Conductivity Cells and Toroidal Sensors for Water Analysers
Product Specifications
Product Overview
Conductivity is measured by putting a voltage across two electrodes and reading the current, so the geometry of those electrodes is part of the answer. That geometry is folded into one number, the cell constant, and every reading is the measured conductance multiplied by it. A cell with k = 0.1 is built for pure water, a cell with k = 1.0 covers drinking water and general process lines, and a cell with k = 10 belongs on brine or on a concentrated acid line where the conductance is high and the resistance is low.
The practical limit of a contacting cell is fouling. Graphite is cheap and good enough for clean water, but a graphite cell in a stream carrying oil or scale reads low and keeps reading low until it is cleaned. Stainless steel with a PEEK insulator survives higher temperature and pressure, and the 316 body can be removed and cleaned without breaking. Neither of them works in a stream that coats everything, and that is where a toroidal sensor comes in.
A toroidal sensor has no electrodes in contact with the liquid. Two coils sit inside a moulded body, one drives a current loop through the liquid and the other reads it, so the only thing in the sample is the body material. That construction is why a toroidal sensor reaches 2000 mS/cm and why it keeps working in slurry, brine and caustic where a contacting cell would polarise or plug. The price is a nominal cell constant with a wider tolerance, so the sensor is calibrated in the loop rather than trusted from the label.
Temperature compensation is not an accessory on this measurement, it is part of the reading. Conductivity rises about 2 percent per degree C for most aqueous solutions, so a 10 C error in the temperature element shows up as a 20 percent error in the displayed value. Pt100 and Pt1000 elements are the stable choices, a 10 k thermistor is the common one on light duty sensors, and the transmitter has to be set to the element that is actually fitted.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Cell constants | k 0.1, k 1.0 and k 10 per cm | k 0.1 for pure water, k 1.0 for general process, k 10 for brine and concentrated acids |
| Cell constant tolerance | plus or minus 10 percent on a graphite cell | the label value is nominal until the cell is checked in a standard solution |
| Measuring range, contacting | 0 to 20 mS/cm on k 0.1, up to 200 mS/cm on k 10 | the practical range follows the cell constant, not the transmitter |
| Measuring range, toroidal | 0 to 2000 mS/cm (2 S/cm) | the reason a toroidal sensor is used on brine and caustic |
| Wetted materials | graphite with epoxy, 316 stainless steel with PEEK, glass filled PEEK, glass filled Tefzel | the body decides which reagents the sensor can see |
| Temperature range | 0 to 70 C on graphite, 0 to 100 C on the stainless family, 0 to 250 C on the high temperature stainless version | each rating is paired with its own pressure figure |
| Pressure | 100 psig on graphite, 200 psig at 100 C on the stainless family, 250 psig at 200 C, 600 psig at 250 C | the highest ratings apply to one option only |
| Temperature element | Pt1000, Pt100 or 10 k thermistor | the thermistor option caps the temperature rating at 70 C |
| Temperature compensation | about 2 percent per degree C for aqueous solutions | a wrong element type looks like a wrong cell constant |
| Fittings | 1/2 inch MNPT, 3/4 inch MNPT, 12 mm gland, flange and sanitary options | the fitting and the insertion length are ordered together |
| Cable | two wire without compensation, four wire with a temperature element | the fourth wire is the element, not a spare |
| Response time | 0.5 s on the inductive family | a slow response on a contacting cell usually means fouling |
| Flow limit | 3 m/s maximum on the inductive family | higher flow erodes the body and shifts the reading |
Included Models and Order Codes
The contacting cells come first because they cover most water lines, and the toroidal sensors come after because they answer a different problem. A graphite cell with k = 0.1 and a stainless cell with k = 10 are the two ends of the same family: the first is for water that is almost an insulator, the second for a line that is almost a conductor. The toroidal parts are listed with the range they are built for, because a nominal cell constant of this type carries a wider tolerance and the loop calibration is what makes the number trustworthy.
| Part | Type | Cell constant | Range | Temperature | Pressure | Note |
|---|---|---|---|---|---|---|
| CS150 K=0.1 | contacting, graphite, 12 mm | 0.1 per cm | low range water | 0 to 70 C | 100 psig | minimum immersion 53 mm, epoxy body |
| CS150 K=1.0 | contacting, graphite, 12 mm | 1.0 per cm | general water | 0 to 70 C | 100 psig | minimum immersion 12.7 mm |
| CS650TC | contacting, 316 stainless, 1/2 inch NPT | 0.1 or 1.0 per cm | pure water and general | 0 to 100 C | 200 psig | PEEK insulator, aflas O-rings |
| CS675TC | contacting, 316 stainless, 3/4 inch NPT | 0.1, 1.0 or 10 per cm | general to brine | 0 to 100 C | 250 psig | with temperature element, PVC jacketed four wire cable |
| CS675HTTC | contacting, 316 stainless, high temperature | 0.1, 1.0 or 10 per cm | boiler and hot process | 0 to 200 C | 250 psig | PTFE coated leads, no PVC jacket |
| CS675HPTC | contacting, 316 stainless, high pressure | 0.1, 1.0 or 10 per cm | high pressure lines | 0 to 100 C | 300 psig | extended insertion length available |
| CS676HTTC | contacting, 316 stainless, extended length | 0.1, 1.0 or 10 per cm | boiler and deep tee | 0 to 200 C | 250 psig | longer body for a thick pipe wall |
| Rosemount 228, PEEK | toroidal, flow through | 2.7 per cm nominal | up to 2000 mS/cm | up to 200 C on the high temperature option | up to 295 psig on the high temperature option | glass filled PEEK body, Pt100 inside, 20 ft cable |
| Rosemount 228, Tefzel | toroidal, flow through | 2.7 per cm nominal | up to 2000 mS/cm | standard range | standard range | Tefzel body for oxidising reagents |
| Hach 3700, PEEK | inductive, convertible mount | not quoted as a fixed value | 0 to 2000 mS/cm | process range of the body | process range of the body | accuracy 0.01 percent of reading, response 0.5 s, flow limit 3 m/s |
Choosing the Cell Constant Before the Sensor
The cell constant is chosen from the conductivity the line is expected to run at, and the choice is made before the body material and the fitting. The table is the short version of that decision, with the error that shows up when the wrong constant is used.
