Peristaltic Pump Tubing and Pump Heads for Analysers
Product Specifications
Product Overview
A peristaltic pump moves liquid by squeezing a tube, so the tube is both the wetted path and the measuring chamber. That is why a tube change can shift a calibration: the same nominal bore from two makers can differ in wall thickness by a tenth of a millimetre, and a thicker wall in the same head reduces the occlusion and the volume per revolution. On a reagent line that dispenses 50 microlitres per cycle, that difference shows up as a bias long before it shows up as a leak.
Bore is the first number to fix. Volume per revolution rises with the square of the bore, so a step from 1.6 to 3.2 mm is not twice the flow but roughly four times, and a step from 3.2 to 4.8 mm is a little over twice. The published figures for one family run 0.004 to 9.0 mL/min at 0.5 mm across 0.1 to 220 rpm, 0.04 to 92 mL/min at 1.6 mm, 0.17 to 370 mL/min at 3.2 mm, and 0.38 to 830 mL/min at 4.8 mm. Those ranges are the ones to use when a pump is selected, because a tube chosen on bore alone can be correct at one speed and outside the head rating at another.
Wall thickness is the second number. A 1.6 mm wall is the common rating for laboratory and analyser heads, and the same bore in a 1.0 mm wall is used on panel mount heads where the tube is held in a fixed channel. The two are not interchangeable: a thin wall tube in a head cut for a thick wall will not occlude, and a thick wall tube in a narrow channel will creep and shorten its life. When a tube is ordered as a replacement, the wall and the head gap are confirmed together, not just the bore.
Material closes the choice. Thermoplastic elastomer tube such as Bioprene gives a long squeeze life and low permeability and runs 5 to 80 C. Platinum-cured silicone gives the smoothest bore and the most stable flow, and runs -20 to 80 C, which is why it is the default for reagent and sample lines that are cleaned in place. Marprene is the choice for oxidising reagents and carries 7 bar (100 psi). Reinforced silicone such as STA-PURE PCS holds 100 psi and stays flexible to -20 C. Neoprene is the low cost option for water and mild chemistry. For a line that runs 24 hours a day, tube life is the cost driver, and the material choice matters more than the pump speed.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Bore range | 0.5 to 9.6 mm, which is 0.02 to 0.38 in | 0.5, 0.8, 1.6, 3.2, 4.8, 6.4 and 8.0 mm are the stocked steps |
| Wall thickness | 1.6 mm on the 913, 933, 941, 960, 965, 902 and 920 families; 1.0 mm on the panel mount family | wall and head gap are confirmed together |
| Flow at 220 rpm, 0.5 mm bore | 0.004 to 9.0 mL/min across 0.1 to 220 rpm | maker rating for Marprene, Bioprene and PureWeld XL |
| Flow at 220 rpm, 1.6 mm bore | 0.04 to 92 mL/min across 0.1 to 220 rpm | about ten times the 0.5 mm bore at the same speed |
| Flow at 220 rpm, 3.2 mm bore | 0.17 to 370 mL/min across 0.1 to 220 rpm | the common step for reagent and sample lines |
| Flow at 220 rpm, 4.8 mm bore | 0.38 to 830 mL/min across 0.1 to 220 rpm | used where a transfer step has to finish in seconds |
| Volume per revolution, panel tubing | 0.085 mL/rev at 1.14 mm bore, 0.34 mL/rev at 2.79 mm bore | colour coded, ten bores from 1.14 to 2.79 mm |
| Pressure, Marprene | 7 bar, which is 100 psi | the highest rating in the thermoplastic group |
| Pressure, STA-PURE PCS | 100 psi | reinforced silicone, flexible to -20 C |
| Temperature, Bioprene and Marprene | 5 to 80 C | 23 to 176 F |
| Temperature, Pumpsil and STA-PURE | -20 to 80 C | -4 to 176 F |
| Materials | Bioprene, Pumpsil platinum-cured silicone, PureWeld XL, STA-PURE PCS, STA-PURE PFL, Marprene, Neoprene | six families cover the analyser range |
| Head families | 114 head with seven tube sizes to 300 mL/min, 300 series heads, 520R head for 0.5 to 9.6 mm | the 520R covers the full bore range of the 500 series |
| Tube format | continuous coils and tube elements | elements are cut to the head length and fitted with the same wall |
Included Models and Order Codes
Part numbers below are the maker codes for a 1.6 mm wall tube, which is the standard element for a laboratory or analyser head. The suffix 016 is the wall in millimetres, so the same family in another wall carries a different suffix and is not a drop-in replacement. Flow is quoted at the top speed of the 500 series, 220 rpm, for the bore in that row.
