IEC 60601 Power Supplies for Medical Diagnostic Equipment
Product Specifications
Product Overview
A patient connected device has a current path to earth that an industrial device does not, so the supply is selected from the leakage budget first. Two means of patient protection between primary and secondary, and one means to earth, is the arrangement that lets a device pass a 60601-1 assessment without reworking the transformer. That is a construction decision inside the power supply, not something that can be added at the harness, which is why a medical project starts with a supply that already carries the 2xMOPP classification.
The second number is earth leakage. The RPS-200 family holds it below 130 uA at 264 VAC with a touch current below 40 uA, the RPS-120 below 190 uA with a touch current below 100 uA, and the RPS-300 below 150 uA on the PCB version and below 200 uA in the enclosed version, with a touch current below 70 uA. Those figures are what the system leakage calculation starts from, and they are quoted at 264 VAC because that is the worst case in the range.
The third decision is mechanical. The 65 W and 120 W units sit on a 3 by 2 inch or 4 by 2 inch footprint and are open frame, so they mount inside the instrument on standoffs with the mains terminals covered. The 300 W unit needs 20.5 CFM of air for its full rating and 200 W without a fan, so the enclosure ventilation is part of the power budget rather than an afterthought. All series offer an enclosed version with the suffix C, which is the one used where the supply is a field replaceable module.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Rated power | 65 W, 120 W, 200 W and 300 W series | the 200 W and 300 W ratings need forced air for the top of the range |
| Convection rating | 140 W on the 200 W series, 200 W on the 300 W series | without a fan, the series derates to the convection figure |
| Input voltage | 80 to 264 VAC on the 65, 120 and 200 W series, 90 to 264 VAC on the 300 W series | 113 to 370 VDC on the 120 and 200 W series |
| Input frequency | 47 to 63 Hz | single phase, two wire input |
| Output voltages | 12, 15, 24, 27 and 48 V | suffix on the model code, plus C for the enclosed version |
| Isolation, primary to secondary | 2xMOPP, 4 kVAC withstand | the figure the 60601-1 assessment starts from |
| Isolation, primary to earth | 1xMOPP, 2 kVAC withstand | secondary to earth 1.5 kVAC on the 200 and 300 W series |
| Earth leakage current | below 130 uA on the 200 W series, below 150 uA on the 300 W PCB version, below 190 uA on the 120 W series | measured at 264 VAC |
| Touch current | below 40 uA on the 200 W series, below 70 uA on the 300 W series, below 100 uA on the 120 W series | measured at 264 VAC |
| Safety standards | IEC 60601-1, ANSI/AAMI ES60601-1, TUV EN 60601-1, CAN/CSA C22.2 No. 60601-1 | medical electrical equipment, edition 3 family |
| EMC | EN 55011 Class B conducted and radiated on the PCB versions | Class B holds with the frame earthed or floating |
| Efficiency | 89 to 91 percent on the 120 W series, 93 to 95 percent on the 200 W series, 90 to 93 percent on the 300 W series | at nominal load |
| No load power | below 0.3 W on the 120 W series, below 0.5 W on the 200 W series | standby consumption |
| Operating temperature | minus 30 to plus 70 C | derating above 50 C per the maker curve |
| Operating altitude | 4000 m on the 120 W series, 5000 m on the 200 W series | derating above 2000 m |
| Cooling | convection, plus a 12 V 0.5 A fan output on the 120, 200 and 300 W series | the fan output is not a general purpose rail |
| Warranty | 3 years | per maker terms |
Included Models and Order Codes
The series name gives the power and the suffix gives the voltage, with a C added for the enclosed version. RPS-65 is the 65 W frame on 3 by 2 inches, with the 12 V model rated 5.42 A, and it is the one used for a small monitor or a handheld instrument where the board has to stay compact. RPS-120 is the 120 W frame on 4 by 2 inches, with 10 A at 12 V, 5 A at 24 V and 2.5 A at 48 V, a leakage below 190 uA and a no load draw below 0.3 W. RPS-200 keeps the same 4 by 2 inch footprint and doubles the power with 10 CFM of air, 140 W without a fan, leakage below 130 uA and 4 kVAC isolation. RPS-300 moves to a larger frame with a 5 V standby output, power good and power fail signals and a remote on off input, which is the arrangement a diagnostic cart uses when the supply has to be controlled by the instrument.
