewatts equal flow in litres per second times 4.18 times the rise in kelvin times 1000 for water10 mL per second and 1000 W gives about 24 K
Chiller capacity150 W and 300 W in the MRC family, 400 W in the benchtop unit, 140 W in the liquid to air assemblycapacity is quoted at a stated temperature difference
Chiller control rangeminus 12 to plus 40 C at the liquid outletbelow ambient the loop needs a glycol mix
Chiller stabilityplus or minus 0.05 C at steady state on the benchtop unitstability is a setpoint figure, not a range figure
Chiller interfaceRS232 or RS485 on the recirculating units, 115 or 230 VAC supplythe controller has to speak the same protocol

Included Models and Order Codes

The heater table is the maker standard offering, so every line is a part that exists at that voltage, wattage and length. Read the density column with the length, because two heaters of the same wattage at different lengths do not behave the same way in the same flow. The chiller table follows, with the recirculating thermoelectric units first and the liquid to air assembly last, since that one cools a loop into room air rather than holding a setpoint on its own.

PartVoltagePowerHeated lengthZonesWatt densityNote
FLUENT 240 V, 500 W240 V500 W3.00 inch, 76 mm1210 W per square inchshortest body in the standard range
FLUENT 120 V, 250 W120 V250 W4.25 inch, 108 mm157 W per square inchlowest density, for a slow flow
FLUENT 240 V, 1000 W240 V1000 W4.25 inch, 108 mm1228 W per square inchabout 24 K rise at 10 mL per second
FLUENT 120 V, 375 W120 V375 W5.25 inch, 133 mm162 W per square inchlong body, low density
FLUENT 240 V, 1500 W240 V1500 W5.25 inch, 133 mm1247 W per square inchmid range single zone
FLUENT 120 V, 500 W120 V500 W6.50 inch, 165 mm163 W per square inchlong low density body
FLUENT 240 V, 2000 W240 V2000 W6.50 inch, 165 mm1250 W per square inchsingle zone, high power
FLUENT 120 V, 750 W120 V750 W6.50 inch, 165 mm294 W per square inchtwo zones on a 120 V supply
FLUENT 240 V, 3000 W240 V3000 W6.50 inch, 165 mm2375 W per square inchhighest density at this length
FLUENT 120 V, 1000 W120 V1000 W7.75 inch, 197 mm2100 W per square inchlongest 120 V body
FLUENT 240 V, 4000 W240 V4000 W7.75 inch, 197 mm2400 W per square inchtop of the standard range
MRC150-DH2-DV115 or 230 VAC150 W coolingclosed loop1not applicableRS232, cooling only
MRC150-DH2-HT-DV115 or 230 VAC150 W coolingclosed loop1not applicableRS232, cooling and heating
MRC300-DH2-DV115 or 230 VAC300 W coolingclosed loop1not applicableRS232, cooling only
MRC300-DH4-HT-DV115 or 230 VAC300 W coolingclosed loop1not applicableRS485, cooling and heating
NRC400benchtop, mains400 W coolingclosed loop1not applicableplus or minus 0.05 C at steady state, about 21 kg
SuperCool liquid to air24 VDC140 W at zero temperature differenceliquid heat exchanger1not applicable230 x 150 x 109 mm, cools a loop into room air

Sizing a Heater or a Chiller from the Flow

The two decisions are made from the same two numbers, the flow and the temperature change wanted, and the table is the short form of both.

DutyFirst number to fixThen checkWhat fails otherwise
Heat water 10 mL per second by 30 Kabout 1250 W of useful powerwatt density at the length chosen, so the wall stays weta short body at high density trips its high limit
Heat a slow flow of 1 mL per second by 50 Kabout 210 W of useful powera low density long body, because the flow cannot carry heat awaya 3 inch 500 W body boils the film at the wall
Heat a gas streamthe mass flow and the specific heat of the gasthe air side density of 23 W per square centimetre and the bafflea water rated density destroys the element
Hold a detector at 15 C in a 25 C roomthe heat load and 10 K of liftchiller capacity at that lift and the pump pressure dropthe loop settles above setpoint and never reaches it
Hold a sample loop at 4 Cthe heat load plus the lift from ambientglycol mix, insulation and condensation controlthe setpoint is reachable on paper and not in the room

