Reflectance and Transmission Probes for Spectrometers
Product Specifications
Product Overview
A reflectance probe is two optical paths inside one ferrule. Six fibers carry light down to the sample, and one fiber in the middle carries the returning light to the spectrometer. Because the same spot is lit and read from the same direction, the measurement is stable on a moving web, a painted panel, a tablet or a powder bed, where a cuvette cannot be used at all.
Core size sets the trade between spot size and signal. A 200 um bundle keeps a small spot and holds resolution on a fiber coupled instrument, a 400 um bundle roughly doubles the light for dark or scattering samples, and a 600 um bundle is for very low light levels and short working distances. The fiber type then decides the range: a solarization resistant fiber for 200 to 1100 nm, a high OH fiber for 300 to 1100 nm, a low OH fiber for 400 to 2500 nm. One fiber cannot cover the whole span well, which is why mixed 12 fiber bundles put both ranges in the same ferrule.
A transmission probe answers a different question. The dip probe sends light across a fixed gap and back off a mirror at the tip, so the reading is absorbance through a known pathlength rather than reflectance off a surface. The tip sets the pathlength, 2 mm for concentrated dye and 10 mm for dilute solutions, and the body sets where it can be used: a 6.35 mm stainless body fits a bioreactor head port, while a 12.7 mm industrial body with a quartz window takes the temperature and pressure of a process line.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Fiber bundle | 6 illumination fibers around 1 read fiber | Read fiber at the centre of the ferrule |
| Core options | 200, 400 and 600 um | Avantes standard bundle 200 or 400 um |
| Numerical aperture | 0.22 | Holds through the bundle |
| UV and visible range | 300 to 1100 nm, high OH fiber | Solarization resistant option 200 to 1100 nm |
| Visible and NIR range | 400 to 2500 nm, low OH fiber | Low OH keeps the water bands out |
| Probe ferrule | 6.35 mm OD x 76.2 mm, or 3.18 mm OD x 74.2 mm | Small tip probes 1.5 or 2.5 mm OD x 100 mm |
| Fiber length | 2 m standard | Split point 1 m from the probe on the R series |
| Connectors | SMA 905, two legs | One leg to the source, one to the spectrometer |
| Operating temperature | -20 to 80 C, -30 to 100 C on Avantes | High temperature build to 200 C |
| Pressure, reflectance head | 50 bar on a 200 um head, 10 bar on a 400 um head | At 25 C, Avantes 316 stainless body |
| Pressure, dip probe | 100 psi on the 6.35 mm body, 250 psi on the 12.7 mm body | Quartz window and O ring seal at the tip |
| Bend radius | 8, 16 or 24 cm long term for 200, 400 and 600 um | Short term is half the long term value |
| Pathlength, transmission | 2, 5, 10, 25 or 50 mm | Tip and body ordered as separate items |
Included Models and Order Codes
The first table is the reflectance family: R series laboratory probes, QR premium probes, mixed 12 fiber probes for two channel work, and the Avantes FCR probes with a 316 stainless head that takes 50 bar. The second table is the transmission family, where the body and the pathlength tip are two separate order lines, so both have to be quoted together.
