Contact Us

Tube Fittings and Ferrules for Reagent Lines

Flangeless fittings and ferrules for analyser and HPLC reagent lines. 1/4-28 flat bottom, 10-32 coned and M6 ports for 1/16 and 1/8 inch tubing. ETFE ferrules to 2000 psi, PEEK and stainless two piece sets to 9000 psi. Nuts and ferrules listed separately so a single worn part can be replaced.
quote now

Product Specifications

Product Overview

A reagent line that drips is almost never a pump problem. It is a ferrule that was tightened twice on the same tubing, a nut that does not match the port, or a fitting reused after the tube was cut back. On an analyser the leak shows up as a drifting baseline or a bubble in the cell, and by then the fitting has usually been replaced once already with the wrong type.

This page separates the two things that are ordered together: the nut and the ferrule. Flangeless fittings use an ETFE ferrule that seals on the outside of the tube and can be reused a few times; two piece fingertight fittings use a PEEK or stainless ferrule that grips harder and holds to 7000 psi and above. Both are listed with the port they fit, because a 1/4-28 flat bottom nut will not seal in a 10-32 coned port however hard it is turned.

Pressure rating is the second trap. A standard flangeless fitting with an ETFE ferrule is rated to 2000 psi, while the same style in 1/8 inch tubing is rated to 500 psi because of the larger ferrule area. Mixing the two on one manifold leaves the 1/8 inch line as the weak point, and the leak appears at the fitting rather than at the pump.

Key Specifications

ParameterValueNotes
Port threads1/4-28 flat bottom, 10-32 coned, M6 flat bottomChosen to match the port, not the tube
Tubing outside diameter1/16 in and 1/8 inAlso 1/32 in on some nuts
Nut materialsPEEK, ETFE, Delrin, PPS, stainless steelColour codes the ferrule size
Ferrule materialsETFE, PEEK, stainless steelETFE is reusable, PEEK grips harder
Flangeless rating2000 psi (138 bar)1/16 in with ETFE ferrule
1/8 in fitting rating500 psi (34 bar)Larger ferrule area lowers the rating
Two piece rating7000 to 9000 psi (483 to 620 bar)Fingertight PEEK or stainless
One piece rating1000 psi (69 bar)Super flangeless, 1/16 in
Small valve ferrule2500 psi (172 bar)P-240 for 1/4-28 or 10-32
Head stylesKnurl, headless knurl, hex, FlushNutHeadless for tight manifolds

Included Models and Order Codes

Nuts and ferrules are listed in separate groups so a worn ferrule can be replaced without buying a full fitting. The number after the dash on an XP part is the pack size, where x is ten pieces. Where a fitting is supplied with a ferrule the included ferrule is named, because the pressure rating belongs to the pair and not to the nut.

Part NumberTypePortTubing ODIncluded FerruleMax Pressure
XP-230Flangeless fitting1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-235Flangeless fitting, short1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-201Flangeless fitting1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-202Flangeless fitting, red1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-218Flangeless fitting, ETFE1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-245Flangeless fitting, PFA1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-286Flangeless fitting, headless1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XF-358Flangeless fitting, FlushNut1/4-28 flat bottom1/16 inP-200 ETFE2000 psi
XP-330Flangeless fitting1/4-28 flat bottom1/8 inP-300 ETFE500 psi
XP-335Flangeless fitting, short1/4-28 flat bottom1/8 inP-300 ETFE500 psi
P-200Replacement ferrule1/4-28 flat bottom1/16 inNot applicable2000 psi
P-200NReplacement ferrule, natural1/4-28 flat bottom1/16 inNot applicable2000 psi
P-240Ferrule for small valve1/4-28 or 10-321/16 inNot applicable2500 psi
P-250-10Super flangeless ferrule1/4-28 flat bottom1/16 inNot applicable2500 psi
P-259-10Super flangeless ferrule, ETFE1/4-28 flat bottom1/16 inNot applicable1350 psi
F-190-10SealTight two piece fitting10-32 coned1/16 inF-192 PEEK7000 to 9000 psi
F-192-10SealTight replacement ferrule10-32 or M6 coned1/16 inNot applicable7000 to 9000 psi

Key Features

Nut and ferrule separate – a worn ferrule can be replaced on its own instead of scrapping a fitting that is still good

ETFE seals and reseals – the flangeless ferrule can be retightened a few times, which suits lines that are opened during service

Colour coded nuts – the nut colour identifies the ferrule size, so a 1/16 inch line is not fitted with a 1/8 inch ferrule by mistake

Headless option – a headless knurl fits manifolds where there is no room to swing a full knob

Two pressure classes – 2000 psi for 1/16 inch and 500 psi for 1/8 inch are stated apart so one line does not set the limit for another

7000 psi two piece set – the PEEK and stainless SealTight pair holds high pressure without a flanging tool

Small valve ferrule – P-240 seats in a 1/4-28 or 10-32 valve port where a standard ferrule bottoms out

Ten piece packs – packs of ten match the way fittings are actually consumed on a service visit

Applications

Fittings are chosen by port and tube, then by the pressure the line has to hold.

HPLC and reagent manifolds – flangeless fittings connect PEEK tubing to valves and tees without a flanging tool

Analyser sample and waste lines – colour coded nuts keep the 1/16 and 1/8 inch runs from being mixed during service

High pressure LC paths – two piece fingertight sets hold to 7000 psi and above on narrow bore columns

Fluidic fittings are supplied with SUPERB PCB and SUPERB PCBA orders when an analyser build needs both the manifold and its control board.

Sourcing and Cross-Reference

Every fitting order starts with the port. 1/4-28 flat bottom, 10-32 coned and M6 flat bottom are not interchangeable, and a nut that matches the thread but not the seat will tighten and still leak. We ask for the port and the tube outside diameter, and only then pick the style.

The ferrule decides the pressure rating, not the nut. An ETFE ferrule in a 1/16 inch fitting holds 2000 psi; the same nut with a PEEK ferrule from a two piece set holds 7000 psi and above. Substituting one for the other to save cost moves the weak point, and it usually moves it to a line that is harder to reach.

Because ferrules are the part that wears, we quote them in packs alongside the nuts and name the included ferrule on every fitting line. That way a repeat order can be a ferrule pack rather than a full fitting, which is what most service kits actually need.

Manifold plates that carry these fittings are assembled through SUPERB Assembly before the line is pressure tested.

Q: Can you supply the ferrule on its own when only that part has worn out?

A: Yes. SUPERB Tech supplies ferrules as separate parts, so a worn ferrule does not mean a new fitting.

Why Choose SUPERB for Fluidic Fittings

  1. Port matched first: the nut is selected on thread and seat, which is where leaks begin
  2. Rating per pair: pressure is quoted for the nut and ferrule together, not for the nut alone
  3. Ferrules sold apart: a worn ferrule can be replaced without buying the fitting again
  4. Consistent colour coding: nut colour and ferrule size stay paired so a service kit cannot be assembled wrongly

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Needle and Metering Valves, Inline Filters and Frits, Sample Loops and Injection Loops, Peristaltic Pump Tubing and Pump Heads, Rotary and Injection Valves, Obsolete Part Cross Reference Hub.