NDIR and Photoacoustic CO2 Sensor Modules for Medical and Building Use
Product Specifications
Product Overview
A CO2 sensor works by measuring absorption at 4.26 micrometre, where carbon dioxide has a strong band and almost nothing else in air does. A non dispersive infrared module puts the gas in a tube between a lamp and a filter detector and measures how much light is absorbed. A photoacoustic module puts the gas in a small closed cavity, pulses the light, and listens to the pressure wave with a microphone. The second method is far smaller and uses less power, which is why it now sits in wall sensors and in portable instruments.
The specification that matters most is not the accuracy at 1,000 ppm. It is the behaviour of the baseline. Both methods drift as the lamp ages and as the detector ages, and both correct that drift by assuming the sensor sees fresh outdoor air at some point, which is about 420 ppm. That assumption is written into the automatic self calibration routine. In a room, a duct or a greenhouse it holds well enough. In an incubator, a capnograph or a closed chamber it never holds, and the routine then walks the reading away from the truth.
The second specification is the range and the band inside it. A module specified for 400 to 2,000 ppm is not a 5,000 ppm module with a wider scale: the accuracy is a percentage of reading on top of a fixed offset, and above the specified range only the output range remains. For an incubator at 5 percent carbon dioxide, which is 50,000 ppm, a room air module is not the part. For an anaesthesia circuit the range needed is different again.
The third specification is the environment. A photoacoustic cavity is sensitive to a change in pressure and to a change in temperature, and the readings follow both. A module with an integrated humidity and temperature sensor can compensate for part of that, and the compensation is only as good as the humidity sensor. Condensation on the optical window is the other failure mode, and it reads as a high concentration because water also absorbs in the infrared.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Measurement methods | non dispersive infrared and photoacoustic | infrared uses a long cell, photoacoustic a small cavity |
| Output range | 0 to 40,000 ppm on the photoacoustic family, 0 to 40,000 ppm on the infrared family over I2C | the output range is wider than the specified range |
| Specified range | 400 to 2,000 ppm, 400 to 5,000 ppm or 400 to 10,000 ppm depending on part | accuracy is only claimed inside the specified range |
| Accuracy | from plus or minus 30 ppm plus 3 percent of reading to plus or minus 50 ppm plus 5 percent | the band is stated per part, not as one figure for the family |
| Repeatability | plus or minus 10 ppm | the figure that matters for a control loop |
| Response time | 60 s on the photoacoustic family, 20 s on the infrared family | a 63 percent step response, not a full settling time |
| Drift per year | plus or minus 5 ppm plus 0.5 percent of reading on the photoacoustic family | with the self calibration routine enabled |
| Supply voltage | 2.4 to 5.5 V on the photoacoustic family | 3.3 V is the usual rail |
| Current | 19 mA at one measurement every 2 s on the infrared family, under 0.4 mA average in a low power mode on the photoacoustic family | duty cycle sets the average, not the part |
| Interface | I2C on both families, UART and PWM on the infrared family | I2C is the interface a host board usually wants |
| Humidity and temperature | integrated on both families, plus or minus 6 percent RH and 0.1 C repeatability | the compensation is only as good as the humidity sensor |
| Pressure dependence | reading follows absolute pressure, about 0.1 percent per mbar | a module behind a regulator needs a pressure input or a vent |
| Self calibration | automatic baseline correction assuming fresh air at about 420 ppm, or a forced recalibration at a known value | automatic correction is invalid in a closed chamber |
| Operating temperature | 0 to 50 C on the infrared family, wider on the photoacoustic parts | condensation on the window reads as high CO2 |
Included Models and Order Codes
Four modules cover the range from a room monitor to an incubator. The SCD40 is the base part at 400 to 2,000 ppm with an accuracy of plus or minus 50 ppm plus 5 percent, which suits a room or a duct. The SCD41 extends the specified range to 5,000 ppm and tightens the band in three steps, with a single shot mode that lets a battery instrument take one reading and sleep. The SCD43 has the tightest band of the three at plus or minus 30 ppm plus 3 percent over 400 to 5,000 ppm. The SCD30 is the dual channel infrared part at 400 to 10,000 ppm with a 20 s response and dual channel compensation, and it is the one to reach for when the response time matters more than the size.
| Module | Method | Specified range | Accuracy | Interface | Note |
|---|---|---|---|---|---|
| SCD40-D-R2 | photoacoustic | 400 to 2,000 ppm | plus or minus 50 ppm plus 5 percent of reading | I2C, 2.4 to 5.5 V | base accuracy, room and duct monitoring |
| SCD41-D-R2 | photoacoustic | 400 to 5,000 ppm | plus or minus 40 ppm plus 5 percent at the top band | I2C, 2.4 to 5.5 V | single shot mode for battery instruments |
| SCD43-D-R2 | photoacoustic | 400 to 5,000 ppm | plus or minus 30 ppm plus 3 percent of reading | I2C, 2.4 to 5.5 V | tightest band of the three, building standard work |
| SCD30 | dual channel infrared | 400 to 10,000 ppm | plus or minus 30 ppm plus 3 percent of reading | I2C, UART, PWM | 20 s response, humidity and temperature integrated |
Where the Automatic Baseline Correction Stops Working
Every module in the table corrects its own drift by assuming that it sees fresh air at about 420 ppm at some point in its life. That assumption is true in a room and false in a closed instrument, so the setting has to be chosen with the application. The table is the short version.
