Integrating Spheres and Diffuse Reflectance Standards

Spheres that collect every reflected angle, and the white tiles that calibrate them. PTFE and Spectralon spheres from 30 to 200 mm, three or four ports. Calibrated standards from 2 to 99 percent, data every 50 nm from 250 to 2500 nm.
quote now

Product Specifications

Product Overview

An integrating sphere turns direction into quantity. Light entering the sphere bounces until the inside surface is evenly lit, so a detector at any port reads the total flux instead of one reflected angle. That is why a sphere is used for laser power, for the reflectance of a rough or curved sample, and for the colour of a textured surface where a plain probe gives a reading that depends on where it is pointed.

Sphere size, port size and coating are one decision. A 50 mm sphere with an 8 mm port suits a fibre coupled spectrometer and a sample that fills the port; a 200 mm sphere with 12.7 mm ports spreads the light further and gives a more uniform signal on a large detector, at the cost of throughput. A PTFE based wall holds above 98 percent between 400 and 1500 nm and above 95 percent over 250 to 2500 nm, which is enough for most UV to NIR work.

The standard is the other half of the measurement. A reflectance reading is only a ratio against something assumed to be known, and that something is a sintered PTFE tile with a certificate: 99 percent white, or a grey step at 75, 50, 20, 10 or 2 percent to check linearity. The certificate reports the hemispherical reflectance factor every 50 nm from 250 to 2500 nm against a NIST scale, so the calibration chain is stated rather than assumed.

Key Specifications

ParameterValueNotes
Sphere diameter30, 50, 80 mm fibre coupled, 50.8 mm (2 in) bench mountLabsphere general purpose 1 to 6 in
Port diameter6 or 8 mm on fibre spheres, 12.7 mm (0.5 in) on 2 in spheresPort reducers and plugs are separate items
Port count3 ports at 0, 90 and top, or 4 ports adding 180Unused ports closed with a coated plug
Wall materialSintered PTFE, or Spectralon machined bodySpectraflect barium sulphate coating on request
Reflectance, Thorlabs 2 inAbout 99 percent from 350 to 1500 nm, above 95 percent from 250 to 2500 nmWall, not the sample
Reflectance, ISP seriesAbove 98 percent from 400 to 1500 nm, above 95 percent from 250 to 2000 nmPTFE based diffusing material
Detector port3 mm recessed port for SM05PD photodiodesRecess keeps direct light off the active area
Thermal limitUp to 250 C on the 2 in sphereCoating sets the real limit
Laser damage2 kW per sq cm continuous, 7 J per sq cm pulsedAt the 2 in sphere port
Standard reflectance levels2, 5, 10, 20, 25, 40, 50, 60, 75, 80 and 99 percentIndividual tiles or matched sets
Standard calibration8 degree illumination, hemispherical collectionReported every 50 nm, 1 nm in the data file
Standard sizes31.8 mm (1.25 in) and 50.8 mm (2 in) optical surfaceTargets from 2 x 2 in to 24 x 24 in

Included Models and Order Codes

The first table lists the spheres, grouped by the instrument they attach to: 2 in bench spheres with SM05 ports, fibre coupled spheres with SMA 905, and the Labsphere general purpose bodies with three or four ports. The second table lists the standards, where the part number fixes the reflectance level and the size, and the order number fixes the certificate.

Part NumberSphere SizePortsCoating or WallDetectorInterfaceUse
IS20050.8 mm (2 in)3 at 0, 90, topPTFE based wallNoneSM05 and 30 mm cageLaser power and flux
IS200-450.8 mm (2 in)4 at 0, 90, 180, topPTFE based wallNoneSM05 and 30 mm cageReflectance with a reference port
IS236A50.8 mm (2 in)3 at 0, 90, topPTFE based wallSM05PD1B siliconSM05 and 30 mm cage350 to 1100 nm radiometry
IS236A-450.8 mm (2 in)4 at 0, 90, 180, topPTFE based wallSM05PD1B siliconSM05 and 30 mm cageTwo detector comparison
IS210C50.8 mm (2 in)3 at 0, 90, topPTFE based wallSM05PD4A InGaAsSM05 and 30 mm cage800 to 1700 nm work
ISP-30-6-R30 mmSample port 6 mmSintered PTFENone2 x SMA 905Small sample reflectance
ISP-50-8-R50 mmSample port 8 mmSintered PTFENone2 x SMA 905Fibre coupled reflectance
ISP-50-8-R-GT50 mmSample port 8 mm, gloss trapSintered PTFENone2 x SMA 905Specular included or excluded
ISP-80-8-R80 mmSample port 8 mmSintered PTFENone2 x SMA 905Larger or textured samples
3P-GPS-020-SL50.8 mm (2 in)3 at 0, 90, topSpectralonNone1 in port framesColour and radiance work
4P-GPS-020-SL50.8 mm (2 in)4 at 0, 90, 180, topSpectralonNone1 in port framesReference port in use
3P-GPS-020-SL-KIT50.8 mm (2 in)3 ports with plugsSpectralonNonePost and holder includedBench setup in one order
4P-GPS-020-SL-KIT50.8 mm (2 in)4 ports with plugsSpectralonNonePost and holder includedLab ready kit

Diffuse Reflectance Standards and Targets

Every standard carries two numbers: the model, which states the reflectance level and the size, and the order number, which states the certificate. The 1.25 in tiles suit a sphere with an 8 mm port, the 2 in tiles suit a bench instrument, and the 5 to 10 in targets are for camera and imaging work. A set is worth quoting when linearity across the range has to be shown at one time.

