Wavelength Calibration Lamps for Spectrometers

Atomic lines, not a continuous spectrum. Five lamps covering 253 to 1984 nm, fibre coupled through SMA 905. Mercury argon for UV and visible, argon and xenon for the NIR.
quote now

Product Specifications

Product Overview

A spectrometer drifts. The grating expands with temperature, the detector array shifts by a fraction of a pixel and the calibration coefficients written when the instrument was built gradually stop describing it. A calibration lamp is how that is measured, because a low pressure gas discharge emits lines at wavelengths fixed by atomic physics rather than by a manufacturing tolerance.

The choice of gas is the choice of wavelength band. Mercury argon covers 253 to 1700 nm and carries the familiar 253.652, 365.015, 404.656, 435.833 and 546.074 nm lines, which is why it is the standard for UV and visible work. Krypton covers 427 to 893 nm, neon 540 to 754 nm, argon 696 to 1704 nm and xenon 916 to 1984 nm, so a NIR method is usually checked with argon or xenon rather than with mercury.

The lamps are small housings with a fibre port, a 3 mm aperture and an SMA 905 connector, so the same lamp can be used on a bench instrument and on a probe based system. Warm up is about a minute and the source life is around 3500 hours, which means a lamp left on permanently is a consumable item and one switched on for a monthly check lasts for years.

Key Specifications

ParameterValueNotes
Gases availableMercury argon, krypton, neon, argon, xenonOne lamp per gas
Coverage across the set253 to 1984 nmPick the gas for the band
Mercury argon lines253.652, 296.728, 302.150, 313.155, 334.148, 365.015, 404.656, 435.833, 546.074 nmPlus argon lines above 600 nm
Argon lines696.543 to 922.450 nm in first order, to 1704 nm in second orderNIR calibration
Emission typeLow pressure gas dischargeAtomic lines, not a continuum
Output couplingSMA 905, 3 mm apertureFibre or direct
Power, current family5 V DC, 2 A maximumHG-2 and the matching models
Power, earlier family12 V DC, 250 mAHG-1 and the matching models
Warm upAbout 1 minute to stable amplitudeSwitch on before the check
Source life3500 hours typicalAt rated current
Dimensions167 x 106 x 34 mm, 180 gCurrent family without feet

Included Models and Order Codes

The -2 models are the current family, running from a 5 V DC supply and sharing one housing and one fibre port. The -1 models are the earlier family with a 12 V supply and a smaller housing, still in service in many laboratories and still supplied for instruments already calibrated against them. A laboratory working across UV and NIR normally holds mercury argon plus argon, or mercury argon plus xenon.

Part NumberGasWavelength RangePowerCouplingUse
HG-2Mercury argon253 to 1700 nm5 V DC, 2 ASMA 905UV and visible calibration
KR-2Krypton427 to 893 nm5 V DC, 2 ASMA 905Visible band check
NE-2Neon540 to 754 nm5 V DC, 2 ASMA 905Dense visible lines
AR-2Argon696 to 1704 nm5 V DC, 2 ASMA 905NIR calibration
XE-2Xenon916 to 1984 nm5 V DC, 2 ASMA 905Long wave NIR
HG-1Mercury argon253 to 1700 nm12 V DC, 250 mASMA 905Earlier instruments
KR-1Krypton427 to 893 nm12 V DC, 250 mASMA 905Earlier instruments
NE-1Neon540 to 754 nm12 V DC, 250 mASMA 905Earlier instruments
AR-1Argon696 to 1704 nm12 V DC, 250 mASMA 905Earlier instruments
XE-1Xenon916 to 1984 nm12 V DC, 250 mASMA 905Earlier instruments

Key Features

Lines fixed by physics – the wavelengths are atomic and do not age, only the intensity does

Five gases on one page – mercury argon for UV and visible, krypton and neon for visible, argon and xenon for NIR

Fibre port as standard – SMA 905 with a 3 mm aperture, so the lamp works on a bench or through a probe

Two supply families – 5 V and 12 V versions, so a lamp can replace one already written into a procedure

One minute warm up – stable enough for a monthly check without leaving the lamp on

Consumable priced as one – 3500 hours of life stated, which sets the replacement interval in a quality system

Applications

The lamp is used wherever the wavelength axis has to be proven, which is a quality activity as much as a technical one.

Method validation and QC – a mercury argon lamp measured before a batch, with the 253.652 and 546.074 nm lines checked against the calibration file

NIR method checks – an argon or xenon lamp where the mercury lines run out above 600 nm

Service and field calibration – a battery capable 12 V lamp carried to an installed analyser, with a fibre patch cord to the sample port

Calibration lamps ship with the SUPERB PCB and SUPERB PCBA boards that drive them, so a check fixture can be built and wired as one item.

Sourcing and Cross-Reference

The useful habit is to keep the calibration spectrum, not just the numbers. If the spectrum of the lamp is stored with the instrument file, a later shift can be compared against a real measurement rather than against a remembered value. Both the lamp model and the fibre used for the check belong in that record, because a different core diameter changes the line shape, though not the wavelength.

Lamp ageing shows up as intensity, not as position. A lamp with 3000 hours on it still puts the lines in the right place, but the signal to noise of the check falls, and a weak line such as 407.783 nm may disappear into the noise before the strong ones do. Replacing on hours, rather than on a failed check, keeps the check meaningful.

The 407.783 nm line is a useful example of an instrument limit rather than a lamp limit. It is not visible on a spectrometer with a 300 or 600 lines per mm grating, and the 576.96 and 579.07 nm pair only separates at 1200 lines per mm and above. A calibration procedure that names those lines has to name the grating as well, or it will fail on a correctly working instrument.

Lamp, supply and fibre patch cord are kitted through SUPERB Assembly with the line list printed on the case.

Q: Can you supply a mercury argon calibration lamp with an SMA 905 port and the 12 V supply that our procedure specifies?

A: Yes. SUPERB Tech supplies calibration lamps by gas and by supply voltage, and the strong line list is quoted with the lamp.

Why Choose SUPERB for Wavelength Calibration Lamps

  1. Gas chosen by band: the mercury argon, krypton, neon, argon and xenon ranges are stated side by side
  2. Line list published: the strong lines with their wavelengths, so a procedure can reference real values
  3. Two supply families: 5 V and 12 V versions, which keeps existing procedures valid
  4. Consumable interval clear: 3500 hours and the warm up time given, so the check is planned rather than improvised

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Xenon and Deuterium Lamps, Photomultiplier Tubes and PMT Housings, Photodiode Arrays and Detector Modules, Monochromator Gratings and Bench Parts, Optical Slits and Apertures, Neutral Density Filters and Reference Standards.