pH and ORP Electrodes for Process Analysers
Product Specifications
Product Overview
A pH loop on an analyser fails in three ways, and only one of them is the glass. The bulb can crack, which is obvious. The reference can be poisoned by the process, which shows up as a slow drift that keeps coming back after every calibration. Or the junction can plug, which shows up as a reading that is correct in buffer and wrong in the sample. The second and third failures are the common ones on a plant, and both are answered by the reference construction rather than by the sensing glass.
The reference is a small electrochemical cell of its own. A gel reference is sealed and needs no maintenance, which suits a sampling line or a bench check, but it cannot be refilled and its junction is the only path between the process and the silver wire. A pressurised liquid reference pushes electrolyte out through the junction, so the process cannot push back in. That is why a steam sterilizable in-line sensor can survive repeated 140 C cycles while a sealed gel electrode used at the same temperature dies in a week.
Temperature is the second number to read. A glass electrode has a theoretical slope of 59.16 mV per pH unit at 25 C, and that slope changes with temperature, so a sensor with no built-in temperature element depends on the transmitter to compensate. Above pH 12.3 the glass also starts to answer sodium ions as if they were hydrogen, which is why the maker prints an error figure at that point instead of promising 0 to 14 everywhere.
An ORP electrode is a different measurement with a different answer. It is a bare platinum or gold surface with no glass membrane, so it does not need calibration in the pH sense and it does not drift with temperature in the same way. What it does need is a reference, and the reference decides the number. An Ag/AgCl reference reads about 40 mV higher than a calomel reference in the same solution, so a setpoint copied from a plant using calomel will be wrong by that offset until it is corrected.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| pH range | 0 to 14 pH on sealed laboratory electrodes, 0 to 12 pH on the steam sterilizable in-line family | sodium error starts above pH 12.3 on the glass used here |
| ORP range | plus or minus 2000 mV | the range belongs to the transmitter as much as to the electrode |
| Slope | 97 percent of theoretical or higher | 59.16 mV per pH unit at 25 C is the theoretical figure |
| Offset | plus or minus 0.20 pH | checked in pH 7 buffer before a two point calibration |
| Isopotential point | pH 7.00 at 0 mV | the point where temperature has least effect on the reading |
| Response | 95 percent in 1 s on a bulb, 95 percent in 5 s on a flat surface | a slow answer after cleaning means a plugged junction |
| Temperature range | 0 to 60 C on polycarbonate bodies, 0 to 80 C on Ultem bodies, 0 to 100 C on the pressurised in-line family | sterilization is rated separately at up to 140 C |
| Pressure | 70 psig on the Ryton submersion family, 100 psig at 75 C on the threaded Ryton family, 4 barg (58 psig) on the in-line family | pressure and temperature are one rating, not two |
| Reference system | sealed gel 3.5 M KCl, or pressurised liquid electrolyte with a ceramic diaphragm and a silver ion trap | the pressurised reference is what survives steam cycles |
| Junctions | single Pellon, double Pellon, HDPE with Pellon, ceramic | a double junction keeps the process out of the reference wire |
| Body materials | polycarbonate, Ultem, epoxy, Ryton, glass | Ultem and Ryton for solvent and hot service |
| Process connections | Pg 13.5 thread, 1/2 inch MNPT, 3/4 inch MNPT, DIN 13.5 cap, 12 mm gland fittings | the gland has to match the 12 mm body |
| Connectors | BNC, K8S watertight, tinned leads, DIN cap with mating plug | K8S is the one that stays dry on a washdown skid |
| Insertion length | 120, 225, 325 and 425 mm on the in-line family; 150 mm on the 12 mm laboratory body | length follows the pipe tee, not the cable run |
| ORP sensor metal | platinum or gold | platinum for general redox, gold where sulphur compounds coat platinum |
| ORP reference offset | about 40 mV higher for Ag/AgCl than for calomel in the same solution | correct the setpoint when swapping reference types |
Included Models and Order Codes
The table is split by duty rather than by brand. The in-line family at the top is the one for a pipe tee on a skid, with a pressurised reference, a watertight connector and a sterilizable membrane. The sealed laboratory electrodes in the middle are the ones for a sampling port or a bench meter, and they are the cheapest way to keep a second measurement point without a second transmitter. The threaded Ryton and epoxy bodies at the bottom are the submersion and inline options for water and waste lines, where the junction has to be replaced long before the glass wears out.
