Inline Degassing Modules and Vacuum Degassers for Liquid Lines
Product Specifications
Product Overview
A degasser is a tube of membrane inside a vacuum chamber. Liquid flows through the membrane, the gas partial pressure outside is held low by a pump, and dissolved gas crosses the membrane into the vacuum. The driving force is the difference between the dissolved concentration and the near zero concentration on the other side, so the removal rate depends on the membrane area, the flow and the vacuum level, and not on the pressure of the liquid.
The number that sizes the chamber is the flow at which half of the dissolved oxygen is removed. A chamber rated at 20 millilitres per minute at 50 percent removal will remove less than half at 40 millilitres per minute, because the liquid spends less time against the membrane. A method that needs 90 percent removal needs a flow well below the 50 percent figure, and the curve in the data sheet is the only honest way to read that.
The vacuum side has its own limits. The recommended level is around 16.7 kilopascals absolute, and the pumpdown time tells how long the chamber takes to reach it. A large chamber takes hundreds of seconds to pump down, so a system that cycles the vacuum pump on and off spends much of its time re establishing the level instead of degassing. That is why the small chambers are used in high duty cycle instruments and the large ones in continuous flow.
The membrane and the housing decide chemical compatibility. The membrane is a vinyl methyl silicone, which passes dissolved gases readily and is not a barrier against every solvent. The housing runs from polypropylene and PVC on the high flow chambers to PPS on the small one, and the fluid contacts are listed per part, because a chamber is chosen from the solvent list before it is chosen from the flow.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Removal principle | permeation through a solid silicone membrane into a vacuum | no chemical scavenger and nothing consumed |
| Membrane material | vinyl methyl silicone | passes dissolved gases, not a barrier against every solvent |
| Flow for 50 percent removal | 20 to 800 millilitres per minute depending on the chamber | the figure falls as the flow rises |
| Recommended vacuum | 16.7 kilopascals absolute | the same level on every chamber in the range |
| Maximum pressure difference | 100 kilopascals between fluid and vacuum | the membrane is the part that sees it |
| Maximum fluid temperature | 40 C | above that the membrane ages quickly |
| Maximum humidity | 70 percent non condensing | condensation on the housing is the usual site fault |
| Fluid volume inside the chamber | 2 to 205 millilitres | a large volume costs a longer wash step |
| Pumpdown time | under 10 s on the smallest chamber to 250 s on the largest | a cycling pump spends this time on every restart |
| Connections | 1/4-28 UNF, M6 by 1 female, or 11 mm barb | the fitting follows the tubing already in the instrument |
| Housing materials | polyphenylene sulfide, polypropylene, PVC | chosen from the solvent list |
| Mounting | upright with the connections up, or horizontal with them to the side | a chamber mounted upside down traps gas at the outlet |
| Channels | single channel, or two independent channels in one body | a two channel body degasses two reagents in the same footprint |
Included Models and Order Codes
The chambers are grouped by flow, from a 2 millilitre body for a micro flow to a 205 millilitre body for a high throughput line. The flow quoted is the one at which half the dissolved oxygen is removed, so the working flow is usually below the figure on the part. The two channel body is listed with the others because it changes the plumbing rather than the chemistry: two reagents are degassed in one footprint, with one vacuum connection.
| Part | Flow at 50 percent | Fluid volume | Channels | Connection | Housing | Note |
|---|---|---|---|---|---|---|
| DGC-020-153881 | 20 millilitres per minute, 333 microlitres per second | 2 millilitres | 1 | 1/4-28 UNF or manifold mount | polyphenylene sulfide | pumpdown under 10 s, 75.7 x 49 x 22 mm |
| DGC-080W | 40 millilitres per minute per channel | 8.9 millilitres | 2 | M6 by 1 female | PVC | two independent channels, 145 x 60 x 60 mm |
| ERC-250W | 350 millilitres per minute | 97 millilitres | 1 | 11 mm barb | polypropylene and PVC | maximum flow 500 millilitres per minute, pumpdown 155 s |
| DGC-500W | 500 millilitres per minute | 59 millilitres | 1 | barb | as listed by the maker | pumpdown 10 s, absolute maximum flow 4600 millilitres per minute |
| DGC-600W | 800 millilitres per minute | 205 millilitres | 1 | 4.7 mm barb | sturdy housing for membrane protection | pumpdown 250 s, 985 cubic centimetres of vacuum volume |
Matching a Chamber to a Pump and a Method
The chamber and the pump are one system, and the method sets both. The table is the short version of that match.
