Flow Cell Holders and Cuvette Changers for Photometric Instruments
Product Specifications
Product Overview
A cell holder does three jobs: it locates the cell in the beam, it controls the temperature, and it keeps the cell clean and dry between readings. The first job is mechanical. A 10 mm cell sits in a block with a fixed stop, and the beam height is set at 8.5 or 15 mm depending on the instrument. If the holder allows the cell to tilt, the path length through the liquid changes and the absorbance changes with it, which is why a holder with a spring or a screw clamp is used rather than a simple slot.
The second job is thermal. A Peltier block holds a cell between roughly -5 and 110 C, and the useful figure is not the range but the stability: a block that holds 0.02 C over a ten minute reading gives a repeatable absorbance, while a block that drifts a tenth of a degree moves the reading on a temperature sensitive reaction. For kinetics and for anything near a melting point, the block, the controller and the stir bar are one system, and the stir bar is not optional, because a cell with a temperature gradient across it reads the average of the gradient.
The third job is the flow path. A flow cell turns the holder into a flow injection detector, and the numbers that matter are the cell volume and the swept path. A 10 mm cell with a 390 microlitre volume is a standard for a benchtop spectrophotometer but is too large for a 1 mm column, where a cell of 18 or 80 microlitres is the right size. The connector matters as well: an M6 screw port with a 1.9 mm outside diameter FEP tube is a different fitting from a 1/4-28 flat bottom port, and the two do not seal into each other.
Changers add the fourth job, which is throughput. A six cell Peltier changer presents six 10 mm cells to the same beam in sequence, with the same temperature and the same purge, so a batch of six samples is read without opening the sample compartment. The trade is that each position has to be mechanically repeatable, and a changer that is a few tenths of a millimetre out between positions shows up as a position to position offset. That offset is measured with the same cell in all six positions, not with six different cells.
Key Specifications
| Parameter | Value | Note |
|---|---|---|
| Cell format | 10 x 10 mm standard, 12.5 x 12.5 mm outer on the Hellma flow cells | the outside dimension sets the block, the path sets the reading |
| Path length | 0.1, 0.2, 0.5, 1, 2, 5, 10 and 50 mm | shorter paths need a spacer in a 10 mm block |
| Cell volume | 6.2 microlitre at 0.1 mm to 3300 microlitre at 10 mm | volume follows path length and aperture |
| Beam height, Z | 8.5 mm and 15 mm | the two common instrument heights |
| Aperture | from 1.5 mm round to 11 x 3.5 mm rectangular | the aperture sets stray light and the minimum volume |
| Positions | 1, 2 in a twin holder, 6 in a changer | a changer needs position to position repeatability |
| Peltier range, PIKE holders | -5 to 110 C | with the matching controller and liquid recirculator |
| Peltier range, six cell changer | -20 to 110 C | continuous ramp, N2 purge and stirring at every position |
| Temperature precision | 0.02 C on a single cell Peltier unit | the figure that matters for kinetics |
| Stirring | magnetic stir bar 6.35 mm long x 3 mm diameter | fits inside a 10 mm cell |
| Fiber coupling | external holder with SMA collimators, two fiber cables | for a measurement away from the instrument |
| Flow cell connections | M6 x 1 screw ports with 1.9 mm OD, 1.1 mm ID FEP tube, 500 mm long | a 1/4-28 port is a different fitting |
| Purge | N2 purge on the six cell holder | keeps condensation off a cell below ambient |
| Materials | anodised aluminium block, quartz or optical glass cell | quartz for 200 to 2500 nm, glass for the visible range |
Included Models and Order Codes
The table is split between holders and changers, because they are bought for different reasons. A single holder is bought to replace a worn block or to add temperature control; a changer is bought for throughput. The cell numbers are on the second table, since the cell and the holder are usually replaced on different schedules.