| Expected conductivity | Cell to use | Why | What goes wrong otherwise |
|---|---|---|---|
| Below 10 microsiemens per cm, pure and ultrapure water | k 0.1 contacting cell | the resistance is high, so a small constant keeps the signal readable | a k 1.0 cell reads near zero and the noise dominates |
| 10 microsiemens to 20 millisiemens per cm | k 1.0 contacting cell | the middle of the range for drinking water and general process | a k 0.1 cell polarises and drifts |
| Above 20 millisiemens per cm, brine and concentrated acid | k 10 contacting cell or a toroidal sensor | the resistance is low, so a large constant keeps the current in range | a k 1.0 cell saturates the transmitter |
| Slurry, caustic, oil bearing or scaling streams | toroidal sensor | nothing metal touches the sample, so there is nothing to coat | a contacting cell reads low within days and cannot be cleaned in place |
| Line above 100 C | stainless cell with the matching rating, or a toroidal body rated for it | the ratings are paired, temperature with pressure | the O-ring and the insulator are the parts that fail first |
Key Features
Cell constants from 0.1 to 10 – one family covers pure water, drinking water and brine, so the range is answered by the constant rather than by a different sensor type
Graphite for cost, stainless for duty – graphite cells suit clean water, and 316 stainless with a PEEK insulator carries the temperature and pressure
Toroidal body with no electrodes – two coils read the current loop through the liquid, so slurry and caustic do not foul a measuring surface
Pt1000, Pt100 or 10 k element – compensation is ordered with the cell, and the transmitter is set to the element actually fitted
Paired temperature and pressure ratings – the 200 C version is rated at 250 psig and the 250 C version at 600 psig, and the two are never separated
Standard pipe fittings – 1/2 and 3/4 inch NPT plus 12 mm glands, so the cell goes into the tee the line already has
Four wire cable with compensation – the element has its own pair, which keeps the temperature reading out of the conductivity loop
Insertion length options – standard and extended bodies for thick pipe walls and for a tee that sits deep in insulation
Applications
Conductivity cells sit on water treatment, on rinse stages and on concentration control. The three cases below are where the specification is decided.
Water treatment and reverse osmosis – feed, permeate and concentrate conductivity on a skid. Permeate runs at a few microsiemens and needs a k 0.1 cell, while the concentrate needs a k 10 cell or a toroidal sensor, and the two are rarely the same part.
Boiler and cooling water – conductivity as a blowdown control. The stainless family with a high temperature rating is the one for a sample that leaves the boiler hot, and the 200 C option exists for exactly that line.
Brine, caustic and chemical concentration – acid, base and salt concentration read through conductivity. A toroidal sensor is used here because the metal electrodes of a contacting cell would polarise, and because the body survives the reagent that the line carries.
The analyser boards behind these sensors are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the electrode input to the isolated current loop.
Sourcing and Cross-Reference
Cell constant, temperature element and cable length are three separate choices on the same order, and a sensor that arrives with the wrong element is a sensor that reads wrong until the transmitter is reprogrammed. We take the three together and quote them as one line.
For a line that is already running, the old cell constant and the old fitting matter more than the brand. A k 1.0 cell with a 3/4 inch thread can be replaced by a different body only if the thread and the insertion depth match, and we check that before naming a part.
Toroidal sensors are the ones that go obsolete quietly, because the transmitter and the sensor are often a pair. Where the sensor is no longer listed, we cross reference the body material and the cell constant against a current part and say plainly when the transmitter also has to change.
A sensor, its cable and the gland are ordered as one set when a line is down, which is how SUPERB Assembly ships a replacement.
Q: Can you supply a conductivity cell with the same cell constant and thread as the one already on the line?
A: SUPERB Tech can confirm the cell constant, the thread and the temperature element against the old part before an order is released.
Why Buy Conductivity Cells from SUPERB
- Cell constant and fitting quoted together: the constant, the thread and the insertion length are confirmed before a part is named, so the replacement drops into the existing tee
- Range checked against the process: we ask for the expected conductivity before recommending k 0.1, k 1.0, k 10 or a toroidal body
- Temperature element matched to the transmitter: Pt1000, Pt100 and 10 k thermistors are not interchangeable, and the cable is ordered with the element the transmitter expects
- Toroidal options for dirty streams: slurry, caustic and brine are answered with a body that has no metal in contact with the sample
- Paired ratings stated plainly: temperature and pressure are quoted as one rating per part number, not as two separate maxima
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Water Quality Sensor Circuits for pH, Conductivity and Dissolved Oxygen, pH and ORP Electrodes for Process Analysers, Liquid Flow Sensors for Medical and Analytical Instruments, Level Sensors and Float Switches for Tanks and Sumps, Turbidity and Photometric Detector Modules for Water Analysis, Sensors and Detectors, Component Lead Time Solutions.