ModelOrder NumberReflectanceSizeFormCalibration
SRS-99-010AS-01160-06099 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-99-020AS-01161-06099 percent50.8 mm (2 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-80-010AS-01160-16080 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-75-010AS-01160-26075 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-60-010AS-01160-36060 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-50-010AS-01160-46050 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-40-010AS-01160-56040 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-20-010AS-01160-66020 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-10-010AS-01160-76010 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-05-010AS-01160-8605 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
SRS-02-010AS-01161-9602 percent31.8 mm (1.25 in)Tile in holder250 to 2500 nm, every 50 nm
RSS-04-010AS-01163-06099, 75, 50 and 2 percent31.8 mm (1.25 in)Set of four tiles250 to 2500 nm, every 50 nm
RSS-08-010AS-01163-16099, 80, 60, 40, 20, 10, 5 and 2 percent31.8 mm (1.25 in)Set of eight tiles250 to 2500 nm, every 50 nm
SRT-99-050AA-00821-00099 percent127 mm (5 in)Mounted target250 to 2500 nm, every 50 nm
SRT-99-100AA-00822-00099 percent254 mm (10 in)Mounted target250 to 2500 nm, every 50 nm
USRT-99-050AS-02627-00099 percent127 mm (5 in)Mounted target, uncalibratedNominal value only

Key Features

Ports closed with the same wall – each unused port ships with a plug coated in the sphere material, so the wall stays uniform

Recessed detector port – the photodiode sits behind a 3 mm recess, so a divergent beam cannot reach the active area directly

Three port sizes on one page – 30, 50 and 80 mm fibre spheres plus the 2 in bench sphere, chosen by sample size

Gloss trap option – specular included or excluded with two inserts, which is how colour work is compared between labs

Standards with certificates – reflectance reported every 50 nm from 250 to 2500 nm, traceable to NIST

Sets for linearity – four or eight grey steps in one box, so a whole calibration runs in one session

Applications

A sphere and a standard are bought as a pair: the sphere collects the light and the tile tells the instrument what 100 percent looks like.

Laser and LED power – a 2 in sphere with an SM05PD photodiode, read directly with no lens in front of the detector

Colour of textured surfaces – a 50 mm sphere with an 8 mm port and a 99 percent tile as the white reference

Reflectance of coatings and glass – a four port sphere with the sample at 0 degrees and a reference port at 180 degrees

Spheres and tiles ship alongside SUPERB PCB and SUPERB PCBA work when an instrument build moves from the board to the optics.

Sourcing and Cross-Reference

Sphere quotes are settled by three numbers: inner diameter, port diameter and wall material. The same 2 in body with a Spectralon wall and with a Spectraflect coating differs in reflectance below 400 nm, and a customer moving a validated method cannot swap one for the other without repeating the calibration.

Standards are often ordered a year after the sphere, when the first audit comes. A 99 percent tile drifts slowly but the holder and the certificate matter, so we keep the order number rather than the model in the quotation, and we state whether the tile is calibrated or nominal. An uncalibrated target costs less and is honest for a go or no go check, but it cannot be used to set the scale.

Two practical points decide the size. A tile has to cover the sphere port completely, or light leaks past the edge and the reading falls short; and a tile handled by its face collects skin oil that shows first in the UV. We quote a spare tile with the first order where the method runs daily.

Optical kits for a new instrument are packed through SUPERB Assembly, so each sphere arrives with its plugs, tile and certificate in one box.

Q: Can you supply a 50 mm sphere with an 8 mm port together with a calibrated 99 percent tile of the same size?

A: Yes. SUPERB Tech quotes the sphere and the matching tile together, and states the port diameter and the certificate level on the quotation.

Why Choose SUPERB for Spheres and Reflectance Standards

  1. Wall material stated per part: PTFE, Spectralon and Spectraflect are separated, because the UV end of the curve depends on it
  2. Port sizes and plugs together: the quote lists the plugs and reducers the sphere needs, so nothing arrives open
  3. Certificate named by order number: calibrated and nominal tiles are never mixed on one line
  4. Sets for linearity: four and eight step sets quoted with the sphere when a method needs a full check

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Neutral Density Filters and Photometric Standards, Reflectance and Transmission Probes, Optical Windows and Sapphire Discs, Photodiode Arrays and Detector Modules, Lenses and Mirrors for Optical Benches, Photomultiplier Tubes and Housings.