Photodiode Arrays and Detector Modules for Spectrometers
Product Specifications
Product Overview
A linear photodiode array replaces the exit slit and the scanning grating of a monochromator with a row of diodes on which the whole spectrum falls at once. The readout is a sequence of charges, one per pixel, so the array does three jobs that a single detector cannot: it removes the moving parts, it shortens the measurement to one integration period, and it makes the wavelength axis a fixed geometric property of the bench rather than a function of a motor.
Pixel pitch sets the optical resolution and the pixel height sets the light budget. A 25 micrometre pitch array puts 1024 pixels on 25.6 mm, which is a typical focal plane for a compact spectrograph, while a 1.0 mm pitch array puts 16 elements across 16 mm and is used where the spectrum is wide, the light is low and the resolution requirement is coarse. The 2.5 mm pixel height of the NMOS family is a light gathering feature, not a resolution feature: a taller pixel collects more photons from a tall slit image without widening the spectral band.
Readout architecture decides the support circuit. Current output NMOS arrays deliver a current per pixel, which is integrated by an external amplifier; this gives the best linearity for absorbance work but needs an integrator and a timing generator. Voltage output arrays and CMOS sensors with an internal hold deliver a rail to rail voltage and are read with a simple analog to digital converter and a clock. A sensor with a hold, such as the 128 pixel CMOS part, freezes all pixels at the same instant, which matters when the light source is pulsed or when the sample moves.
Spectral range follows the window material, not the silicon. A resin potted array covers 340 to 1100 nm, while the same die behind a quartz window reaches 190 nm, which is the difference between a visible spectrometer and one that can see the deuterium line at 254 nm. A separate family trades peak sensitivity in the near infrared for a lower capacitance, which is the version chosen when the readout speed matters more than the responsivity at 900 nm.
Detector modules shrink the whole bench into a package. A fingertip spectrometer contains the slit, the grating, the CMOS array and a sealed housing, and is quoted with a spectral resolution rather than a pixel count. That resolution is a property of the module, 12 to 15 nm across the visible range for a 256 pixel unit and 5.5 to 8 nm for a 288 pixel ultraviolet unit, so the module is chosen on the resolution and the range, and the pixel count only explains why the resolution is what it is.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Array formats | 16, 35 and 46 elements in ceramic DIP; 256, 512 and 1024 elements in NMOS; 128 pixels in CMOS | the format is chosen from the focal plane length and the resolution target |
| Spectral range, resin window | 340 to 1100 nm | S4111-16R, peak sensitivity 960 nm |
| Spectral range, quartz window | 190 to 1100 nm on the S4111Q family, 190 to 1000 nm on the S4114 family | the quartz window is what opens the ultraviolet range |
| Spectral range, NMOS | 200 to 1000 nm | peak at 600 nm, quartz window as standard |
| Pixel size, 16 element array | 1.45 x 0.9 mm, 1.305 mm2 effective | 0.1 mm gap, 1.0 mm pitch |
| Pixel size, 35 and 46 element array | 4.4 x 0.9 mm, 3.96 mm2 effective | 0.1 mm gap, 1.0 mm pitch, 35 or 46 elements |
| Pixel size, NMOS | 25 x 2500 um | 25.6 mm long for 1024 pixels, 12.8 mm for 512, 6.4 mm for 256 |
| Pixel size, CMOS linear sensor | 400 DPI, 8 um gap, 3524.3 um2 per pixel | 128 pixels over 8.12 mm of active length |
| Dark current, 16 element | 5 pA maximum at 10 mV, 25 pA at 10 V | the reverse bias raises the dark current and the speed together |
| Dark current, 35 and 46 element | 10 pA at 10 mV and 50 pA at 10 V | ten times the 16 element figure because the pixel is larger |
| Dark current, NMOS | 0.1 pA typical, 0.3 pA maximum | the low figure is what allows a long integration at room temperature |
| Shunt resistance | 2.0 GOhm minimum on the 16 element, 1.0 GOhm on the 35 and 46 element | 250 GOhm and 30 GOhm typical respectively |
| Terminal capacitance | 200 pF at 0 V and 50 pF at 10 V on the 16 element | 550 pF and 120 pF on the 35 and 46 element, 10 pF on the NMOS |
| Rise time | 0.5 us at 0 V and 0.1 us at 10 V on the 16 element, 1.2 us and 0.3 us on the larger arrays | measured with a 1 kOhm load at 655 nm |
| Line rate, NMOS 1024 | 1950 lines per second maximum | sets the shortest integration time in a scanning instrument |
| Clock rate, CMOS linear sensor | 5 kHz to 8 MHz | the analog to digital converter is usually the limit below 8 MHz |
| Integration time, CMOS linear sensor | 0.03375 to 100 ms | 33.75 us is the floor at the maximum clock rate |
| Dynamic range, CMOS linear sensor | 4000 to 1, which is 72 dB | image lag 0.5 percent typical |
| Supply, CMOS linear sensor | 3.0 to 5.5 V | single supply, output referenced to ground |
| Operating temperature | -20 to +60 C on the silicon arrays, -40 to +65 C on the NMOS, -25 to +85 C on the CMOS sensor | the module family is narrower, +5 to +50 C for the visible unit |
| Module resolution | 12 nm typical and 15 nm maximum on C12666MA and C12880MA | 5.5 nm typical and 8 nm maximum on the 190 to 440 nm module |
Included Models and Order Codes
Part numbers below are grouped by readout architecture. The suffix Q on a Hamamatsu array means a quartz window and the suffix R means a resin potting, so S4111-16R and S4111-16Q share a die and differ in the window, which is the difference between 340 nm and 190 nm at the short end. The NMOS parts share a 25 micrometre pitch and a 2.5 mm pixel height, so the pixel count alone tells the focal plane length.
