Bubble Traps and Debubblers for Analyser Lines

A bubble in a flow cell reads as a spike, and a bubble in a pump head reads as a lost prime. Both are easier to prevent than to clear, which is why a trap sits at the point where gas collects. This page covers inline traps from 25 microlitre to 300 microlitre, process traps from 0.4 to 4.6 litre, and vacuum degassing chambers from 480 to 925 microlitre. Each entry gives the volume, the membrane or the vent, the pressure and the mounting, because a trap that is mounted horizontally will not release the gas it collects. The micro traps remove free bubbles only. Dissolved gas has to be removed by vacuum or by helium sparging, and the page says which part does which.
quote now

Product Specifications

Product Overview

Gas enters a liquid line in three ways: it comes out of solution when the pressure drops, it is drawn in at a fitting, or it is generated by a reaction. A trap handles the first and the third by giving the gas somewhere to collect and a path to leave. The construction is simple, a chamber with a vent at the top and a membrane that passes gas but not liquid, and the performance depends on the volume, the membrane pore size and the mounting.

Volume sets the residence time. A 25 microlitre trap is small enough to sit between a syringe pump and a valve without adding a measurable delay, and it is used where a single bubble would spoil a 10 microlitre injection. A 300 microlitre trap holds more gas before it re entrain, and it is the size used on a perfusion loop or a bioreactor line where the flow is millilitres per minute. At process scale the trap is a litre scale vessel with a Tri-Clamp connection, and the volume is chosen against the flow: 0.4 litre for 220 litre per hour, 4.6 litre for 3200 litre per hour.

The membrane is what makes a trap a trap rather than a chamber. A 10 micrometre PTFE membrane is hydrophobic, so liquid does not wet it and gas passes through while the liquid is held back. The membrane is a wear item, and a membrane that has been wetted by a solvent with a low surface tension stops venting. That is why the membrane is replaceable and why the trap is mounted with the vent at the top: a trap on its side gives the gas no path to the membrane, and the most common report of a trap that does not work is a trap that was mounted horizontally.

Dissolved gas is a different problem. A vacuum degassing chamber pulls gas out of the liquid before it reaches the pump, and the useful figures are the swept volume, the flow it can treat and the pressure it holds. A 925 microlitre chamber at 6 millilitres per minute is a typical analytical size, and it holds 5 bar on the liquid side. A trap downstream of the pump cannot remove dissolved gas, because the gas comes out of solution after the pressure drop, so the two parts are used together rather than as alternatives.

Key Specifications

ParameterValueNote
Inline trap volume25 microlitre and 300 microlitrePEEK body, 1/4-28 UNF ports
Process trap volume0.4, 1.7, 2.7 and 4.6 litreglass or acrylic tube
Process flow rating220, 600, 1000 and 3200 litre per hourmatched to the trap volume
Process pressure6 bar on the first three sizes, 5 bar on 4.6 litre87 psi and 72 psi at 30 C
Process temperature3 to 60 Cthe acrylic tube has the same rating as the glass tube
Membrane10 micrometre PTFE, hydrophobicreplaceable, 5 per pack
Vacuum on the dry sideto -14.5 psispeeds venting on a high flow line
Degassing chamber volume480, 670 and 925 microlitrePEEK body, 1.14 mm internal diameter
Degassing chamber flow3, 4 and 6 mL/minnominal capacity per chamber
Degassing chamber pressure5 bar, which is 70 psiliquid side
Wetted materials, process trapstainless steel 1.4404, Ra below 0.4 micrometreglass, acrylic or stainless tube
Connections, process trapTri-Clamp 25 or 50, hose 1500 mmupper vent 3/4 in, bottom inlet 3/4 in
Sterilisationthe PEEK micro trap is autoclavablesterile versions are supplied sealed
Mountingvertical, both fluid ports at the lowest point, vent upa horizontal trap does not vent

Included Models and Order Codes

The table is split by scale: an inline trap for a microfluidic or analytical line, a process trap for a chromatography column, and a vacuum chamber for dissolved gas. The part number, the volume and the flow are given together, because a trap is chosen on the gas load and not on the volume alone.

PartTypeVolumeFlow or pressureMaterialNote
PG-BT-STD25ULInline bubble trap25 microlitrevacuum to -14.5 psiPEEK, 10 micrometre PTFE membranefor a microfluidic or narrow bore line
PG-BT-STD300ULInline bubble trap300 microlitrevacuum to -14.5 psiPEEK, 10 micrometre PTFE membranefor a perfusion or bioreactor loop
739-2410Process bubble trap, glass0.4 litre220 litre per hour, 6 barglass tube, 1.4404 wetted partsTri-Clamp 25
739-2412Process bubble trap, glass1.7 litre600 litre per hour, 6 barglass tube, 1.4404 wetted partsTri-Clamp 25
739-2414Process bubble trap, glass2.7 litre1000 litre per hour, 6 barglass tube, 1.4404 wetted partsTri-Clamp 50
739-2416Process bubble trap, glass4.6 litre3200 litre per hour, 5 barglass tube, 1.4404 wetted partsTri-Clamp 50
739-2420Process bubble trap, acrylic0.4 litre220 litre per hour, 6 baracrylic tube, 1.4404 wetted partsvisible gas collection
739-2422Process bubble trap, acrylic1.7 litre600 litre per hour, 6 baracrylic tube, 1.4404 wetted partsvisible gas collection
739-2424Process bubble trap, acrylic2.7 litre1000 litre per hour, 6 baracrylic tube, 1.4404 wetted partsvisible gas collection
739-2426Process bubble trap, acrylic4.6 litre3200 litre per hour, 5 baracrylic tube, 1.4404 wetted partsvisible gas collection
9000-1004Vacuum degassing chamber480 microlitre3 mL/min, 5 barPEEK, 1.14 mm IDanalytical, biocompatible
9000-1006Vacuum degassing chamber670 microlitre4 mL/min, 5 barPEEK, 1.14 mm IDanalytical, biocompatible
9000-1009Vacuum degassing chamber925 microlitre6 mL/min, 5 barPEEK, 1.14 mm IDthe largest analytical chamber
9000-1184Vacuum degassing chamber480 microlitre3 mL/min, 5 barPPS with stainless supportGPC version, not biocompatible
9000-1189Vacuum degassing chamber925 microlitre6 mL/min, 5 barPPS with stainless supportsemi preparative, not biocompatible