| Series | Rated power | Outputs | Leakage current | Isolation |
|---|---|---|---|---|
| RPS-65 | 65 W | 12 V 5.42 A, 15 V, 24 V, 48 V | per maker sheet | 2xMOPP, 4 kVAC |
| RPS-120 | 120 W | 12 V 10 A, 15 V 8 A, 24 V 5 A, 27 V 4.5 A, 48 V 2.5 A | below 190 uA | 2xMOPP, 4 kVAC |
| RPS-200 | 200 W with 10 CFM, 140 W convection | 12 V, 15 V, 24 V, 27 V, 48 V | below 130 uA | 2xMOPP, 4 kVAC |
| RPS-300 | 300 W with 20.5 CFM, 200 W convection | 12 V 25 A, 15 V 20 A, 24 V 12.5 A, 27 V 11.12 A, 48 V 6.25 A | below 150 uA PCB, below 200 uA enclosed | 2xMOPP, 4 kVAC |
| RPS-120-C | 120 W, enclosed | same outputs as RPS-120 | below 190 uA | 2xMOPP, 4 kVAC |
| RPS-300-C | 300 W, enclosed | same outputs as RPS-300 | below 200 uA | 2xMOPP, 4 kVAC |
Key Features
2xMOPP construction – two means of patient protection between primary and secondary and one to earth, which is the arrangement a 60601-1 assessment expects to find inside the supply.
Leakage current quoted at 264 VAC – 130 to 190 uA on earth leakage and 40 to 100 uA on touch current, at the top of the input range rather than at nominal.
4 kVAC primary to secondary – the withstand voltage is stated with the MOPP classification, so the isolation barrier is a documentable figure and not an assumption.
Class B EMC on the PCB versions – conducted and radiated emissions hold at Class B with the frame earthed or floating, which is what a compact instrument enclosure needs.
No load draw below 0.3 W – the 120 W series holds standby below 0.3 W, which matters for a device that stays plugged in on a ward.
Convection or forced air stated – 140 W and 200 W are given as the convection ratings of the 200 W and 300 W series, so the enclosure ventilation is part of the power budget.
Enclosed version in the same family – the C suffix gives a field replaceable module with the same electrical data, so a service part does not need a new approval.
Fan output and control signals – a 12 V 0.5 A fan rail, power good, power fail and a remote on off input on the larger series, which is what a cart with a display needs.
Applications
Three device classes where the supply is chosen from the leakage budget and the enclosure, not from the wattage alone.
Patient monitors and diagnostic carts – a 120 W or 200 W unit inside the cart feeds the board and the display, and the leakage figure is carried into the system calculation for the applied part.
Laboratory analysers – a 300 W unit with a 5 V standby rail keeps the controller alive while the main output is switched, which is how an analyser sequences its pumps and its reader.
Home care devices – the enclosed version is the field replaceable module, and the no load draw below 0.3 W keeps the device inside the standby budget it was certified against.
A diagnostic instrument is usually built as a board plus a harness plus a supply, so the SUPERB PCBA and the SUPERB PCB are checked against the same leakage budget and the earth stud is reviewed once for both.
Sourcing and Cross-Reference
The first question is the applied part classification, because it decides the MOPP level and the leakage budget the supply has to fit. The second question is the enclosure: an open frame unit needs a standoff pattern and a mains cover inside the instrument, and the enclosed version is a different mechanical interface with the same electrical data.
The third question is the airflow. A 300 W unit asked for 300 W inside a sealed box will derate to 200 W, and the instrument then browns out during a pump cycle. Reading the convection figure before the enclosure is signed off avoids the rework.
Where a series has been superseded, the usual route is the same footprint at the next power level with the leakage figure rechecked, since the mounting holes and the connector set often stay. See lead time and availability for a specific model, and BOM shortage recovery where a build is waiting on one supply.
The SUPERB Assembly team bonds the supply earth stud to the chassis at one point and keeps the mains and the secondary routed apart, because a stray return path is what turns a compliant supply into a failed system test.
Q: Can you confirm the leakage current and the MOPP classification of the supply you quote, and whether the enclosed version is available, because our applied part assessment depends on both?
A: Yes. Send the applied part classification, the output voltage and current, the enclosure type, the ambient inside the box and the standards the device is assessed against, and the SUPERB Tech team returns the series and the suffix.
Why Medical Device Builders Source Power Supplies Through SUPERB
- Leakage and MOPP quoted together: the isolation class and the measured leakage current are given per series, because one without the other is not a usable number for a 60601-1 assessment.
- Convection rating stated: the fan free rating sits beside the rated power, so the enclosure is sized from the figure the box can actually dissipate.
- Enclosed and open frame in one family: the C suffix keeps the same electrical data, which means a service module does not restart the approval process.
- Safety documents named: the standards list is part of the offer, so the compliance file can be assembled from the supply datasheet rather than from a phone call.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Isolated DC DC Modules and Brick Converters, LDO Linear Regulators for Low Noise Power Rails, PWM Controller ICs for Switching Power Supplies, Temperature and Humidity Sensors, Liquid Flow Sensors for Medical and Analytical Instruments, Component Lead Time Solutions.