Inline Static Mixers for Reagent Streams

A static mixer has no moving parts, so its performance is set by the element geometry and by the volume it adds to the flow path. Too small and a gradient arrives as a series of steps; too large and a peak broadens before it reaches the detector. This page covers in-line assemblies, binary and ternary tees, replacement cartridges and process elements from 0.5 microlitre to 15 millilitres. Body materials are 316 stainless steel to 6,000 psi, PEEK to 3,000 psi and a UHPLC range to 18,000 psi, with a standard temperature of 40 C and a high temperature option to 80 C. The volume and the fitting are listed together, because a mixer chosen on volume alone can arrive with a port that does not match the union on the line.
quote now

Product Specifications

Product Overview

A static mixer works by splitting and recombining the flow, so the number of elements and the element geometry set how many times the stream is divided. The result is a pressure drop and a mixing length, and both scale with flow rate. At a low flow a small mixer gives good mixing with little delay; at a high flow the same mixer is a restriction and the mixing is incomplete. That is why the same family is offered from 0.5 microlitre to 15 millilitres rather than as one universal part.

Volume is the number that matters for chromatography. A gradient pump delivers a step change in solvent composition, and the mixer turns that step into a ramp. A 0.5 microlitre mixer gives a short ramp and a short delay, which is what a narrow bore column needs. A 1.5 millilitre mixer gives a smoother ramp and a longer delay, which is what a standard bore column with a noisy pump needs. The rule used in practice is to match the mixer volume to the system dwell volume and then choose the smaller of the two if peak shape matters more than baseline noise.

Pressure is the second constraint. A 316 stainless body carries 6,000 psi, which covers a standard HPLC system. A UHPLC system runs to 18,000 psi, and the mixer in that system has to be a high pressure part rather than a standard part with a different label. A PEEK body carries 3,000 psi and gives a metal free flow path, which is used where a stainless union would release iron into a protein or a chelator stream.

Temperature and connections close the choice. The standard rating is 40 C, and a high temperature version covers 80 C for the small volumes, which matters for a reagent stream that is heated to speed a reaction. The connection is a 10-32 coned port or a 1/4-28 flat bottom port, and the two are not interchangeable. A mixer with the wrong port will either leak at the thread or, worse, seal on the ferrule and leave a dead volume at the bottom of the port. For process scale, a helical element in a pipe spool does the same job at a much larger volume, and the Kenics element is the reference for that geometry.

Key Specifications

ParameterValueNote
Volume range, standard HPLC0.5 microlitre to 15 mL316 stainless steel body
Volume range, UHPLC0.5 microlitre to 1.5 mL18,000 psi rating
Volume range, PEEK0.5 microlitre to 1.5 mL3,000 psi rating, metal free path
Pressure, standard HPLC6,000 psi316 stainless body
Pressure, UHPLC18,000 psicomplete assembly, not a standard body relabelled
Pressure, PEEK3,000 psifor protein and chelator streams
Pressure, micro high temperature12,000 psi for 0.5 to 25 microlitrehigh temperature and high pressure together
Temperature, standard40 C across all volumesthe rating for the standard and PEEK ranges
Temperature, high temperature option80 C maximum for 0.5 microlitre to 1.5 mLstandard HPLC and UHPLC
Configurationsin-line assembly, binary tee, ternary tee, replacement cartridgethe tee versions add the second or third stream at the mixer body
Port types10-32 coned and 1/4-28 flat bottoma flat bottom port needs a flat seat ferrule
Materials316 stainless steel, PEEKPEEK for a metal free path
Process elementhelical element in a pipe spool, 3/16 to 120 incarbon steel, 304 or 316 stainless
Element countselected on the number of flow divisions neededmore elements give better mixing and a higher pressure drop

Included Models and Order Codes

Part numbers are grouped by body and volume. A letter X in a part number is the fitting code for the port on that assembly, so the same volume is ordered with a different letter for a different union. Where a family is rated for both a standard and a high pressure version, the two are listed as separate rows rather than as one row with a range.

PartTypeVolumePressureMaterialNote
401-000.5In-line assembly0.5 microlitre6,000 psi316 stainless steelsmallest in-line body in the standard range
402-000.5Binary tee assembly0.5 microlitre6,000 psi316 stainless steeltwo streams meet at the mixer body
403-000.5Ternary tee assembly0.5 microlitre6,000 psi316 stainless steelthree streams, for a modifier and a reagent
400-000.5Replacement cartridge0.5 microlitre6,000 psi316 stainless steelthe element is changed without changing the body
40X-000.5Standard HPLC mixer0.5 microlitre6,000 psi316 stainless steelX is the fitting code
40X-000.5HPUHPLC mixer0.5 microlitre18,000 psi316 stainless steelfor a sub 2 micron column system
40X-000.5BBiocompatible mixer0.5 microlitre3,000 psiPEEKmetal free path
43X-1500HPUHPLC mixer, large volume1.5 mL18,000 psi316 stainless steelthe largest volume in the UHPLC range
43X-1500BBiocompatible mixer, large volume1.5 mL3,000 psiPEEKfor a preparative PEEK line
U-466Static mixing tee2.2 microlitre swept volumeper the maker data316 stainless steelfor 0.5 to 3 mL/min microbore and analytical gradients
U-466SStatic mixing tee, small bore2.2 microlitre swept volumeper the maker data316 stainless steelsame body, smaller port bore
Kenics KMProcess static mixerpipe volume, 3/16 to 120 inper the pipe schedulecarbon steel, 304 or 316 stainlesshelical element, for a reagent or a dilution stream at process scale

Where the Mixer Sits and What It Fixes

The same mixer body does a different job in four positions. This table is the short form used when a flow path is reviewed, and it is also the one to check when a baseline problem is blamed on the pump.

