Optical Power Meter Heads and Sensors
Product Specifications
Product Overview
A power meter reading is only as good as the head in front of it. The console supplies the electronics and the display, but the detector material decides the wavelength range, the attenuator decides the maximum power, and the aperture decides how much of a non uniform beam is actually measured. Choosing a head on maximum power alone is the most common way to get a reading that looks stable and is quietly wrong.
Silicon covers 200 to 1100 nm and germanium or indium gallium arsenide covers the near infrared to 1800 nm. Both are limited by the photodiode current, so an attenuator is placed in front of the diode to reach tens or hundreds of milliwatts, and the attenuation is calibrated into the head. Thermal heads work differently: a black absorber heats up and a thermopile measures the rise, which gives a flat response from 190 nm to 20 um at the cost of a slower response time and a lower resolution.
The aperture is the third variable. A 9.5 mm aperture is right for a collimated beam that fits inside it, and a 25 or 40 mm aperture is needed for a diverging beam or a large spot, where a smaller head would read only the centre and under-report the total. Where the beam overfills the aperture, the reading becomes a function of position, which is the point at which a head with an integrating sphere is the better answer.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Wavelength range | 200 nm to 1800 nm on photodiode heads, 190 nm to 20 um on thermal heads | Depends on the detector material |
| Power range | 50 nW to 500 mW on photodiode heads, 10 uW to 200 W on thermal heads | Attenuator sets the upper end |
| Aperture | 9.5 mm standard, 5 mm to 40 mm across the family | Beam must fit inside the aperture |
| Detector materials | Silicon, UV extended silicon, germanium, thermopile | Thermopile for broadband and high power |
| Linearity | Plus or minus 0.5 per cent on photodiode heads, plus or minus 1 per cent on some thermal heads | At a stated power level |
| Measurement uncertainty | Plus or minus 3 per cent at 1064 nm, plus or minus 5 per cent elsewhere | NIST or PTB traceable calibration |
| Resolution | 1 nW on the most sensitive heads, 100 uW to 5 mW on thermal heads | Depends on gain range |
| Response time | 0.6 s on the mid power thermal heads, 9 s on the 40 W head | Thermal heads are slower by nature |
| Head temperature | NTC thermistor, 4.7 kOhm | Used for compensation and over temperature alert |
| Cable | 1.5 m | Fixed to the head |
| Mounting | 8-32 and M4 taps, SM1 thread on the aperture | Posts and fiber adapters sold separately |
| Consoles | PM100A, PM100D, PM400 and PM5020 | Interfaces PM100USB and PM103 also supported |
Included Models and Order Codes
The table is grouped by detector. The first rows are silicon and germanium heads for low power and high resolution, the middle rows are thermal heads up to 10 W, and the last rows are the larger thermal heads for tens or hundreds of watts. A head is always ordered with the console or interface it will run on, because the calibration data is loaded through that connection.