Bubble Size, Volume and Why the Two Are Confused

A trap is rated in microlitres, and the rating is the volume of the gas bubble it can hold, not the length of tube the bubble occupies. The two only agree at one internal diameter, which is why the same trap can look oversized on one line and marginal on another.

Bubble diameterSphere volumeLength in a 0.8 mm tubeLength in a 4.3 mm tubePractical meaning
1.0 mm0.52 microlitre1.0 mm0.03 mma 25 microlitre trap holds about 48 of these
2.0 mm4.19 microlitre8.3 mm0.29 mma 25 microlitre trap holds about 6 of these
3.5 mm22.4 microlitre44.7 mm1.5 mmone bubble fills most of a 25 microlitre trap
5.0 mm65.4 microlitre130 mm4.5 mma 300 microlitre trap holds about 4 of these
8.0 mm268 microlitre533 mm18.5 mma process trap is the only practical size
19 mm3590 microlitre7.1 m248 mmthe gas pocket on a litre scale trap

Key Features

Inline traps from 25 microlitre – small enough to sit between a pump and a valve without adding a measurable delay

300 microlitre for a perfusion loop – holds a gas pocket that a 25 microlitre trap would pass on

10 micrometre PTFE membrane – hydrophobic, passes gas and holds liquid, and is replaceable in the field

Vacuum assist to -14.5 psi – vents a high flow line faster than atmosphere alone

Process traps to 4.6 litre – 3200 litre per hour with a Tri-Clamp connection and 1.4404 wetted parts

Vacuum chambers from 480 microlitre – removes dissolved gas before the pump rather than after the pressure drop

Autoclavable PEEK micro trap – a sterile version is supplied sealed for a cell culture line

Mounting instruction on the page – vertical with the vent up, because a horizontal trap is the usual reason a trap appears not to work

Applications

A trap is placed where gas collects, not where it is convenient. The three cases below set the volume and the membrane.

HPLC and UHPLC mobile phase – a vacuum degasser removes dissolved gas before the pump, and a 25 microlitre inline trap catches the bubble that forms at the detector. The two parts work as a pair, because the trap cannot remove gas that is still in solution.

Clinical chemistry and immunoassay – a reagent line draws from a pack that is pierced once and used for days. A 300 microlitre PEEK trap with a PTFE membrane holds the gas that enters at the pierce point, and the autoclavable body suits a line that is cleaned between batches.

Process chromatography – a 1.7 or 2.7 litre glass trap sits on top of a column and vents the gas that comes out of a buffer at low pressure. The rated flow and the Tri-Clamp size are chosen together, and the acrylic version is used where the gas pocket has to be seen.

The pump and detector boards behind these lines are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the bubble detector input to the pump drive.

Sourcing and Cross-Reference

A trap is ordered by volume, flow and connection, and the connection is what goes wrong on a replacement. A Tri-Clamp 25 and a Tri-Clamp 50 look similar in a catalogue photo and do not fit the same ferrule, and a 1/4-28 UNF port will not accept a 10-32 fitting. The flange or the port on the existing line is confirmed before a part is named.

Where a trap is no longer supplied, the substitute is chosen on volume and flow first and on material second. A 300 microlitre trap in place of a 25 microlitre trap adds 275 microlitres to the delay volume, which matters on a narrow bore line. Where the volume has to be kept, we say so rather than offering the next size up.

Membranes are a consumable and should be ordered with the trap. A 10 micrometre PTFE membrane is the standard, and a membrane that has been in contact with a low surface tension solvent is replaced rather than cleaned. We ship the membrane pack with the trap and note the solvent groups that shorten its life.

Where a fluidic panel is assembled with its traps and degassers, the parts are picked as one set, which is how SUPERB Assembly ships a reagent panel.

Q: Can you supply a replacement membrane and the same volume trap when the original part is discontinued?

A: SUPERB Tech can confirm the volume, the connection and the membrane against the old trap before the order is placed.

Why Buy Bubble Traps from SUPERB

  1. Volume, flow and pressure together: a trap is not offered on volume alone, because the flow rating and the pressure fall as the volume rises
  2. Free and dissolved gas separated: the page says which part removes which, so a trap is not bought for a job it cannot do
  3. Membrane listed as a consumable: the 10 micrometre PTFE membrane is ordered with the trap rather than after it wets out
  4. Connection confirmed: Tri-Clamp 25 or 50 and 1/4-28 UNF are checked against the line before the part is named
  5. Mounting stated: vertical with the vent up, which is the part of the installation that decides whether it works

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Air Bubble Detectors for Infusion and Reagent Lines, Inline Degassing Modules and Vacuum Degassers for Liquid Lines, Inline Static Mixers for Reagent Streams, Syringe Pumps and Precision Dispensers for Analysers, Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Component Lead Time Solutions.