| Part | Type | pH range | Temperature | Pressure | Connection | Note |
|---|---|---|---|---|---|---|
| InPro3253i/SG/120 | in-line, pressurised reference, digital | 0 to 12 | 0 to 100 C | 0 to 4 barg | Pg 13.5, K8S | 120 mm shaft, platinum solution ground, sterilizable to 140 C |
| InPro3253i/SG/225 | in-line, pressurised reference, digital | 0 to 12 | 0 to 100 C | 0 to 4 barg | Pg 13.5, K8S | 225 mm shaft, same membrane and reference |
| InPro3253i/SG/325 | in-line, pressurised reference, digital | 0 to 12 | 0 to 100 C | 0 to 4 barg | Pg 13.5, K8S | 325 mm shaft for a larger tee |
| InPro3253i/SG/425 | in-line, pressurised reference, digital | 0 to 12 | 0 to 100 C | 0 to 4 barg | Pg 13.5, K8S | 425 mm shaft, longest of the family |
| S200C/BNC | laboratory, sealed gel, single junction | 0 to 14 | 0 to 80 C | not rated for line pressure | BNC, 1 m cable | 12 x 150 mm Ultem body, 97 percent slope |
| S200CD/BNC | laboratory, sealed gel, double junction | 0 to 14 | 0 to 80 C | not rated for line pressure | BNC, 1 m cable | double junction for samples of unknown composition |
| S150C/BNC | laboratory, sealed gel, single junction | 0 to 14 | 0 to 60 C | not rated for line pressure | BNC, 1 m cable | polycarbonate body, lowest cost of the group |
| S265C | submersion or in-line, Ryton body | 0 to 14 | 0 to 80 C | 70 psig | 1/2 or 3/4 inch MNPT, tinned leads | 10 ft cable, threaded nipple body |
| S271CD | submersion, Ryton body, double junction | 0 to 14 | 75 C | 100 psig | 1/2 or 3/4 inch MNPT | 10 ft waterproof cable |
| S272CD | submersion, Ryton body, double junction | 0 to 14 | 81 C | 85 psig | 1/2 or 3/4 inch MNPT | the pressure figure drops as the temperature rises |
| S354CD | in-line, flat glass, DIN cap | 0 to 14 | 0 to 80 C | line pressure of the fitting | DIN 13.5 thread, DIN cap | flat surface resists coating in slurry |
| S224CD | in-line, spherical glass, DIN 13.5 cap | 0 to 14 | 0 to 80 C | line pressure of the fitting | DIN 13.5, S6 cap | mounts without a separate gland |
ORP Electrodes and Redox Reference
ORP parts are listed apart because the choice turns on the metal and the junction, not on a glass membrane. Platinum is the general purpose surface and gold is the one for streams where sulphur compounds coat platinum. The reference behind that metal sets the number on the display, so the last column notes what changes when the reference type changes.
