Bandpass and Interference Filters for Photometric Channels
Product Specifications
Product Overview
A bandpass filter does two jobs at once: it passes the light the assay uses and it rejects everything else. The passband is described by its centre and its width, and the width appears twice in a data sheet with two different meanings. The guaranteed minimum bandwidth is the span over which the average transmission exceeds the quoted figure, and it is the number printed in the part number. The full width at half maximum is usually a little wider, by about 1 percent of the centre wavelength.
Blocking is the half of the specification that gets read last and matters most. A filter that passes 93 percent at 470 nm but leaks at 700 nm will show a background in a channel that is supposed to see only blue, and the leak is invisible in a transmission curve printed on a linear scale. That is why the blocking levels are quoted in optical density: OD 4 means one part in ten thousand passes, OD 6 one in a million, and the bands over which each figure holds are listed separately.
Angle is the third variable. A thin film filter shifts to shorter wavelengths as the incidence angle grows, so a part quoted at 0 degrees plus or minus 5 degrees with a 7 degree cone half angle will behave differently in a converging beam. An instrument that puts the filter in front of a lens sees a spread of angles, and the effective passband is wider and slightly blue shifted compared with the collimated case.
Hard coated filters are built as a single substrate with the coating deposited on it, rather than as two glass plates cemented together with a dyed layer. The difference shows up in two places: the transmission is higher because there is no cement absorption, and the filter survives cleaning and high light levels without delaminating. The laser damage note on the data sheet is a statement about that construction, and it is quoted for a specific lamp and spot size rather than as a general promise.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Centre wavelength range | 340 to 1055 nm on the parts listed | ultraviolet through near infrared |
| Guaranteed bandwidth | 6 to 90 nm | the second number in the part number, over which the average transmission is guaranteed |
| FWHM | about the guaranteed bandwidth plus 1 percent of the centre wavelength | a 470/22 part gives about 27 nm at half maximum |
| Average transmission | 75 percent at 340 nm, 90 to 95 percent from 390 nm upward | an average over the stated band, not a peak figure |
| Blocking | OD 2 to OD 8 depending on the band | OD 4 passes one part in ten thousand, OD 6 one in a million |
| Angle of incidence | 0 degrees plus or minus 5 degrees | a tilted filter shifts to shorter wavelengths |
| Cone half angle | 7 degrees | a converging beam widens the effective passband |
| Housing | 25 mm diameter black anodised aluminium ring | the industry standard holder diameter |
| Clear aperture | at least 21 mm | the coated area is smaller than the ring |
| Thickness | 3.5 mm or 5.0 mm housed, with 2.0 to 3.5 mm of glass inside | the holder depth has to match the thickness |
| Substrate construction | single substrate hard coating | no cement layer, so no absorption and no delamination |
| Damage note | no degradation after 6.0 W from an unfiltered xenon arc over a 25 mm spot for 500 hours | that is 1.2 W per square centimetre |
| Surface quality | not quoted as a scratch number, the transmission curve is the specification | a scratched filter shows as a drop in transmission |
Included Models and Order Codes
The table follows the wavelengths that photometric and fluorescence channels actually use, from a 340 nm ultraviolet channel through the visible bands to a 1055 nm near infrared part. The narrow parts at 6, 11 and 13 nm are for separating two dyes that sit close together, and the wide parts at 64 and 90 nm are for collecting as much light as possible from a broad emitter.
| Part | Centre | Guaranteed band | FWHM | Transmission | Housing | Note |
|---|---|---|---|---|---|---|
| FF01-340/12-25 | 340 nm | 12 nm | 17.9 nm | above 75 percent over 334 to 346 nm | 25 mm, 5.0 mm | ultraviolet channel, lowest transmission of the group |
| FF01-390/18-25 | 390 nm | 18 nm | about 22 nm | above 90 percent | 25 mm, 5.0 mm | violet channel |
| FF01-470/22-25 | 470 nm | 22 nm | about 27 nm | above 93 percent | 25 mm, 5.0 mm | blue channel, common fluorophore pair |
| FF01-530/11-25 | 530 nm | 11 nm | about 16 nm | above 90 percent | 25 mm, 5.0 mm | narrow green channel |
| FF01-590/20-25 | 590 nm | 20 nm | about 26 nm | above 93 percent | 25 mm, 5.0 mm | amber channel |
| FF01-625/90-25 | 625 nm | 90 nm | 97.2 nm | above 93 percent over 580 to 670 nm | 25 mm, 3.5 mm | wide red channel, blocking OD 5 from 275 to 567 nm |
| FF01-660/13-25 | 660 nm | 13 nm | about 20 nm | above 93 percent | 25 mm, 5.0 mm | far red channel |
| FF01-680/13-25 | 680 nm | 13 nm | about 20 nm | above 93 percent | 25 mm, 5.0 mm | far red, narrow |
| FF01-700/13-25 | 700 nm | 13 nm | about 20 nm | above 93 percent | 25 mm, 5.0 mm | edge of the visible range |
| FF01-792/64-25 | 792 nm | 64 nm | 71.5 nm | above 93 percent over 760 to 824 nm | 25 mm, 3.5 mm | blocking OD 4 to 8 in bands from 280 to 925 nm |
| FF01-1055/6-25 | 1055 nm | 6 nm | 10.2 nm | above 90 percent over 1052 to 1058 nm | 25 mm, 3.5 mm | narrow near infrared channel |
Reading a Filter Data Sheet
Four numbers decide whether a filter works in a channel, and three of them are easy to misread. The table is the short version.
