A grating sets the resolution, the range and the stray light of a monochromator, and the three are traded against each other. More grooves per millimetre give more dispersion and a shorter usable range. This page covers ruled reflective gratings at 300, 600 and 1200 grooves per millimetre, with 400 or 500 nm blaze, in 12.7, 25 and 50 mm sizes. The blaze angle, the dispersion and the efficiency at the blaze wavelength are listed for each density, so a grating is chosen on the range that has to be covered rather than on the groove count alone. Mounting and bench hardware are covered as a group, because a grating that is not held at the right angle loses more light than a lower cost grating that is.
quote now

Product Specifications

Product Overview

A ruled grating works by interference between light reflected from a large number of parallel grooves. The groove spacing sets the angle at which a wavelength leaves the grating, so more grooves per millimetre means a larger angle change per nanometre, which is more dispersion. That is why a 1200 groove grating resolves a narrow band and a 300 groove grating covers a wide band: the same rotation of the grating sweeps a shorter wavelength range when the dispersion is high.

The blaze is the second number. A ruled groove has a triangular profile, and the blaze wavelength is the wavelength at which the reflection from the groove facet is specular, which is where the efficiency peaks. A 500 nm blaze suits the visible range, and a 400 nm blaze is used where the ultraviolet end matters more. The efficiency falls away from the blaze, and the published figure for one family is 60 to 80 percent at the blaze wavelength, which is the number to use when the light budget is calculated.

The blaze angle follows from the groove spacing and the blaze wavelength. A 300 groove grating with a 500 nm blaze has a 4 degree 18 minute blaze angle and a dispersion of 3.32 nanometres per milliradian. A 600 groove grating has 8 degrees 37 minutes and 1.65, and a 1200 groove grating has 17 degrees 27 minutes and 0.80. The dispersion is the figure that converts a slit width and a focal length into a bandpass, so it is the one to use when a monochromator is specified.

The substrate and the coating decide the range and the handling. A soda lime glass substrate with an aluminium coating covers the visible and the near infrared, and an optional magnesium fluoride overcoat protects the aluminium, while a gold coating is used for the infrared. A bare aluminium coating is soft, so the grating is handled by the edges and never wiped. Ghosting from a ruled grating is specified below 0.5 percent of the parent line, which is acceptable for a monochromator and is the reason a holographic grating is chosen where a ghost would sit on a weak line.

Key Specifications

ParameterValueNote
Typeruled reflective, blazeda triangular groove profile, not holographic
Groove density300, 600 and 1200 grooves per mmthe visible range family
Blaze wavelength500 nm on the three densities, 400 nm on a 1200 groove partthe wavelength of peak efficiency
Blaze angle4 degrees 18 minutes at 300, 8 degrees 37 minutes at 600, 17 degrees 27 minutes at 1200follows from the spacing and the blaze
Dispersion3.32, 1.65 and 0.80 nm per mradthe figure that converts slit width into bandpass
Groove spacing at 1200 per mm833.3 nm1 divided by 1200 per mm
Diffraction order-1the order used in a monochromator
Efficiency at blaze60 to 80 percentmeasured in the Littrow configuration
Ghostingbelow 0.5 percent of the parent linethe reason a holographic grating is chosen for a weak line
Sizes12.7 x 12.7 x 6 mm, 25 x 25 x 6 mm, 50 x 50 x 9.5 mmthe 50 mm part is thicker
Substratesoda lime glassa low expansion substrate is used for a large part
Coatingaluminium, with optional magnesium fluoride or goldgold for the infrared, fluoride for protection
Wavelength range, visible family380 to 700 nm at 1200 grooves per mm with a 500 nm blazethe range narrows as the density rises
Handlingby the edges only, never wipeda bare aluminium coating is soft

Included Models and Order Codes

The table is grouped by groove density, because that is the choice that sets the range and the resolution. Each row gives the size, the blaze wavelength, the dispersion and the efficiency class, so a grating can be matched to an existing monochromator without a trial.

PartGrooves per mmBlazeSizeDispersionNote
GR13-0305300500 nm12.7 x 12.7 x 6 mm3.32 nm/mradwidest range, lowest dispersion
GR25-0305300500 nm25 x 25 x 6 mm3.32 nm/mradthe common monochromator size at this density
GR50A-0305300500 nm50 x 50 x 9.5 mm3.32 nm/mradlarge aperture, thicker substrate
GR13-0605600500 nm12.7 x 12.7 x 6 mm1.65 nm/mradthe general purpose density
GR25-0605600500 nm25 x 25 x 6 mm1.65 nm/mradgeneral purpose, larger aperture
GR50A-0605600500 nm50 x 50 x 9.5 mm1.65 nm/mradlarge aperture at 600 per mm
GR13-12051200500 nm12.7 x 12.7 x 6 mm0.80 nm/mradhighest dispersion, narrow range
GR25-12051200500 nm25 x 25 x 6 mm0.80 nm/mradhigh resolution monochromator
GR25-12041200400 nm25 x 25 x 6 mm0.80 nm/mradthe ultraviolet end of the same density

Groove Density and What It Costs

The density is the first choice and the blaze is the second. This table is the short form used when a monochromator is reviewed, and it shows the trade between range, resolution and light throughput.

