Rotary and Injection Valves for Analyser Sample Paths

A sample path is only as repeatable as the valve that switches it, and the two numbers that decide that are the internal volume and the pressure rating. The parts below cover six port injection valves with 20 microlitre loops rated to 345 bar, and multiposition selector valves from 4 to 16 ports. Each entry gives port count, bore or loop volume, maximum pressure, maximum temperature and the fitting, so the valve is matched to the pump and to the tubing. Loop volume tolerance and the filling method are stated as well, because a nominal 20 microlitre loop does not deliver 20 microlitres.
quote now

Product Specifications

Product Overview

An injection valve does one thing: it traps a known volume of sample in a loop and then puts that loop into the mobile phase stream. The volume is set by the loop, and the loop is not exact. The maker quotes a tolerance of about 10 percent on a 20 microlitre loop and about 5 percent on a 10 millilitre loop, so a method that needs a precise volume either uses partial filling with a syringe or accepts the loop tolerance. Complete filling, where two to five loop volumes are pushed through, gives the best repeatability because the loop is overfilled and the excess goes to waste.

The switching action is the second specification. A make before break valve connects the new path before it disconnects the old one, so the flow is never interrupted and the pump never sees a dead end. That is what allows a manual injector to be turned while a pump is running at pressure, and it is also why the valve has a maximum pressure instead of a burst pressure. Above the rated pressure the rotor seal extrudes and the valve starts to leak at the stator face.

A multiposition selector valve answers a different question. Instead of trapping a volume, it chooses which of several lines is connected to a common port, so a single pump or detector can be switched between reagents, columns or sample vials. Port counts run from 4 to 16, the bore is chosen from the flow, and the pressure rating drops as the bore grows, because the same seal area carries more force at a larger diameter.

The material of the stator and the rotor decides what the valve can carry. A PAEK stator with a Valcon E rotor covers most aqueous and dilute acid work up to 50 C, and a stainless body with a PTFE based rotor covers higher temperature at a lower pressure. Where the sample is aggressive, the wetted path is the limit rather than the pressure rating, and the valve is chosen from the chemical list before the port count.

Key Specifications

ParameterValueNote
Injection valve ports6 ports, 2 positionsload and inject, with the loop between two of them
Standard loop volume20 microlitres on the analytical valvea 2 microlitre internal loop is available for the same body
Loop toleranceplus or minus 10 percent on a 20 microlitre loop, plus or minus 5 percent on a 10 millilitre loopa nominal loop is not a delivered volume
Filling methodpartial filling with a syringe, or complete filling with 2 to 5 loop volumescomplete filling gives the better repeatability
Switching actionmake before breakthe flow is not interrupted while the handle turns
Maximum pressure345 bar (5000 psi) on the analytical valve, 483 bar (7000 psi) on the high pressure bodythe rotor seal is the part that limits it
Maximum temperature80 C on the analytical valve, 50 C on the high temperature and microvolume bodiesthe rotor material sets the limit
Fitting10-32 coned ports for 1/16 inch outside diameter tubingthe ferrule and the port are a matched pair
Internal passages0.3 mm and 0.5 mm on the microvolume bodya smaller passage keeps the swept volume down
Selector port counts4, 6, 8, 10, 12 and 16 positions on the 1/16 inch familythe count follows the number of lines to switch
Selector bore0.75 mm on the dead end and trapping configurations, 1.0 mm on the common outlet and flow throughbore follows the flow, not the pressure
Selector pressure400 psi gas on the 1/16 inch dead end body, 200 psi on the common outlet and flow through bodiesthese are gas ratings, liquid ratings differ
Selector temperature200 C on the 1/16 inch familythe rotor, not the body, is what ages
Selector stator materialsPAEK composite with a Valcon E rotor, or stainless steel with a PTFE based rotorthe chemical list decides which

Included Models and Order Codes

The injection valves come first because they set the sample volume, and the selector valves follow because they set the path. The four digit numbers are the manual analytical bodies: the first one is the standard 5000 psi body with a position switch, the next is the same body at a lower temperature limit, and the high pressure one carries 7000 psi with smaller internal passages. The selector valves are listed by configuration, because the same port count behaves differently depending on whether the ports are dead ended or flow through.

PartTypePortsVolume or borePressureTemperatureNote
7725imanual injection valve620 microlitre loop, 2 microlitre internal option345 bar, 5000 psi80 Cmake before break, position switch, 10-32 coned
9725imanual injection valve620 microlitre loop345 bar, 5000 psi50 Csame body, lower temperature rotor
3725imanual injection valve, high pressure6loop as ordered483 bar, 7000 psi50 C0.3 and 0.5 mm passages, lower swept volume
9125manual injection valve6loop as ordered345 bar, 5000 psi50 Csix port sample injection, front loading
VICI C5M, 8 positionselector, Cheminert8150 micrometre bore345 bar, 5000 psi liquid50 CPAEK stator, Valcon E rotor, 360 micrometre fittings
VICI SD, 16 positionselector, dead ended160.75 mm bore400 psi gas200 C1/16 inch fittings, the widest port count
VICI SC, 8 positionselector, common outlet81.0 mm bore200 psi200 Cone common port, eight inlets
VICI SF, 8 positionselector, flow through81.0 mm bore200 psi200 Call ports on a through path
VICI ST, 8 positionselector, trapping80.75 mm bore200 psi200 Cfor trapping a sample on a short column

Where the Sample Volume Really Comes From

The loop is only part of the volume that reaches the column, and the rest is the passages between the needle port and the loop. The table is what changes the delivered volume.

