  • Gas Mass Flow Sensors for Ventilators and Analysers

Gas Mass Flow Sensors for Ventilators and Analysers

A ventilator or a gas analyser has to know how much gas moved, not how fast the pressure changed. The sensors below are thermal mass flow meters with a calibrated digital output, so the host reads standard litres per minute instead of a raw bridge voltage. Flow ranges run from 20 slm on a gas mixing sensor to 300 slm on an inspiratory channel, and each one is quoted with its pressure drop at the flow that matters. Ranges are listed with accuracy per band, because the error grows at the top of a range and a single percentage hides that.
quote now

Product Specifications

Product Overview

A thermal mass flow sensor puts a heater in the gas path and measures how much heat the flow carries away, which makes the reading a mass flow rather than a volume flow. That matters in a ventilator, because the gas that reaches the patient is defined at body temperature and ambient pressure, while the gas that leaves the blender is cold and compressed. A sensor calibrated in standard litres per minute removes that correction from the firmware.

The first figure to look at is the pressure drop at the working flow. A sensor that drops 600 Pa at 200 slm is asking the turbine or the blower for that pressure, and on a spontaneous breathing circuit it also changes the work of breathing. The second is accuracy at the bottom of the range, because a ventilator spends most of its time at low flow and a sensor with a 1.5 percent of reading error still has an offset that dominates at 5 slm. The third is the response time, which decides whether the sensor can follow a breath at all.

The third point is calibration gas. A sensor calibrated for air reads oxygen correctly only if it was calibrated for oxygen or for a mixture, and the same is true for carbon dioxide and nitrous oxide. The selection tables publish which gases each family was calibrated for, and that line decides whether a sensor can sit in an inspiratory limb, an expiratory limb or a gas mixing manifold. A sensor calibrated for air and used in oxygen is a common and expensive mistake.

The fourth point is the fluidic interface, because it sets the leak path. Medical cones in 22 mm and 15 mm mate with the breathing circuit, an O-ring face seals onto a manifold, and a push-in fitting takes a tube. A sensor with the wrong interface needs an adapter, and every adapter is a place where a leak can start, which is why the interface is chosen together with the flow range and not after it.

Key Specifications

ParameterValueNote
Measurement principlethermal mass flow with a calibrated digital outputreads standard litres per minute, not a raw bridge voltage
Flow ranges in the range20, 50, 150, 200, 240, 250 and 300 slm depending on familybidirectional versions read negative flow on inspiration and expiration
Accuracy1.5 to 4 percent of reading, stated per flow banderror grows at the top of a range, so the band matters
Resolution0.4 sccm on the 20 slm gas mixing sensor, 1.0 sccm on the 50 slm versionresolution is the smallest change the reading can show
Pressure dropunder 100 Pa at 60 slm and under 600 Pa at 200 slm on the low drop familythe blower has to supply this pressure, and the patient feels it
Response timeflow step response under 5 ms on the mixing sensors, 0.5 ms sampling on the ventilator familydecides whether a breath waveform can be resolved
Gases calibratedair, oxygen, air and oxygen mixtures, carbon dioxide and nitrous oxide depending on modela sensor calibrated for air is not an oxygen sensor
Supply voltage5 V on the SFM3000 and SFM3200 families, 3.3 V on the SFM3119, SFM3505 and SFM43003.3 V parts still accept 5 V tolerant logic in most builds
InterfaceI2C on the compact families, RS485 on the manifold familyI2C needs a short cable, RS485 tolerates a long one
Operating pressure0.7 to 1.3 bar absolute on the SFM3000, up to 7 bar on the SFM4300a sensor behind a regulator needs the high pressure version
Temperature range-20 to +80 C on the SFM3000, +10 to +50 C on the SFM3200condensation is the limit on an expiratory channel
Fluidic interface22 mm and 15 mm medical cones, O-ring face, push-in fitting or down mountthe interface is chosen with the flow range, not after it

Included Models and Order Codes

Seven sensors cover the ventilator and analyser range. The SFM3000 is the general purpose part at plus or minus 200 slm with a 1.5 percent of reading accuracy and a pressure drop under 600 Pa at full flow. The SFM3200 is the expiratory part, bidirectional from minus 100 to plus 250 slm, with a 22 mm medical cone and an integrated heater for condensation. The SFM3003-300-CL and SFM3505-300 reach 300 slm for an inspiratory channel, the SFM3119 gives a 240 slm digital part on 3.3 V, and the SFM4300 covers gas mixing at 20 and 50 slm with calibration for oxygen, carbon dioxide and nitrous oxide.

SensorFlow rangeAccuracyPressure dropInterfaceNote
SFM3000-200 to +200 slm1.5 percent of reading typicalunder 100 Pa at 60 slm, under 600 Pa at 200 slmI2C, 5 Vbidirectional, air, nitrogen and oxygen
SFM3200-100 to +250 slm3 percent of reading at 100 slm100 Pa at 60 slm, 750 Pa at 200 slmI2C, 5 V22 mm medical cone, heater for condensation
SFM3003-300-CL-30 to +300 slm2 percent of reading at 200 slm80 Pa at 60 slm, 500 Pa at 200 slmI2C, 2.7 to 5.5 V22 mm ISO 5356 cone, inspiratory channel
SFM3505-300-150 to +300 slm2.3 percent of reading over the range85 Pa at 60 slm, 750 Pa at 200 slmI2C, 3.3 Vextended negative range for expiration
SFM3119-10 to +240 slm2 percent of reading at 100 slm200 Pa at 60 slm, 1,600 Pa at 200 slmdigital, 3.3 Vcompact body with a NTC in the gas path
SFM4300-20-O0 to 20 slm2 percent of reading for air and oxygen2,500 Pa at 20 slmI2C, 3.3 Vgas mixing, calibrated for oxygen, carbon dioxide and nitrous oxide
SFM4300-50-O0 to 50 slm4 percent of reading at 50 slm10,000 Pa at 50 slmI2C, 3.3 Voperating pressure up to 7 bar

