Medical Isolation Transformers for Diagnostic and Laboratory Equipment
Product Specifications
Product Overview
A medical isolation transformer does not make an instrument safer by magic. It puts a known, low capacitance barrier between the mains and the secondary, so the current that can flow through a patient or a touch current path is small enough to meet a limit. A toroidal construction helps because the windings are short and close, which keeps the leakage capacitance and the stray field low. That is why a 100 VA toroid can sit inside a monitor without a steel box around it.
The first figure is the isolation voltage, 5 kVrms between primary and secondary in this family, which is a test the part has to survive rather than a working voltage. The second is the leakage current, quoted under 100 microampere, which is the number that gets added to the leakage of the power supply and the leakage of the applied part circuit. The sum has to stay under the IEC 60601-1 limit for the equipment class, and for a type CF applied part that limit is 10 microampere in normal conditions.
The third figure is the creepage distance, 8.0 mm minimum here, which is the physical gap along the insulation surface between the windings. A transformer with a good isolation test and a short creepage will fail a humid environment, because a film of moisture along the surface bridges the gap. The fourth is the thermal class, B in this family, which sets how hot the winding may run and therefore how much of the rating is usable inside a closed enclosure.
The fifth point is the winding arrangement, because dual primary and dual secondary windings are what make one part cover 100, 120, 200, 220 and 240 V inputs. The two primary windings go in parallel for 100 or 120 V and in series for 200 to 240 V, and the two secondary windings give either two independent 120 V supplies or one 240 V supply. That flexibility is why the same part number appears in instruments sold in several markets, with only the link bars changed.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Power range | 50 VA to 3,000 VA in the same family | a toroid in this range fits a desktop instrument chassis |
| Isolation | 5 kVrms primary to secondary | a withstand test, not a working voltage |
| Leakage current | under 100 microampere | added to the supply and applied part leakage before the 60601 limit is checked |
| Creepage distance | 8.0 mm minimum | a short creepage fails in a humid enclosure even with a good isolation test |
| Insulation class | class B | sets the winding temperature rise and the usable rating in a closed box |
| Primary windings | dual 0-100-120 V, connected in parallel or series for 100, 120, 200, 220 and 240 V | one part covers several markets |
| Secondary windings | dual 0-120 V, series or parallel, or two independent 120 V supplies | two rails or one 240 V rail from the same part |
| Frequency | 50 and 60 Hz | a 50 Hz part on 60 Hz runs cooler, not the other way round |
| Regulation | plus or minus 3 percent on load | the secondary voltage falls by this much from no load to full load |
| Recognition | UL 60601-1 recognised, categories XORU2 and XORU8, file E251176 | the file number is what a safety reviewer asks for |
| Thermal protection | self resetting thermal switch available as an option | mandatory in some enclosure classes |
| Magnetic shield | available as an option | for a part that sits next to a sensitive amplifier |
| Mounting | various methods, including a potted centre and a bracket | the mounting has to be chosen with the chassis cut out |
| Construction | open style toroidal, RoHS and REACH compliant | an open part needs a finger guard in the enclosure |
Included Models and Order Codes
Eight ratings cover the range that a diagnostic or laboratory instrument normally needs. The 50 VA and 100 VA parts are the ones that fit a hand held or a bench top enclosure, the 200 to 500 VA range covers a monitor or an analyser with a display, and the 1,000 VA and above parts are for a rack instrument with a motor or a compressor. Each part is a dual primary, dual secondary toroid with 5 kVrms isolation and leakage under 100 microampere, and the outline, the height and the weight are listed because a 26 kg toroid is a mechanical problem before it is an electrical one.
| Rating | Outline | Height | Weight | Secondary | Note |
|---|---|---|---|---|---|
| 50 VA | 83 by 40 mm | 6.35 mm centre | 1.0 kg | 2 by 0-120 V, or 0-240 V | smallest part, hand held and bench enclosures |
| 100 VA | 101 by 43 mm | 6.35 mm centre | 1.3 kg | 2 by 0-120 V, or 0-240 V | single board instrument |
| 200 VA | 119 by 58 mm | 6.35 mm centre | 2.5 kg | 2 by 0-120 V, or 0-240 V | monitor with a display |
| 300 VA | 130 by 65 mm | 10 mm centre | 3.0 kg | 2 by 0-120 V, or 0-240 V | analyser with a pump |
| 500 VA | 137 by 73 mm | 10 mm centre | 4.5 kg | 2 by 0-120 V, or 0-240 V | largest part that still mounts on a bracket |
| 1,000 VA | 173 by 85 mm | 12.7 mm centre | 8.5 kg | 2 by 0-120 V, or 0-240 V | rack instrument, two rails |
| 2,000 VA | 220 by 110 mm | 12.7 mm centre | 16.2 kg | 2 by 0-120 V, or 0-240 V | instrument with a compressor |
| 3,000 VA | 270 by 125 mm | 15 mm centre | 26.8 kg | 2 by 0-120 V, or 0-240 V | largest rating, needs a chassis base plate |
Leakage Limits the Transformer Has to Fit Inside
The transformer is one contributor to the leakage current of the finished instrument, and the limit it has to fit is set by the equipment class and the applied part. The table is the IEC 60601-1 set that the quotation is normally checked against.