Key Features

Layered heater on a thin wall – the circuit is sprayed onto a 444 stainless substrate, so the mass is low and the response is measured in seconds rather than minutes

Density quoted for water and for air – 450 W per square inch on water and 150 on air, which stops a water figure being used on a gas line

One or two heating zones – a distributed profile along the tube, useful where the inlet is cold and the outlet must not overshoot

High limit thermocouple inside – the control loop reads the wall temperature inside the heated section, not a sensor bolted to the outside

Vertical mounting on liquid duty – inlet at the bottom so a bubble leaves the tube instead of sitting on the element

Thermoelectric cooling without refrigerant – a Peltier stack holds a setpoint without compressor cycling and without a refrigerant charge

Control range from minus 12 to plus 40 C – the recirculating units cover chilled sample loops and warm rinse lines with one body

Serial interface on the chiller – RS232 or RS485 lets the instrument controller read and set the loop instead of running it open loop

Applications

These parts sit where a liquid has to arrive at a temperature, either hot for a reaction or cold for a detector. The three cases below are the ones that fix the specification.

Reagent and sample preheating – a colorimetric or enzymatic method runs at 37 C and the sample arrives at room temperature. A 250 to 1000 W heater on a low flow line gives the rise without a hot spot, and the wall thermocouple is what the controller regulates.

Detector and laser temperature control – a photometric detector or a diode laser drifts with case temperature. A thermoelectric chiller holds the loop within a fraction of a degree, and the serial interface lets the instrument log the setpoint alongside the reading.

Rinse and wash stations – wash water has to arrive hot enough to dissolve a residue and cool enough not to distort a cell. A two zone heater allows the inlet section to take the bulk of the power and the outlet section to trim it.

The control boards that drive these heaters and chillers are the sort of assembly SUPERB PCB and SUPERB PCBA programmes build, from the thermocouple input to the solid state output.

Sourcing and Cross-Reference

A heater is ordered by voltage, wattage, length and fitting, and the fitting is the part that goes wrong most often on a replacement. We confirm the inlet and outlet fitting and the lead length before naming a part, because a 1/4 inch tube fitting and a 1/4 inch NPT fitting are not the same connection.

For a chiller, the loop is the deciding factor. The coolant, the pump pressure drop, the hose bore and the heat load all change the capacity that reaches the setpoint, and a unit chosen from its nameplate alone usually settles a few degrees high.

Where the original heater or chiller is discontinued, the electrical rating is the constraint. Resistance tolerance and voltage have to match the controller that drives it, and we state plainly when a substitute needs a different control output rather than quietly shipping a part that fits the thread.

A heater with its fitting, lead and thermocouple is shipped as one set when a line is down, which is how SUPERB Assembly handles a replacement.

Q: Can you match a heater by voltage, wattage and fitting when the original part number is no longer listed?

A: SUPERB Tech can confirm voltage, wattage and fitting against the old part before a substitute is quoted.

Why Buy Heaters and Chillers from SUPERB

  1. Duty read before the part: flow, temperature rise and the liquid are asked for first, because wattage alone does not size a heater
  2. Density and length quoted together: the same wattage at two lengths behaves differently, so both numbers appear on every line
  3. Fittings and leads confirmed: tube fittings, NPT fittings and lead length are checked against the assembly before an order is released
  4. Chiller capacity at the working lift: capacity is quoted at a stated temperature difference, and the loop pressure drop is part of the question
  5. Documentation with the shipment: maker data sheet and the electrical rating travel with the part, which is what a validation file needs

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Thermoelectric Coolers and Peltier Modules for Precision Cooling, Medical Isolation Transformers for Diagnostic and Laboratory Equipment, Turbidity and Photometric Detector Modules for Water Analysis, Peristaltic and Diaphragm Micro Pumps for Analysers, Gas Mass Flow Sensors for Ventilators and Analysers, Passive Components, BOM Shortage Recovery.