Pressure Drop and Accuracy by Flow Band

A single accuracy figure hides the band where the error is largest. The table below is the short version for the three flow classes an analyser usually needs, taken from the published accuracy bands of the families above.

Flow classTypical useAccuracy band to readPressure drop to check
Low flow, 0 to 5 slmgas mixing, anaesthesia agent deliveryzero point accuracy dominates, read the offset figureunder 2,500 Pa on a mixing sensor
Mid flow, 5 to 60 slmspontaneous breathing, sidestream sampling1.5 to 2 percent of readingunder 100 Pa on the low drop family
High flow, 60 to 300 slminspiratory limb, high frequency ventilation2 to 4 percent of reading, band by band500 to 750 Pa at 200 slm
Bidirectional flowexpiratory limb with a leakread the negative flow band separatelymatters more on expiration than on inspiration
Gases other than airoxygen, carbon dioxide, nitrous oxideonly the calibrated gas list appliesdensity changes the drop, so check per gas

Key Features

Standard litres per minute – the reading is already a mass flow, so the host does not correct for temperature and pressure.

Accuracy given per band – the error at 5 slm and at 250 slm are listed separately instead of one figure covering the whole range.

Pressure drop stated – the drop at 60 slm and at 200 slm is written down, because the blower and the patient both see it.

Calibrated gas list – air, oxygen, carbon dioxide and nitrous oxide are listed per part, which is the line that decides where a sensor can sit.

Medical cone options – 22 mm and 15 mm cones mate with the breathing circuit without an adapter, and every adapter is a potential leak.

Heater on the expiratory part – the SFM3200 carries an integrated heater, which is what keeps condensation off the sensor in a warm humid expiratory limb.

Fast step response – under 5 ms on the mixing sensors, enough to resolve a breath waveform rather than an average.

Two pressure classes – the SFM3000 works from 0.7 to 1.3 bar absolute, the SFM4300 up to 7 bar, so a sensor behind a regulator has a version.

Applications

Three gas paths where the flow range, the gas list and the pressure class decide the sensor together.

Critical care ventilator – an inspiratory sensor at 300 slm and an expiratory sensor with a heater and a 22 mm cone, both read on the same I2C bus with different addresses.

Anaesthesia gas mixing – oxygen, carbon dioxide and nitrous oxide are blended at 20 to 50 slm, and only a sensor calibrated for those three gases can sit in that manifold.

Gas analyser sampling – a low flow sensor meters sample into a cell while a second one watches the bypass, so the two readings are taken at the same pressure.

A flow channel is a small analogue board, so the SUPERB PCBA and the SUPERB PCB keep the heater return and the I2C lines apart in the same hand held housing.

Sourcing and Cross-Reference

The first question on a flow sensor build is which gas the calibration was made for. A sensor calibrated in air reads high in oxygen and low in carbon dioxide, and the error is large enough to fail a ventilator test. The selection tables list the calibrated gases per family, so the short list is made from that line before the flow range is compared.

The second question is the interface and the leak path. A 22 mm cone mates with a breathing circuit, an O-ring face seals onto a manifold, and a push-in fitting takes a tube, and each of those is a different sealing problem. A sensor with the wrong interface costs an adapter, and an adapter is a place where a leak starts.

The third question is condensation. An expiratory limb carries warm saturated gas, and a cold sensor in that path collects water until the reading drifts. A heated sensor, or a sensor placed after a water trap, is the answer, and the heater is listed on the part rather than added later. See lead time and availability for a specific sensor and BOM shortage recovery where a build is held up by one part.

The SUPERB Assembly team seals the cone joint before the sensor is screwed down, because a leak at the cone reads as a flow error at low rates.

Q: Can you supply the flow sensor with the gas calibration already matched to the mixture in our breathing circuit?

A: Yes. Send the gas mixture, the flow range, the interface and the supply, and the SUPERB Tech team returns the sensors and the calibrated gases that fit.

Why Ventilator Builders Source Flow Sensors Through SUPERB

  1. Band accuracy: the error at low flow is listed separately, because a ventilator spends most of a breath below 20 slm.
  2. Pressure drop published: the drop at the working flow is written down, so the blower is not sized against a sensor with an unknown restriction.
  3. Gas list per part: air, oxygen, carbon dioxide and nitrous oxide are stated, which is the first filter in any ventilator verification review.
  4. Interface options: cones, O-ring faces and push-in fittings are quoted with the part, so no adapter is added after the fact.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Differential Pressure Sensors for Ventilators, Oxygen Sensor Cells for Ventilators, CO2 Sensor Modules for Medical and Building Use, Liquid Flow Sensors for Medical and Analytical Instruments, Peristaltic and Diaphragm Micro Pumps, BOM Shortage Recovery.