| Part | Type | Positions | Temperature | Beam or fitting | Note |
|---|---|---|---|---|---|
| 013-000-71 | Cell holder for type 106 demountable cells | 1 | ambient | 10 mm block | the holder is ordered separately from the cell |
| 020-001-761 | Cell holder for type 124 and 201/202 cells | 1 | ambient | circular cell | for cylindrical and demountable cells |
| 013-101-71 | Aluminium spacer, 1 mm | not applicable | ambient | fits a 10 mm block | puts a 1 mm cell at the right height |
| 013-102-71 | Aluminium spacer, 2 mm | not applicable | ambient | fits a 10 mm block | 2 mm path in a 10 mm block |
| 013-105-71 | Aluminium spacer, 5 mm | not applicable | ambient | fits a 10 mm block | 5 mm path in a 10 mm block |
| 325.000 | CleanAssist plastic holder | 4 | ambient | 10 mm cells | holds four cells for cleaning and drying |
| 171-70XXX | Peltier cell holder, single | 1 | -5 to 110 C | 8.5 or 15 mm Z height | XXX is the instrument code, controller required |
| 171-80XXX | Peltier cell holder, twin | 2 | -5 to 110 C | 8.5 or 15 mm Z height | single and dual beam instruments |
| 076-1510 | Temperature control module, single channel | not applicable | reads the block RTD | USB and LCD | one module per holder |
| 076-1515 | Temperature control module, dual channel | not applicable | reads two blocks | USB and LCD | for the twin holder |
| 162-1201 | Spacer, 1 mm path | not applicable | ambient | Peltier holder | with the 162-1202 and 162-1205 set |
| 162-1202 | Spacer, 2 mm path | not applicable | ambient | Peltier holder | same set |
| 162-1205 | Spacer, 5 mm path | not applicable | ambient | Peltier holder | same set |
| 162-1905 | Micro stir bar | not applicable | ambient | 6.35 mm x 3 mm | fits a 10 mm cell |
| 170-1100 | Liquid recirculator | not applicable | heat sink for the Peltier | hose to the holder | required for operation below ambient |
| CS/PC6 | Six cell Peltier autochanger | 6 | -20 to 110 C | 10 x 10 mm cells | N2 purge and stirring at every position |
| CS/SP6 | Spacer kit for CS/PC6 | not applicable | ambient | shorter path cells | lets a 1 or 5 mm cell sit in the changer |
Flow Cells That Fit These Holders
A flow cell is ordered by path length, volume and connector, and the volume is what decides whether it suits a column. The table gives the maker figures for the cell body, so a cell is matched to the holder and to the flow before it is ordered.
| Cell | Path length | Volume | Aperture | Outside size | Note |
|---|---|---|---|---|---|
| 176.700-QS, 5 mm | 5 mm | 195 microlitre | 11 x 3.5 mm | 35 x 12.5 x 12.5 mm | M6 ports, FEP tube fitted |
| 176.700-QS, 10 mm | 10 mm | 390 microlitre | 11 x 3.5 mm | 35 x 12.5 x 12.5 mm | the standard flow cell for a benchtop unit |
| 176.700-QS, 50 mm | 50 mm | 1950 microlitre | 11 x 3.5 mm | 35 x 12.5 x 52.5 mm | long path for low concentration work |
| 178.710-QS | 10 mm | 80 microlitre | 3 mm round | 35 x 12.5 x 12.5 mm | narrow aperture, lower volume |
| 178.712-QS | 10 mm | 18 microlitre | 1.5 mm round | 35 x 12.5 x 12.5 mm | for a 1 mm column |
| 170.700-QS, 0.1 mm | 0.1 mm | 6.2 microlitre | 17.5 x 3.5 mm | 35 x 12.5 x 12.5 mm | highest absorbance, thinnest path |
| 170.700-QS, 1 mm | 1 mm | 62 microlitre | 17.5 x 3.5 mm | 35 x 12.5 x 12.5 mm | concentrated samples |
| 176.760-QS | 5 and 10 mm | 550 microlitre | 11 x 6 or 11 x 5 mm | 35 x 12.5 x 12.5 mm | two path lengths in one cell |
| 176.762-QS | 1.5 and 3 mm | 50 microlitre | 11 x 2.5 or 11 x 1.5 mm | 35 x 12.5 x 12.5 mm | smallest dual path cell |
| 176.765-QS | 1 and 10 mm | 110 microlitre | 11 x 6 or 11 x 1 mm | 35 x 12.5 x 12.5 mm | turns 90 degrees to change path, tubes stay connected |
Key Features
Holders and changers separated – a single holder is bought to replace a worn block, a six cell changer is bought for throughput, and the two are not priced or specified the same way
Beam height 8.5 mm and 15 mm – the two common instrument heights are named on every holder, so a holder is not matched on cell size alone
Peltier blocks from -5 to 110 C – with the controller, the stir bar and the recirculator listed as one system
Six cell changer with purge and stirring – six 10 mm cells at the same temperature, which is what a batch of six samples needs
Spacers for 1, 2 and 5 mm cells – a short path cell sits correctly in a 10 mm block instead of resting on the bottom
Fiber coupled external holder – SMA collimators and two fibers move the measurement to a fume hood without moving the instrument
Flow cells from 6.2 to 3300 microlitre – 18 microlitre for a 1 mm column and 390 microlitre for a benchtop unit, so the cell is chosen against the column
M6 ports with FEP tube – the fitting is named with the cell, because a 1/4-28 port will not seal into an M6 thread
Applications
These holders sit between the light source and the detector, in three settings that put different demands on the block.