| Part Number | Bundle | Core | Range | Ferrule | Jacket | Use |
|---|---|---|---|---|---|---|
| R200-7-UV-VIS | 6 around 1 | 200 um | 300 to 1100 nm | 6.35 x 76.2 mm | PVDF zip | Small spots and thin films |
| R400-7-UV-VIS | 6 around 1 | 400 um | 300 to 1100 nm | 6.35 x 76.2 mm | PVDF zip | General UV and visible work |
| R400-7-VIS-NIR | 6 around 1 | 400 um | 400 to 2500 nm | 6.35 x 76.2 mm | PVDF zip | Colour and NIR methods |
| R400-7-SR | 6 around 1 | 400 um | 200 to 1100 nm | 6.35 x 76.2 mm | PVDF zip | Deep UV without solarization |
| R600-7-UV-125F | 6 around 1 | 600 um | 300 to 1100 nm | 3.18 x 74.2 mm | PVDF zip | Low light levels |
| R600-7-VIS-125F | 6 around 1 | 600 um | 400 to 2500 nm | 3.18 x 74.2 mm | PVDF zip | Low light in the NIR |
| R600-7-SR-125F | 6 around 1 | 600 um | 200 to 1100 nm | 3.18 x 74.2 mm | PVDF zip | UV source monitoring |
| RP200-7-UV-VIS | 6 around 1 | 200 um | 300 to 1100 nm | 6.35 mm PEEK | PVDF zip | Corrosive liquid samples |
| R400-7-ANGLE-VIS | 6 around 1, 30 degree window | 400 um | 400 to 2100 nm | 6.35 x 76.2 mm | PVDF zip | Powders and slurries |
| R200-12-MIXED | 6 plus 6 around 1 plus 1 | 200 um | 300 to 1100 and 400 to 2100 nm | 6.35 x 76.2 mm | PVDF zip | Two channel instruments |
| R200-REF-UV-VIS | 6 around 1 plus reference leg | 200 um | 300 to 1100 nm | 6.35 x 76.2 mm | PVDF zip | Source drift correction |
| QR400-7-UV-VIS | 6 around 1, premium grade | 400 um | 300 to 1100 nm | 6.35 x 76.2 mm | Silicone monocoil | Plant and field use |
| FCR-7UVIR200-2-BX | 6 around 1 | 200 um | 200 to 2500 nm | 6.35 x 50 mm | Stainless BX, 6.0 mm OD | Broadband with pressure |
| FCR-7UVIR400-2-ME | 6 around 1 | 400 um | 200 to 2500 nm | 6.35 x 50 mm | Chrome brass, 5.0 mm OD | Flexible runs on a bench |
| FCR-7UVIR200-2-45-BX | 6 around 1, 45 degree window | 200 um | 200 to 2500 nm | 6.0 x 140 mm | Stainless BX | Powders and thick fluids |
| FCR-7UVIR200-2-1.5x100 | 6 around 1, small tip | 200 um | 200 to 2500 nm | 1.5 x 100 mm | PTFE with PVC sleeve | Medical and wafer probes |
| FCR-7UVIR200-2-2.5x100 | 6 around 1, small tip | 200 um | 200 to 2500 nm | 2.5 x 100 mm | PTFE with PVC sleeve | Tight fixtures |
| FCR-COL | Collimating lens | For 6.35 mm probes | 200 to 2500 nm | SMA at both ends | Anodised body | Working distance above 20 mm |
| FCR-90-Option | 90 degree reflector | For 6.35 mm probes | 200 to 2500 nm | Slip on tip | Mirror at 45 degrees | Downward looking heads |
Transmission Dip Probes and Pathlength Tips
A transmission measurement needs two order lines: the body with its fiber legs, and the tip that sets the gap. The 6.35 mm bodies are for laboratory vessels and bioreactor ports up to 100 C, while the 12.7 mm industrial body carries a quartz window and reaches 300 C and 250 psi. Tips are not interchangeable between the two body sizes.
