Inline Degassing Modules and Vacuum Degassers for Liquid Lines

Dissolved gas leaves a liquid when the pressure drops, so a degasser works by giving the gas somewhere to go before the liquid reaches the detector. The parts below are membrane degassing chambers from 20 to 800 millilitres per minute, with the flow at which each one removes half the dissolved oxygen stated on the line. Each entry gives the flow for 50 percent removal, the fluid volume inside the chamber, the vacuum level and the connections, so a chamber is matched to the pump rather than to the box size. The vacuum side is quoted as a pressure and a volume, because the pump and the chamber are one system and neither works at its rating alone.
quote now

Product Specifications

Product Overview

A degasser is a tube of membrane inside a vacuum chamber. Liquid flows through the membrane, the gas partial pressure outside is held low by a pump, and dissolved gas crosses the membrane into the vacuum. The driving force is the difference between the dissolved concentration and the near zero concentration on the other side, so the removal rate depends on the membrane area, the flow and the vacuum level, and not on the pressure of the liquid.

The number that sizes the chamber is the flow at which half of the dissolved oxygen is removed. A chamber rated at 20 millilitres per minute at 50 percent removal will remove less than half at 40 millilitres per minute, because the liquid spends less time against the membrane. A method that needs 90 percent removal needs a flow well below the 50 percent figure, and the curve in the data sheet is the only honest way to read that.

The vacuum side has its own limits. The recommended level is around 16.7 kilopascals absolute, and the pumpdown time tells how long the chamber takes to reach it. A large chamber takes hundreds of seconds to pump down, so a system that cycles the vacuum pump on and off spends much of its time re establishing the level instead of degassing. That is why the small chambers are used in high duty cycle instruments and the large ones in continuous flow.

The membrane and the housing decide chemical compatibility. The membrane is a vinyl methyl silicone, which passes dissolved gases readily and is not a barrier against every solvent. The housing runs from polypropylene and PVC on the high flow chambers to PPS on the small one, and the fluid contacts are listed per part, because a chamber is chosen from the solvent list before it is chosen from the flow.

Key Specifications

ParameterValueNote
Removal principlepermeation through a solid silicone membrane into a vacuumno chemical scavenger and nothing consumed
Membrane materialvinyl methyl siliconepasses dissolved gases, not a barrier against every solvent
Flow for 50 percent removal20 to 800 millilitres per minute depending on the chamberthe figure falls as the flow rises
Recommended vacuum16.7 kilopascals absolutethe same level on every chamber in the range
Maximum pressure difference100 kilopascals between fluid and vacuumthe membrane is the part that sees it
Maximum fluid temperature40 Cabove that the membrane ages quickly
Maximum humidity70 percent non condensingcondensation on the housing is the usual site fault
Fluid volume inside the chamber2 to 205 millilitresa large volume costs a longer wash step
Pumpdown timeunder 10 s on the smallest chamber to 250 s on the largesta cycling pump spends this time on every restart
Connections1/4-28 UNF, M6 by 1 female, or 11 mm barbthe fitting follows the tubing already in the instrument
Housing materialspolyphenylene sulfide, polypropylene, PVCchosen from the solvent list
Mountingupright with the connections up, or horizontal with them to the sidea chamber mounted upside down traps gas at the outlet
Channelssingle channel, or two independent channels in one bodya two channel body degasses two reagents in the same footprint

Included Models and Order Codes

The chambers are grouped by flow, from a 2 millilitre body for a micro flow to a 205 millilitre body for a high throughput line. The flow quoted is the one at which half the dissolved oxygen is removed, so the working flow is usually below the figure on the part. The two channel body is listed with the others because it changes the plumbing rather than the chemistry: two reagents are degassed in one footprint, with one vacuum connection.

PartFlow at 50 percentFluid volumeChannelsConnectionHousingNote
DGC-020-15388120 millilitres per minute, 333 microlitres per second2 millilitres11/4-28 UNF or manifold mountpolyphenylene sulfidepumpdown under 10 s, 75.7 x 49 x 22 mm
DGC-080W40 millilitres per minute per channel8.9 millilitres2M6 by 1 femalePVCtwo independent channels, 145 x 60 x 60 mm
ERC-250W350 millilitres per minute97 millilitres111 mm barbpolypropylene and PVCmaximum flow 500 millilitres per minute, pumpdown 155 s
DGC-500W500 millilitres per minute59 millilitres1barbas listed by the makerpumpdown 10 s, absolute maximum flow 4600 millilitres per minute
DGC-600W800 millilitres per minute205 millilitres14.7 mm barbsturdy housing for membrane protectionpumpdown 250 s, 985 cubic centimetres of vacuum volume

Matching a Chamber to a Pump and a Method

The chamber and the pump are one system, and the method sets both. The table is the short version of that match.

