28 mm side on tubes for spectroscopy and fluorescence. Multialkali to 900 nm, bialkali to 650 nm, gain of 1.0 x 10^7. Socket assemblies with the divider built in, plus D type sockets.
quote now

Product Specifications

Product Overview

A photomultiplier tube is still the lowest noise detector available for a scanning spectrometer, and the reason is the gain happens before the first amplifier. A photon releases an electron at the photocathode, and nine dynode stages multiply that single electron into a pulse of about ten million electrons, which arrives at the anode well above the thermal noise of the following circuit.

The tube and its housing are one decision. The socket carries the voltage divider that sets the gradient along the dynodes, and the housing sets the magnetic and light shielding around the tube. A 28 mm side on tube with a 9 stage circular cage dynode is the standard shape for a scanning monochromator, and the 11 pin base is what the sockets on this page fit.

The choice inside that family is the photocathode and the window. A multialkali cathode covers 185 to 900 nm with a peak at 400 nm and is the general purpose spectroscopy tube; a bialkali cathode has lower dark current but stops at 650 nm and suits blue and UV fluorescence; a quartz window pushes the short wavelength limit to 160 nm for work below the UV glass cutoff.

Key Specifications

ParameterValueNotes
Tube size28 mm, 1-1/8 inSide on type, 11 pin base
Spectral range160 to 900 nm depending on windowUV glass cuts at about 185 nm
PhotocathodeMultialkali or bialkaliEffective area 8 x 24 mm
DynodeCircular cage, 9 stagesLinear focused versions on request
Gain1.0 x 10^7 at 1000 V typicalRises with supply voltage
Anode dark current3 nA typical, 50 nA maximumAfter 30 minutes in darkness
Anode pulse rise time2.2 nsTransit time 22 ns, jitter 1.2 ns
Maximum supply1250 V anode to cathode250 V anode to last dynode
Anode current limit0.1 mA averageExceeding it ages the tube
Operating temperatureMinus 30 to plus 50 CStorage the same
HV socket assembly0 to minus 1250 V from a plus 15 V inputC6270, for 28 mm side on tubes
Socket assembliesD type with divider, 11 pinE717-63 and E717-74

Included Models and Order Codes

The tube rows give the cathode, the window and the spectral span, which is the first cut in a selection. The socket rows are the electrical half: E678-11A is the bare 11 pin socket, E717-63 and E717-74 are socket assemblies with the divider built in, and C6270 is a complete assembly with a regulated high voltage supply that runs from plus 15 V and sets the tube voltage from a potentiometer or a 0 to 5 V signal.

Part NumberTypeCathode or FunctionWindowSpectral RangeBase or MountUse
R928PMT, 28 mmMultialkali, 9 stageUV glass185 to 900 nm11 pinScanning spectroscopy
R928PPMT, 28 mmMultialkali, photon countingUV glass185 to 900 nm11 pinLow level counting
R955PMT, 28 mmMultialkali, 9 stageQuartz160 to 900 nm11 pinDeep UV work
R955PPMT, 28 mmMultialkali, photon countingQuartz160 to 900 nm11 pinDeep UV counting
R11568PMT, 28 mmBialkali, 9 stageUV glass185 to 650 nm11 pinLow dark current
E678-11ASocket11 pin socket onlyNot applicableNot applicable11 pinBare socket, divider external
E717-63Socket assemblyD type dividerNot applicableNot applicable11 pinStandard divider assembly
E717-74Socket assemblyD type divider, alternativeNot applicableNot applicable11 pinDivider for long runs
C6270HV socket assemblyRegulated supply, 0 to minus 1250 VNot applicableNot applicable11 pin, 28 mmSelf contained tube operation

Key Features

Gain before the amplifier – ten million electrons per photoelectron, so the signal leaves the tube above the noise floor

Cathode matched to the band – multialkali for the full 185 to 900 nm span, bialkali where dark current matters more than red response

Window matched to the UV limit – UV glass for 185 nm, quartz for 160 nm

Divider built into the socket – E717-63 and E717-74 carry the dynode chain, so the external wiring is a supply and a signal

Regulated supply in one package – C6270 takes plus 15 V in and delivers 0 to minus 1250 V, set by a potentiometer or a 0 to 5 V signal

Dark current stated, not implied – 3 nA typical and 50 nA maximum at the stated voltage, which is what a detection limit calculation needs

Applications

A tube is chosen with the monochromator and the cathode together, and the housing follows from the tube base.

UV to visible absorbance – a multialkali tube behind a scanning monochromator, with the voltage set once and the gain left alone

Fluorescence and chemiluminescence – a bialkali tube where the emission sits below 650 nm and dark counts set the detection limit

Photon counting – the P versions of both cathodes, with the divider and the discriminator matched to the pulse height

Detector assemblies are quoted with the SUPERB PCB and SUPERB PCBA boards that supply the divider and read the anode, so the high voltage side is settled once.

Sourcing and Cross-Reference

Two specifications decide whether a tube will do the job, and neither is the gain. The first is the cathode radiant sensitivity at the wavelength of interest, because a tube with a 74 mA/W peak at 400 nm may be half that at 600 nm. The second is the anode dark current, because a detection limit is set by the noise, not by the signal.

Tubes age, and the way they age is worth knowing before a replacement is ordered. Gain falls slowly with accumulated charge, so a tube that needs 1100 V to reach the gain it gave at 900 V when new is simply older, not faulty. Running a tube at a lower voltage and accepting a lower gain extends its life more than any other single change.

The housing is where magnetic shielding and light tightness are settled. A side on tube in an unshielded housing picks up mains hum and the baseline wanders with nearby equipment; a mu metal shield and a light tight cover fix both. The socket assemblies on this page are the electrical part, and the mechanical housing, cover and shield are quoted with the drawing of the monochromator flange.

Tube, socket and shield are kitted through SUPERB Assembly with the flange drilling matched to the monochromator.

Q: Can you supply the R928 tube together with a high voltage socket assembly that runs from a plus 15 V rail?

A: Yes. SUPERB Tech quotes the tube with its socket assembly and states the cathode sensitivity at the wavelengths the method uses.

Why Choose SUPERB for Photomultiplier Tubes and Housings

  1. Tube and socket together: the 11 pin base, the divider and the supply are quoted as one set
  2. Cathode data per wavelength: radiant sensitivity given at 254, 400, 633 and 852 nm, not only at the peak
  3. Dark current stated: typical and maximum anode dark current, so a detection limit can be calculated
  4. Housing points named: shield, cover and flange settled with the drawing before the tube is ordered

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Photodiode Arrays and Detector Modules, Filter Wheels and Automated Changers, Monochromator Gratings and Bench Parts, Lamp Housings and Shutters, Xenon and Deuterium Lamps, Neutral Density Filters and Reference Standards.