PositionJobVolume to chooseSymptom it removesTrade
After the pump, before the injection valveturns a gradient step into a rampclose to the system dwell volumestepped or stair case gradienta larger volume adds delay
At a reagent junctionmixes sample and reagent before the detectorlarger, up to ten times the rate limiting flowspikes or striations at the detectoradds dispersion
Ahead of a narrow bore columnkeeps the delay short0.5 to 25 microlitrebroad injection pluga small volume may not smooth a noisy pump
On a process dilution lineblends a concentrate into a carrierpipe volume, element count by flowconcentration streaks downstreampressure drop rises with flow
In a PEEK or metal free linemixes without a stainless wetted part0.5 microlitre to 1.5 mLmetal uptake in a protein streampressure limited to 3,000 psi

Key Features

0.5 microlitre to 15 mL – one family covers a narrow bore gradient and a preparative line, so the volume is chosen rather than the supplier

Three pressure classes – 6,000 psi standard, 18,000 psi UHPLC and 3,000 psi PEEK, named on the part rather than implied

In-line, binary and ternary bodies – the second and third stream can meet at the mixer instead of at a separate tee

Replacement cartridges – the element is changed without cutting the tubing or replacing the body

High temperature option to 80 C – for a reagent stream that is heated to speed a reaction

10-32 and 1/4-28 ports – the port is named with the volume, because the two do not seal into each other

Process helical elements – a pipe spool element covers a flow that is far above a laboratory mixer

Metal free PEEK range – 3,000 psi with no stainless in the wetted path

Applications

Static mixers are used wherever two liquids have to be combined without a stirred vessel. The three cases below set the volume and the material.

Gradient HPLC and UHPLC – a binary pump delivers two solvents as a step. A 0.5 to 25 microlitre mixer turns the step into a ramp; on a standard bore system a 1.5 mL mixer smooths the pump ripple instead. Choosing the volume against the dwell volume is what keeps the retention time repeatable.

Clinical chemistry reagent mixing – sample and reagent meet at a tee and have to be mixed before the detector. A 50 to 500 microlitre body with a ternary tee adds the reagent at the mixer, and a PEEK body keeps iron out of a reaction that is sensitive to it.

Process dilution and pH adjustment – a concentrate is blended into a carrier at litres per minute. A helical element in a pipe spool gives complete mixing with no moving part, and the element count is set by the flow and the allowed pressure drop.

The pump and detector boards behind these flow paths are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the gradient control to the detector output.

Sourcing and Cross-Reference

A mixer is ordered by volume, pressure class and port, and the port is the field that is missed. A 1/4-28 flat bottom port and a 10-32 coned port look similar in a photo and do not seal into each other, so the union on the line is confirmed before a part is named. Where the old mixer is still fitted, the port is measured rather than guessed from the tubing.

Where a volume is no longer supplied, the substitute is chosen on the same volume first and on the same body second. A 2.2 microlitre tee and a 2 microlitre in-line body are close enough for a gradient, but a 25 microlitre body in place of a 2.2 microlitre tee changes the delay volume and the retention time. Where no equivalent exists at the same volume, we say so before shipping a mixer that fits the thread but shifts the chromatogram.

The process side is a different selection. A helical element is sized on the pipe diameter and the flow, and the number of elements is set by the mixing requirement and the allowed pressure drop. We quote the element and the spool as a pair and state the pressure drop at the rated flow, because a mixer that is correct at half flow can be a restriction at full flow.

Where a flow path is assembled as a module, the mixer, the tee and the tubing are picked as one set, which is how SUPERB Assembly ships a fluidic manifold.

Q: Can you supply a static mixer at the same volume when the original part is no longer listed?

A: SUPERB Tech can confirm the volume, the pressure class and the port against the old mixer before the order is placed.

Why Buy Static Mixers from SUPERB

  1. Volume and pressure quoted together: a mixer is never offered on volume alone, because the same body has a different rating in stainless and in PEEK
  2. Port confirmed against the line: the 10-32 and 1/4-28 ports are checked before the part is named
  3. Tee and in-line bodies on one list: the second stream can meet at the mixer rather than at a separate fitting
  4. Replacement cartridges: the element is changed without cutting the tubing
  5. Process and laboratory elements: the same page covers a 0.5 microlitre gradient mixer and a pipe spool element

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Bubble Traps and Debubblers for Analyser Lines, Peristaltic and Diaphragm Micro Pumps for Analysers, Inline Filters and Frits for Reagent Lines, Sample Loops and Injection Loops for Valve Injection, Solenoid and Pinch Valves for Fluid Handling in Analysers, Component Lead Time Solutions.