| Part | Type | Elements | Pixel and pitch | Spectral range | Package |
|---|---|---|---|---|---|
| S4111-16R | Si photodiode array, current output | 16 | 1.45 x 0.9 mm, 1.0 mm pitch | 340 to 1100 nm | 18 pin ceramic DIP, resin potting |
| S4111-16Q | Si photodiode array, current output | 16 | 1.45 x 0.9 mm, 1.0 mm pitch | 190 to 1100 nm | 18 pin ceramic DIP, quartz window |
| S4111-35Q | Si photodiode array, current output | 35 | 4.4 x 0.9 mm, 1.0 mm pitch | 190 to 1100 nm | 40 pin ceramic DIP, quartz window |
| S4111-46Q | Si photodiode array, current output | 46 | 4.4 x 0.9 mm, 1.0 mm pitch | 190 to 1100 nm | 48 pin ceramic DIP, quartz window |
| S4114-35Q | Si photodiode array, near infrared | 35 | 4.4 x 0.9 mm, 1.0 mm pitch | 190 to 1000 nm | 40 pin ceramic DIP, quartz window |
| S4114-46Q | Si photodiode array, near infrared | 46 | 4.4 x 0.9 mm, 1.0 mm pitch | 190 to 1000 nm | 48 pin ceramic DIP, quartz window |
| S3904-256Q | NMOS linear image sensor, current output | 256 | 25 x 2500 um, 25 um pitch | 200 to 1000 nm | ceramic, quartz window |
| S3904-512Q | NMOS linear image sensor, current output | 512 | 25 x 2500 um, 25 um pitch | 200 to 1000 nm | ceramic, quartz window |
| S3904-1024Q | NMOS linear image sensor, current output | 1024 | 25 x 2500 um, 25 um pitch | 200 to 1000 nm | ceramic, quartz window, 1950 lines/s |
| TSL1401CL | CMOS linear sensor array with hold | 128 | 400 DPI, 8 um gap | 400 to 1000 nm | CL-8, 3.0 to 5.5 V, 8 MHz |
| C12666MA | Fingertip spectrometer module | 256 | 12.5 x 1000 um, slit 50 x 750 um | 340 to 780 nm | 20.1 x 12.5 x 10.1 mm, 5 g |
| C12880MA | Fingertip spectrometer module | 288 | 14 x 200 um, slit 50 x 500 um | 340 to 850 nm | 20.1 x 12.5 x 10.1 mm, 5 g |
| C16767MA | Fingertip spectrometer module, ultraviolet | 288 | 14 x 200 um, slit 15 x 500 um | 190 to 440 nm | 20.1 x 12.5 x 10.1 mm, 5 g |
Choosing the Readout Architecture
The four architectures below are not interchangeable at the circuit level, even when two of them look identical on a datasheet. The choice is made once, at the point where the analog front end is fixed, because changing it later means a new board and a new timing generator.
| Architecture | Output per pixel | Support circuit | Best for | Limitation |
|---|---|---|---|---|
| Current output NMOS, S3904 family | Photocurrent, integrated externally | Integrator, timing generator, 5 V pulses | Absorbance and low light work where linearity matters | Needs a low noise integrator and careful reset timing |
| Voltage output NMOS, S3921 to S3924 family | Voltage ramp, internally buffered | Clock and analog to digital converter | Compact instruments that cannot carry an integrator | Linearity is slightly lower than the current output type |
| CMOS linear sensor with hold, TSL1401CL | Rail to rail voltage, all pixels held | Serial input, clock, analog to digital converter | Pulsed sources and moving samples | 128 pixels only, 400 DPI |
| Fingertip spectrometer module, C12666MA and C12880MA | Digital spectrum after the internal array | Clock, start pulse, analog to digital converter | Handheld and inline instruments with a fixed bench | Resolution is fixed by the module, 12 to 15 nm |
Key Features
Pixel height for light budget – The NMOS family uses a 2.5 mm pixel height, which collects more light from a tall slit image without widening the spectral band.
