Neutral Density Filters and Photometric Reference Standards
Product Specifications
Product Overview
Two different jobs are done with a grey filter. The first is attenuation: dropping the light level so a detector stays in its linear range, which is what a neutral density filter does. The second is qualification: proving that the absorbance scale and the wavelength scale are right, which needs a certified reference and a certificate that names the wavelengths. Both are on this page, and they are not interchangeable.
An absorptive ND filter attenuates across a band rather than at a line, so it does not create the interference fringes that a thin metal film can add to a spectrum. The optical density is quoted at a stated wavelength, and the tolerance widens at higher density: an OD 4.0 filter is a very small transmission, so a small manufacturing spread becomes a large relative uncertainty. That is why high density filters are used for stray light checks rather than for quantitative work.
A photometric reference is a different article. It has a certificate tied to NIST practice, it is measured at a stated spectral bandwidth, and it is mounted stress free so the glass does not become birefringent in its holder. The references listed here follow the wavelengths in the NIST technical specification for spectrophotometric reference materials, which means the values in the certificate can be compared with the values in the method.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| ND filter material | Schott NG glass or metal on quartz | Absorptive, low fringe |
| Optical density range | 1.0 to 4.0 | 25 mm mounted filters |
| Absorbance range of references | 0.04 A to 3.5 A | Nineteen filters in the standard format |
| Certified wavelengths | 440, 465, 546.1, 590, 635 nm | Visible reference set |
| Certified bandwidth | 1.0 nm | Other bandwidths on request |
| Reference material | NG type glass, stress free mount | Anodised aluminium holder |
| Traceability | NIST SP 260-140 | SRM 930e, 1930 and 2930 equivalents |
| Wavelength references | Holmium 360-640 nm, didymium 430-890 nm | Bonded to a 1 A neutral filter |
| ND blocking range | 400 to 650 nm | Absorptive 25 mm filters |
| Surface quality | 40-20 scratch dig | Mounted ND filters |
| Filter diameter | 25 mm | Mounted in an SM1 thread |
Included Models and Order Codes
The first group is attenuation, the second and third are qualification. On the Starna glass filters the catalogue number carries the nominal transmission, so 1N is 10% and 9N is 90%, and the same numbers combine into the SRM equivalent sets. Metal on quartz parts use the RM- prefix because they are reference materials rather than working filters.
| Part Number | Type | Nominal Value | Wavelengths | Mount | Use |
|---|---|---|---|---|---|
| NE10A | Absorptive ND filter | OD 1.0, 10% T | 400 to 650 nm | 25 mm, SM1 thread | Attenuation |
| NE20A | Absorptive ND filter | OD 2.0, 1% T | Visible | 25 mm, SM1 thread | Attenuation |
| NE30A | Absorptive ND filter | OD 3.0, 0.1% T | Visible | 25 mm, SM1 thread | Attenuation |
| NE40A | Absorptive ND filter | OD 4.0, 0.01% T | Visible | 25 mm, SM1 thread | Stray light check |
| 1N | NG glass reference | 10% T, 1.0 A | 440 to 635 nm | Aluminium holder | Absorbance check |
| 2N | NG glass reference | 20% T, 0.7 A | 440 to 635 nm | Aluminium holder | Absorbance check |
| 3N | NG glass reference | 30% T, 0.5 A | 440 to 635 nm | Aluminium holder | Absorbance check |
| 5N | NG glass reference | 50% T, 0.3 A | 440 to 635 nm | Aluminium holder | Absorbance check |
| 9N | NG glass reference | 90% T, 0.04 A | 440 to 635 nm | Aluminium holder | Absorbance check |
| RM-1Q | Metal on quartz reference | 10% T, 1.0 A | 250 to 635 nm | Aluminium holder | UV and visible |
| RM-3Q | Metal on quartz reference | 30% T, 0.5 A | 250 to 635 nm | Aluminium holder | UV and visible |
| RM-5Q | Metal on quartz reference | 50% T, 0.3 A | 250 to 635 nm | Aluminium holder | UV and visible |
| RM-9Q | Metal on quartz reference | 90% T, 0.04 A | 250 to 635 nm | Aluminium holder | UV and visible |
| RM-Q3 | Metal on quartz reference | 3% T, 1.5 A | 250 to 635 nm | Aluminium holder | UV and visible |
| RM-Q1 | Metal on quartz reference | 1% T, 2.0 A | 250 to 635 nm | Aluminium holder | UV and visible |
| RM-HG | Holmium glass filter | Wavelength reference | 360 to 640 nm | Aluminium holder | Wavelength check |