| Leakage path | Type B | Type BF | Type CF |
|---|---|---|---|
| Earth leakage, normal condition | 500 microampere | 500 microampere | 500 microampere |
| Earth leakage, single fault | 1 mA | 1 mA | 1 mA |
| Enclosure leakage, normal condition | 100 microampere | 100 microampere | 100 microampere |
| Enclosure leakage, single fault | 500 microampere | 500 microampere | 500 microampere |
| Patient leakage, normal condition | 100 microampere | 100 microampere | 10 microampere |
| Patient leakage, single fault | 500 microampere | 500 microampere | 50 microampere |
Key Features
Leakage under 100 microampere – the figure that is added to the supply and applied part leakage, so the sum can be checked against the class limit.
5 kVrms isolation – a withstand test between primary and secondary, which is what a safety reviewer looks for on the part.
8.0 mm creepage – the surface gap is stated, because a humid enclosure will bridge a short one.
Dual primary and dual secondary – one part covers 100 to 240 V inputs and gives two 120 V rails or one 240 V rail.
Outline and weight per rating – a 3,000 VA toroid is 26.8 kg, and that is a chassis question before it is an electrical one.
Class B insulation – the winding rise is bounded, which is what a closed enclosure without a fan can use.
Thermal switch option – a self resetting switch is available, which some enclosure classes require.
Magnetic shield option – a toroid already has low stray field, and the shield option takes it further for a sensitive front end.
Applications
Three instruments where the rating and the leakage figure are decided by the applied part rather than by the load.
Patient monitor – a 200 VA part feeds the display and the front end, and the 100 microampere leakage leaves room under the type CF patient limit of 10 microampere for the isolated front end.
Laboratory analyser – a 500 VA part feeds pumps, a heater and the control board, and the dual secondary gives the analogue rail a separate winding from the motor rail.
Rack instrument with a compressor – a 2,000 VA or 3,000 VA part feeds the compressor and the electronics, and the weight drives the base plate of the chassis.
A mains inlet, a transformer and a control board share one chassis, so the SUPERB PCBA and the SUPERB PCB keep the primary side creepage clear of the low level analogue area.
Sourcing and Cross-Reference
The first question on a transformer build is the applied part class. A type B instrument has 100 microampere of patient leakage to spend and a type CF instrument has 10, and that difference decides how much of the budget the transformer may use. The second question is the enclosure, because a class B winding in a closed box without a fan cannot use the full rating continuously.
The second question is the input voltage set. A dual primary part covers 100 to 240 V with link bars, which is what a product sold in several markets needs, and the link bar arrangement has to be right before the part is energised. The third question is the mechanical fit, because the outline and the weight of a 1,000 VA part and a 3,000 VA part are different problems.
The third question is the recognition file. A safety reviewer asks for the UL file number and the category, and a part without recognition becomes a test item on the finished instrument. See lead time and availability for a specific rating and obsolete part cross reference where a family has been replaced.
The SUPERB Assembly team routes the primary wiring in its own duct and ties it away from the signal harness, because a mains lead lying across an input reads as noise.
Q: Can you supply the transformer with the link bars set for the input voltage of each market, and the UL file number on the certificate?
A: Yes. Send the load, the applied part class, the input voltages and the chassis space, and the SUPERB Tech team returns the ratings that fit.
Why Instrument Builders Source Isolation Transformers Through SUPERB
- Leakage figure per part: the under 100 microampere figure is stated, so the budget against the class limit can be worked out before the enclosure is closed.
- Outline and weight listed: the mechanical envelope of every rating is on the page, because a toroid is a chassis problem as well as an electrical one.
- Recognition quoted: the UL 60601-1 category and file number are given, which is what a safety submission needs.
- Options named: thermal switch and magnetic shield are listed as options rather than discovered at the specification stage.
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: IEC 60601 Power Supplies for Medical Diagnostic Equipment, DIN Rail Power Supplies for Control Cabinets, Water Quality Sensor Circuits for pH and Conductivity, Oxygen Sensor Cells for Ventilators, Current Transformers and AC Current Sensing, Component Lead Time Solutions.