DensityDispersionRange coveredResolutionBest for
300 per mm3.32 nm/mradthe widest range in the familylowesta scanning instrument that covers a full spectrum
600 per mm1.65 nm/mradabout half the range of 300mediumgeneral photometry and colorimetry
1200 per mm, 500 nm blaze0.80 nm/mrad380 to 700 nm at this blazehighesta narrow band around a visible line
1200 per mm, 400 nm blaze0.80 nm/mradshifted to the ultraviolet endhighestHPLC and clinical detectors at 214 or 280 nm
Holographic, same spacingsimilar dispersionsame rangeno ghosts, lower efficiencya weak line next to a strong one

Key Features

Three densities on one page – 300, 600 and 1200 grooves per mm, so the range and the resolution are traded deliberately

400 and 500 nm blaze options – the ultraviolet end and the visible end are both covered at the highest density

Dispersion stated per part – 3.32, 1.65 and 0.80 nm per mrad, which is the figure a slit width is calculated from

60 to 80 percent efficiency at blaze – a realistic light budget rather than a peak figure

Ghosting below 0.5 percent – known and stated, so a weak line next to a strong one is planned for

Three standard sizes – 12.7, 25 and 50 mm, with the 50 mm part on a thicker substrate

Aluminium coating with optional overcoat – magnesium fluoride for protection, gold for the infrared

Handle by the edges – a bare aluminium coating is soft, and a fingerprint is a permanent loss

Applications

A grating is chosen with the slit and the detector, in three settings that put the emphasis on a different number.

Scanning UV visible spectrophotometers – a 1200 groove grating with a 500 nm blaze covers 380 to 700 nm at 0.80 nanometres per milliradian. The 25 mm size is the common choice, and the 50 mm part is used where a larger aperture is needed to collect more light.

HPLC and clinical detectors – a 1200 groove grating with a 400 nm blaze puts the efficiency peak at the ultraviolet end, which is where a 214 or 280 nanometre detection channel sits. A 12.7 mm part fits a compact detector and a 25 mm part fits a bench instrument.

Colorimeters and process photometers – a 600 groove grating gives enough dispersion for a 10 nanometre bandpass over the visible range, and the lower density keeps the light throughput up on a low power lamp. A 300 groove part is used where the bandpass can be wider and the range has to be full.

The detector and lamp drive boards behind these instruments are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the CCD output to the shutter timing.

Sourcing and Cross-Reference

A grating is ordered by density, blaze, size and coating, and the size is the field that decides whether the existing mount can hold it. A 12.7 mm part and a 25 mm part do not share a mount, and a 50 mm part is 9.5 mm thick rather than 6 mm. We ask for the mount and the existing part number before naming a replacement.

Where a density is no longer supplied, the substitute is chosen on the dispersion first, because that is what sets the bandpass. A 1200 groove part and a 600 groove part at twice the focal length give the same bandpass in theory, but the second needs a longer instrument, so in practice the replacement keeps the density and changes the blaze. Where no equivalent exists at the same size, we say so rather than shipping a part that has to be re mounted.

Coatings and handling are part of the specification. A bare aluminium grating that has been wiped is a loss, and a magnesium fluoride overcoat is worth having on an instrument that is cleaned often. We ship in the maker package with the part number and the blaze marked, because two gratings of the same size and different blaze look identical.

Where a spectrometer is assembled with its grating, slit and detector, the optics are picked as a set, which is how SUPERB Assembly ships an optical bench.

Q: Can you supply a grating with the same density and a different blaze when the original part is discontinued?

A: SUPERB Tech can confirm the density, the blaze and the size against the old grating before the order is placed.

Why Buy Gratings from SUPERB

  1. Density, blaze and dispersion together: a grating is not offered on groove count alone, because the dispersion is what sets the bandpass
  2. Efficiency and ghosting stated: 60 to 80 percent at blaze and below 0.5 percent ghosting, so the light budget is realistic
  3. Sizes matched to the mount: 12.7, 25 and 50 mm, with the thickness given, so a replacement fits the existing holder
  4. Coating options: aluminium, magnesium fluoride or gold, chosen for the range and the cleaning routine
  5. Maker part and blaze on the delivery file: the two are written down, because parts of the same size and different blaze look the same

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Lamp Housings and Shutters for Photometric Instruments, Xenon and Deuterium Lamps for Photometric Instruments, Bandpass and Interference Filters for Photometric Channels, Flow Cell Holders and Cuvette Changers for Photometric Instruments, Turbidity and Photometric Detector Modules for Water Analysis, Component Lead Time Solutions.