| Part Number | Body | Fiber | Range | Pathlength | Pressure | Temperature |
|---|---|---|---|---|---|---|
| T300-RT-UV-VIS | 6.35 mm x 127 mm | 300 um UV | 200 to 1100 nm | Tip ordered separately | 100 psi | 100 C without sleeve |
| T300-RT-VIS-NIR | 6.35 mm x 127 mm | 300 um VIS NIR | 400 to 2100 nm | Tip ordered separately | 100 psi | 100 C without sleeve |
| T200-RT-VIS-NIR | 6.35 mm x 127 mm | 200 um VIS NIR | 400 to 2500 nm | Tip ordered separately | 100 psi | 100 C without sleeve |
| RT-2MM | Tip for T200 and T300 | Quartz | 200 to 2500 nm | 2 mm | 100 psi | 100 C |
| RT-5MM | Tip for T200 and T300 | Quartz | 200 to 2500 nm | 5 mm | 100 psi | 100 C |
| RT-10MM | Tip for T200 and T300 | Quartz | 200 to 2500 nm | 10 mm | 100 psi | 100 C |
| T300SLEEVE | Spare stainless sleeve | None | Not applicable | Not applicable | Not applicable | Not applicable |
| TI300-UV-VIS | 12.7 mm x 127 mm | 300 um UV | 200 to 1100 nm | Tip ordered separately | 250 psi | 300 C body |
| RT-TI-2MM | Tip for TI300 | Quartz | 200 to 2500 nm | 2 mm | 250 psi | 300 C |
| RT-TI-5MM | Tip for TI300 | Quartz | 200 to 2500 nm | 5 mm | 250 psi | 300 C |
| RT-TI-10MM | Tip for TI300 | Quartz | 200 to 2500 nm | 10 mm | 250 psi | 300 C |
| RT-TI-25MM | Tip for TI300 | Quartz | 200 to 2500 nm | 25 mm | 250 psi | 300 C |
| RT-TI-50MM | Tip for TI300 | Quartz | 200 to 2500 nm | 50 mm | 250 psi | 300 C |
Key Features
Lit and read at one spot – six illumination fibers ring a single read fiber, so the viewing angle stays fixed when the sample moves
Fiber chosen by range – solarization resistant for 200 to 1100 nm, high OH for 300 to 1100 nm, low OH for 400 to 2500 nm
Ferrule by access – 6.35 mm for standard holders, 3.18 mm for tight fixtures, 1.5 and 2.5 mm for wafer and medical work
Reference leg on request – a second read fiber watches the lamp, so drift shows up as a ratio rather than as a false shift in the sample
Angled windows for powders – a 30 or 45 degree window keeps the surface reflection out of the read path on slurries and tablets
Tip sets the pathlength – 2 mm for strong absorbers, 10 mm for dilute solutions, 25 and 50 mm on the industrial body
Applications
Probes are chosen together with the light source and the spectrometer slit, so an enquiry usually names the range and the sample rather than the part on its own.
Coating and film thickness – a 400 um probe at 90 degrees over a coated web, with a reference leg to hold the lamp constant
Colour on plastic and paint – a 45 degree ferrule on a bench stand, read against a white tile before each batch
In line absorbance – a 6.35 mm dip probe in a bioreactor head port, 5 mm tip for protein concentration at 280 nm
Probes and fiber assemblies ship alongside SUPERB PCB and SUPERB PCBA work when an analyser build moves from the board to the optics.
Sourcing and Cross-Reference
Two probes can carry the same core size and behave differently, because the jacket, the ferrule and the fiber type all sit in the price. We quote the range and the ferrule first, then the jacket that suits the plant: PVDF zip tube for a bench, stainless BX for a cable tray, silicone monocoil where the cable flexes all day.
The common ordering error is a probe without its tip. On the T200 and T300 bodies the pathlength tip is a separate line, and a body on its own cannot make a transmission reading. The same applies to the industrial body, where the quartz window and the O ring seal come with the tip, and the tip for one body size will not fit the other.
Spare sleeves and replacement tips are worth quoting with the first order. A dip probe in a process line loses its window to deposits long before the fiber fails, and a spare tip on the shelf turns a shutdown into a ten minute change. Lead time on the fiber assemblies is normally stock to two weeks, and we state it per line on the quotation.
Optical kits for a new instrument are packed through SUPERB Assembly, so each probe arrives with its tip, cap and label in one bag.
Q: Can you supply a 400 um reflectance probe with a reference leg, and a 5 mm tip dip probe for the same instrument?
A: Yes. SUPERB Tech quotes the body and the pathlength tip as a pair, and states the fiber range and ferrule size before the order is placed.
Why Choose SUPERB for Spectrometer Probes
- Range matched to the fiber: the fiber type is named per part, so a 2500 nm method is not quoted on a visible probe
- Body and tip quoted together: transmission probes are checked for the pathlength tip before the order is confirmed
- Pressure and temperature stated: 50 bar heads and 250 psi industrial bodies are separated from laboratory parts
- Ferrule sizes in one place: 1.5, 2.5, 3.18, 6.0 and 6.35 mm options on one page for fixture work
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Fiber Bundles and Light Guides, Fiber Collimators and Patch Cords, Quartz Flow Cells and Cuvettes, Integrating Spheres and Reflectance Standards, Photodiode Arrays and Detector Modules, Optical Filter Mounts and Lens Holders.