Gas Mass Flow Sensors for Ventilators and Analysers
Product Specifications
Product Overview
A thermal mass flow sensor puts a heater in the gas path and measures how much heat the flow carries away, which makes the reading a mass flow rather than a volume flow. That matters in a ventilator, because the gas that reaches the patient is defined at body temperature and ambient pressure, while the gas that leaves the blender is cold and compressed. A sensor calibrated in standard litres per minute removes that correction from the firmware.
The first figure to look at is the pressure drop at the working flow. A sensor that drops 600 Pa at 200 slm is asking the turbine or the blower for that pressure, and on a spontaneous breathing circuit it also changes the work of breathing. The second is accuracy at the bottom of the range, because a ventilator spends most of its time at low flow and a sensor with a 1.5 percent of reading error still has an offset that dominates at 5 slm. The third is the response time, which decides whether the sensor can follow a breath at all.
The third point is calibration gas. A sensor calibrated for air reads oxygen correctly only if it was calibrated for oxygen or for a mixture, and the same is true for carbon dioxide and nitrous oxide. The selection tables publish which gases each family was calibrated for, and that line decides whether a sensor can sit in an inspiratory limb, an expiratory limb or a gas mixing manifold. A sensor calibrated for air and used in oxygen is a common and expensive mistake.
The fourth point is the fluidic interface, because it sets the leak path. Medical cones in 22 mm and 15 mm mate with the breathing circuit, an O-ring face seals onto a manifold, and a push-in fitting takes a tube. A sensor with the wrong interface needs an adapter, and every adapter is a place where a leak can start, which is why the interface is chosen together with the flow range and not after it.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Measurement principle | thermal mass flow with a calibrated digital output | reads standard litres per minute, not a raw bridge voltage |
| Flow ranges in the range | 20, 50, 150, 200, 240, 250 and 300 slm depending on family | bidirectional versions read negative flow on inspiration and expiration |
| Accuracy | 1.5 to 4 percent of reading, stated per flow band | error grows at the top of a range, so the band matters |
| Resolution | 0.4 sccm on the 20 slm gas mixing sensor, 1.0 sccm on the 50 slm version | resolution is the smallest change the reading can show |
| Pressure drop | under 100 Pa at 60 slm and under 600 Pa at 200 slm on the low drop family | the blower has to supply this pressure, and the patient feels it |
| Response time | flow step response under 5 ms on the mixing sensors, 0.5 ms sampling on the ventilator family | decides whether a breath waveform can be resolved |
| Gases calibrated | air, oxygen, air and oxygen mixtures, carbon dioxide and nitrous oxide depending on model | a sensor calibrated for air is not an oxygen sensor |
| Supply voltage | 5 V on the SFM3000 and SFM3200 families, 3.3 V on the SFM3119, SFM3505 and SFM4300 | 3.3 V parts still accept 5 V tolerant logic in most builds |
| Interface | I2C on the compact families, RS485 on the manifold family | I2C needs a short cable, RS485 tolerates a long one |
| Operating pressure | 0.7 to 1.3 bar absolute on the SFM3000, up to 7 bar on the SFM4300 | a sensor behind a regulator needs the high pressure version |
| Temperature range | -20 to +80 C on the SFM3000, +10 to +50 C on the SFM3200 | condensation is the limit on an expiratory channel |
| Fluidic interface | 22 mm and 15 mm medical cones, O-ring face, push-in fitting or down mount | the interface is chosen with the flow range, not after it |
Included Models and Order Codes
Seven sensors cover the ventilator and analyser range. The SFM3000 is the general purpose part at plus or minus 200 slm with a 1.5 percent of reading accuracy and a pressure drop under 600 Pa at full flow. The SFM3200 is the expiratory part, bidirectional from minus 100 to plus 250 slm, with a 22 mm medical cone and an integrated heater for condensation. The SFM3003-300-CL and SFM3505-300 reach 300 slm for an inspiratory channel, the SFM3119 gives a 240 slm digital part on 3.3 V, and the SFM4300 covers gas mixing at 20 and 50 slm with calibration for oxygen, carbon dioxide and nitrous oxide.
| Sensor | Flow range | Accuracy | Pressure drop | Interface | Note |
|---|---|---|---|---|---|
| SFM3000 | -200 to +200 slm | 1.5 percent of reading typical | under 100 Pa at 60 slm, under 600 Pa at 200 slm | I2C, 5 V | bidirectional, air, nitrogen and oxygen |
| SFM3200 | -100 to +250 slm | 3 percent of reading at 100 slm | 100 Pa at 60 slm, 750 Pa at 200 slm | I2C, 5 V | 22 mm medical cone, heater for condensation |
| SFM3003-300-CL | -30 to +300 slm | 2 percent of reading at 200 slm | 80 Pa at 60 slm, 500 Pa at 200 slm | I2C, 2.7 to 5.5 V | 22 mm ISO 5356 cone, inspiratory channel |
| SFM3505-300 | -150 to +300 slm | 2.3 percent of reading over the range | 85 Pa at 60 slm, 750 Pa at 200 slm | I2C, 3.3 V | extended negative range for expiration |
| SFM3119 | -10 to +240 slm | 2 percent of reading at 100 slm | 200 Pa at 60 slm, 1,600 Pa at 200 slm | digital, 3.3 V | compact body with a NTC in the gas path |
| SFM4300-20-O | 0 to 20 slm | 2 percent of reading for air and oxygen | 2,500 Pa at 20 slm | I2C, 3.3 V | gas mixing, calibrated for oxygen, carbon dioxide and nitrous oxide |
| SFM4300-50-O | 0 to 50 slm | 4 percent of reading at 50 slm | 10,000 Pa at 50 slm | I2C, 3.3 V | operating pressure up to 7 bar |
Pressure Drop and Accuracy by Flow Band
A single accuracy figure hides the band where the error is largest. The table below is the short version for the three flow classes an analyser usually needs, taken from the published accuracy bands of the families above.