| Part | Material | Bore and wall | Flow at 220 rpm | Temperature | Note |
|---|---|---|---|---|---|
| 913.A005.016 | Pumpsil, platinum-cured silicone | 0.5 mm, 1.6 mm wall | 9.0 mL/min | -20 to 80 C | smoothest bore, lowest flow in the family |
| 913.A008.016 | Pumpsil | 0.8 mm, 1.6 mm wall | not stated for this bore | -20 to 80 C | between the 0.5 and 1.6 mm steps |
| 913.A016.016 | Pumpsil | 1.6 mm, 1.6 mm wall | 92 mL/min | -20 to 80 C | the default reagent line element |
| 913.A032.016 | Pumpsil | 3.2 mm, 1.6 mm wall | 370 mL/min | -20 to 80 C | sample and wash lines |
| 913.A048.016 | Pumpsil | 4.8 mm, 1.6 mm wall | 830 mL/min | -20 to 80 C | transfer and drain steps |
| 913.A064.016 | Pumpsil | 6.4 mm, 1.6 mm wall | not stated for this bore | -20 to 80 C | larger head, lower speed |
| 913.A080.016 | Pumpsil | 8.0 mm, 1.6 mm wall | not stated for this bore | -20 to 80 C | highest bore in the 500 series |
| 933.0005.016 | Bioprene, thermoplastic elastomer | 0.5 mm, 1.6 mm wall | 9.0 mL/min | 5 to 80 C | long squeeze life, low permeability |
| 933.0032.016 | Bioprene | 3.2 mm, 1.6 mm wall | 370 mL/min | 5 to 80 C | same flow as Pumpsil at this bore |
| 941.0016.016 | PureWeld XL | 1.6 mm, 1.6 mm wall | 92 mL/min | as Pumpsil | welded flow path for disposable assemblies |
| 960.0016.016 | STA-PURE PCS, reinforced silicone | 1.6 mm, 1.6 mm wall | 92 mL/min | -20 to 80 C | 100 psi rating |
| 965.0032.016 | STA-PURE PFL | 3.2 mm, 1.6 mm wall | 370 mL/min | -20 to 80 C | reinforced, for higher pressure lines |
| 902.0016.016 | Marprene, thermoplastic elastomer | 1.6 mm, 1.6 mm wall | 92 mL/min | 5 to 80 C | 7 bar rating, oxidising reagents |
| 920.0032.016 | Neoprene | 3.2 mm, 1.6 mm wall | 370 mL/min | 5 to 80 C | water and mild chemistry |
| 049.EF6D.N05 | Bioprene, panel mount | 0.5 mm bore, 1.0 mm wall | not stated for this family | 5 to 80 C | fixed channel head, not for a 520R |
| 049.EH6D.N10 | Silicone, panel mount | 1.0 mm bore, 1.0 mm wall | not stated for this family | -20 to 80 C | the same channel as the 0.5 mm tube |
| 049.EF6D.N20 | Bioprene, panel mount | 2.0 mm bore, 1.0 mm wall | not stated for this family | 5 to 80 C | between the 1.0 and 3.0 mm steps |
| 049.EH6D.N30 | Silicone, panel mount | 3.0 mm bore, 1.0 mm wall | not stated for this family | -20 to 80 C | largest bore in this family |
Colour Coded Panel Tubing, Volume per Revolution
Panel mount tubing is matched to a fixed channel rather than to a roller head, and it is identified by colour rather than by bore. The table gives the bore, the volume per revolution and the flow at four common speeds, which is the form used when a dispense volume is set in software.