FactorEffect on delivered volumeHow to control itSymptom when it is wrong
Loop nominal sizesets the volume within plus or minus 10 percent on a 20 microlitre loopuse complete filling, or calibrate the loop by weightpeaks vary between valves of the same part number
Filling methodpartial filling delivers what the syringe pushed, complete filling delivers the looppick one method and keep it for the methoda method transferred between labs gives a different peak height
Passage volumeadds a small dead volume between the needle port and the loopuse the microvolume body where the loop is under 5 microlitresa small peak tails because the dead volume mixes
Sample viscositya viscous sample fills the loop slowly and can leave a bubblepull the sample through with two loop volumes and pausethe peak is low and repeats badly
Pressure at the moment of switchinga loop above the rotor rating extrudes the sealkeep the loop pressure below the rated figure before turninga leak appears at the stator face after a few injections

Key Features

Make before break switching – the new path connects before the old one opens, so the pump never sees a dead end and the flow stays continuous

Position switch on the handle – the injector sends a start signal when the rotor reaches the inject position, so the data system gets a repeatable time zero

Front loading needle port – the syringe connects directly to the loop, so the sample does not pass through a transfer passage on the way in

20 microlitre standard loop – the loop that most analytical methods are written around, with a 2 microlitre internal option for small volumes

Microvolume body at 0.3 mm – smaller passages keep the swept volume down where a 2 microlitre loop is used

Selector valves from 4 to 16 ports – one body switches reagents, columns or sample lines without replumbing

PAEK stator and Valcon E rotor – the pair that covers aqueous, dilute acid and dilute base work to 50 C at high pressure

Dead end, common outlet, flow through and trapping – four flow configurations on the same port count, chosen from how the lines are used

Applications

These valves sit between the sample and the separation or the detector. The three cases below decide the port count and the pressure rating.

Liquid chromatography sample injection – a fixed volume of sample has to enter a pressurised stream without stopping the pump. The six port body with a 20 microlitre loop and make before break switching is the standard answer, and the position switch gives the data system its start time.

Analyser reagent and wash switching – one detector has to see several reagents and a wash line in sequence. A multiposition selector valve with a common outlet does the switching, and the port count follows the number of lines rather than the flow.

Automated sample changers – a rack of vials is addressed one at a time. A dead ended selector with 8 or 16 ports connects the sampler to the pump, and the bore is chosen from the flow the pump needs.

The controller boards that drive these valves are the kind of assembly SUPERB PCB and SUPERB PCBA programmes build, from the position switch input to the stepper output.

Sourcing and Cross-Reference

A valve is ordered by port count, bore and pressure, and then by the rotor material, which is the part that decides chemical compatibility. Two valves with the same body and different rotors are not interchangeable in a solvent line, and the difference only shows after a few hundred injections.

Rotor seals and stator faces are consumables on this family. We supply the seal as a separate line with the valve, because a valve that is out of service for a seal is a line that is down for a part that costs a fraction of the body.

Where an older injector is no longer listed, the loop volume and the fitting are the two things to carry across. A new body with the same 10-32 ports and the same loop keeps the method, and a body with a different swept volume changes the peak shape even when the loop is identical.

A valve with its loop, rotor seal and fittings is shipped as one set when a method is being transferred, which is how SUPERB Assembly handles it.

Q: Can you supply a replacement injection valve that keeps the same loop volume and 10-32 fittings?

A: SUPERB Tech can confirm port count, loop volume and fitting against the valve already in the instrument before quoting.

Why Buy Valves from SUPERB

  1. Loop volume and tolerance stated: the delivered volume and the filling method are quoted with the part, not left to the label
  2. Pressure and temperature as one rating: the rotor material sets both, and both appear on every line
  3. Rotor and stator available as spares: the consumable is quoted with the body so a seal change does not become a valve order
  4. Bore chosen from the flow: selector valves are matched to the line rather than to the highest pressure on the data sheet
  5. Wetted materials listed: PAEK, Valcon E and stainless options are named, so the valve can be checked against the solvent list

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Peristaltic and Diaphragm Micro Pumps for Analysers, Syringe Pumps and Precision Dispensers for Analysers, Solenoid and Pinch Valves for Fluid Handling in Analysers, Fluid Check Valves and Shuttle Valves for Reagent and Pneumatic Lines, Inline Degassing Modules and Vacuum Degassers, Sensors and Detectors, BOM Shortage Recovery.