Fiber Bundles and Light Guides for Spectrometers
Product Specifications
Product Overview
A single fiber into a slit is a compromise. The slit is 20 um by 2 mm and the fiber end is a circle, so most of the light misses the entrance and lands on the inside of the monochromator. A bundle rearranges the same glass into a line at the instrument end and a circle at the sample end, and the coupling efficiency jumps before anything else in the chain changes.
Three numbers decide which bundle is right. The core size sets how much light enters and how much resolution is lost, the hydroxyl content sets the wavelength span, and the layout sets where the light lands. A 105 um solarization resistant fiber covers 180 to 1200 nm and suits UV work; a 200 um low OH fiber covers 400 to 2400 nm and is the usual choice for NIR methods; a 400 or 600 um bifurcated pair is what a reflectance or backscatter probe needs.
The light guide is the other half of this family. Where a bundle carries a signal into a detector, a light guide carries illumination out of a lamp housing to a bench, a microscope or a cure station. These are glass fiber bundles in a flexible metal hose with a machined ferrule at each end, sized by bundle diameter and length, and they are quoted by the pair of ends the illuminator and the fixture need.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Fiber core options | 50, 105, 200, 400 and 600 um | Multimode step index glass |
| Numerical aperture | 0.22 plus or minus 0.02 | A bundle keeps the NA of the single fiber |
| Solarization resistant range | 180 to 1200 nm | Below 300 nm the fiber ages with use |
| High OH range | 250 to 1200 nm | UV to visible work |
| Low OH range | 400 to 2400 nm | Visible to NIR work |
| Bundle layout | 7 fibers in a line at one end, round at the other | The linear end matches the slit |
| Linear end size | 0.9 x 0.1 mm for 105 um, 1.6 x 0.2 mm for 200 um | Per bundle, measured at the face |
| Round end effective core | 380 to 700 um depending on fiber | Per bundle |
| Connectors | SMA 905 at both ends | Ferrule keyed for the slit axis |
| Short term bend radius | 16 to 24 mm | Long term roughly double |
| Jacket options | Stainless steel tubing, PVC monocoil, silicone monocoil, stainless BX | Temperature and chemical resistance differ |
| Light guide bundle size | 0.375 in and 0.5 in | Ferrule 14.27 mm on the 0.5 in guide |
Included Models and Order Codes
The first block is the round to linear family, which is the one to quote when a slit is involved. The second is the bifurcated Y family, used where one source feeds two paths or two detectors watch one sample. The last two rows are flexible light guides for illumination, quoted with the illuminator end and the fixture end named.
| Part Number | Type | Core or Bundle | Range | Length | Jacket | Use |
|---|---|---|---|---|---|---|
| BFL105RS2 | Round to linear, 7 fibers | 105 um | 180 to 1200 nm | 2 m | Stainless tubing | UV to NIR coupling |
| BFL105HS02 | Round to linear, 7 fibers | 105 um | 250 to 1200 nm | 2 m | Stainless tubing | UV and visible |
| BFL105LS02 | Round to linear, 7 fibers | 105 um | 400 to 2400 nm | 2 m | Stainless tubing | Visible to NIR |
| BFL200RS2 | Round to linear, 7 fibers | 200 um | 180 to 1200 nm | 2 m | Stainless tubing | More light at the slit |
| BFL200HS02 | Round to linear, 7 fibers | 200 um | 250 to 1200 nm | 2 m | Stainless tubing | UV and visible |
| BFL200LS02 | Round to linear, 7 fibers | 200 um | 400 to 2400 nm | 2 m | Stainless tubing | Visible to NIR |
| QBIF50-VIS-NIR | Bifurcated Y | 50 um | 400 to 2100 nm | 2 m | Silicone monocoil | Reference leg |
| QBIF200-VIS-NIR | Bifurcated Y | 200 um | 400 to 2100 nm | 2 m | Silicone monocoil | Two spectrometers |
| QBIF200-VIS-BX | Bifurcated Y | 200 um | 400 to 2100 nm | 2 m | Stainless BX | Field and plant use |
| QBIF400-VIS-NIR | Bifurcated Y | 400 um | 400 to 2100 nm | 2 m | Silicone monocoil | Reflectance pairs |
| QBIF400-VIS-BX | Bifurcated Y | 400 um | 400 to 2100 nm | 2 m | Stainless BX | Process lines |