| Situation | Chamber to use | Why | What goes wrong otherwise |
|---|---|---|---|
| Micro flow under 50 microlitres per minute | DGC-020 at 2 millilitres of fluid volume | the wash volume is small, so a reagent change clears quickly | a large chamber keeps the old reagent for minutes |
| Two reagents on one instrument | DGC-080W two channel body | one footprint, one vacuum connection, two independent paths | two single chambers need twice the space and two pumps |
| HPLC solvent line at 1 millilitre per minute | ERC-250W or DGC-500W | the 50 percent flow is above the working flow, so removal is well over half | a small chamber passes gas through at that flow |
| High throughput line at 5 millilitres per minute | DGC-600W | the largest membrane area in the range | the pressure drop rises and the pump starves |
| Vacuum pump cycles on and off | the smallest chamber that meets the flow | pumpdown time is lost time on every restart | the vacuum level never settles and removal falls |
Key Features
Fully swept inline path – the fluid path has no dead leg, so a reagent change clears in the chamber volume rather than in a corner
Flow for half removal stated – the figure that actually sizes the chamber, rather than a maximum flow that means little
Vacuum level and pumpdown time – the chamber and the pump are quoted together, so a cycling pump can be assessed
Two channel body – two reagents degassed in one footprint with one vacuum connection
Silicone membrane – dissolved oxygen and nitrogen cross it readily, and the material is listed on every part
Housing material per part – polyphenylene sulfide, polypropylene and PVC are named, so the chamber can be checked against the solvent
Barb, M6 and 1/4-28 connections – the chamber fits the tubing the instrument already uses instead of forcing a new fitting
Mounting orientation stated – upright or horizontal with the connections to the side, which is what stops gas collecting at the outlet
Applications
A degasser sits between the reservoir and the pump, or between the pump and the detector. The three cases below decide the chamber volume and the flow.
Liquid chromatography and mass spectrometry – dissolved air forms bubbles in the detector cell and shows up as a drifting baseline. A chamber on the solvent line removes the gas before the pump, and the fluid volume sets how long a solvent change takes to clear.
Clinical chemistry and immunoassay – a bubble in a dispense line is a short sample. A small two channel chamber degasses the reagent and the wash buffer close to the probe, and the low fluid volume keeps the reagent change fast.
Bubble sensitive detection – an air bubble detector downstream needs the line to be free of entrained gas, not only dissolved gas. The degasser removes the dissolved fraction, and the bubble detector catches what is left, which is why the two are specified together.
The control boards that run the vacuum pump and read the pressure are the kind of assembly SUPERB PCB and SUPERB PCBA programmes build, from the pressure sensor input to the pump drive.
Sourcing and Cross-Reference
A degasser is ordered with its vacuum pump, because the two are a system. A chamber with a pump that cannot hold 16.7 kilopascals removes less than its rating, and the shortfall appears as a noisy baseline rather than as a fault code.
The fluid volume is the part that affects the method. A 205 millilitre chamber on a line that changes solvent twice an hour costs minutes of equilibration each time, and a 2 millilitre chamber on a high flow line passes gas through. The volume and the flow are read together.
Where a chamber is discontinued, the connection and the volume are what have to carry across. A substitute with a different fitting needs new tubing, and one with a larger volume changes the wash time even when the removal rate is the same.
A chamber with its vacuum pump and tubing is shipped as one set when a line is rebuilt, which is how SUPERB Assembly handles it.
Q: Can you supply a degassing chamber with the same fluid volume and connection as the one in the instrument?
A: SUPERB Tech can match fluid volume, connection and vacuum level against the chamber already in the instrument before quoting.
Why Buy Degassing Modules from SUPERB
- Flow at half removal quoted: the number that sizes a chamber, instead of a maximum flow that says little about performance
- Vacuum level and pumpdown time listed: the chamber is assessed with the pump that will run it
- Fluid volume stated: the volume decides the wash time, and it appears on every line
- Connections matched to the instrument: 1/4-28, M6 and barb options are checked before an order
- Housing materials named: the chamber can be checked against the solvent list before it is bought
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Air Bubble Detectors for Infusion and Reagent Lines, Syringe Pumps and Precision Dispensers for Analysers, Peristaltic and Diaphragm Micro Pumps for Analysers, Rotary and Injection Valves for Analyser Sample Paths, Liquid Flow Sensors for Medical and Analytical Instruments, Sensors and Detectors, BOM Shortage Recovery.