| Colour code | Bore | mL/rev | 12 rpm | 25 rpm | 60 rpm | 100 rpm |
|---|---|---|---|---|---|---|
| Red/Red | 1.14 mm | 0.085 | 1.0 mL/min | 2.1 mL/min | 5.1 mL/min | 8.5 mL/min |
| Grey/Grey | 1.30 mm | 0.11 | 1.3 mL/min | 2.8 mL/min | 6.6 mL/min | 11 mL/min |
| Yellow/Yellow | 1.42 mm | 0.13 | 1.6 mL/min | 3.3 mL/min | 7.8 mL/min | 13 mL/min |
| Yellow/Blue | 1.52 mm | 0.14 | 1.7 mL/min | 3.5 mL/min | 8.4 mL/min | 14 mL/min |
| Blue/Blue | 1.65 mm | 0.17 | 2.0 mL/min | 4.3 mL/min | 10.2 mL/min | 17 mL/min |
| Green/Green | 1.85 mm | 0.20 | 2.4 mL/min | 5.0 mL/min | 12.0 mL/min | 20 mL/min |
| Purple/Purple | 2.06 mm | 0.24 | 2.9 mL/min | 6.0 mL/min | 14.4 mL/min | 24 mL/min |
| Purple/Black | 2.29 mm | 0.27 | 3.2 mL/min | 6.8 mL/min | 16.2 mL/min | 27 mL/min |
| Purple/Orange | 2.54 mm | 0.31 | 3.7 mL/min | 7.8 mL/min | 18.6 mL/min | 31 mL/min |
| Purple/White | 2.79 mm | 0.34 | 4.1 mL/min | 8.5 mL/min | 20.4 mL/min | 34 mL/min |
Key Features
Bore range 0.5 to 9.6 mm – eight stocked steps cover a 1 microlitre per revolution line and a 9 litre per hour transfer line from the same head family
Wall thickness stated with the bore – the 016 suffix is 1.6 mm and the panel family is 1.0 mm, so a replacement is not ordered on bore alone
Six materials on one list – Bioprene, Pumpsil, PureWeld XL, two STA-PURE grades, Marprene and Neoprene, with the temperature and pressure of each
Platinum-cured silicone bore – the smooth bore gives a stable volume per revolution, which is what a reagent dispense is calibrated against
Marprene to 7 bar – 100 psi covers a pressurised reagent line without a reinforced tube
Reinforced silicone to 100 psi – STA-PURE PCS stays flexible to -20 C, for a line that is chilled or cleaned in place
Colour coded panel tubing – ten bores from 1.14 to 2.79 mm with volume per revolution, so a dispense volume is set without a calibration run
Head matching in the table – the 114 head to 300 mL/min, the 300 series heads and the 520R head for the full 0.5 to 9.6 mm range are named next to the tube
Applications
Peristaltic tubing appears wherever a liquid has to be moved without touching the pump. The three cases below set the material and the wall.
Clinical chemistry reagent lines – a reagent pack is pierced once and dispensed for a week. Pumpsil in 1.6 mm bore at a 1.6 mm wall gives the stable volume per revolution that a 50 microlitre dispense needs, and the smooth bore resists the protein film that builds on a rougher tube.
Water and process analysers – sample is drawn from a sample pot that carries grit and algae. A 3.2 mm bore in Marprene carries the flow at 7 bar without a reinforced wall, and the tube is changed on a schedule rather than on a failure, because a worn tube changes the sample volume before it splits.
Benchtop and panel mount instruments – a compact analyser uses a fixed channel head and colour coded tubing. The 1.14 mm red tube at 0.085 mL/rev allows a 100 microlitre dispense in about 1.2 revolutions, which is the smallest dispense that stays repeatable without a syringe pump.
The analyser boards that drive these pumps are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the motor drive to the fluidic control stage.
Sourcing and Cross-Reference
Tubing is bought by bore, wall and material, and the wall is the field that gets dropped. A 3.2 mm bore in a 1.6 mm wall and the same bore in a 1.0 mm wall carry different part numbers and are not interchangeable, so the head and the channel are confirmed before a part is named. Where the old tube is still on the pump, the wall is measured rather than read from a catalogue.
Where a maker has retired a tube grade, the substitute is chosen on pressure and temperature first and on bore second. A silicone tube can replace a thermoplastic one on a reagent line, but it will not hold the same pressure at 80 C, and a line that runs near the pump limit will show the difference as a slow delivery rather than as a leak.
Tube life is the hidden cost. A line that runs 24 hours a day can consume a tube a month, and the decision between a low cost Neoprene tube and a longer life Bioprene tube is a cost per litre question. We quote both and let the maintenance interval decide, rather than defaulting to the cheapest part.
When a line is rebuilt, the tube, the head and the fittings are picked as one set, which is how SUPERB Assembly ships a pump overhaul.
Q: Can you supply the same bore in a different wall thickness when the original tube is no longer listed?
A: SUPERB Tech can confirm the bore, the wall and the head gap against the old tube before the order is placed.
Why Buy Peristaltic Tubing from SUPERB
- Wall thickness confirmed, not assumed: the head gap and the wall are checked together, because the same bore in two walls gives two different dispense volumes
- Flow per revolution on the panel family: a dispense volume is set from a published figure rather than from a trial and error calibration
- Six materials with pressure and temperature: the choice between silicone, thermoplastic and reinforced tube is made on the line conditions, not on availability
- Head families named in the table: the tube and the head are quoted as a pair, which is what avoids a tube that fits the pump but not the channel
- Maker code on the delivery file: the part number and the wall are both written down, so the next replacement is the same part
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Peristaltic and Diaphragm Micro Pumps for Analysers, Inline Degassing Modules and Vacuum Degassers for Liquid Lines, Quick Disconnect Couplings and Luer Fittings for Reagent Lines, Inline Filters and Frits for Reagent Lines, Gas Mass Flow Sensors for Ventilators and Analysers, Component Lead Time Solutions.