| QBIF600-VIS-NIR | Bifurcated Y | 600 um | 400 to 2100 nm | 2 m | Silicone monocoil | Low light levels |
| QBIF600-VIS-BX | Bifurcated Y | 600 um | 400 to 2100 nm | 2 m | Stainless BX | Harsh sites |
| 38-659 | Flexible light guide | 0.5 in bundle | Visible | 12 in | PVC over metal hose | Illuminator to bench |
| 54-214 | Flexible light guide | 0.375 in bundle | Visible | 48 in | PVC over metal hose | Remote illumination |
Key Features
Linear end matches the slit – 7 fibers stacked into a line, so the light enters the monochromator instead of hitting the wall beside the entrance
Fiber by wavelength – solarization resistant for 180 to 1200 nm, high OH for 250 to 1200 nm, low OH for 400 to 2400 nm
No dark fibers – the vendor screens each bundle, so a dead channel does not quietly cut the signal by a seventh
Bifurcated legs in matched pairs – common end fibers sit side by side at the same diameter, which keeps source and reference paths comparable
Jacket by environment – silicone monocoil for bench and laboratory, stainless BX for cable trays and washdown areas
Light guides by the pair of ends – ferrule size, bundle diameter and length quoted against the illuminator and the fixture, not on their own
Applications
These parts are chosen at the same time as the slit, the grating and the detector, so the application usually names all four.
UV to NIR absorbance – a 105 or 200 um round to linear bundle between a cuvette holder and the spectrometer entrance
Reflectance and backscatter – a bifurcated 400 or 600 um pair with one leg to the lamp and one to the detector
Illumination for microscopy and cure – flexible light guides from a xenon or LED housing to the fixture, sized by bundle diameter and length
Bundles and light guides ship alongside SUPERB PCB and SUPERB PCBA work when an analyser build moves from the board to the optical bench.
Sourcing and Cross-Reference
Bundles are ordered by geometry, not by part number alone. The same 200 um fiber is sold in a 7 fiber line, a 19 fiber line, a Y pair and a fan out, and the price and the coupling efficiency change with the layout. We quote the layout that matches the slit height first, then the fiber type, then the jacket.
Two failure modes are worth naming. A bundle used below 300 nm without the solarization resistant fiber goes brown over a few hundred hours and the throughput falls off first in the UV, which reads as a drifting lamp rather than a fiber problem. A keyed ferrule that is turned 90 degrees in the holder loses a large part of the signal, which is why the sleeve carries an alignment mark.
Length is usually fixed at 2 m for the standard assemblies because the fiber attenuation is small over that run and the handling is easier. Longer runs are possible, but step index multimode fiber loses light at the bends more than in a straight line, so a 5 m run should be quoted with the bend radius in mind. Both points are settled at quotation, before the cable is made.
Optical kits for a new instrument are packed through SUPERB Assembly, so each bundle arrives with its ferrule caps and label in one bag.
Q: Can you supply a bundle with a linear end that matches our 50 um slit and a round end for a standard SMA probe?
A: Yes. SUPERB Tech quotes bundles against the slit height and the fiber range, and states the linear end dimensions before the order is placed.
Why Choose SUPERB for Fiber Bundles and Light Guides
- Slit first, fiber second: the layout is chosen against the entrance slit, then the fiber type and jacket follow
- Three fiber ranges on one page: UV, visible and NIR options with the wavelength span stated per part
- Y pairs quoted together: common end diameter and both leg lengths stated, so source and reference paths match
- Light guides with both ends named: ferrule, bundle diameter and length fixed against the illuminator and the fixture
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Fiber Collimators and Patch Cords, Photodiode Arrays and Detector Modules, Optical Slits and Apertures, Dichroic Filters and Beam Splitters, Quartz Flow Cells and Cuvettes, Monochromator Gratings and Bench Parts.