Medical Isolation Transformers for Diagnostic and Laboratory Equipment

A medical isolation transformer separates the mains from the instrument, and the figure that proves it did its job is the leakage current, not the isolation voltage. The parts below are toroidal transformers from 50 VA to 3,000 VA with dual primary and dual secondary windings, 5 kVrms isolation and leakage under 100 microampere. Each rating is listed with its outline, its height and its weight, because a 3,000 VA toroid has to fit a chassis before it can be ordered. The leakage limits from IEC 60601-1 are listed next to the parts, because the transformer is only one contributor to the figure the finished instrument has to meet.
quote now

Product Specifications

Product Overview

A medical isolation transformer does not make an instrument safer by magic. It puts a known, low capacitance barrier between the mains and the secondary, so the current that can flow through a patient or a touch current path is small enough to meet a limit. A toroidal construction helps because the windings are short and close, which keeps the leakage capacitance and the stray field low. That is why a 100 VA toroid can sit inside a monitor without a steel box around it.

The first figure is the isolation voltage, 5 kVrms between primary and secondary in this family, which is a test the part has to survive rather than a working voltage. The second is the leakage current, quoted under 100 microampere, which is the number that gets added to the leakage of the power supply and the leakage of the applied part circuit. The sum has to stay under the IEC 60601-1 limit for the equipment class, and for a type CF applied part that limit is 10 microampere in normal conditions.

The third figure is the creepage distance, 8.0 mm minimum here, which is the physical gap along the insulation surface between the windings. A transformer with a good isolation test and a short creepage will fail a humid environment, because a film of moisture along the surface bridges the gap. The fourth is the thermal class, B in this family, which sets how hot the winding may run and therefore how much of the rating is usable inside a closed enclosure.

The fifth point is the winding arrangement, because dual primary and dual secondary windings are what make one part cover 100, 120, 200, 220 and 240 V inputs. The two primary windings go in parallel for 100 or 120 V and in series for 200 to 240 V, and the two secondary windings give either two independent 120 V supplies or one 240 V supply. That flexibility is why the same part number appears in instruments sold in several markets, with only the link bars changed.

Key Specifications

ParameterValueNote
Power range50 VA to 3,000 VA in the same familya toroid in this range fits a desktop instrument chassis
Isolation5 kVrms primary to secondarya withstand test, not a working voltage
Leakage currentunder 100 microampereadded to the supply and applied part leakage before the 60601 limit is checked
Creepage distance8.0 mm minimuma short creepage fails in a humid enclosure even with a good isolation test
Insulation classclass Bsets the winding temperature rise and the usable rating in a closed box
Primary windingsdual 0-100-120 V, connected in parallel or series for 100, 120, 200, 220 and 240 Vone part covers several markets
Secondary windingsdual 0-120 V, series or parallel, or two independent 120 V suppliestwo rails or one 240 V rail from the same part
Frequency50 and 60 Hza 50 Hz part on 60 Hz runs cooler, not the other way round
Regulationplus or minus 3 percent on loadthe secondary voltage falls by this much from no load to full load
RecognitionUL 60601-1 recognised, categories XORU2 and XORU8, file E251176the file number is what a safety reviewer asks for
Thermal protectionself resetting thermal switch available as an optionmandatory in some enclosure classes
Magnetic shieldavailable as an optionfor a part that sits next to a sensitive amplifier
Mountingvarious methods, including a potted centre and a bracketthe mounting has to be chosen with the chassis cut out
Constructionopen style toroidal, RoHS and REACH compliantan open part needs a finger guard in the enclosure

Included Models and Order Codes

Eight ratings cover the range that a diagnostic or laboratory instrument normally needs. The 50 VA and 100 VA parts are the ones that fit a hand held or a bench top enclosure, the 200 to 500 VA range covers a monitor or an analyser with a display, and the 1,000 VA and above parts are for a rack instrument with a motor or a compressor. Each part is a dual primary, dual secondary toroid with 5 kVrms isolation and leakage under 100 microampere, and the outline, the height and the weight are listed because a 26 kg toroid is a mechanical problem before it is an electrical one.

