Lamp Housings and Shutters for Photometric Instruments
Product Specifications
Product Overview
The first job of a housing is to keep the arc in the same place. A xenon arc is a few millimetres long and its position moves as the lamp ages, so a housing with a fixed socket and an adjustable mirror keeps the output stable while a housing that relies on the lamp base does not. The specification that matters is not the lamp wattage but the stability of the output, which for one precision housing is 0.3 percent peak to peak with a xenon lamp and 1.0 percent with a mercury xenon lamp.
The second job is collection. A rear reflector doubles the light that reaches the condenser, and the condenser sets the beam. An F/1 two element plano convex condenser gives a 33 mm collimated beam at a 1.5 inch flange, an F/2.2 asphere gives the same beam diameter with a narrower acceptance angle and less aberration, and an F/1 asphere gives the highest collection with a single element. The choice depends on the lamp arc size and on what the beam has to illuminate: a monochromator slit wants an F/1 beam, a fibre bundle wants a spot, and a sample cuvette wants a collimated beam that fills the aperture.
The third job is heat. An arc lamp puts most of its power into heat, and the housing has to remove it without putting a fan airstream across the optical path. A forced air housing with a baffled path keeps the outer surface below 40 C at a 25 C ambient, which is the figure that matters for a laboratory instrument. A housing without an interlock is a safety problem: the interlock stops the lamp when the cover is opened or when the internal temperature rises, and both conditions are part of the specification rather than an accessory.
The shutter sits at the exit or at an intermediate focus and its job is exposure, not dimming. A uni stable shutter needs power to stay open, so a power failure closes it, which is the safe state. The numbers that matter are the aperture, the transfer time on opening and the transfer time on closing. A 14 mm shutter opens in 1.5 milliseconds and closes in 3.5 milliseconds; a 6 mm laser shutter opens in 0.7 milliseconds and closes in 0.8; a 45 mm shutter takes 14 milliseconds. A shutter that is much faster than the measurement adds no value, and a shutter that is slower than the exposure sets the minimum exposure.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Lamp types | xenon, mercury and mercury xenon DC arc lamps | deuterium lamps are covered on a separate page |
| Lamp wattage, arc housings | 50 to 500 W | the 450 to 1000 W range uses a larger housing family |
| Condenser options | F/1 two element plano convex, F/1.5 plano convex, F/1 asphere, F/2.2 asphere | the condenser sets the beam and the collection angle |
| Output beam | 33 mm collimated at the 1.5 inch flange | the flange size determines the coupled beam |
| Lens transmission | 200 to 2500 nm on fused silica | below 200 nm the lens absorbs |
| Housing construction | aluminium body with igniter, condenser, rear reflector and cooling fan | the igniter is inside the housing |
| Output stability, precision housing | 0.3 percent peak to peak with a xenon lamp, 1.0 percent with a mercury xenon lamp | the figure that decides a photometric baseline |
| Lamp trigger | 25 kV typical on the precision housing | built in starter |
| Surface temperature | below 40 C at a 25 C ambient | forced air cooling, baffled path |
| Guaranteed ambient | 5 to 35 C, operating 0 to 40 C | precision housing |
| Interlock | stops the lamp on an open cover or an over temperature | part of the housing, not an add on |
| Fiber adapter | SMA adapter for a precision housing | A11096 |
| Shutter aperture range | 2 to 150 mm | matched to the beam diameter |
| Shutter transfer time | 0.3 ms at 2 mm, 1.5 ms at 14 mm, 14 ms at 45 mm | opening time, closing is slower on the larger sizes |
| Shutter driver | VCM-D1 for most uni stable shutters, D880C or VED24 for the alternative coils | the driver is named with the shutter |
Included Models and Order Codes
The first block is housings, the second is shutters and the third is the adapter and driver. A housing is ordered by lamp wattage and condenser, and a shutter by aperture and speed, so the two are quoted as a pair when they leave the same instrument.
| Part | Type | Aperture or beam | Speed or rating | Note |
|---|---|---|---|---|
| 66902 | Arc lamp housing, F/1 two element | 33 mm collimated beam | 50 to 500 W | 1.5 inch flange, fused silica 200 to 2500 nm |
| 66901 | Arc lamp housing, F/1.5 plano convex | 33 mm collimated beam | 50 to 500 W | narrower acceptance, lower cost |
| 67001 | Arc lamp housing, F/2.2 asphere | 33 mm collimated beam | 50 to 500 W | lowest collection, least aberration |
| 67005 | Arc lamp housing, F/1 asphere | 33 mm collimated beam | 50 to 500 W | single element, highest collection |
| E7536 | Precision lamp housing | 30 mm lens and R30 mirror | 0.3 percent stability with xenon | 25 kV starter, SMA adapter A11096 |
| A11096 | SMA fiber adapter | for the E7536 exit | not applicable | couples a fiber bundle to the housing |
| LS2 | Laser shutter | 2 mm | 0.3 ms opening | fastest in the range |
| LS3 | Laser shutter | 3 mm | 0.5 ms opening | between LS2 and LS6 |
| LS6 | Laser shutter | 6 mm | 0.7 ms opening, 0.8 ms closing | 5 W per square mm, 20 Hz continuous |
| ES6B | Laser shutter, long life | 6 mm | 1.9 ms opening | for a continuous duty laser line |
| VS14 | Optical shutter | 14 mm | 1.5 ms opening, 3.5 ms closing | 12 ohm coil, +65 V, 0 to 80 C |
| VS25 | Optical shutter | 25 mm | 3.0 ms opening | the mid size laboratory shutter |
| VS35 | Optical shutter | 35 mm | 13 ms opening | for a wider beam |
| CS25 | Optical shutter, long life | 25 mm | 9 ms opening | longer life than the VS series |
| CS35 | Optical shutter, long life | 35 mm | 13 ms opening | for a monochromator exit |
| CS45 | Optical shutter, long life | 45 mm | 14 ms opening | larger beam, higher moving mass |
| VCM-D1 | Shutter driver | not applicable | +65 V open, +5 V hold | the standard driver for the uni stable shutters |
Shutter Selection by Aperture
The aperture has to be larger than the beam and not much larger. This table gives the opening time for each aperture in the range, which is the figure that decides whether a shutter suits an exposure or only a dark period.
