An electrode that drifts is usually blamed on the glass, but on a process analyser the reference is the part that fails first. The parts below cover steam sterilizable in-line pH sensors with a pressurised reference, sealed gel electrodes for sampling lines, and platinum or gold ORP electrodes for redox control. Each entry lists pH or mV range, temperature and pressure limits, junction type and connector, so the electrode is matched to the process before it is matched to the meter. Slope, offset and response time are given where the maker publishes them, because a two point buffer check only proves the electrode still answers.
quote now

Product Specifications

Product Overview

A pH loop on an analyser fails in three ways, and only one of them is the glass. The bulb can crack, which is obvious. The reference can be poisoned by the process, which shows up as a slow drift that keeps coming back after every calibration. Or the junction can plug, which shows up as a reading that is correct in buffer and wrong in the sample. The second and third failures are the common ones on a plant, and both are answered by the reference construction rather than by the sensing glass.

The reference is a small electrochemical cell of its own. A gel reference is sealed and needs no maintenance, which suits a sampling line or a bench check, but it cannot be refilled and its junction is the only path between the process and the silver wire. A pressurised liquid reference pushes electrolyte out through the junction, so the process cannot push back in. That is why a steam sterilizable in-line sensor can survive repeated 140 C cycles while a sealed gel electrode used at the same temperature dies in a week.

Temperature is the second number to read. A glass electrode has a theoretical slope of 59.16 mV per pH unit at 25 C, and that slope changes with temperature, so a sensor with no built-in temperature element depends on the transmitter to compensate. Above pH 12.3 the glass also starts to answer sodium ions as if they were hydrogen, which is why the maker prints an error figure at that point instead of promising 0 to 14 everywhere.

An ORP electrode is a different measurement with a different answer. It is a bare platinum or gold surface with no glass membrane, so it does not need calibration in the pH sense and it does not drift with temperature in the same way. What it does need is a reference, and the reference decides the number. An Ag/AgCl reference reads about 40 mV higher than a calomel reference in the same solution, so a setpoint copied from a plant using calomel will be wrong by that offset until it is corrected.

Key Specifications

ParameterValueNote
pH range0 to 14 pH on sealed laboratory electrodes, 0 to 12 pH on the steam sterilizable in-line familysodium error starts above pH 12.3 on the glass used here
ORP rangeplus or minus 2000 mVthe range belongs to the transmitter as much as to the electrode
Slope97 percent of theoretical or higher59.16 mV per pH unit at 25 C is the theoretical figure
Offsetplus or minus 0.20 pHchecked in pH 7 buffer before a two point calibration
Isopotential pointpH 7.00 at 0 mVthe point where temperature has least effect on the reading
Response95 percent in 1 s on a bulb, 95 percent in 5 s on a flat surfacea slow answer after cleaning means a plugged junction
Temperature range0 to 60 C on polycarbonate bodies, 0 to 80 C on Ultem bodies, 0 to 100 C on the pressurised in-line familysterilization is rated separately at up to 140 C
Pressure70 psig on the Ryton submersion family, 100 psig at 75 C on the threaded Ryton family, 4 barg (58 psig) on the in-line familypressure and temperature are one rating, not two
Reference systemsealed gel 3.5 M KCl, or pressurised liquid electrolyte with a ceramic diaphragm and a silver ion trapthe pressurised reference is what survives steam cycles
Junctionssingle Pellon, double Pellon, HDPE with Pellon, ceramica double junction keeps the process out of the reference wire
Body materialspolycarbonate, Ultem, epoxy, Ryton, glassUltem and Ryton for solvent and hot service
Process connectionsPg 13.5 thread, 1/2 inch MNPT, 3/4 inch MNPT, DIN 13.5 cap, 12 mm gland fittingsthe gland has to match the 12 mm body
ConnectorsBNC, K8S watertight, tinned leads, DIN cap with mating plugK8S is the one that stays dry on a washdown skid
Insertion length120, 225, 325 and 425 mm on the in-line family; 150 mm on the 12 mm laboratory bodylength follows the pipe tee, not the cable run
ORP sensor metalplatinum or goldplatinum for general redox, gold where sulphur compounds coat platinum
ORP reference offsetabout 40 mV higher for Ag/AgCl than for calomel in the same solutioncorrect the setpoint when swapping reference types