Pharmaceutical dissolution testing – a six cell changer with a 5 mm or 10 mm flow cell reads six vessels in sequence. The path length is chosen for the concentration, and the purge keeps condensation off the cell when the bath runs at 37 C in a cold room.
Enzyme kinetics – a single Peltier holder with a stir bar holds one cell at 37 C for a ten minute run. The temperature precision matters more than the range, because the reaction rate changes by about seven percent per degree for a typical enzyme.
Process photometry on a flow line – a flow cell with an 80 microlitre volume sits after a debubbler and before the detector. The swept volume has to be small enough not to broaden a peak, and the connector has to match the 1/16 in tubing on the line.
The detector boards behind these cells are the kind of build SUPERB PCB and SUPERB PCBA programmes cover, from the photodiode front end to the lamp drive.
Sourcing and Cross-Reference
A holder is ordered by instrument, cell format and beam height, and the instrument code is the field that decides the part. A Peltier holder for one spectrophotometer does not fit another even when the cell is the same 10 mm cell, because the block height and the mounting screws differ. We ask for the instrument model and the existing holder number before naming a part.
Where a holder is no longer supplied, the substitute is chosen on the cell format and the beam height first and on the temperature range second. A holder that fits the cell but sits 3 mm lower moves the beam off the aperture and the reading falls without any fault code. Where no equivalent exists, we say so rather than shipping a block that needs a machined adapter.
Flow cells are a wear item on a process line. The cell window etches, the port thread wears and the volume grows with a repaired connector. A replacement cell is matched on path length, volume and port, and where the original cell is a two path length type, the replacement is chosen so the tubes do not have to be re routed.
Where an instrument is assembled with its cell changer and flow cell, the set is picked together, which is how SUPERB Assembly ships an optics module.
Q: Can you supply a Peltier cell holder for an instrument whose original holder is discontinued?
A: SUPERB Tech can confirm the instrument code, the beam height and the cell format against the old holder before the order is placed.
Why Buy Cell Holders from SUPERB
- Instrument code confirmed first: the holder is matched to the spectrophotometer, not only to the cell, because the block height and the mounting differ
- Cell and holder quoted together: the path length, volume and port are checked as a set, which is what avoids a cell that fits the block but not the flow
- Spacers and stir bars on the same list: a 1 mm cell and a 10 mm block are supplied as a working pair
- Peltier system complete: holder, controller, recirculator and stir bar are named together, so a temperature controlled cell is not ordered in three parts
- Maker part number on the delivery file: the holder, the cell and the spacer are written down, so the next replacement is the same part
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Quartz Flow Cells and Cuvettes for Photometric Instruments, Bandpass and Interference Filters for Photometric Channels, UV and Visible LED Sources for Photometric Detection, Lamp Housings and Shutters for Photometric Instruments, Turbidity and Photometric Detector Modules for Water Analysis, Component Lead Time Solutions.