Xenon and Deuterium Lamps for Photometric Instruments
Product Specifications
Product Overview
Two lamp types cover most photometric instruments, and they are not interchangeable in either spectrum or drive. A deuterium lamp runs an arc in deuterium gas at low pressure and emits a molecular continuum that rises through the UV and fades past 400 nm. It is the standard source for absorbance between 190 and 400 nm, which is where most of the organic chromophores and most of the pharmacopoeial methods sit. A xenon short arc lamp runs a high pressure arc in xenon and emits a continuum from 185 nm to about 2,000 nm with a very small source. It is the standard source where a compact, bright point is needed and where the near infrared also matters.
The window is the first thing to read on a data sheet. A deuterium lamp with UV glass transmits from 185 nm, one with synthetic silica from 160 nm, and one with a magnesium fluoride window from 115 nm. The arc inside is the same; the envelope is what changes. That matters when an instrument is specified at 190 nm and then fitted with a lamp that is quoted from 185 nm, because the margin at the short end is then a window question rather than a lamp question. The same logic applies at the long end of a xenon lamp, where the fused silica envelope and the low ozone version of it differ in the deep UV.
Drive is the second thing, and it is where replacement lamps go wrong. A deuterium lamp is run at a fixed anode current, 300 mA on most analytical bodies, with a tube drop of 75 to 85 V, and it needs a starter that can deliver 350 to 500 V to break down the gas. It also has a filament that is warmed at 2.5 V for about 20 seconds and then held at 1.0 V while the lamp runs. A xenon lamp needs a trigger of 15 to 20 kV and a supply that limits current after the arc strikes. A lamp with the right wattage and the wrong trigger simply does not light, and a supply with the right voltage and the wrong current shortens the arc life.
Life is quoted in a way that surprises people. The guaranteed figure is the time at which output at 230 nm falls to half of its initial value, or the time at which the short term fluctuation exceeds a stated limit. A lamp at that point still lights and still looks bright to the eye; it is simply half as bright as it was in the region the method uses. For a ratio instrument the change is partly corrected, for a single beam instrument it is not, so the replacement interval is usually set from lamp hours rather than from a reading. Warm up is the other half of the same story: a deuterium body needs 20 to 30 minutes to reach thermal equilibrium, and a measurement taken at five minutes is a measurement of a lamp that is still moving.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Spectral range, deuterium | 115 to 400 nm on a fluoride window, 160 to 400 nm on synthetic silica, 185 to 400 nm on UV glass | the arc is the same, the envelope sets the short end |
| Spectral range, xenon | 185 to 2,000 nm | low ozone silica extends the deep UV at the cost of some output |
| Deuterium anode current | 300 mA dc, plus or minus 30 mA on the L2D2 bodies | the current is regulated, not the voltage |
| Deuterium tube drop | 75 to 85 V dc depending on the body | a 300 mA lamp therefore runs at about 24 W |
| Deuterium starting voltage | 350 V dc on most bodies, 500 V dc on the water cooled 200 W type | the starter has to reach this before the arc strikes |
| Filament sequence | 2.5 V at 4 to 5.8 A for about 20 s, then 1.0 V at 1.8 A, or 3 to 6 V at 0.3 to 0.45 A on the long life body | a supply built for one sequence will not run the other |
| Aperture | 0.5 mm or 1.0 mm on the deuterium bodies | 0.5 mm suits a narrow slit and a small detector |
| Xenon ratings | 75 W, 100 W, 150 W and 300 W | arc length runs 1.3 mm to 3.0 mm with the rating |
| Xenon voltage and current | 15 to 20 V dc at 5.4 to 15 A | the supply is a current limited source, not a fixed voltage |
| Xenon trigger | 15 kV on the 75 W body, 20 kV on the 150 W bodies | a 20 kV starter will not harm a 15 kV lamp, the reverse is not true |
| Output noise | 0.005 percent peak to peak on a long life deuterium body, 1 percent peak to peak maximum on xenon | noise is often the supply rather than the lamp |
| Drift | plus or minus 0.5 percent per hour on the long life deuterium bodies and on xenon | a ratio instrument corrects part of it, a single beam does not |
| Guaranteed life | 1,000 to 2,000 h on the analytical bodies, 4,000 h on the long life series, 500 h on the vacuum UV types | end of life is 50 percent of initial output at 230 nm |
| Warm up | 20 to 30 minutes to thermal equilibrium on a deuterium body | a reading taken earlier is taken on a moving lamp |
| Cooling | convection on the 75 W to 150 W xenon bodies, forced air on the 300 W body, water on the 200 W deuterium type | the cooling method is part of the lamp rating, not an accessory |
Included Models and Order Codes
The table is split by lamp type, deuterium first and xenon second, because the two are chosen for different ranges and cannot be compared line by line. Deuterium lines carry the window material and the aperture, which are the two facts that decide whether a body fits an existing housing. Xenon lines carry the arc length and the trigger voltage, which decide whether the optics and the starter can accept it.