Included Models and Order Codes

The table is split by duty rather than by brand. The in-line family at the top is the one for a pipe tee on a skid, with a pressurised reference, a watertight connector and a sterilizable membrane. The sealed laboratory electrodes in the middle are the ones for a sampling port or a bench meter, and they are the cheapest way to keep a second measurement point without a second transmitter. The threaded Ryton and epoxy bodies at the bottom are the submersion and inline options for water and waste lines, where the junction has to be replaced long before the glass wears out.

PartTypepH rangeTemperaturePressureConnectionNote
InPro3253i/SG/120in-line, pressurised reference, digital0 to 120 to 100 C0 to 4 bargPg 13.5, K8S120 mm shaft, platinum solution ground, sterilizable to 140 C
InPro3253i/SG/225in-line, pressurised reference, digital0 to 120 to 100 C0 to 4 bargPg 13.5, K8S225 mm shaft, same membrane and reference
InPro3253i/SG/325in-line, pressurised reference, digital0 to 120 to 100 C0 to 4 bargPg 13.5, K8S325 mm shaft for a larger tee
InPro3253i/SG/425in-line, pressurised reference, digital0 to 120 to 100 C0 to 4 bargPg 13.5, K8S425 mm shaft, longest of the family
S200C/BNClaboratory, sealed gel, single junction0 to 140 to 80 Cnot rated for line pressureBNC, 1 m cable12 x 150 mm Ultem body, 97 percent slope
S200CD/BNClaboratory, sealed gel, double junction0 to 140 to 80 Cnot rated for line pressureBNC, 1 m cabledouble junction for samples of unknown composition
S150C/BNClaboratory, sealed gel, single junction0 to 140 to 60 Cnot rated for line pressureBNC, 1 m cablepolycarbonate body, lowest cost of the group
S265Csubmersion or in-line, Ryton body0 to 140 to 80 C70 psig1/2 or 3/4 inch MNPT, tinned leads10 ft cable, threaded nipple body
S271CDsubmersion, Ryton body, double junction0 to 1475 C100 psig1/2 or 3/4 inch MNPT10 ft waterproof cable
S272CDsubmersion, Ryton body, double junction0 to 1481 C85 psig1/2 or 3/4 inch MNPTthe pressure figure drops as the temperature rises
S354CDin-line, flat glass, DIN cap0 to 140 to 80 Cline pressure of the fittingDIN 13.5 thread, DIN capflat surface resists coating in slurry
S224CDin-line, spherical glass, DIN 13.5 cap0 to 140 to 80 Cline pressure of the fittingDIN 13.5, S6 capmounts without a separate gland

ORP Electrodes and Redox Reference

ORP parts are listed apart because the choice turns on the metal and the junction, not on a glass membrane. Platinum is the general purpose surface and gold is the one for streams where sulphur compounds coat platinum. The reference behind that metal sets the number on the display, so the last column notes what changes when the reference type changes.

PartMetalJunctionBodyTemperatureNote
S500C-ORP/BNCplatinum, polished tipsingle PellonUltem0 to 80 Cplus or minus 2000 mV, 12 x 150 mm, BNC
S500C-ORP-AU/BNCgold, polished tipsingle PellonUltem0 to 80 Cgold surface for sulphur-bearing streams
S550CD-ORP/BNCplatinum, extended tipdouble HDPE and PellonUltem0 to 100 Chighest temperature of the laboratory group, teeth guard
ORP1000platinumsingle Pellonpolycarbonate0 to 60 Cbasic cell, swimming pool and aquarium duty
ORP2000platinumdouble PellonUltem0 to 80 Cfor samples of unknown composition
ORP3000platinumdouble PellonUltem0 to 80 Cremovable bulb guard and strain relief
S150C-ORP/BNCplatinumsingle Pellonpolycarbonate0 to 60 C30 inch cable, slot guard

