Rotary and Injection Valves for Analyser Sample Paths
Product Specifications
Product Overview
An injection valve does one thing: it traps a known volume of sample in a loop and then puts that loop into the mobile phase stream. The volume is set by the loop, and the loop is not exact. The maker quotes a tolerance of about 10 percent on a 20 microlitre loop and about 5 percent on a 10 millilitre loop, so a method that needs a precise volume either uses partial filling with a syringe or accepts the loop tolerance. Complete filling, where two to five loop volumes are pushed through, gives the best repeatability because the loop is overfilled and the excess goes to waste.
The switching action is the second specification. A make before break valve connects the new path before it disconnects the old one, so the flow is never interrupted and the pump never sees a dead end. That is what allows a manual injector to be turned while a pump is running at pressure, and it is also why the valve has a maximum pressure instead of a burst pressure. Above the rated pressure the rotor seal extrudes and the valve starts to leak at the stator face.
A multiposition selector valve answers a different question. Instead of trapping a volume, it chooses which of several lines is connected to a common port, so a single pump or detector can be switched between reagents, columns or sample vials. Port counts run from 4 to 16, the bore is chosen from the flow, and the pressure rating drops as the bore grows, because the same seal area carries more force at a larger diameter.
The material of the stator and the rotor decides what the valve can carry. A PAEK stator with a Valcon E rotor covers most aqueous and dilute acid work up to 50 C, and a stainless body with a PTFE based rotor covers higher temperature at a lower pressure. Where the sample is aggressive, the wetted path is the limit rather than the pressure rating, and the valve is chosen from the chemical list before the port count.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Injection valve ports | 6 ports, 2 positions | load and inject, with the loop between two of them |
| Standard loop volume | 20 microlitres on the analytical valve | a 2 microlitre internal loop is available for the same body |
| Loop tolerance | plus or minus 10 percent on a 20 microlitre loop, plus or minus 5 percent on a 10 millilitre loop | a nominal loop is not a delivered volume |
| Filling method | partial filling with a syringe, or complete filling with 2 to 5 loop volumes | complete filling gives the better repeatability |
| Switching action | make before break | the flow is not interrupted while the handle turns |
| Maximum pressure | 345 bar (5000 psi) on the analytical valve, 483 bar (7000 psi) on the high pressure body | the rotor seal is the part that limits it |
| Maximum temperature | 80 C on the analytical valve, 50 C on the high temperature and microvolume bodies | the rotor material sets the limit |
| Fitting | 10-32 coned ports for 1/16 inch outside diameter tubing | the ferrule and the port are a matched pair |
| Internal passages | 0.3 mm and 0.5 mm on the microvolume body | a smaller passage keeps the swept volume down |
| Selector port counts | 4, 6, 8, 10, 12 and 16 positions on the 1/16 inch family | the count follows the number of lines to switch |
| Selector bore | 0.75 mm on the dead end and trapping configurations, 1.0 mm on the common outlet and flow through | bore follows the flow, not the pressure |
| Selector pressure | 400 psi gas on the 1/16 inch dead end body, 200 psi on the common outlet and flow through bodies | these are gas ratings, liquid ratings differ |
| Selector temperature | 200 C on the 1/16 inch family | the rotor, not the body, is what ages |
| Selector stator materials | PAEK composite with a Valcon E rotor, or stainless steel with a PTFE based rotor | the chemical list decides which |
Included Models and Order Codes
The injection valves come first because they set the sample volume, and the selector valves follow because they set the path. The four digit numbers are the manual analytical bodies: the first one is the standard 5000 psi body with a position switch, the next is the same body at a lower temperature limit, and the high pressure one carries 7000 psi with smaller internal passages. The selector valves are listed by configuration, because the same port count behaves differently depending on whether the ports are dead ended or flow through.
| Part | Type | Ports | Volume or bore | Pressure | Temperature | Note |
|---|---|---|---|---|---|---|
| 7725i | manual injection valve | 6 | 20 microlitre loop, 2 microlitre internal option | 345 bar, 5000 psi | 80 C | make before break, position switch, 10-32 coned |
| 9725i | manual injection valve | 6 | 20 microlitre loop | 345 bar, 5000 psi | 50 C | same body, lower temperature rotor |
| 3725i | manual injection valve, high pressure | 6 | loop as ordered | 483 bar, 7000 psi | 50 C | 0.3 and 0.5 mm passages, lower swept volume |
| 9125 | manual injection valve | 6 | loop as ordered | 345 bar, 5000 psi | 50 C | six port sample injection, front loading |
| VICI C5M, 8 position | selector, Cheminert | 8 | 150 micrometre bore | 345 bar, 5000 psi liquid | 50 C | PAEK stator, Valcon E rotor, 360 micrometre fittings |
| VICI SD, 16 position | selector, dead ended | 16 | 0.75 mm bore | 400 psi gas | 200 C | 1/16 inch fittings, the widest port count |
| VICI SC, 8 position | selector, common outlet | 8 | 1.0 mm bore | 200 psi | 200 C | one common port, eight inlets |
| VICI SF, 8 position | selector, flow through | 8 | 1.0 mm bore | 200 psi | 200 C | all ports on a through path |
| VICI ST, 8 position | selector, trapping | 8 | 0.75 mm bore | 200 psi | 200 C | for trapping a sample on a short column |
Where the Sample Volume Really Comes From
The loop is only part of the volume that reaches the column, and the rest is the passages between the needle port and the loop. The table is what changes the delivered volume.
