Needle and Metering Valves for Analyser Lines
Product Specifications
Product Overview
Somewhere on almost every analyser line there is a restriction that has to be adjustable: a split to a mass spectrometer, a back pressure set so a degasser works, or a bleed that keeps a peristaltic pump from pulsing into the cell. A fixed orifice cannot do that job, and a ball valve has too little resolution. That is where a needle or metering valve belongs, and the two numbers that decide the part are the port thread and the tubing outside diameter.
The valves below are PEEK bodies with PTFE seals and a 0.020 inch thru hole, rated to 800 psi for the standard metering and splitter bodies. The graduated splitter raises that to 4000 psi and adds a scale so a working split setting can be written down and returned to, which matters on a method that has already been validated. Shut off valves are listed alongside because they are usually specified with the same port in the same block.
Port style is the most common ordering error. A 1/4-28 flat bottom port and a 10-32 coned port will not interchange, and a valve body built for 1/16 inch tubing will not seal on 1/8 inch tubing even with a different ferrule. Confirm the port and the tubing first, then choose between metering, splitter and shut off.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Body material | PEEK with PTFE seals | Biocompatible flow path |
| Thru hole | 0.020 in (0.50 mm) | All adjustable valves listed |
| Port thread | 1/4-28 flat bottom or 10-32 coned | Must be matched to the existing port |
| Tubing outside diameter | 1/16 in or 1/8 in | Fixed by the valve body |
| Pressure rating | 800 psi (55 bar) | Standard metering and splitter valves |
| High pressure splitter | 4000 psi (276 bar) | Graduated splitter, stainless needle |
| Minimum controllable flow | 3.5 uL/min | From 1.0 mL/min inlet with water |
| Internal volume | 1.2 uL to 7.7 uL | Measured at full open |
| Needle material | PEEK, or stainless steel on the splitter | Stainless on high pressure models |
| Adjustment | Thumbscrew, graduated scale on the splitter | Repeatable split settings |
Included Models and Order Codes
The table is grouped by function. Metering valves set a continuous restriction; splitter valves divide one stream into two and are used ahead of a mass spectrometer; shut off valves stop flow quickly and are not meant for adjustment. Included fittings are listed because a valve ordered without them will not connect to the line.
| Part Number | Function | Port Thread | Tubing OD | Internal Volume | Max Pressure |
|---|---|---|---|---|---|
| P-445 | Metering valve | 1/4-28 flat bottom | 1/16 in | 7.7 uL | 800 psi |
| P-446 | Metering valve | 10-32 coned | 1/16 in | 7.2 uL | 800 psi |
| P-447 | Metering valve | 1/4-28 flat bottom | 1/8 in | 7.7 uL | 800 psi |
| P-450 | Splitter valve | 1/4-28 flat bottom | 1/16 in | 2.1 uL closed | 800 psi |
| P-451 | Splitter valve | 10-32 flat bottom | 1/16 in | 1.2 uL closed | 800 psi |
| P-470 | Graduated splitter | 10-32 coned | 1/16 in | 1.2 uL closed | 4000 psi |
| P-732 | Shut off valve | 1/4-28 flat bottom | 1/16 in | 2.5 uL | 500 psi |
| P-733 | Shut off valve | 1/4-28 flat bottom | 1/8 in | 10.0 uL | 500 psi |
| P-721 | Shut off valve, ETFE | 1/4-28 flat bottom | 1/8 in | 10.0 uL | 500 psi |
| P-782 | Shut off valve, ETFE | 1/4-28 flat bottom | 1/16 in | 2.5 uL | 500 psi |
| P-783 | Shut off valve, ETFE | 1/4-28 flat bottom | 1/8 in | 10.0 uL | 500 psi |
Key Features
Fine control at low flow – flow can be trimmed to 3.5 uL per minute from a 1 mL per minute inlet, which is the range needed ahead of a mass spectrometer
PEEK and PTFE flow path – no metal in the wetted path on the standard bodies, so metal sensitive samples stay clean
0.020 inch thru hole – the same restriction across the range keeps pressure drop predictable when a valve is swapped
Graduated splitter – the P-470 scale lets a working split setting be recorded and returned to after a service
Two port styles – 1/4-28 flat bottom and 10-32 coned bodies cover most analyser and LC ports
High pressure option – 4000 psi rating on the graduated splitter with a stainless needle for higher pressure lines
Shut off to the same ports – stop valves use the same port and tubing so the block does not need a second fitting family
Fixed fittings listed – each valve states which fittings are included, so the line can be connected on arrival
Applications
Valves are placed where flow has to be trimmed or stopped without changing the tubing run.
LC and MS interfaces – a splitter valve takes a controlled fraction of the column effluent to the source and sends the rest to waste
Peristaltic pump lines – a metering valve absorbs the pulse and sets a steady flow into the mixing coil
Degasser and sparging lines – a needle valve sets gas flow so the vacuum degasser is not overrun
Valves for these lines ship alongside SUPERB PCB and SUPERB PCBA builds when the analyser also needs a new control board.
Sourcing and Cross-Reference
Replacement is decided by port, tubing and function. A 10-32 coned port cannot take a 1/4-28 flat bottom body, and a body for 1/16 inch tubing cannot be made to seal on 1/8 inch tubing. Once the port and the tube are fixed, the choice between metering, splitter and shut off follows from what the line has to do.
Internal volume matters more than it looks on a low flow system. A valve with 7.7 uL swept volume ahead of a narrow column adds delay and mixes peaks; the same job done with a 1.2 uL body keeps the injection sharp. We state the volume at full open because that is the figure that changes with the setting.
For methods that have already been validated, the graduated splitter is the safer replacement: the setting can be read off the scale and restored, which avoids a revalidation. We keep both the standard and the graduated versions available so the choice can be made on the line rather than in a catalogue.
Where a valve block sits on its own manifold board, the assembly is built through SUPERB Assembly.
Q: Can you supply a metering valve that matches the port and swept volume of the one we are replacing?
A: Yes. SUPERB Tech matches port thread, tubing diameter and swept volume against the valve being replaced before quoting.
Why Choose SUPERB for Analyser Valves
- Port checked first: the body is selected on port thread and tubing diameter, which is where most wrong orders start
- Swept volume stated: internal volume at full open is quoted so the effect on peak shape can be judged before ordering
- PEEK wetted path: standard bodies keep metal out of the flow path for ion sensitive work
- Graduated option held: a documented split setting can be restored without revalidating the method