HPLC Columns for Reversed Phase Method Development
Product Specifications
Product Overview
A reversed phase column is described by six numbers: chemistry, particle size, pore size, dimensions, pH range and pressure. The chemistry decides selectivity, the particle size decides efficiency and back pressure, the pore size decides which molecules can reach the surface, and the pH range decides whether the method can use the mobile phase it needs. Everything else is packaging.
The pH range is the first cut for most methods, and it is where the chemistries separate. A non endcapped C18 on a silica support with bulky side chains is stable from pH 0.8 to 8 and is the choice for acidic methods and trifluoroacetic acid gradients. A hybrid particle with a trifunctional C18 bond holds pH 1 to 12, which allows basic mobile phases and the different selectivity that comes with them.
Dimensions follow the method rather than the other way round. A 4.6 x 150 mm column at 5 um is the standard analytical format for a conventional HPLC, a 4.6 x 250 mm column gives more plates where the sample is complex, and a 2.1 x 50 mm column is used where the sample is limited or the detector needs a narrow peak. The same chemistry is available in each, so a method can be scaled without changing selectivity.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Formats | 4.6 x 150 mm, 4.6 x 250 mm, 2.1 x 50 mm | Analytical and narrow bore |
| Chemistries | C18 and C8, non endcapped silica and hybrid | Selectivity differs as well as pH range |
| Particle size | 5 um and 3.5 um | Smaller particles raise efficiency and pressure |
| Pore size | 80 A and 130 A | Small molecule and peptide work |
| pH range | 0.8 to 8 for the silica C18, 1 to 12 for the hybrid C18 | The method pH decides the chemistry |
| Pressure rating | 400 bar and 415 bar | HPLC, not UHPLC pressures |
| Carbon load | 10 percent for the silica C18, 18 percent for the hybrid C18 | Retention and loadability |
| Endcapping | No for the silica C18, yes for the hybrid C18 | Peak shape for basic analytes |
| USP code | L1 for the C18 phases | Reference for pharmacopoeial methods |
| Temperature limit | Up to 90 C at pH 1 to 6 | 40 C at pH 6 to 8 on the silica phase |
| Surface area | 185 m2/g on the hybrid C18 | Capacity and retention |
| Shipping solvent | Methanol and water, or acetonitrile and water | Flush before the method solvent |
Included Models and Order Codes
The Agilent rows are the StableBond family on an 80 A silica, which is the low pH workhorse. The Waters rows are the XBridge hybrid family at 130 A, which carries the wider pH range and a higher carbon load. The last row is a C8 in the same silica family, useful where a C18 holds too long or the method needs a different selectivity at the same pH.
| Part Number | Chemistry | Particle | Pore | Dimensions | pH Range | Pressure | Use |
|---|---|---|---|---|---|---|---|
| 883975-902 | SB-C18, non endcapped | 5 um | 80 A | 4.6 x 150 mm | 0.8 to 8 | 400 bar | Acidic methods, TFA gradients |
| 883975-906 | SB-C8, non endcapped | 5 um | 80 A | 4.6 x 150 mm | 0.8 to 8 | 400 bar | Less retention than C18 |
| 186003116 | BEH C18, endcapped | 5 um | 130 A | 4.6 x 150 mm | 1 to 12 | 415 bar | Wide pH method development |
| 186003117 | BEH C18, endcapped | 5 um | 130 A | 4.6 x 250 mm | 1 to 12 | 415 bar | Complex samples, more plates |
| 186003034 | BEH C18, endcapped | 3.5 um | 130 A | 4.6 x 150 mm | 1 to 12 | 415 bar | Faster run at the same length |
| 186003021 | BEH C18, endcapped | 3.5 um | 130 A | 2.1 x 50 mm | 1 to 12 | 415 bar | Limited sample, LC-MS |
| 186003055 | BEH C8, endcapped | 3.5 um | 130 A | 4.6 x 150 mm | 1 to 12 | 415 bar | Hydrophobic analytes |
Key Features
pH range as the first filter – 0.8 to 8 on the silica phase against 1 to 12 on the hybrid, stated before the price
Endcapping declared – no on the StableBond C18, yes on the hybrid, which is what basic peak shape turns on
Three formats of one chemistry – 150 mm, 250 mm and 2.1 x 50 mm, so a method can be scaled without changing selectivity
Particle and pore together – 5 and 3.5 um particles on 80 and 130 A pores, matched to the molecule size
USP code given – L1 for both C18 phases, which is the reference a pharmacopoeial method uses
Pressure rating stated – 400 and 415 bar, so a method is not pushed onto a column that cannot take it
Applications
A column is picked with the method, and the method names the pH, the sample and the detector.
Pharmaceutical assay and purity – a hybrid C18 at pH 1 to 12 for a method that needs a basic mobile phase, with the guard and the batch record kept together
Acidic compounds and peptides – a non endcapped silica C18 at low pH with trifluoroacetic acid, where the hybrid phase would give a different selectivity
LC-MS with limited sample – a 2.1 x 50 mm narrow bore column at 3.5 um, where the peak volume has to suit the source
Column and instrument consumables ship with the SUPERB PCB and SUPERB PCBA builds for the detector and pump control boards, so a method transfer has one supplier.
Sourcing and Cross-Reference
Column choice is usually fixed by a validated method, so the sourcing question is about equivalence rather than improvement. Two C18 columns with the same dimensions and particle size can differ in carbon load, endcapping and pore size, and any of those changes retention and selectivity. We quote the part number the method names, and where that number is discontinued we quote the maker replacement and state what changed.
The consumable that protects the column is the guard. A guard cartridge of the same chemistry catches the matrix and is replaced at a fraction of the column cost, and a method that skips it usually shows up as a rising back pressure and a splitting peak after a few hundred injections. Guard holders are specific to the maker, which is worth knowing before a column family is chosen.
Storage and flushing decide how long a column lasts. A silica C18 kept in a high aqueous mobile phase can lose retention as the C18 chains fold, and a hybrid phase kept in buffer can grow a microbial film. Flushing into the shipping solvent before storage, and never letting a column dry out, is most of the maintenance a column needs.
Column, guard and fittings are kitted through SUPERB Assembly and labelled with the chemistry and the batch.
Q: Can you supply the same C18 chemistry in 4.6 x 150 mm and 2.1 x 50 mm so we can scale the method down?
A: Yes. SUPERB Tech quotes the column by the part number the method names, and states the carbon load, endcapping and pH range of any replacement.
Why Choose SUPERB for HPLC Columns
- Method part number first: the column the method names is quoted, with any replacement stated as a change
- pH and endcapping visible: the two specifications that most often decide whether a method transfers
- Scalable formats: 150 mm, 250 mm and narrow bore in the same chemistry and particle size
- Guard and consumables together: holder, cartridge and fittings quoted with the column so the method is complete
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Autosampler Vials and Septa, Syringe Filters and Filter Paper, Column Ovens and Heater Blocks, Sample Loops and Injection Loops, Calibration Buffers and Conductivity Standards, Peristaltic Pump Tubing and Heads.