| Lamp | Type | Spectral range | Aperture or arc | Drive | Life | Note |
|---|---|---|---|---|---|---|
| Hamamatsu L6565 | Deuterium, UV glass | 185 to 400 nm | 1.0 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | standard body for a bench spectrophotometer |
| Hamamatsu L6301 | Deuterium, synthetic silica | 160 to 400 nm | 1.0 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | deeper UV than the glass window body |
| Hamamatsu L6302 | Deuterium, synthetic silica | 160 to 400 nm | 0.5 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | small aperture for a narrow slit |
| Hamamatsu L7293 | Deuterium, magnesium fluoride | 115 to 400 nm | 1.0 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | vacuum UV, 350 V dc starting |
| Heraeus DX 202 R, 80001131 | Deuterium, fused silica | 185 to 400 nm | 1.0 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | 0.005 percent noise, 2.5 V 5.8 A warm up |
| Heraeus DX 201/05 RJ, 80008685 | Deuterium, fused silica | 185 to 400 nm | 0.5 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | same body as DX 202 R with a small aperture |
| Heraeus DX 246, 80006934 | Deuterium, fused silica | 185 to 400 nm | 1.0 mm aperture | 300 mA, 80 V | 2,000 h guaranteed | 10 to 12 V 0.8 to 1.0 A warm up supply |
| Heraeus DO 605 T, 80052430 | Deuterium, synthetic silica | 160 to 400 nm | 1.0 mm aperture | 300 mA, 75 V | 1,000 h guaranteed | higher intensity, shorter life than DX |
| Heraeus DO 660/05 J, 45002349 | Deuterium, synthetic silica | 160 to 400 nm | 0.5 mm aperture | 300 mA, 85 V | 1,000 h guaranteed | narrow slit version of the high intensity body |
| Heraeus D 200 F, 56001671 | Deuterium, water cooled | 160 to 400 nm | 1.0 mm aperture | 0.9 to 1.8 A, 110 V | 1,000 h guaranteed | 200 W class, needs the cooling jacket |
| Hamamatsu L2174 | Xenon short arc | 185 to 2,000 nm | 1.3 mm arc | 5.4 A, 15 V | 1,000 h guaranteed, 2,000 h average | 75 W, 15 kV trigger, 15 g, convection cooled |
| Hamamatsu L2175 | Xenon short arc | 185 to 2,000 nm | 2.5 mm arc | 7.5 A, 19 V | 1,200 h guaranteed, 2,500 h average | 150 W, 20 kV trigger, 45 g |
| Hamamatsu L2273 | Xenon short arc, quiet cathode | 185 to 2,000 nm | 2.0 mm arc | 8.5 A, 17 V | 1,800 h guaranteed, 3,000 h average | 150 W, 20 kV trigger, 45 g, stable arc point |
| Hamamatsu L11307 | Xenon short arc | 185 to 2,000 nm | 1.3 mm arc | 7.0 A, 15 V | not published in the sheet read | 100 W, small source for a compact bench |
| Hamamatsu L2479 | Xenon short arc | 185 to 2,000 nm | 3.0 mm arc | 15 A, 20 V | not published in the sheet read | 300 W, forced air cooling |
Reading the Life and Output Numbers
Four numbers decide whether a lamp will behave in an instrument: the life definition, the noise, the aperture and the trigger. This table is the short form used when a data sheet is read against a method.
