Oxygen Sensor Cells for Ventilators and Oxygen Concentrators
Product Specifications
Product Overview
A ventilator does not measure oxygen flow, it measures partial pressure at the cell face. The cell converts that pressure into a small voltage, and the machine converts the voltage into FiO2 with a calibration made at two known points. Everything that goes wrong later shows up in that same small voltage: a cell that has lost electrolyte reads low, a cell with a blocked diffusion path reads slow, and a cell exposed to a dry circuit reads high for a while before it settles. The datasheet numbers that matter are therefore the output in air, the tolerance on that output and the time it takes to reach 90 percent of a step.
There are two families on this page. The galvanic cell, which covers the KE series and the MAX-250, generates its own voltage and needs no supply at all. It behaves like a battery with a very long life and a very high internal impedance, so the load on it has to be in the megohm range. The optical cell, which covers the LuminOx range, excites a luminescent dye and measures the decay, so it does not consume oxygen and it does not dry out. It needs 4.5 to 5.5 V, it delivers the reading as a serial message, and its range stops at 25 percent oxygen, which fits ambient and ventilator monitoring but not a cylinder purity check at 100 percent.
The third decision is mechanical and it is the one that delays a replacement most often. The KE-25 is sold as a plain body, as a body without a flange, as a body with a threaded top and as a body with an O ring top. They share the same cell inside, so the electrical data is identical, and only the way the cell seals into the sensor block changes. A cell ordered in the wrong top will not seat, and a cell forced into the wrong seat leaks around the O ring and reads ambient air instead of the sample gas.
The fourth decision is life, and life in a galvanic cell is counted in percent hours rather than years. A cell that spends its service at 21 percent oxygen and one that spends it at 100 percent draw down the same reservoir at very different rates, which is why a spare kept in a drawer still ages once the seal is opened. The numbered life on the datasheet assumes 20 degrees C in normal air and ends when the output falls to 70 percent of its first value.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Cell type | galvanic (KE, MAX-250) or optical (LuminOx) | galvanic cells generate their own voltage, optical cells need 4.5 to 5.5 V |
| Measurement range | 0 to 100 percent O2 on galvanic, 0 to 25 percent O2 on optical | 0 to 300 mbar ppO2 on the optical range |
| Output in air at 21 percent O2 | 10.0 to 15.5 mV on KE-25 and MAX-250, 19.0 to 35.0 mV on KE-12, 47 to 65 mV on KE-50 | optical versions report percent, ppO2, temperature and pressure over UART |
| Zero offset | below 0.5 mV on KE-25 and MAX-250, below 6.0 mV on KE-50 | measured in nitrogen or in a certified zero gas |
| Accuracy | plus or minus 1 percent of full scale on KE-12, KE-25 and MAX-250, plus or minus 2 percent on KE-50 | stated after a two point calibration at 0 and 100 percent |
| Linearity | (Va - V0) / (V100 - V0) = 0.21 plus or minus 0.02 | the ratio is what makes a one point air calibration possible |
| Response time T90 | about 5 s KE-12, about 15 s KE-25, about 60 s KE-50, under 30 s LOX-02, under 10 s LOX-02-F | measured with 200 to 300 ml per minute flowing across the cell face |
| Operating temperature | 5 to 40 C on galvanic, 0 to 45 C on optical | storage -15 to 50 C on MAX-250, -30 to 60 C on LuminOx |
| Humidity | 10 to 90 percent RH without condensation on galvanic, 0 to 99 percent RH on optical | condensation on the cell face blocks the diffusion path |
| Atmospheric pressure | 811 to 1216 hPa on galvanic, 500 to 1200 mbar barometric on optical | the optical cell compensates internally for pressure |
| Supply | none on galvanic, 4.5 to 5.5 V at under 7.5 mA on optical | optical peak current under 20 mA while streaming |
| Interface | analogue millivolt on galvanic, 3.3 V TTL UART on optical | optical message carries percent, ppO2, pressure, temperature and status |
| Life | about 2.5 years KE-12, 5 years KE-25, 10 years KE-50, over 1,500,000 percent hours MAX-250 | end of life is 70 percent of the first output |
| Cross sensitivity | no effect from CO, CO2, H2S, NOx or H2 on the KE series | optical cells show negligible response to other gases |
| Package | KE body 22.7 mm diameter, LOX body 20 mm diameter by 12.5 mm | KE-25F3 has a threaded top, KE-25F4 an O ring top |
Included Models and Order Codes
The KE series is one cell in three sizes of reservoir. KE-12 has the shortest life and the fastest response, KE-25 is the general purpose cell with 10.0 to 15.5 mV in air, and KE-50 carries the largest reservoir with 47 to 65 mV output and a 60 second response. KE-25F1, KE-25F3 and KE-25F4 are the same cell as KE-25 with a different top: F1 has no flange, F3 has a threaded top and F4 seals with an O ring, so the choice is made by the socket in the instrument, not by the electrical data. MAX-250 and MAX-250E are the same 10.0 to 15.5 mV cell as KE-25 in a housing that drops into instruments built around the older R29 pattern, and they are the usual route where a therapy device is out of support but still in service. The LuminOx range is the optical alternative: LOX-02 is the immersion body, LOX-02-S has a gas tight rear so only the face sees the sample, and LOX-02-F is the flow through body with the faster T90 under 10 seconds.
