Cuvette Stirrers and Magnetic Stir Bars
Product Specifications
Product Overview
A spectrophotometer reads the average of whatever sits in the light path, so a sample that is not mixed reads as a moving number. The usual answer is a small stir bar in the bottom of the cell, driven by a magnet under the cell holder. The bar has to be small enough to turn inside a 10 mm cell, heavy enough not to be lifted by the beam stirring the liquid, and coated so that it does not react with the sample.
The awkward part is the drive. A conventional magnetic stirrer is too large to sit inside a sample compartment, so the cell stirrers in this family use a flat block, 5 mm thick, that takes the place of the cell in the holder. The cell sits on top of the block and the bar turns inside the cell. A ribbon cable runs up the non transmitting side of the cell to a controller outside the compartment, thin enough that the lid still closes.
The alternative is a cell with a recess in the base. A circular well holds the bar in a fixed position, which stops it wandering into the light path and lets a semi micro cell be stirred with a bar of 6 by 1.5 mm. The cells are supplied with a PTFE lid or stopper, and the bars are sold in packs because they are the part that gets lost.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Cell format | 10 mm path, 12.5 x 12.5 mm outside | Standard rectangular cell |
| Stirrer thickness | 5 mm | Sits under the cell in the holder |
| Mixing time | About 4 seconds for a standard 10 mm cell | About 5 seconds to 30 mm of liquid in a micro cell |
| Controller supply | 110 V or 220 to 240 V versions | Speed adjustable from the controller |
| Ribbon cable | 1 m | Thin enough for the compartment lid to close |
| Stir bar sizes | 5 x 2, 6 x 1.5 and 6 x 3 mm | 10 per pack |
| Stir bar coating | PTFE | Chemically inert, low friction |
| Hellma bar size | About 3 to 4 mm diameter, 6 to 7 mm long | For 109 and 119 series cells |
| Stirring cell volumes | 0.9, 1.5, 1.8, 3.5 and 4 ml | Micro, semi micro and macro |
| Cell inside width | 4 mm or 9.5 mm | Narrow cells need the 6 x 1.5 mm bar |
| Cell base | Circular recess for the bar, 5 mm base thickness | Keeps the bar out of the light path |
| Window material | Far UV quartz, 170 to 2700 nm | Standard for stirring cells |
Included Models and Order Codes
The first block is the stirring hardware: the drive block, the controller and the bars. The second block is the Hellma bars and stirring cells, whose article numbers carry the cell volume, and the third is the Starna stirring cell range where the suffix MS marks a recessed base.
| Part Number | Type | Size or Volume | Material | Notes |
|---|---|---|---|---|
| 9400 | Cell stirrer | 5 mm block, 10 mm cell | Electronic magnet block | 110 V controller, ribbon cable, one bar included |
| 9400 (220 V) | Cell stirrer | 5 mm block, 10 mm cell | Electronic magnet block | 220 to 240 V controller version |
| MSB-5X2 | Stir bar | 5 x 2 mm | PTFE coated | 10 per pack, for standard 10 mm cells |
| MSB-6X1.5 | Stir bar | 6 x 1.5 mm | PTFE coated | 10 per pack, for narrow micro cells |
| MSB-6X3 | Stir bar | 6 x 3 mm | PTFE coated | 10 per pack, stronger stirring |
| 332-300-VE10 | Stir bar | 3 to 4 mm x 6 to 7 mm | PTFE coated | 10 per pack, for 109 and 119 series cells |
| 109-000-10-40 | Stirring cell | 3,500 ul, 10 mm path | Far UV quartz | 45 x 12.5 x 12.5 mm, 9.5 mm inside width |
| 109-004-10-40 | Stirring cell | 1,500 ul, 10 mm path | Far UV quartz | Semi micro, 4 mm inside width |
| 119-10-40 | Stirring cell | 3,500 ul, 10 mm path | Far UV quartz | Open top version, 49.5 mm high |
| 119-004-10-40 | Stirring cell | 1,500 ul, 10 mm path | Far UV quartz | Semi micro, open top |
| 1-Q-10-MS | Stirring cell | 3.5 ml, 10 mm path | Far UV quartz | Standard rectangular, PTFE lid |
| 3-Q-10-MS | Stirring cell | 3.5 ml, 10 mm path | Far UV quartz | Fluorometer cell, four windows |
| 9-Q-10-MS | Stirring cell | 1.8 ml, 10 mm path | Far UV quartz | Semi micro with recessed base |
| 18-Q-10-MS | Stirring cell | 0.9 ml, 10 mm path | Far UV quartz | Micro cell, 6 x 1.5 mm bar |
Key Features
Drive block fits a standard holder – 5 mm thick, so the cell sits at the normal height and the compartment lid closes
Mixing stated in seconds – about four seconds for a 10 mm cell, not a stirrer speed in arbitrary units
Bars sized to the cell – 6 x 1.5 mm for a 4 mm inside width, 5 x 2 mm for a standard cell
Recessed base on stirring cells – the bar turns in a well, so it cannot drift into the light path
PTFE on every wetted part – bar coating and cell lid, for solvent and acid samples
Controller outside the compartment – speed set where it can be reached, with a 1 m ribbon cable to the block
Applications
Stirring is bought when a sample changes during the measurement, which is why the cell and the stirrer are usually ordered together.
Kinetic measurements – a stirred cell following an absorbance change over minutes, with the bar turning throughout
Dissolution and suspension – a 10 mm cell keeping a powder suspended while a spectrum is taken
Titration and flow work – a stirred micro cell mixing a reagent added through the cell cap
Photometric instruments that use these cells are built as SUPERB PCB and SUPERB PCBA hardware when the method moves to production.
Sourcing and Cross-Reference
The bar size follows the cell, not the stirrer. A 6 x 3 mm bar will not turn in a cell with a 4 mm inside width, and a 5 x 2 mm bar in a wide cell can be lifted off the base at high speed and end up in the light path. The inside width and the base recess are the two figures that decide this, and both are quoted with the cell.
The drive is the part that is easy to overlook. A flat block under the cell needs a cable route out of the compartment, and on some instruments the only gap is at the front edge. The ribbon cable in this family is 1 m long and thin enough to pass under a lid, but the route has to be checked before the instrument is closed up.
Stirring cells are also a way to reduce the sample volume. A semi micro cell with a 4 mm inside width holds 1.5 ml, and a micro cell holds 0.9 ml, both with a bar that still turns. Where a method is limited by sample volume rather than by absorbance, the cell is the part that changes, and the stirrer block stays the same.
Cell and stirrer sets for a method are packed through SUPERB Assembly, with the bars counted into the same box as the cell.
Q: Can you supply a cell stirrer for a 10 mm cuvette together with the 6 x 1.5 mm bars for a micro cell?
A: Yes. SUPERB Tech quotes the stirrer, the controller and the bars together, and states the cell inside width the bar is meant for.
Why Choose SUPERB for Cell Stirrers and Stir Bars
- Bar size matched to the cell: inside width and bar dimensions on the same page, so the bar turns in the cell it was bought for
- Drive block thickness stated: 5 mm, so the cell height in the holder does not change
- Stirring cells with a recess: the bar stays in the well rather than drifting into the beam
- Bars in packs: ten per pack, which is the quantity a working lab actually uses
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Quartz Flow Cells and Cuvettes, Flow Cell Holders and Cuvette Changers, Column Ovens and Heater Blocks, Autosampler Vials and Septa, HPLC Columns for Method Development, Lamp Housings and Shutters.