| Part | Metal | Junction | Body | Temperature | Note |
|---|---|---|---|---|---|
| S500C-ORP/BNC | platinum, polished tip | single Pellon | Ultem | 0 to 80 C | plus or minus 2000 mV, 12 x 150 mm, BNC |
| S500C-ORP-AU/BNC | gold, polished tip | single Pellon | Ultem | 0 to 80 C | gold surface for sulphur-bearing streams |
| S550CD-ORP/BNC | platinum, extended tip | double HDPE and Pellon | Ultem | 0 to 100 C | highest temperature of the laboratory group, teeth guard |
| ORP1000 | platinum | single Pellon | polycarbonate | 0 to 60 C | basic cell, swimming pool and aquarium duty |
| ORP2000 | platinum | double Pellon | Ultem | 0 to 80 C | for samples of unknown composition |
| ORP3000 | platinum | double Pellon | Ultem | 0 to 80 C | removable bulb guard and strain relief |
| S150C-ORP/BNC | platinum | single Pellon | polycarbonate | 0 to 60 C | 30 inch cable, slot guard |
Key Features
Pressurised reference for steam cycles – electrolyte is held above process pressure so the junction keeps flowing outward during a 140 C sterilization cycle
Silver ion trap in the reference – the Argenthal cartridge keeps silver from the wire out of the junction, which is what stops the drift that follows a hot cycle
Double junction option – an HDPE or Pellon second junction keeps sulphide, protein and solvent out of the reference wire on dirty samples
Body material matched to the reagent – polycarbonate for clean water, Ultem and Ryton for solvents and hot caustic, glass where nothing else survives
Digital temperature element inside – the in-line family carries its own temperature sensor, so compensation follows the process instead of a neighbouring probe
Platinum dot as solution ground – the same metal that measures redox also grounds the solution, which removes the electrostatic noise that shows up as a wandering last digit
Flat and spear surfaces – a flat glass surface on the S354CD resists coating in slurry, and a spear tip on the S175CD reaches into semi solids
Standard 12 mm body – the laboratory and submersion electrodes share a 12 mm diameter, so one gland fits the whole set
Applications
These electrodes sit on the measurement point of an analyser, close to the sample and often close to the pump that moves it. The three cases below are the ones that decide the specification.
Water and waste analysers – pH on an aeration basin or a final effluent line, where the junction fouls with biofilm and the pressure is low. A Ryton body with a double junction and a 3/4 inch thread is the working choice, and the flat surface version is used where the sample carries grit.
Bioreactor and sterilizable skids – pH inside a fermentation vessel that is cleaned and sterilized in place. The pressurised reference and the 140 C rating are the reason this family exists, and the digital temperature element removes the separate probe from the vessel head.
Sampling and reagent lines – pH or redox checked at a sampling port before a colorimetric step. A sealed gel electrode with a BNC plug does the job at low cost, and a double junction version is used when the sample contains protein or solvent that would poison a single junction.
These electrodes are sourced for the analyser builds that SUPERB PCB and SUPERB PCBA programmes support, from the sample port to the transmitter board.
Sourcing and Cross-Reference
Electrode bodies, cables and connectors are the items that stop a line, because a cracked bulb cannot be repaired and a K8S plug cannot be replaced with a BNC. We hold the electrode, the gland, the cable and the buffer as one group so a replacement arrives as a set rather than as one part and three questions.
For analysers that are no longer supported, the electrode is often the only part still available while the transmitter is obsolete. We cross reference the old part number against the current body and connector, and we say plainly when the connector has changed and the cable has to be replaced with it.
Redox setpoints copied from an older plant need the reference type checked before anything else. A setpoint that came from a calomel reference reads about 40 mV away from an Ag/AgCl reference, and that difference is bigger than most control bands.
A replacement electrode is often ordered with the cable and gland as one kit, which is how SUPERB Assembly ships it when a line is down.
Q: Can you supply a replacement electrode for an analyser whose original part is discontinued?
A: SUPERB Tech can cross reference the discontinued number against the current body, junction and connector before the order is placed.
Why Buy pH and ORP Electrodes from SUPERB
- One group, one delivery: electrode, gland, cable, buffer and the transmitter connector are quoted together, so the line restarts with one shipment
- Range read before the quote: we ask for pH or mV range, temperature, pressure and the reagent before we name a part, because a 0 to 14 laboratory electrode is the wrong answer for a 100 C line
- Connector checked against the meter: BNC, K8S, DIN cap and tinned leads are not interchangeable, and the cable is ordered with the plug the transmitter already has
- Obsolete and hard to find parts: discontinued electrode numbers are cross referenced against the current body and junction rather than replaced with a lookalike
- Documentation as shipped: maker data sheet and the lot number travel with the electrode, which is what an audit asks for on a pH loop
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen, Conductivity Cells and Toroidal Sensors, Liquid Flow Sensors for Medical and Analytical Instruments, Air Bubble Detectors for Infusion and Reagent Lines, Peristaltic and Diaphragm Micro Pumps for Analysers, Sensors and Detectors, Obsolete Part Cross Reference.