| Number | What it means | Where it misleads | How to check it |
|---|---|---|---|
| Guaranteed bandwidth | the span over which the average transmission is at least the quoted figure | it is not the width where the filter is at half maximum | compare it with the FWHM on the same sheet |
| FWHM | the width at half the peak transmission | it is wider than the guaranteed band by about 1 percent of the centre | a 470/22 part is about 27 nm at half maximum |
| Average transmission | an average across the band, not a peak | a single high point at the centre hides a weak edge | look at the band limits as well as the average |
| Blocking in OD | OD 4 means one part in ten thousand, OD 6 one in a million | the OD figure only holds inside the bands listed | check the wavelengths just outside the passband |
| Centre wavelength at angle | the passband moves to shorter wavelengths as the angle grows | a filter specified at 0 degrees is used in a converging beam | keep the cone half angle at 7 degrees or less |
Key Features
Single substrate hard coating – the coating sits on one glass substrate with no cement layer, so there is nothing to absorb or delaminate
Average transmission above 93 percent – the figure is an average across the guaranteed band, which is what the detector actually integrates
Blocking stated in optical density – OD 4, 5, 6 and 8 are quoted band by band, so the leak path outside the passband is visible
Narrow and wide bands in one range – 6 nm parts separate close dyes, 64 and 90 nm parts collect light from broad emitters
Standard 25 mm housing – the black anodised ring drops into the holders that filter wheels and cubes already use
Clear aperture of 21 mm – the coated area is stated, so the beam is not clipped at the edge of the ring
Angle tolerance stated – 0 degrees plus or minus 5 degrees with a 7 degree cone half angle, so a converging beam can be assessed
Damage note for lamp sources – a measured lamp and spot size rather than a vague power limit, which is what a xenon source needs
Applications
These filters sit in the emission or excitation path of a photometric instrument. The three cases below decide the bandwidth and the blocking requirement.
Clinical chemistry absorbance channels – a narrow band at the absorption peak of a chromophore. A 10 to 20 nm band separates the analyte from the reagent blank, and the blocking figure matters because the lamp also emits where the chromophore does not absorb.
Fluorescence and immunoassay optics – an excitation filter and an emission filter with the passbands chosen so that they do not overlap. A wide emission band collects more light, and the blocking level is what keeps the excitation light out of the detector.
Near infrared and flow cytometry channels – a 792 or 1055 nm channel reads a dye or a scatter signal beyond the visible range. Narrow bands at these wavelengths need a hard coated filter, because a dyed glass filter does not hold a 6 nm band.
The photometer boards behind these filters are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the photodiode front end to the channel switching.
Sourcing and Cross-Reference
A filter is ordered by centre wavelength and bandwidth first, and by blocking second, because two parts with the same passband can differ by four orders of magnitude outside it. We ask for the lamp spectrum and the detector range before recommending a part, since a filter that is correct at the peak can still leak into a silicon detector.
The housing thickness is the detail that stops a filter going into an existing wheel. A 3.5 mm ring and a 5.0 mm ring do not interchange in every holder, and the clear aperture has to clear the beam at the same time.
Where an instrument uses a discontinued filter, the passband and the blocking band are the two things to match. A substitute with the same centre and a wider band changes the calibration, and a substitute with the same band and weaker blocking raises the background.
A filter set with its holder is shipped as one kit when an instrument is being serviced, which is how SUPERB Assembly handles it.
Q: Can you match a filter by centre wavelength, bandwidth and blocking when the original part is discontinued?
A: SUPERB Tech can compare a substitute filter on centre wavelength, bandwidth and blocking before it is quoted.
Why Buy Optical Filters from SUPERB
- Band and blocking quoted together: a passband without a blocking figure is half a specification, and both appear on every line
- Housing and aperture confirmed: 25 mm ring, thickness and clear aperture are checked against the holder before an order
- Angle behaviour stated: incidence angle and cone half angle are part of the data sheet, so a converging beam can be assessed
- Narrow bands from a hard coating: 6 and 11 nm parts are single substrate coatings, not cemented dyed glass
- Substitutes assessed on the full curve: a replacement is compared on band, transmission and blocking rather than on centre wavelength alone
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: UV and Visible LED Sources for Photometric Detection, Quartz Flow Cells and Cuvettes for Photometric Instruments, Turbidity and Photometric Detector Modules for Water Analysis, RF Detector Diodes and Zero Bias Schottky Detectors, Passive Components, Obsolete Part Cross Reference, Component Lead Time Solutions.