| Number | What it means | What to check before ordering |
|---|---|---|
| Guaranteed life | hours until output at 230 nm falls to 50 percent of the initial value, or until fluctuation exceeds the stated limit | the lamp still lights at that point, so a bench test will not show it |
| Average life | typical value across a population, usually 1.5 to 2 times the guaranteed figure | plan replacement from lamp hours, not from the average |
| Noise, peak to peak | short term variation, 0.005 percent on a long life body and up to 1 percent on xenon | a noisy trace is often the supply or the arc position, not an old lamp |
| Drift | long term change per hour, plus or minus 0.5 percent on the bodies listed here | a ratio instrument corrects it, a single beam instrument shows it |
| Aperture | the emitting area that the optics see, 0.5 mm or 1.0 mm on deuterium | a 0.5 mm body in a wide slit loses light; a 1.0 mm body in a narrow slit wastes it |
| Arc length | the source size for imaging, 1.3 mm to 3.0 mm on xenon | a 3 mm arc does not image into a 1 mm slit without loss |
| Trigger voltage | 15 or 20 kV on xenon, 350 to 500 V on deuterium | the starter is matched to the lamp, not to the wattage |
| Warm up | 20 to 30 minutes to thermal equilibrium on deuterium, minutes on xenon | a method run immediately after power on is run on a moving baseline |
Key Features
Window chosen for the short end – UV glass at 185 nm, synthetic silica at 160 nm and magnesium fluoride at 115 nm on the same arc body, so the range is a window decision
Aperture of 0.5 mm or 1.0 mm – two emitting areas in one family, which lets a narrow slit instrument and a wide slit instrument share a housing
Long life body at 2,000 h and a 4,000 h series – the analytical bodies are quoted at 2,000 h guaranteed, and the extended series doubles that
Xenon from 75 W to 300 W – arc length from 1.3 mm to 3.0 mm, so a compact bench and a larger instrument are covered by one range
Trigger voltage on the sheet – 15 kV or 20 kV stated per body, because a starter that cannot reach it leaves the lamp dark
Two filament sequences supported – 2.5 V warm up at 4 to 5.8 A, or 10 to 12 V at 0.8 to 1.0 A, so an existing supply can be matched
Water cooled 200 W deuterium type – a high power body with a cooling jacket for instruments that need more UV output than a 30 W arc gives
Life defined against a wavelength – the 50 percent figure is at 230 nm and 300 mA, which is the region a UV method actually uses
Applications
These lamps sit at the front of an optical bench, and the bench is built around the source size and the spectrum. The three cases below are the ones that fix the lamp type, the window and the drive.
UV visible spectrophotometry – a deuterium lamp for 190 to 400 nm and a tungsten halogen lamp for the visible, switched at about 360 nm. The 2,000 h body is the usual replacement, and the aperture is chosen against the slit width of the monochromator.
HPLC and flow analysis detectors – a deuterium body with a 0.5 mm aperture and a low volume flow cell. The lamp is run at a fixed 300 mA and the drift figure matters more than the absolute output, because the detector measures a ratio.
Water, gas and process photometers – a xenon flash or continuous arc where a compact source and a near infrared path are needed, or a deuterium body where the method is a UV absorbance at a single wavelength. The trigger voltage and the cooling method decide whether the housing can take the lamp.
The lamp driver and the detector board behind these sources are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the high voltage starter to the transimpedance stage.
Sourcing and Cross-Reference
A lamp is ordered by body number, not by wattage, because the same wattage exists in several envelopes with different windows and apertures. The housing, the slit and the starter are the fixed points, and the lamp is chosen to fit them rather than the other way round.
Where a lamp is obsolete, the drive is the constraint. A body with the same spectrum and a different filament sequence will not run on the existing supply, and a xenon lamp with the same wattage and a different trigger will not strike. We compare the current, the warm up voltage and the trigger before naming a replacement.
Ageing stock is a real risk with lamps, so the date code and the storage are part of the question. A lamp that has been on a shelf for years is not the same as one from the current batch, and the guaranteed life runs from first strike rather than from the invoice date. We state the batch and the date code on the delivery file.
A lamp, its housing gasket and the connector are ordered as one set when an instrument is down, which is how SUPERB Assembly ships a replacement.
Q: Can you match a deuterium lamp by window material, aperture and filament sequence when the original body is discontinued?
A: SUPERB Tech can compare the window, the aperture and the filament sequence against the old body before an order is released.
Why Buy Lamps from SUPERB
- Window and aperture asked first: the short wavelength limit and the emitting area decide whether a body fits the instrument, so both are confirmed before a part is named
- Drive figures on every line: current, voltage, warm up and trigger appear together, which is what stops a lamp that fits the housing and not the supply
- Life defined at a wavelength: the 50 percent point at 230 nm is quoted as the maker states it, not as a round number
- Batch and date code on the paperwork: a lamp is a consumable with a shelf history, and the file says which batch was shipped
- Deuterium and xenon on one list: a single source for the UV body and the xenon arc, which keeps an instrument build on one purchase order
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Bandpass and Interference Filters for Photometric Channels, UV and Visible LED Sources for Photometric Detection, Quartz Flow Cells and Cuvettes for Photometric Instruments, Turbidity and Photometric Detector Modules for Water Analysis, High Power LED Emitters for Illumination and Infrared, Component Lead Time Solutions.