Polarizers and Waveplates for Photometric Instruments
Product Specifications
Product Overview
A photometric measurement that depends on polarization needs two parts, not one. The polarizer transmits a single linear state and rejects the rest, and a wave plate turns that state by a quarter or a half wave before it reaches the sample. Buying the polarizer alone gives a clean beam but no control over its orientation, and buying the plate alone gives a rotation of whatever polarization happened to arrive.
The polarizers here are calcite prisms. A Glan-Taylor prism is air spaced, which raises the damage threshold to a few joules per square centimetre at 532 nm, and passes 350 nm to 2.3 um with an extinction ratio of 100,000 to 1. A Glan-Thompson prism is cemented, which limits it to about 1 W per square centimetre continuous, but accepts a beam that converges or diverges over 14 to 16 degrees, which is the reason to choose it in front of a lamp rather than a laser.
Wave plates are the second half. A zero order plate is made of two pieces stacked with their axes crossed, so the retardance barely changes with wavelength or temperature. That is what makes a quarter wave plate at 532 nm usable at 530 or 534 nm, and it is why a multi order plate, which is cheaper, drifts enough to matter in a temperature controlled instrument.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Extinction ratio | 100,000 to 1 on the calcite prisms | Specified for the transmitted beam only |
| Wavelength range | 350 nm to 2.3 um on calcite, 250 nm to 4 um on the wire grid part | AR coating narrows the band further |
| Clear aperture | 6.5, 12, 17 and 19 mm across the family | The beam has to fit inside it |
| Acceptance angle | 14 to 16 degrees full angle on the Glan-Thompson prism | Suits converging and diverging beams |
| Damage threshold | 5 J per square cm at 532 nm, 10 ns, on the coated Glan-Taylor parts | 20 J per square cm uncoated at 1064 nm |
| Continuous power | 1 W per square cm on the cemented Glan-Thompson prism | Glan-Taylor is air spaced and takes more |
| Wave plate retardance | Quarter wave and half wave | Zero order, two plates stacked |
| Retardance accuracy | Better than lambda by 300 on the quartz zero order plates | Holds across the specified band |
| Waveplate clear aperture | 10 mm on the mounted half inch plates | Fits an SM1 rotation mount |
| Waveplate mount | SM1 threaded, 1 inch outside diameter | Rotates in a standard mount with graduations |
| Wire grid polarizer | 10 to 1 extinction, 19 mm clear aperture, fused silica | Broadband, lower purity, higher power |
| Surface quality | 20-10 scratch dig on the calcite prisms | 80-50 on the wire grid part |
Included Models and Order Codes
The calcite prisms come first, ordered by clear aperture, and the Newport prism follows with its own acceptance angle. The wave plates are next, at the 532 nm line where most photometric work sits, and the wire grid part is last because it covers the widest band at the lowest extinction ratio.
| Part Number | Type | Clear Aperture | Range | Extinction | Notes |
|---|---|---|---|---|---|
| GT5 | Glan-Taylor prism | 6.5 mm | 350 nm to 2.3 um | 100,000 to 1 | Air spaced calcite, 12.7 mm across flats |
| GT10 | Glan-Taylor prism | 12 mm | 350 nm to 2.3 um | 100,000 to 1 | Most common size for a 10 mm beam |
| GT15 | Glan-Taylor prism | 17 mm | 350 nm to 2.3 um | 100,000 to 1 | Largest calcite prism in the family |
| 10GT04 | Glan-Thompson prism | 10 mm | 350 to 2300 nm | 100,000 to 1 | Cemented, 14 to 16 degrees acceptance, 1 W per square cm |
| WPQ05M-532 | Quarter wave plate | 10 mm | 532 nm | Not applicable | Zero order quartz, SM1 mount, better than lambda by 300 |
| WPH05M-532 | Half wave plate | 10 mm | 532 nm | Not applicable | Zero order quartz, SM1 mount |
| WP25M-UB | Wire grid polarizer | 19 mm | 250 nm to 4 um | 10 to 1 | Fused silica substrate, 1.53 mm thick, 80-50 surface |
Key Features
Extinction stated for the transmitted beam – 100,000 to 1 on the prisms, with the escape beam described as unpolarised rather than quoted as a second output
Air spaced against cemented – the Glan-Taylor prism takes 5 J per square cm at 532 nm, the Glan-Thompson takes 1 W per square cm continuous
Acceptance angle given – 14 to 16 degrees on the cemented prism, which is what a converging lamp beam needs
Zero order wave plates – two stacked plates, so retardance holds across the specified band rather than at one wavelength
SM1 mounted plates – the 532 nm quarter and half wave plates drop into a standard rotation mount with graduations
Wire grid for broadband – 250 nm to 4 um at 10 to 1, where a calcite prism would be limited to 2.3 um
Applications
Polarizers and wave plates are bought as a pair when a measurement depends on polarization, because one cleans the beam and the other sets its angle.
Spectrophotometer polarization – a calcite prism before the sample and a wave plate after it, to separate polarized absorbance from scattering
Laser power control – a half wave plate followed by a prism, turning the plate to set the transmitted power without changing the beam profile
Ellipsometry and thin films – a quarter wave plate at 532 nm setting the phase before a sample, with the prism cleaning the source
Polarization kits that sit in an instrument are built alongside the SUPERB PCB and SUPERB PCBA boards that drive the measurement.
Sourcing and Cross-Reference
The aperture is chosen from the beam, and the beam is usually wider than expected. A 12 mm prism takes a 10 mm collimated beam with a millimetre of margin, and a 6.5 mm prism does not. Where a beam is converging, the Glan-Thompson prism is the safer choice because its 14 to 16 degree acceptance angle tolerates the cone that a Glan-Taylor prism would clip.
Power decides the construction. An air spaced Glan-Taylor prism is the right part for a pulsed laser at a few joules per square centimetre, while a cemented Glan-Thompson prism is limited to about 1 W per square centimetre continuous and will delaminate above it. The number is not a safety margin, it is the point at which the cement fails, so a 2 W beam needs the air spaced part.
For the wave plate, the wavelength has to be stated. A zero order quarter wave plate at 532 nm will still work at 530 nm, but a plate bought for 633 nm will not give a quarter wave at 532 nm at all. Where a method runs at two lines, two plates are cheaper than one plate and a correction that nobody applies.
Prisms and wave plates shipped with a mount and a case are consolidated through SUPERB Assembly into one box.
Q: Can you supply a 10 mm calcite polarizer and a quarter wave plate at 532 nm that fits the same rotation mount?
A: Yes. SUPERB Tech quotes the prism, the wave plate and the mount together, and states the beam diameter the clear aperture will take.
Why Choose SUPERB for Polarizers and Waveplates
- Extinction and aperture on the page: 100,000 to 1 with the clear aperture that goes with it, so a wide beam is not clipped
- Construction explained: air spaced for pulsed power, cemented for a wide acceptance angle, with the limit stated
- Wave plates at the working line: 532 nm quarter and half wave parts, zero order, mounted for an SM1 rotation stage
- One quote for the pair: prism, plate and mount quoted together, since the pair is what sets the polarization
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Dichroic Filters and Beam Splitters, Optical Windows and Sapphire Discs, Lenses and Mirrors for Optical Benches, Neutral Density Filters and Standards, Optical Filter Mounts and Lens Holders, Microscope Objectives and Tube Lenses.