| RM-DG | Didymium glass filter | Wavelength reference | 430 to 890 nm | Aluminium holder | Wavelength check |
| RM-HG/1ND | Holmium with neutral filter | Wavelength and 1 A | 360 to 640 nm | Aluminium holder | One pass check |
| RM-DG/1ND | Didymium with neutral filter | Wavelength and 1 A | 430 to 890 nm | Aluminium holder | One pass check |
| 1N3N5DHG | Qualification set | 1.0, 0.5, 0.25 A plus holmium | 440 to 635 nm | Holder set | Instrument qualification |
| RM-1N2N3N | SRM 930e equivalent set | 10, 20, 30% T | 440 to 635 nm | Holder set | Absorbance scale |
| RM-N1N35N | SRM 1930 equivalent set | 1, 3, 50% T | 440 to 635 nm | Holder set | Absorbance scale |
| RM-1Q3Q9Q | SRM 2013a equivalent set | 10, 30, 90% T | 250 to 635 nm | Holder set | UV and visible scale |
Key Features
Absorptive, not reflective – NG glass attenuates through the bulk, so it does not add interference fringes to a spectrum
Four densities – OD 1.0 to 4.0 covers routine attenuation through to a stray light check
Certified at five wavelengths – 440, 465, 546.1, 590 and 635 nm match the wavelengths used in pharmacopoeial methods
One nanometre bandwidth – the certificate figures are measured at the bandwidth the method specifies
Stress free mounting – the glass sits in an aluminium holder without strain, so it does not become birefringent
Holmium and didymium – rare earth glass gives sharp peaks for a wavelength accuracy check in the visible
Bonded 1 A versions – a rare earth filter bonded to a neutral filter checks wavelength and absorbance in one pass
SRM equivalent sets – sets matched to SRM 930e, 1930 and 2013a let a lab work to the published specification
Applications
Attenuation and qualification are used at different points in an instrument check, so the part follows the check.
Spectrophotometer qualification – a set of glass references checks absorbance accuracy, linearity and wavelength accuracy in one sequence
Stray light and linearity checks – high density filters push the detector into the range where stray light shows up
Detector and lamp level setting – an absorptive filter drops the light level without changing the spectrum shape
Photometric standards are supplied alongside SUPERB PCB and SUPERB PCBA work when an analyser build needs its detector board and its qualification set together.
Sourcing and Cross-Reference
A working filter and a reference are ordered for different reasons, and mixing them is a common mistake. An OD 2.0 filter will attenuate light, but without a certificate naming the wavelength and bandwidth its absorbance value cannot be used to qualify an instrument. If the purpose is qualification, the certificate is the part that matters.
For references, three details decide whether a certificate is usable: the wavelengths, the spectral bandwidth, and the mounting. A value measured at 5 nm bandwidth cannot be compared with a method written for 1 nm, and a filter clamped in a screw holder can read differently from one mounted stress free. Both are stated on the parts listed here.
Replacement references are usually bought as a set rather than a single filter, because linearity is checked by comparing several values. We quote the individual filters and the sets that match the published SRM specifications, so a lab can either extend an existing set or buy into a specification it already works to.
Reference sets that ship with an instrument are packed through SUPERB Assembly with the detector board they qualify.
Q: Can you supply a certified reference set that matches the specification our method quotes?
A: Yes. SUPERB Tech quotes certified sets against the SRM specification named in the method and states the bandwidth on the certificate.
Why Choose SUPERB for Photometric Standards
- Certificate details stated: wavelengths and bandwidth are given, because a value without them cannot qualify an instrument
- Sets and single filters: a set can be extended one filter at a time instead of replaced
- Absorptive working filters: bulk absorbing glass attenuates without adding fringes to the spectrum
- Traceability named: the NIST specification behind each set is quoted so it can be checked against the method
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Bandpass and Interference Filters, Optical Windows and Sapphire Discs, Calibration Buffers and Conductivity Standards, Quartz Flow Cells and Cuvettes, Dichroic Filters and Beam Splitters, Obsolete Part Cross Reference Hub.