Pitch for resolution – A 25 micrometre pitch puts 1024 pixels on 25.6 mm, while a 1.0 mm pitch spreads 16 elements over 16 mm for coarse, low light work.
Quartz window for the ultraviolet – The Q suffix opens 190 nm, which is where a deuterium lamp and a 254 nm line live, and the R suffix stops at 340 nm.
Low dark current for long integration – 0.1 pA typical on the NMOS family allows an integration of seconds at room temperature without a cooler.
Internal hold for pulsed sources – The 128 pixel CMOS sensor freezes all pixels at one instant, so a flash lamp does not smear across the readout.
Line rate for scanning instruments – 1950 lines per second on the 1024 element NMOS array sets the shortest integration time and the maximum scan speed.
Module with a stated resolution – The fingertip modules are quoted at 12 to 15 nm in the visible range and 5.5 to 8 nm in the ultraviolet, which is the number an instrument specification needs.
One die, two windows – S4111-16R and S4111-16Q are the same array with different windows, so a bench can be moved from visible to ultraviolet without a new layout.
Applications
The same arrays appear in three instrument types, and the difference between them is the readout, not the silicon.
Diode array spectrophotometers – A 1024 element NMOS array covers 200 to 1000 nm in one integration, which is what allows a full spectrum to be captured before a reaction moves.
Handheld and inline analysers – A fingertip module with an internal bench removes the optical alignment step, so a colour or concentration reading can be built into a handset.
Emission and fluorescence readers – A 16 or 46 element array with a 1.0 mm pitch reads a wide, low level band, which suits a compact fluorimeter with no moving grating.
Colour and display measurement – A 256 pixel module in the visible range measures a display or a print patch against a reference, with the resolution stated in nanometres.
These arrays are used on SUPERB PCB and SUPERB PCBA spectrometer boards, where the analog front end is laid out around the pixel rate.
Sourcing and Cross-Reference
Array sourcing is driven by the optical bench rather than by the electronics, because the pixel pitch and the package height are fixed by the focal plane that already exists. A substitution is only practical when pitch, pixel height, window material and package footprint all match, and the fourth field is the one that usually fails: two arrays with the same 25 micrometre pitch can sit in packages with different pin rows, which means a new board.
Where a part is marked not recommended for new builds, the practical route is a last time buy against the production forecast and a parallel qualification of the successor. Hamamatsu keeps long product lives on this family, but the quartz window versions and the ultraviolet modules are the lines that move first, so the buffer is held on those rather than on the visible parts.
Spectrometer modules carry a wavelength calibration that is written at the module level, so two modules are not interchangeable in software even when the model number matches. The calibration factors are supplied per serial number, and a replacement module needs its own factors loaded before the instrument is released. That is the reason a module order is placed with the serial numbers recorded, and the reason a spare is kept calibrated rather than in a bare box.
The module and its flex are kitted together for SUPERB Assembly builds, so the calibration sheet travels with the unit.
Q: Can you supply a pin compatible array with the same pixel pitch and a quartz window?
A: SUPERB Tech confirms the window material and the readout type before quoting, because the same array is sold in both versions.
Why Buy Detector Arrays Through SUPERB
- Pitch and package checked together: A cross reference is only proposed when pitch, pixel height, window and footprint match, so the board does not have to change.
- Window option on the same die: The resin and quartz versions of one array are quoted side by side, which is what a range extension from visible to ultraviolet needs.
- Module and array in one order: A bare array for the production bench and a calibrated module for the handheld version can be sourced together.
- Long life parts held for build: The ultraviolet modules and the quartz window arrays are the lines kept in buffer, because they are the ones with the longest qualification tail.
- Serial numbers recorded: Module shipments are booked with the serial number, so the wavelength factors can be matched to the unit that was built.
- Support for the analog front end: The readout type is confirmed before the order, because a current output array and a voltage output array need different integrators.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Monochromator Gratings and Optical Bench Parts, Lamp Housings and Shutters, Xenon and Deuterium Lamps, UV and Visible LED Sources, PIN and Avalanche Photodiodes, Quartz Flow Cells and Cuvettes, Component Lead Time Solutions.