| Part number | Cell type | Range | Output or interface | T90 | Note |
|---|---|---|---|---|---|
| KE-12 | galvanic | 0 to 100 percent O2 | 19.0 to 35.0 mV in air | about 5 s | shortest life of the family at about 2.5 years |
| KE-25 | galvanic | 0 to 100 percent O2 | 10.0 to 15.5 mV in air | about 15 s | general purpose cell, 5 year life |
| KE-25F1 | galvanic | 0 to 100 percent O2 | 10.0 to 15.5 mV in air | about 15 s | same cell as KE-25, body without flange |
| KE-25F3 | galvanic | 0 to 100 percent O2 | 10.0 to 15.5 mV in air | about 15 s | same cell as KE-25, threaded top for a screwed socket |
| KE-25F4 | galvanic | 0 to 100 percent O2 | 10.0 to 15.5 mV in air | about 15 s | same cell as KE-25, O ring top for a sealed seat |
| KE-50 | galvanic | 0 to 100 percent O2 | 47 to 65 mV in air | about 60 s | largest reservoir, about 10 year life, plus or minus 2 percent full scale |
| MAX-250 | galvanic | 0 to 100 percent O2 | 10.0 to 15.5 mV in air | 15 s or less | replacement path for instruments built around the R29 pattern |
| MAX-250E | galvanic | 0 to 100 percent O2 | 10.0 to 15.5 mV in air | 15 s or less | same electrical data, alternate housing for FiO2 measurement |
| LOX-02 | optical | 0 to 25 percent O2 | 3.3 V TTL UART, 4.5 to 5.5 V supply | under 30 s | whole body sits in the sample stream, 20 mm diameter |
| LOX-02-S | optical | 0 to 25 percent O2 | 3.3 V TTL UART, 4.5 to 5.5 V supply | under 30 s | gas tight rear, only the sensing face sees the sample |
| LOX-02-F | optical | 0 to 25 percent O2 | 3.3 V TTL UART, 4.5 to 5.5 V supply | under 10 s | flow through body for a pumped sample line |
Bench Checks Before a Cell Goes Into Service
A galvanic cell is usually replaced during a service visit, and the checks below are the ones that decide whether the new cell is good before the instrument is closed. All of them use the numbers printed by the maker and all of them assume a two point calibration at 0 and 100 percent oxygen.
| Check | Method | Acceptance |
|---|---|---|
| Zero offset | expose the cell to nitrogen or a certified zero gas until the reading is stable | below 0.5 mV on KE-25 and MAX-250, below 6.0 mV on KE-50 |
| Span in air | expose the cell to ambient air at 21 percent O2 with 200 to 300 ml per minute across the face | 10.0 to 15.5 mV on KE-25, 19.0 to 35.0 mV on KE-12, 47 to 65 mV on KE-50 |
| Linearity | take the air reading and compare it with the two calibration points | (Va - V0) / (V100 - V0) falls inside 0.21 plus or minus 0.02 |
| 21 percent cross check | read a known 21 percent gas after calibrating at 0 and 100 percent | within 21 percent plus or minus 1 percent on KE-25, plus or minus 2 percent on KE-50 |
| Response | step the cell from 21 percent to 100 percent and time the rise | T90 of 14 plus or minus 2 s on KE-25, about 60 s on KE-50 |
| Flow | hold the sample flow across the cell face | above 200 to 300 ml per minute, otherwise the response slows and reads behind the breath |
| Leak at the seat | draw a vacuum on the sample side with the cell seated | no drift toward ambient, which is the signature of a bad O ring top |
| Age check | compare the air output with the value written at the last service | replace when the output falls to 70 percent of the first value |
Key Features
Output stated in air – every row gives the millivolt output at 21 percent O2, which is the single number a service technician can compare against a working instrument.