| Factor | Effect on delivered volume | How to control it | Symptom when it is wrong |
|---|---|---|---|
| Loop nominal size | sets the volume within plus or minus 10 percent on a 20 microlitre loop | use complete filling, or calibrate the loop by weight | peaks vary between valves of the same part number |
| Filling method | partial filling delivers what the syringe pushed, complete filling delivers the loop | pick one method and keep it for the method | a method transferred between labs gives a different peak height |
| Passage volume | adds a small dead volume between the needle port and the loop | use the microvolume body where the loop is under 5 microlitres | a small peak tails because the dead volume mixes |
| Sample viscosity | a viscous sample fills the loop slowly and can leave a bubble | pull the sample through with two loop volumes and pause | the peak is low and repeats badly |
| Pressure at the moment of switching | a loop above the rotor rating extrudes the seal | keep the loop pressure below the rated figure before turning | a leak appears at the stator face after a few injections |
Key Features
Make before break switching – the new path connects before the old one opens, so the pump never sees a dead end and the flow stays continuous
Position switch on the handle – the injector sends a start signal when the rotor reaches the inject position, so the data system gets a repeatable time zero
Front loading needle port – the syringe connects directly to the loop, so the sample does not pass through a transfer passage on the way in
20 microlitre standard loop – the loop that most analytical methods are written around, with a 2 microlitre internal option for small volumes
Microvolume body at 0.3 mm – smaller passages keep the swept volume down where a 2 microlitre loop is used
Selector valves from 4 to 16 ports – one body switches reagents, columns or sample lines without replumbing
PAEK stator and Valcon E rotor – the pair that covers aqueous, dilute acid and dilute base work to 50 C at high pressure
Dead end, common outlet, flow through and trapping – four flow configurations on the same port count, chosen from how the lines are used
Applications
These valves sit between the sample and the separation or the detector. The three cases below decide the port count and the pressure rating.
Liquid chromatography sample injection – a fixed volume of sample has to enter a pressurised stream without stopping the pump. The six port body with a 20 microlitre loop and make before break switching is the standard answer, and the position switch gives the data system its start time.
Analyser reagent and wash switching – one detector has to see several reagents and a wash line in sequence. A multiposition selector valve with a common outlet does the switching, and the port count follows the number of lines rather than the flow.
Automated sample changers – a rack of vials is addressed one at a time. A dead ended selector with 8 or 16 ports connects the sampler to the pump, and the bore is chosen from the flow the pump needs.
The controller boards that drive these valves are the kind of assembly SUPERB PCB and SUPERB PCBA programmes build, from the position switch input to the stepper output.
Sourcing and Cross-Reference
A valve is ordered by port count, bore and pressure, and then by the rotor material, which is the part that decides chemical compatibility. Two valves with the same body and different rotors are not interchangeable in a solvent line, and the difference only shows after a few hundred injections.
Rotor seals and stator faces are consumables on this family. We supply the seal as a separate line with the valve, because a valve that is out of service for a seal is a line that is down for a part that costs a fraction of the body.
Where an older injector is no longer listed, the loop volume and the fitting are the two things to carry across. A new body with the same 10-32 ports and the same loop keeps the method, and a body with a different swept volume changes the peak shape even when the loop is identical.
A valve with its loop, rotor seal and fittings is shipped as one set when a method is being transferred, which is how SUPERB Assembly handles it.
Q: Can you supply a replacement injection valve that keeps the same loop volume and 10-32 fittings?
A: SUPERB Tech can confirm port count, loop volume and fitting against the valve already in the instrument before quoting.
Why Buy Valves from SUPERB
- Loop volume and tolerance stated: the delivered volume and the filling method are quoted with the part, not left to the label
- Pressure and temperature as one rating: the rotor material sets both, and both appear on every line
- Rotor and stator available as spares: the consumable is quoted with the body so a seal change does not become a valve order
- Bore chosen from the flow: selector valves are matched to the line rather than to the highest pressure on the data sheet
- Wetted materials listed: PAEK, Valcon E and stainless options are named, so the valve can be checked against the solvent list
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Peristaltic and Diaphragm Micro Pumps for Analysers, Syringe Pumps and Precision Dispensers for Analysers, Solenoid and Pinch Valves for Fluid Handling in Analysers, Fluid Check Valves and Shuttle Valves for Reagent and Pneumatic Lines, Inline Degassing Modules and Vacuum Degassers, Sensors and Detectors, BOM Shortage Recovery.