Polarizers and Waveplates for Photometric Instruments

Polarization is set by two parts: the polarizer that cleans the beam and the wave plate that turns it. Calcite polarizers with 100,000 to 1 extinction from 350 nm to 2.3 um, clear apertures 6.5 to 19 mm. Zero order quarter and half wave plates at 532 nm, and a wire grid part for 250 nm to 4 um.
quote now

Product Specifications

Product Overview

A photometric measurement that depends on polarization needs two parts, not one. The polarizer transmits a single linear state and rejects the rest, and a wave plate turns that state by a quarter or a half wave before it reaches the sample. Buying the polarizer alone gives a clean beam but no control over its orientation, and buying the plate alone gives a rotation of whatever polarization happened to arrive.

The polarizers here are calcite prisms. A Glan-Taylor prism is air spaced, which raises the damage threshold to a few joules per square centimetre at 532 nm, and passes 350 nm to 2.3 um with an extinction ratio of 100,000 to 1. A Glan-Thompson prism is cemented, which limits it to about 1 W per square centimetre continuous, but accepts a beam that converges or diverges over 14 to 16 degrees, which is the reason to choose it in front of a lamp rather than a laser.

Wave plates are the second half. A zero order plate is made of two pieces stacked with their axes crossed, so the retardance barely changes with wavelength or temperature. That is what makes a quarter wave plate at 532 nm usable at 530 or 534 nm, and it is why a multi order plate, which is cheaper, drifts enough to matter in a temperature controlled instrument.

Key Specifications

ParameterValueNotes
Extinction ratio100,000 to 1 on the calcite prismsSpecified for the transmitted beam only
Wavelength range350 nm to 2.3 um on calcite, 250 nm to 4 um on the wire grid partAR coating narrows the band further
Clear aperture6.5, 12, 17 and 19 mm across the familyThe beam has to fit inside it
Acceptance angle14 to 16 degrees full angle on the Glan-Thompson prismSuits converging and diverging beams
Damage threshold5 J per square cm at 532 nm, 10 ns, on the coated Glan-Taylor parts20 J per square cm uncoated at 1064 nm
Continuous power1 W per square cm on the cemented Glan-Thompson prismGlan-Taylor is air spaced and takes more
Wave plate retardanceQuarter wave and half waveZero order, two plates stacked
Retardance accuracyBetter than lambda by 300 on the quartz zero order platesHolds across the specified band
Waveplate clear aperture10 mm on the mounted half inch platesFits an SM1 rotation mount
Waveplate mountSM1 threaded, 1 inch outside diameterRotates in a standard mount with graduations
Wire grid polarizer10 to 1 extinction, 19 mm clear aperture, fused silicaBroadband, lower purity, higher power
Surface quality20-10 scratch dig on the calcite prisms80-50 on the wire grid part

Included Models and Order Codes

The calcite prisms come first, ordered by clear aperture, and the Newport prism follows with its own acceptance angle. The wave plates are next, at the 532 nm line where most photometric work sits, and the wire grid part is last because it covers the widest band at the lowest extinction ratio.

Part NumberTypeClear ApertureRangeExtinctionNotes
GT5Glan-Taylor prism6.5 mm350 nm to 2.3 um100,000 to 1Air spaced calcite, 12.7 mm across flats
GT10Glan-Taylor prism12 mm350 nm to 2.3 um100,000 to 1Most common size for a 10 mm beam
GT15Glan-Taylor prism17 mm350 nm to 2.3 um100,000 to 1Largest calcite prism in the family
10GT04Glan-Thompson prism10 mm350 to 2300 nm100,000 to 1Cemented, 14 to 16 degrees acceptance, 1 W per square cm
WPQ05M-532Quarter wave plate10 mm532 nmNot applicableZero order quartz, SM1 mount, better than lambda by 300
WPH05M-532Half wave plate10 mm532 nmNot applicableZero order quartz, SM1 mount
WP25M-UBWire grid polarizer19 mm250 nm to 4 um10 to 1Fused silica substrate, 1.53 mm thick, 80-50 surface

Key Features

Extinction stated for the transmitted beam – 100,000 to 1 on the prisms, with the escape beam described as unpolarised rather than quoted as a second output

Air spaced against cemented – the Glan-Taylor prism takes 5 J per square cm at 532 nm, the Glan-Thompson takes 1 W per square cm continuous

Acceptance angle given – 14 to 16 degrees on the cemented prism, which is what a converging lamp beam needs

Zero order wave plates – two stacked plates, so retardance holds across the specified band rather than at one wavelength

SM1 mounted plates – the 532 nm quarter and half wave plates drop into a standard rotation mount with graduations

Wire grid for broadband – 250 nm to 4 um at 10 to 1, where a calcite prism would be limited to 2.3 um

Applications

Polarizers and wave plates are bought as a pair when a measurement depends on polarization, because one cleans the beam and the other sets its angle.

Spectrophotometer polarization – a calcite prism before the sample and a wave plate after it, to separate polarized absorbance from scattering

Laser power control – a half wave plate followed by a prism, turning the plate to set the transmitted power without changing the beam profile

Ellipsometry and thin films – a quarter wave plate at 532 nm setting the phase before a sample, with the prism cleaning the source

Polarization kits that sit in an instrument are built alongside the SUPERB PCB and SUPERB PCBA boards that drive the measurement.

Sourcing and Cross-Reference

The aperture is chosen from the beam, and the beam is usually wider than expected. A 12 mm prism takes a 10 mm collimated beam with a millimetre of margin, and a 6.5 mm prism does not. Where a beam is converging, the Glan-Thompson prism is the safer choice because its 14 to 16 degree acceptance angle tolerates the cone that a Glan-Taylor prism would clip.

Power decides the construction. An air spaced Glan-Taylor prism is the right part for a pulsed laser at a few joules per square centimetre, while a cemented Glan-Thompson prism is limited to about 1 W per square centimetre continuous and will delaminate above it. The number is not a safety margin, it is the point at which the cement fails, so a 2 W beam needs the air spaced part.

For the wave plate, the wavelength has to be stated. A zero order quarter wave plate at 532 nm will still work at 530 nm, but a plate bought for 633 nm will not give a quarter wave at 532 nm at all. Where a method runs at two lines, two plates are cheaper than one plate and a correction that nobody applies.

Prisms and wave plates shipped with a mount and a case are consolidated through SUPERB Assembly into one box.

Q: Can you supply a 10 mm calcite polarizer and a quarter wave plate at 532 nm that fits the same rotation mount?

A: Yes. SUPERB Tech quotes the prism, the wave plate and the mount together, and states the beam diameter the clear aperture will take.

Why Choose SUPERB for Polarizers and Waveplates

  1. Extinction and aperture on the page: 100,000 to 1 with the clear aperture that goes with it, so a wide beam is not clipped
  2. Construction explained: air spaced for pulsed power, cemented for a wide acceptance angle, with the limit stated
  3. Wave plates at the working line: 532 nm quarter and half wave parts, zero order, mounted for an SM1 rotation stage
  4. One quote for the pair: prism, plate and mount quoted together, since the pair is what sets the polarization

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Dichroic Filters and Beam Splitters, Optical Windows and Sapphire Discs, Lenses and Mirrors for Optical Benches, Neutral Density Filters and Standards, Optical Filter Mounts and Lens Holders, Microscope Objectives and Tube Lenses.