| Part Number | Detector | Wavelength | Power Range | Aperture | Notes |
|---|---|---|---|---|---|
| S120VC | UV extended silicon | 200 to 1100 nm | 50 nW to 50 mW | 9.5 mm | Reflective ND attenuator, 1 nW resolution |
| S120C | Silicon | 400 to 1100 nm | 50 nW to 50 mW | 9.5 mm | General purpose head, 10 nW resolution |
| S121C | Silicon | 400 to 1100 nm | 500 nW to 500 mW | 9.5 mm | Higher power silicon head, 2 nW resolution |
| S122C | Germanium | 700 to 1800 nm | 50 nW to 40 mW | 9.5 mm | Near infrared, 2 nW resolution |
| S130C | Silicon | 450 to 1100 nm | 500 pW to 0.2 mW, up to 20 mW with filter | 9.5 mm | Slim housing, 150 x 19 x 10 mm |
| S132C | Germanium | 700 to 1800 nm | 5 nW to 5 mW, up to 500 mW with filter | 9.5 mm | Slim housing, near infrared |
| S401C | Thermopile | 190 nm to 20 um | 10 uW to 1 W, 3 W with heat sink | 10 mm | Active background compensation, 1 uW resolution |
| S405C | Thermopile | 190 nm to 20 um | 100 uW to 5 W | 10 mm | Internal SM05 thread, 5 uW resolution |
| S415C | Thermopile | 190 nm to 20 um | 2 mW to 10 W, 20 W with heat sink | 15 mm | Removable heat sink, 0.6 s response |
| S425C | Thermopile | 190 nm to 20 um | 2 mW to 10 W, 20 W with heat sink | 25.4 mm | Larger aperture for diverging beams |
| S350C | Thermopile | 190 nm to 1.1 um and 10.6 um | 10 mW to 40 W, 60 W with heat sink | 40 mm | Large spot sizes, 9 s response |
| S322C | Thermopile | 250 nm to 11 um | 100 mW to 200 W, 250 W with heat sink | 25 mm | Fan cooled, 5 mW resolution |
Key Features
Material matched to wavelength – silicon to 1100 nm, germanium to 1800 nm, thermopile from 190 nm to 20 um
Attenuator calibrated in – the reflective or absorptive filter in front of the diode is part of the calibration, not an add on
Aperture stated per head – 9.5 to 40 mm, so a diverging beam is not measured through a hole that clips it
Flat thermal response – a thermopile reads across the spectrum without a wavelength correction table
Calibration data supplied – NIST or PTB traceable data per head, with a recalibration service offered
Console and interface options – bench consoles PM100A and PM100D, or USB interfaces for a fixed installation
Applications
Heads are bought with the source they will measure, because the wavelength range and the power level are fixed by the laser or lamp rather than by the meter.
Laser diode characterisation – a silicon or germanium head on a rail, measuring the output of a collimated diode across its current range
Photometric instrument service – a thermal head checking lamp output in a spectrophotometer, where the spectrum is broad
Fiber and component test – a head with a fiber adapter measuring launch power before a component under test
Instruments that read these heads are produced as SUPERB PCB and SUPERB PCBA builds when the meter leaves the bench.
Sourcing and Cross-Reference
Reading a head specification starts with the wavelength, then the power, then the aperture. A head that covers the wavelength but saturates at a tenth of the power will need an attenuator, and one with a 9.5 mm aperture will under-read a beam that is 20 mm across. The three numbers are quoted together here precisely because they are chosen together.
Thermal heads are slower than photodiode heads and that is not a fault. A 9 s response time means the absorber has to come to equilibrium, so a thermal head is the wrong instrument for a modulated source but the right one for a steady lamp or a high power laser. Where a reading is needed quickly at high power, a photodiode head with an attenuator is the faster route within its power limit.
Calibration is worth keeping with the head rather than the console. The data is tied to the individual sensor, so a head moved to another console has to carry its calibration with it. We supply the calibration record with each head and can quote a recalibration interval where an instrument is audited.
Heads shipped with a console and adapters are consolidated through SUPERB Assembly, with the calibration sheet in the same box.
Q: Can you supply a thermal head for 20 W at 1064 nm together with the console and a fiber adapter?
A: Yes. SUPERB Tech quotes the head, its console and any fiber adapter together, and states the maximum power density before the order is placed.
Why Choose SUPERB for Power Meter Heads
- Wavelength, power and aperture together: each head lists all three, so a beam is not measured through an aperture that clips it
- Attenuator included in the calibration: the filter in front of the diode is part of the head rather than an accessory
- Thermal heads for broadband work: flat response from 190 nm to 20 um where a photodiode needs a correction table
- Calibration record supplied: traceable data per head, with a recalibration route for audited instruments
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Photodiode Arrays and Detector Modules, PIN and Avalanche Photodiodes, Integrating Spheres and Reflectance Standards, Neutral Density Filters and Standards, Fiber Collimators and Patch Cords, UV and Visible LED Sources.