| Flow class | Typical use | Accuracy band to read | Pressure drop to check |
|---|---|---|---|
| Low flow, 0 to 5 slm | gas mixing, anaesthesia agent delivery | zero point accuracy dominates, read the offset figure | under 2,500 Pa on a mixing sensor |
| Mid flow, 5 to 60 slm | spontaneous breathing, sidestream sampling | 1.5 to 2 percent of reading | under 100 Pa on the low drop family |
| High flow, 60 to 300 slm | inspiratory limb, high frequency ventilation | 2 to 4 percent of reading, band by band | 500 to 750 Pa at 200 slm |
| Bidirectional flow | expiratory limb with a leak | read the negative flow band separately | matters more on expiration than on inspiration |
| Gases other than air | oxygen, carbon dioxide, nitrous oxide | only the calibrated gas list applies | density changes the drop, so check per gas |
Key Features
Standard litres per minute – the reading is already a mass flow, so the host does not correct for temperature and pressure.
Accuracy given per band – the error at 5 slm and at 250 slm are listed separately instead of one figure covering the whole range.
Pressure drop stated – the drop at 60 slm and at 200 slm is written down, because the blower and the patient both see it.
Calibrated gas list – air, oxygen, carbon dioxide and nitrous oxide are listed per part, which is the line that decides where a sensor can sit.
Medical cone options – 22 mm and 15 mm cones mate with the breathing circuit without an adapter, and every adapter is a potential leak.
Heater on the expiratory part – the SFM3200 carries an integrated heater, which is what keeps condensation off the sensor in a warm humid expiratory limb.
Fast step response – under 5 ms on the mixing sensors, enough to resolve a breath waveform rather than an average.
Two pressure classes – the SFM3000 works from 0.7 to 1.3 bar absolute, the SFM4300 up to 7 bar, so a sensor behind a regulator has a version.
Applications
Three gas paths where the flow range, the gas list and the pressure class decide the sensor together.
Critical care ventilator – an inspiratory sensor at 300 slm and an expiratory sensor with a heater and a 22 mm cone, both read on the same I2C bus with different addresses.
Anaesthesia gas mixing – oxygen, carbon dioxide and nitrous oxide are blended at 20 to 50 slm, and only a sensor calibrated for those three gases can sit in that manifold.
Gas analyser sampling – a low flow sensor meters sample into a cell while a second one watches the bypass, so the two readings are taken at the same pressure.
A flow channel is a small analogue board, so the SUPERB PCBA and the SUPERB PCB keep the heater return and the I2C lines apart in the same hand held housing.
Sourcing and Cross-Reference
The first question on a flow sensor build is which gas the calibration was made for. A sensor calibrated in air reads high in oxygen and low in carbon dioxide, and the error is large enough to fail a ventilator test. The selection tables list the calibrated gases per family, so the short list is made from that line before the flow range is compared.
The second question is the interface and the leak path. A 22 mm cone mates with a breathing circuit, an O-ring face seals onto a manifold, and a push-in fitting takes a tube, and each of those is a different sealing problem. A sensor with the wrong interface costs an adapter, and an adapter is a place where a leak starts.
The third question is condensation. An expiratory limb carries warm saturated gas, and a cold sensor in that path collects water until the reading drifts. A heated sensor, or a sensor placed after a water trap, is the answer, and the heater is listed on the part rather than added later. See lead time and availability for a specific sensor and BOM shortage recovery where a build is held up by one part.
The SUPERB Assembly team seals the cone joint before the sensor is screwed down, because a leak at the cone reads as a flow error at low rates.
Q: Can you supply the flow sensor with the gas calibration already matched to the mixture in our breathing circuit?
A: Yes. Send the gas mixture, the flow range, the interface and the supply, and the SUPERB Tech team returns the sensors and the calibrated gases that fit.
Why Ventilator Builders Source Flow Sensors Through SUPERB
- Band accuracy: the error at low flow is listed separately, because a ventilator spends most of a breath below 20 slm.
- Pressure drop published: the drop at the working flow is written down, so the blower is not sized against a sensor with an unknown restriction.
- Gas list per part: air, oxygen, carbon dioxide and nitrous oxide are stated, which is the first filter in any ventilator verification review.
- Interface options: cones, O-ring faces and push-in fittings are quoted with the part, so no adapter is added after the fact.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Differential Pressure Sensors for Ventilators, Oxygen Sensor Cells for Ventilators, CO2 Sensor Modules for Medical and Building Use, Liquid Flow Sensors for Medical and Analytical Instruments, Peristaltic and Diaphragm Micro Pumps, BOM Shortage Recovery.