Wavelength Calibration Lamps for Spectrometers
Product Specifications
Product Overview
A spectrometer drifts. The grating expands with temperature, the detector array shifts by a fraction of a pixel and the calibration coefficients written when the instrument was built gradually stop describing it. A calibration lamp is how that is measured, because a low pressure gas discharge emits lines at wavelengths fixed by atomic physics rather than by a manufacturing tolerance.
The choice of gas is the choice of wavelength band. Mercury argon covers 253 to 1700 nm and carries the familiar 253.652, 365.015, 404.656, 435.833 and 546.074 nm lines, which is why it is the standard for UV and visible work. Krypton covers 427 to 893 nm, neon 540 to 754 nm, argon 696 to 1704 nm and xenon 916 to 1984 nm, so a NIR method is usually checked with argon or xenon rather than with mercury.
The lamps are small housings with a fibre port, a 3 mm aperture and an SMA 905 connector, so the same lamp can be used on a bench instrument and on a probe based system. Warm up is about a minute and the source life is around 3500 hours, which means a lamp left on permanently is a consumable item and one switched on for a monthly check lasts for years.
Key Specifications
| Parameter | Value | Notes |
|---|---|---|
| Gases available | Mercury argon, krypton, neon, argon, xenon | One lamp per gas |
| Coverage across the set | 253 to 1984 nm | Pick the gas for the band |
| Mercury argon lines | 253.652, 296.728, 302.150, 313.155, 334.148, 365.015, 404.656, 435.833, 546.074 nm | Plus argon lines above 600 nm |
| Argon lines | 696.543 to 922.450 nm in first order, to 1704 nm in second order | NIR calibration |
| Emission type | Low pressure gas discharge | Atomic lines, not a continuum |
| Output coupling | SMA 905, 3 mm aperture | Fibre or direct |
| Power, current family | 5 V DC, 2 A maximum | HG-2 and the matching models |
| Power, earlier family | 12 V DC, 250 mA | HG-1 and the matching models |
| Warm up | About 1 minute to stable amplitude | Switch on before the check |
| Source life | 3500 hours typical | At rated current |
| Dimensions | 167 x 106 x 34 mm, 180 g | Current family without feet |
Included Models and Order Codes
The -2 models are the current family, running from a 5 V DC supply and sharing one housing and one fibre port. The -1 models are the earlier family with a 12 V supply and a smaller housing, still in service in many laboratories and still supplied for instruments already calibrated against them. A laboratory working across UV and NIR normally holds mercury argon plus argon, or mercury argon plus xenon.
| Part Number | Gas | Wavelength Range | Power | Coupling | Use |
|---|---|---|---|---|---|
| HG-2 | Mercury argon | 253 to 1700 nm | 5 V DC, 2 A | SMA 905 | UV and visible calibration |
| KR-2 | Krypton | 427 to 893 nm | 5 V DC, 2 A | SMA 905 | Visible band check |
| NE-2 | Neon | 540 to 754 nm | 5 V DC, 2 A | SMA 905 | Dense visible lines |
| AR-2 | Argon | 696 to 1704 nm | 5 V DC, 2 A | SMA 905 | NIR calibration |
| XE-2 | Xenon | 916 to 1984 nm | 5 V DC, 2 A | SMA 905 | Long wave NIR |
| HG-1 | Mercury argon | 253 to 1700 nm | 12 V DC, 250 mA | SMA 905 | Earlier instruments |
| KR-1 | Krypton | 427 to 893 nm | 12 V DC, 250 mA | SMA 905 | Earlier instruments |
| NE-1 | Neon | 540 to 754 nm | 12 V DC, 250 mA | SMA 905 | Earlier instruments |
| AR-1 | Argon | 696 to 1704 nm | 12 V DC, 250 mA | SMA 905 | Earlier instruments |
| XE-1 | Xenon | 916 to 1984 nm | 12 V DC, 250 mA | SMA 905 | Earlier instruments |
Key Features
Lines fixed by physics – the wavelengths are atomic and do not age, only the intensity does
Five gases on one page – mercury argon for UV and visible, krypton and neon for visible, argon and xenon for NIR
Fibre port as standard – SMA 905 with a 3 mm aperture, so the lamp works on a bench or through a probe
Two supply families – 5 V and 12 V versions, so a lamp can replace one already written into a procedure
One minute warm up – stable enough for a monthly check without leaving the lamp on
Consumable priced as one – 3500 hours of life stated, which sets the replacement interval in a quality system
Applications
The lamp is used wherever the wavelength axis has to be proven, which is a quality activity as much as a technical one.
Method validation and QC – a mercury argon lamp measured before a batch, with the 253.652 and 546.074 nm lines checked against the calibration file
NIR method checks – an argon or xenon lamp where the mercury lines run out above 600 nm
Service and field calibration – a battery capable 12 V lamp carried to an installed analyser, with a fibre patch cord to the sample port
Calibration lamps ship with the SUPERB PCB and SUPERB PCBA boards that drive them, so a check fixture can be built and wired as one item.
Sourcing and Cross-Reference
The useful habit is to keep the calibration spectrum, not just the numbers. If the spectrum of the lamp is stored with the instrument file, a later shift can be compared against a real measurement rather than against a remembered value. Both the lamp model and the fibre used for the check belong in that record, because a different core diameter changes the line shape, though not the wavelength.
Lamp ageing shows up as intensity, not as position. A lamp with 3000 hours on it still puts the lines in the right place, but the signal to noise of the check falls, and a weak line such as 407.783 nm may disappear into the noise before the strong ones do. Replacing on hours, rather than on a failed check, keeps the check meaningful.
The 407.783 nm line is a useful example of an instrument limit rather than a lamp limit. It is not visible on a spectrometer with a 300 or 600 lines per mm grating, and the 576.96 and 579.07 nm pair only separates at 1200 lines per mm and above. A calibration procedure that names those lines has to name the grating as well, or it will fail on a correctly working instrument.
Lamp, supply and fibre patch cord are kitted through SUPERB Assembly with the line list printed on the case.
Q: Can you supply a mercury argon calibration lamp with an SMA 905 port and the 12 V supply that our procedure specifies?
A: Yes. SUPERB Tech supplies calibration lamps by gas and by supply voltage, and the strong line list is quoted with the lamp.
Why Choose SUPERB for Wavelength Calibration Lamps
- Gas chosen by band: the mercury argon, krypton, neon, argon and xenon ranges are stated side by side
- Line list published: the strong lines with their wavelengths, so a procedure can reference real values
- Two supply families: 5 V and 12 V versions, which keeps existing procedures valid
- Consumable interval clear: 3500 hours and the warm up time given, so the check is planned rather than improvised
Related Product Families
This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Xenon and Deuterium Lamps, Photomultiplier Tubes and PMT Housings, Photodiode Arrays and Detector Modules, Monochromator Gratings and Bench Parts, Optical Slits and Apertures, Neutral Density Filters and Reference Standards.