| Application | Fresh air exposure | Baseline setting | What goes wrong otherwise |
|---|---|---|---|
| Room or duct monitor | daily, through ventilation | automatic baseline correction on | nothing, the assumption holds |
| Greenhouse or livestock house | daily, through venting | automatic correction on a longer interval | a sealed night cycle walks the zero |
| Incubator | never, the chamber is sealed | automatic correction off, forced recalibration at a known gas | the reading drifts toward the assumed baseline |
| Capnography or breathing gas | never, the cell sees exhaled gas | automatic correction off, two point check with calibration gas | the waveform offset moves and the end tidal value is wrong |
| Portable battery instrument | irregular | single shot readings with a forced recalibration when serviced | continuous correction never completes |
Key Features
Photoacoustic and infrared side by side – a small cavity and a long cell are quoted together, so size, response and range can be traded off in one place.
Accuracy per band – the figure is stated over the specified range and in bands, not as one number covering the output range.
Response time stated – 60 s on the photoacoustic family and 20 s on the infrared family, which decides whether a waveform can be followed.
Baseline behaviour written down – the automatic correction and its assumption are described, because that is what fails in a sealed chamber.
Humidity and temperature integrated – the compensation comes from a sensor on the same package, so the two readings are taken at the same moment.
Low power modes – under 0.4 mA average is possible with a low duty cycle, which is what a battery instrument needs.
Single shot mode – the SCD41 and SCD43 can take one reading and return to sleep, instead of running a continuous measurement.
Supply range to 5.5 V – a 3.3 V host and a 5 V host can both drive the part without a level shift.
Applications
Three chambers where the specified range and the baseline setting are decided by the application rather than by the module.
Cell incubator – a 5 percent carbon dioxide chamber sits near 50,000 ppm, so a module specified to 5,000 ppm is used with a dilution path or a dedicated high range part, and the automatic correction is switched off.
Capnography and breathing gas – the infrared part with the 20 s response is the one that can follow a breath, and the baseline is set by a calibration gas rather than by fresh air.
Room and duct monitoring – a photoacoustic part at 2,000 ppm with automatic correction runs for years on a 3.3 V rail, which is what a wall sensor needs.
A CO2 channel is a lamp driver, a detector and a temperature sensor, so the SUPERB PCBA and the SUPERB PCB keep the pulse driver away from the microphone input on the same board.
Sourcing and Cross-Reference
The first question on a CO2 build is the range against the chamber. An incubator, a breathing circuit and a room are three different ranges, and a module quoted for the room will saturate in the other two. The second question is whether the chamber ever sees fresh air, because that single fact decides whether the automatic baseline routine can be left on.
The second question is the pressure and the humidity at the sensing point. A module downstream of a regulator reads the pressure it sees, not the pressure in the chamber, and a module in a warm wet gas stream collects condensation on the window. A pressure tap and a hydrophobic filter are the usual fixes, and both have to be planned before the enclosure is closed.
The third question is the calibration route. A forced recalibration at a known gas needs a gas supply and a fixture, and a two point check needs a second gas. Where neither is available, the automatic routine is the only correction and the installation has to guarantee fresh air. See lead time and availability for a specific module and BOM shortage recovery where a build is held up by one part.
The SUPERB Assembly team fits a hydrophobic filter over the gas inlet, because a droplet on the window reads as a concentration step.
Q: Can you supply the CO2 module with the baseline mode set for a sealed chamber, so it does not walk its own zero?
A: Yes. Send the chamber, the range, the gas mixture and the supply, and the SUPERB Tech team returns the modules and the baseline settings that fit.
Why Instrument Builders Source CO2 Modules Through SUPERB
- Range against the chamber: the 2,000, 5,000 and 10,000 ppm parts are quoted separately, because an incubator and a room need different ones.
- Baseline explained: the automatic correction and the assumption behind it are written down, which is the first thing a sealed chamber has to change.
- Response time per part: 20 s and 60 s are listed, so a waveform application is not quoted a part that averages a breath away.
- Pressure and humidity noted: both shift the reading, and both are named next to the specification rather than left to the field.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Gas Mass Flow Sensors for Ventilators and Analysers, Oxygen Sensor Cells for Ventilators, Gas Sensors for Air Quality and Leak Detection, Particulate Matter Sensor Modules, Infrared Thermopile Temperature Sensors, BOM Shortage Recovery.