SituationChamber to useWhyWhat goes wrong otherwise
Micro flow under 50 microlitres per minuteDGC-020 at 2 millilitres of fluid volumethe wash volume is small, so a reagent change clears quicklya large chamber keeps the old reagent for minutes
Two reagents on one instrumentDGC-080W two channel bodyone footprint, one vacuum connection, two independent pathstwo single chambers need twice the space and two pumps
HPLC solvent line at 1 millilitre per minuteERC-250W or DGC-500Wthe 50 percent flow is above the working flow, so removal is well over halfa small chamber passes gas through at that flow
High throughput line at 5 millilitres per minuteDGC-600Wthe largest membrane area in the rangethe pressure drop rises and the pump starves
Vacuum pump cycles on and offthe smallest chamber that meets the flowpumpdown time is lost time on every restartthe vacuum level never settles and removal falls

Key Features

Fully swept inline path – the fluid path has no dead leg, so a reagent change clears in the chamber volume rather than in a corner

Flow for half removal stated – the figure that actually sizes the chamber, rather than a maximum flow that means little

Vacuum level and pumpdown time – the chamber and the pump are quoted together, so a cycling pump can be assessed

Two channel body – two reagents degassed in one footprint with one vacuum connection

Silicone membrane – dissolved oxygen and nitrogen cross it readily, and the material is listed on every part

Housing material per part – polyphenylene sulfide, polypropylene and PVC are named, so the chamber can be checked against the solvent

Barb, M6 and 1/4-28 connections – the chamber fits the tubing the instrument already uses instead of forcing a new fitting

Mounting orientation stated – upright or horizontal with the connections to the side, which is what stops gas collecting at the outlet

Applications

A degasser sits between the reservoir and the pump, or between the pump and the detector. The three cases below decide the chamber volume and the flow.

Liquid chromatography and mass spectrometry – dissolved air forms bubbles in the detector cell and shows up as a drifting baseline. A chamber on the solvent line removes the gas before the pump, and the fluid volume sets how long a solvent change takes to clear.

Clinical chemistry and immunoassay – a bubble in a dispense line is a short sample. A small two channel chamber degasses the reagent and the wash buffer close to the probe, and the low fluid volume keeps the reagent change fast.

Bubble sensitive detection – an air bubble detector downstream needs the line to be free of entrained gas, not only dissolved gas. The degasser removes the dissolved fraction, and the bubble detector catches what is left, which is why the two are specified together.

The control boards that run the vacuum pump and read the pressure are the kind of assembly SUPERB PCB and SUPERB PCBA programmes build, from the pressure sensor input to the pump drive.

Sourcing and Cross-Reference

A degasser is ordered with its vacuum pump, because the two are a system. A chamber with a pump that cannot hold 16.7 kilopascals removes less than its rating, and the shortfall appears as a noisy baseline rather than as a fault code.

The fluid volume is the part that affects the method. A 205 millilitre chamber on a line that changes solvent twice an hour costs minutes of equilibration each time, and a 2 millilitre chamber on a high flow line passes gas through. The volume and the flow are read together.

Where a chamber is discontinued, the connection and the volume are what have to carry across. A substitute with a different fitting needs new tubing, and one with a larger volume changes the wash time even when the removal rate is the same.

A chamber with its vacuum pump and tubing is shipped as one set when a line is rebuilt, which is how SUPERB Assembly handles it.

Q: Can you supply a degassing chamber with the same fluid volume and connection as the one in the instrument?

A: SUPERB Tech can match fluid volume, connection and vacuum level against the chamber already in the instrument before quoting.

Why Buy Degassing Modules from SUPERB

  1. Flow at half removal quoted: the number that sizes a chamber, instead of a maximum flow that says little about performance
  2. Vacuum level and pumpdown time listed: the chamber is assessed with the pump that will run it
  3. Fluid volume stated: the volume decides the wash time, and it appears on every line
  4. Connections matched to the instrument: 1/4-28, M6 and barb options are checked before an order
  5. Housing materials named: the chamber can be checked against the solvent list before it is bought

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Air Bubble Detectors for Infusion and Reagent Lines, Syringe Pumps and Precision Dispensers for Analysers, Peristaltic and Diaphragm Micro Pumps for Analysers, Rotary and Injection Valves for Analyser Sample Paths, Liquid Flow Sensors for Medical and Analytical Instruments, Sensors and Detectors, BOM Shortage Recovery.