Three reservoir sizes – KE-12, KE-25 and KE-50 share one cell and differ in life and response, so the choice is made on the service interval rather than on price alone.
Four body styles – KE-25, F1, F3 and F4 carry the same cell in different tops, which covers a plain socket, a flange mount, a threaded socket and an O ring seat.
Zero supply for galvanic – the cell produces its own voltage, so a replacement needs no rail, no regulator and no change to the instrument firmware.
Optical option with no depletion – the LuminOx cell does not consume oxygen and holds its calibration on the shelf, which suits instruments that run a few weeks a year.
Compensation built in – the optical cell reports pressure and temperature with the oxygen value, so the host does not have to model either one.
Known cross sensitivity – the KE series is documented as unaffected by CO, CO2, H2S, NOx and H2, which matters where the sample passes a CO2 absorber.
Life in percent hours – MAX-250 is rated over 1,500,000 percent hours, which can be converted into a service interval once the duty cycle is known.
Applications
Three places where the difference between a 15 second cell and a 30 second cell, or between an analogue and a serial output, decides the choice.
Ventilator FiO2 monitoring – the cell sits in the inspiratory limb and the alarm limits are set close to the delivered mixture, so the tolerance on the air output and the T90 both matter at every breath.
Oxygen concentrator purity – the cell checks the sieve outlet over a slow sample flow, and here the larger reservoir of the KE-50 is worth more than a fast response.
Incubator and anaesthesia circuits – a cell with a sealed top keeps ambient air out of the sample path, which is the difference between reading the circuit and reading the room.
A ventilator board is reviewed as a whole, so the SUPERB PCBA and the SUPERB PCB are checked against the same cell list and the analogue front end is scaled for the cell that was chosen.
Sourcing and Cross-Reference
Most enquiries on this family start with an instrument that is still in service after its support has ended. The useful facts are the output in air, the body style and the connector, because a cell that matches on range but not on top will not fit the socket. A photo of the old cell next to a rule is usually enough to identify the body style, and the millivolt reading at 21 percent O2 confirms whether the replacement has the same sensitivity class.
The second question is shelf life. A galvanic cell starts working the moment its seal is opened, and a spare that has sat in a warm store for two years is not the same part as one from the current batch. Where a line keeps cells as spares, the batch date on the label matters more than the delivery date, and the instrument record should carry it.
Where the original cell is no longer listed, the route is the same output class in the current family, or the optical cell where the host can accept a serial reading. See obsolete part cross reference for a code that has gone out of support, and lead time and availability where a therapy device needs a cell before the next service window.
The SUPERB Assembly team keeps the cell out of the wash and out of the reflow oven, and fits it after the board has been cleaned, because a wet cell face reads low for days.
Q: Can you supply a replacement cell for a ventilator that is out of production, with the same output at 21 percent O2 and the same threaded top?
A: Yes. Send the old part number or the output in air, the body style and the connector, and the SUPERB Tech team confirms the cell that matches before it ships.
Why Medical Instrument Builders Source Oxygen Cells Through SUPERB
- Output at air on every row: the 21 percent figure is listed per part number, so a replacement is matched on sensitivity and not only on shape.
- Body style named: F1, F3 and F4 are called out by their tops, which is the detail that decides whether the cell seats in the instrument.
- Life and response together: the reservoir size and the T90 are given side by side, so a slow cell is not chosen for a fast alarm loop.
- Optical route where analogue is a problem: the LuminOx range gives a serial reading with pressure and temperature, which removes the analogue front end from a new build.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Gas Sensors for Air Quality and Leak Detection, Liquid Flow Sensors for Medical and Analytical Instruments, Differential Pressure Sensors for Ventilators, IEC 60601 Power Supplies for Medical Diagnostic Equipment, Thermocouple and RTD Front Ends, BOM Shortage Recovery.