| Aperture | Series | Opening time | Typical use |
|---|---|---|---|
| 2 mm | LS2 | 0.3 ms | a focused laser beam |
| 3 mm | LS3 | 0.5 ms | a small beam or a fiber output |
| 6 mm | LS6, ES6B | 0.7 ms, 1.9 ms | a laser line or a fiber bundle |
| 14 mm | VS14 | 1.5 ms | a collimated beam from a small condenser |
| 25 mm | VS25, CS25 | 3.0 ms, 9 ms | a monochromator entrance or exit |
| 35 mm | VS35, CS35 | 13 ms both | a wider collimated beam |
| 45 mm | CS45 | 14 ms | a large aperture instrument |
| 65 mm | CS65 | 45 ms | an imaging or projection path |
| 90 mm | CS90HS | 48 ms | the largest standard optical shutter |
| 150 mm | MS150B | per the maker data | a very large beam or a safety shutter |
Key Features
Arc housings for 50 to 500 W – xenon, mercury and mercury xenon lamps in one housing family, with the socket adapter as the only change
Four condenser options – F/1 doublet, F/1.5 plano convex, F/1 asphere and F/2.2 asphere, so the beam is chosen for the slit or the sample
33 mm collimated output – a defined beam at the 1.5 inch flange, which is what a monochromator or a fiber adapter expects
0.3 percent output stability – the precision housing figure with a xenon lamp, which sets the photometric baseline
Built in starter and interlock – 25 kV trigger inside the housing and a lamp stop on an open cover or an over temperature
Surface below 40 C – forced air cooling with a baffled path, so the housing does not heat the sample compartment
Shutters from 2 to 150 mm – with the opening time on every row, so a shutter is chosen on speed and not only on size
Driver named with the shutter – VCM-D1 and the alternative coils are listed, so a shutter is not ordered without a driver
Applications
A housing and a shutter are chosen together, in three settings that put different demands on the beam.
UV visible spectrophotometers – an F/1 doublet condenser feeds a monochromator slit, and a 25 or 35 mm shutter sits at the exit. The stability of the housing matters more than the wattage, because a one percent drift in lamp output looks like a one percent drift in absorbance.
Fluorescence and microplate readers – a xenon flash lamp or a continuous xenon lamp is coupled to a fiber bundle through an SMA adapter. The shutter has to open faster than the integration time, so a 6 mm laser shutter is used where a 25 mm optical shutter would set the minimum exposure.
Imaging and machine vision – a large aperture shutter blanks a projection path between frames. A 45 or 65 mm shutter is used, and the closing time rather than the opening time sets the frame gap, because a shutter that is still closing when the next exposure starts leaves a smear.
The lamp drive and shutter control boards in these instruments are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the trigger circuit to the exposure timer.
Sourcing and Cross-Reference
A housing is ordered by lamp wattage, condenser and flange, and the flange is the field that decides whether it couples to the rest of the instrument. A 1.5 inch flange with a 33 mm beam and a 2 inch flange are not interchangeable without an adapter, and an adapter that is 10 mm too long moves the focus. We ask for the existing flange and the focal distance before naming a housing.
Where a lamp is no longer supplied, the housing is kept and the socket adapter is changed if the new lamp has the same arc size. If the arc is a different length, the condenser has to change as well, and the substitute is chosen on the beam rather than on the lamp base. Where no equivalent exists, we say so rather than shipping a lamp that starts but does not focus.
Shutters are a life limited part. A shutter that runs at 40 Hz for a year will wear its blades, and the replacement is chosen on the same aperture and the same driver so the wiring does not change. A larger shutter on the same driver draws a different current and can hold open slowly, which shows up as a longer exposure rather than as a failure.
Where an optics module is assembled with its housing, shutter and adapter, the set is picked together, which is how SUPERB Assembly ships a light source module.
Q: Can you supply a replacement shutter with the same aperture and driver when the original part is discontinued?
A: SUPERB Tech can confirm the lamp type, the flange and the shutter aperture against the old assembly before the order is placed.
Why Buy Lamp Housings and Shutters from SUPERB
- Beam and flange quoted together: a housing is matched on the output beam and the flange, not only on the lamp wattage
- Stability stated: the 0.3 percent peak to peak figure is given for a xenon lamp, because it is what a photometric baseline depends on
- Condenser options on one page: F/1, F/1.5, F/2.2 and asphere versions, so the collection angle is chosen for the slit or the sample
- Shutter speed on every row: the opening time is given with the aperture, so a shutter is not chosen on size alone
- Driver and adapter listed: the shutter driver and the fiber adapter are named with the part, so the set is complete
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Xenon and Deuterium Lamps for Photometric Instruments, UV and Visible LED Sources for Photometric Detection, Monochromator Gratings and Optical Bench Parts, Flow Cell Holders and Cuvette Changers for Photometric Instruments, Bandpass and Interference Filters for Photometric Channels, Component Lead Time Solutions.