RatingOutlineHeightWeightSecondaryNote
50 VA83 by 40 mm6.35 mm centre1.0 kg2 by 0-120 V, or 0-240 Vsmallest part, hand held and bench enclosures
100 VA101 by 43 mm6.35 mm centre1.3 kg2 by 0-120 V, or 0-240 Vsingle board instrument
200 VA119 by 58 mm6.35 mm centre2.5 kg2 by 0-120 V, or 0-240 Vmonitor with a display
300 VA130 by 65 mm10 mm centre3.0 kg2 by 0-120 V, or 0-240 Vanalyser with a pump
500 VA137 by 73 mm10 mm centre4.5 kg2 by 0-120 V, or 0-240 Vlargest part that still mounts on a bracket
1,000 VA173 by 85 mm12.7 mm centre8.5 kg2 by 0-120 V, or 0-240 Vrack instrument, two rails
2,000 VA220 by 110 mm12.7 mm centre16.2 kg2 by 0-120 V, or 0-240 Vinstrument with a compressor
3,000 VA270 by 125 mm15 mm centre26.8 kg2 by 0-120 V, or 0-240 Vlargest rating, needs a chassis base plate

Leakage Limits the Transformer Has to Fit Inside

The transformer is one contributor to the leakage current of the finished instrument, and the limit it has to fit is set by the equipment class and the applied part. The table is the IEC 60601-1 set that the quotation is normally checked against.

Leakage pathType BType BFType CF
Earth leakage, normal condition500 microampere500 microampere500 microampere
Earth leakage, single fault1 mA1 mA1 mA
Enclosure leakage, normal condition100 microampere100 microampere100 microampere
Enclosure leakage, single fault500 microampere500 microampere500 microampere
Patient leakage, normal condition100 microampere100 microampere10 microampere
Patient leakage, single fault500 microampere500 microampere50 microampere

Key Features

Leakage under 100 microampere – the figure that is added to the supply and applied part leakage, so the sum can be checked against the class limit.

5 kVrms isolation – a withstand test between primary and secondary, which is what a safety reviewer looks for on the part.

8.0 mm creepage – the surface gap is stated, because a humid enclosure will bridge a short one.

Dual primary and dual secondary – one part covers 100 to 240 V inputs and gives two 120 V rails or one 240 V rail.

Outline and weight per rating – a 3,000 VA toroid is 26.8 kg, and that is a chassis question before it is an electrical one.

Class B insulation – the winding rise is bounded, which is what a closed enclosure without a fan can use.

Thermal switch option – a self resetting switch is available, which some enclosure classes require.

Magnetic shield option – a toroid already has low stray field, and the shield option takes it further for a sensitive front end.

Applications

Three instruments where the rating and the leakage figure are decided by the applied part rather than by the load.

Patient monitor – a 200 VA part feeds the display and the front end, and the 100 microampere leakage leaves room under the type CF patient limit of 10 microampere for the isolated front end.

Laboratory analyser – a 500 VA part feeds pumps, a heater and the control board, and the dual secondary gives the analogue rail a separate winding from the motor rail.

Rack instrument with a compressor – a 2,000 VA or 3,000 VA part feeds the compressor and the electronics, and the weight drives the base plate of the chassis.

A mains inlet, a transformer and a control board share one chassis, so the SUPERB PCBA and the SUPERB PCB keep the primary side creepage clear of the low level analogue area.

Sourcing and Cross-Reference

The first question on a transformer build is the applied part class. A type B instrument has 100 microampere of patient leakage to spend and a type CF instrument has 10, and that difference decides how much of the budget the transformer may use. The second question is the enclosure, because a class B winding in a closed box without a fan cannot use the full rating continuously.

The second question is the input voltage set. A dual primary part covers 100 to 240 V with link bars, which is what a product sold in several markets needs, and the link bar arrangement has to be right before the part is energised. The third question is the mechanical fit, because the outline and the weight of a 1,000 VA part and a 3,000 VA part are different problems.

The third question is the recognition file. A safety reviewer asks for the UL file number and the category, and a part without recognition becomes a test item on the finished instrument. See lead time and availability for a specific rating and obsolete part cross reference where a family has been replaced.

The SUPERB Assembly team routes the primary wiring in its own duct and ties it away from the signal harness, because a mains lead lying across an input reads as noise.

Q: Can you supply the transformer with the link bars set for the input voltage of each market, and the UL file number on the certificate?

A: Yes. Send the load, the applied part class, the input voltages and the chassis space, and the SUPERB Tech team returns the ratings that fit.

Why Instrument Builders Source Isolation Transformers Through SUPERB

  1. Leakage figure per part: the under 100 microampere figure is stated, so the budget against the class limit can be worked out before the enclosure is closed.
  2. Outline and weight listed: the mechanical envelope of every rating is on the page, because a toroid is a chassis problem as well as an electrical one.
  3. Recognition quoted: the UL 60601-1 category and file number are given, which is what a safety submission needs.
  4. Options named: thermal switch and magnetic shield are listed as options rather than discovered at the specification stage.

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: IEC 60601 Power Supplies for Medical Diagnostic Equipment, DIN Rail Power Supplies for Control Cabinets, Water Quality Sensor Circuits for pH and Conductivity, Oxygen Sensor Cells for Ventilators, Current Transformers and AC Current Sensing, Component Lead Time Solutions.