Key Features

Pressurised reference for steam cycles – electrolyte is held above process pressure so the junction keeps flowing outward during a 140 C sterilization cycle

Silver ion trap in the reference – the Argenthal cartridge keeps silver from the wire out of the junction, which is what stops the drift that follows a hot cycle

Double junction option – an HDPE or Pellon second junction keeps sulphide, protein and solvent out of the reference wire on dirty samples

Body material matched to the reagent – polycarbonate for clean water, Ultem and Ryton for solvents and hot caustic, glass where nothing else survives

Digital temperature element inside – the in-line family carries its own temperature sensor, so compensation follows the process instead of a neighbouring probe

Platinum dot as solution ground – the same metal that measures redox also grounds the solution, which removes the electrostatic noise that shows up as a wandering last digit

Flat and spear surfaces – a flat glass surface on the S354CD resists coating in slurry, and a spear tip on the S175CD reaches into semi solids

Standard 12 mm body – the laboratory and submersion electrodes share a 12 mm diameter, so one gland fits the whole set

Applications

These electrodes sit on the measurement point of an analyser, close to the sample and often close to the pump that moves it. The three cases below are the ones that decide the specification.

Water and waste analysers – pH on an aeration basin or a final effluent line, where the junction fouls with biofilm and the pressure is low. A Ryton body with a double junction and a 3/4 inch thread is the working choice, and the flat surface version is used where the sample carries grit.

Bioreactor and sterilizable skids – pH inside a fermentation vessel that is cleaned and sterilized in place. The pressurised reference and the 140 C rating are the reason this family exists, and the digital temperature element removes the separate probe from the vessel head.

Sampling and reagent lines – pH or redox checked at a sampling port before a colorimetric step. A sealed gel electrode with a BNC plug does the job at low cost, and a double junction version is used when the sample contains protein or solvent that would poison a single junction.

These electrodes are sourced for the analyser builds that SUPERB PCB and SUPERB PCBA programmes support, from the sample port to the transmitter board.

Sourcing and Cross-Reference

Electrode bodies, cables and connectors are the items that stop a line, because a cracked bulb cannot be repaired and a K8S plug cannot be replaced with a BNC. We hold the electrode, the gland, the cable and the buffer as one group so a replacement arrives as a set rather than as one part and three questions.

For analysers that are no longer supported, the electrode is often the only part still available while the transmitter is obsolete. We cross reference the old part number against the current body and connector, and we say plainly when the connector has changed and the cable has to be replaced with it.

Redox setpoints copied from an older plant need the reference type checked before anything else. A setpoint that came from a calomel reference reads about 40 mV away from an Ag/AgCl reference, and that difference is bigger than most control bands.

A replacement electrode is often ordered with the cable and gland as one kit, which is how SUPERB Assembly ships it when a line is down.

Q: Can you supply a replacement electrode for an analyser whose original part is discontinued?

A: SUPERB Tech can cross reference the discontinued number against the current body, junction and connector before the order is placed.

Why Buy pH and ORP Electrodes from SUPERB

  1. One group, one delivery: electrode, gland, cable, buffer and the transmitter connector are quoted together, so the line restarts with one shipment
  2. Range read before the quote: we ask for pH or mV range, temperature, pressure and the reagent before we name a part, because a 0 to 14 laboratory electrode is the wrong answer for a 100 C line
  3. Connector checked against the meter: BNC, K8S, DIN cap and tinned leads are not interchangeable, and the cable is ordered with the plug the transmitter already has
  4. Obsolete and hard to find parts: discontinued electrode numbers are cross referenced against the current body and junction rather than replaced with a lookalike
  5. Documentation as shipped: maker data sheet and the lot number travel with the electrode, which is what an audit asks for on a pH loop

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen, Conductivity Cells and Toroidal Sensors, Liquid Flow Sensors for Medical and Analytical Instruments, Air Bubble Detectors for Infusion and Reagent Lines, Peristaltic and Diaphragm Micro Pumps for Analysers, Sensors and Detectors, Obsolete Part Cross Reference.