Arc Lamp Igniters and Xenon Lamp Power Supplies

Power supplies that carry the igniter inside the case. Plus 15 kV trigger pulses, 4 strikes per second, over 80,000 strikes. Cermax xenon and mercury versions, 100 to 300 W constant power.
quote now

Product Specifications

Product Overview

A short arc lamp does not start at its running voltage. The lamp runs at 12 to 25 V once the arc is lit, but the gas has to be broken down first, and that takes a pulse of 17 to 35 kV at the lamp. The igniter is the circuit that makes that pulse, and it has to stop working the moment the arc forms, because the same high voltage across a lit lamp destroys it.

That is why the igniter is nearly always quoted with the supply rather than on its own. The supply charges the boost capacitor, the igniter discharges it through a step up transformer and a spark gap, and both share the sense circuit that decides when to stop pulsing. Splitting the two across different makers means matching the boost voltage, the pulse width and the shutoff threshold, which is a great deal of work to save a small amount of money.

The supplies on this page carry the igniter inside the enclosure. They accept 100 to 240 V AC, deliver 100 to 300 W in constant power mode, and pulse at plus 15 kV nominal until the lamp lights or a six second timer ends the attempt. Xenon and mercury versions differ in output voltage and current, so the lamp type decides the model before anything else.

Key Specifications

ParameterValueNotes
Output power100 to 300 W, constant powerPotentiometer adjustable
Igniter outputPlus 15 kV nominal, plus or minus 3 kVBilateral pulse
Boost voltage before ignitionAbove 110 VFalls to the running voltage after the arc forms
Strike rate4 strikes per second minimumPlus or minus 2
Ignition timer6 plus or minus 1 s of pulsesCan be disabled by jumper
Igniter lifeMore than 80,000 strikesPer supply
Output current, xenon9 to 22 A DCRunning voltage 10 to 25 V
Output current, mercury4 to 10 A DCRunning voltage 10 to 40 V
Output rippleUnder 5 percent at 300 WMeasured DC to 20 MHz
Line input100 to 240 V AC, 50 to 60 Hz5.3 A rms maximum
Trigger needed at the lamp17 to 35 kV for Cermax lampsPulse 20 ns to 1 us
Ground leakage optionsUnder 300 uA, under 30 uA, under 100 uAModel letter sets the leakage

Included Models and Order Codes

The part number carries the lamp type and the ground leakage. The letter before XE or HG sets the leakage: A is under 300 uA, B is under 30 uA and C is under 100 uA. XE models run xenon short arc lamps, HG models run mercury arc lamps, and the fixed power code sets the output the potentiometer is trimmed to at the ends of its travel. CE300BXE is the older 300 W xenon supply, kept on the page because it is still fitted in installed instruments.

Part NumberLamp TypeOutput PowerOutput CurrentGround LeakageIgniterUse
CEP301AXEXenon, Cermax100 to 300 W9 to 22 AUnder 300 uAPlus 15 kV, 6 s timerGeneral xenon lamps
CEP301BXEXenon, Cermax100 to 300 W9 to 22 AUnder 30 uAPlus 15 kV, 6 s timerLow leakage, patient contact
CEP301CXEXenon, Cermax100 to 300 W9 to 22 AUnder 100 uAPlus 15 kV, 6 s timerHospital equipment
CEP301AHGMercury arc100 to 300 W4 to 10 AUnder 300 uAPlus 15 kV, 6 s timerMercury lamp retrofits
CEP301BHGMercury arc100 to 300 W4 to 10 AUnder 30 uAPlus 15 kV, 6 s timerLow leakage mercury
CEP301CHGMercury arc100 to 300 W4 to 10 AUnder 100 uAPlus 15 kV, 6 s timerFluorescence excitation
CE300BXEXenon, XBO and Cermax270 to 315 W17 to 22 AUnder 100 uAPlus 15 kVInstalled instrument replacement

Key Features

Igniter inside the supply – the boost, the pulse and the shutoff are one circuit, so there is nothing to match between makers

Leakage class in the part number – under 300, 30 or 100 uA, which decides whether the unit can sit near a patient connection

Six second ignition timer – the pulses stop if the lamp does not light, instead of running until something fails

Constant power output – lamp ageing changes the voltage, and the supply holds the power so the light output stays steady

Both lamp chemistries – xenon and mercury versions share the case and the igniter, which keeps spares simple

Auxiliary rails included – isolated 5 V and 12 V outputs for fans and CMOS logic inside the same enclosure

Applications

The supply is chosen with the lamp, the reflector and the housing, and the ignition behaviour is what ties the three together.

Medical and industrial illuminators – a 300 W xenon lamp behind a fibre bundle, with the leakage class set by the endoscope or microscope it feeds

Spectroscopy and fluorescence – a xenon or mercury lamp with a stabilised supply, where light output stability sets the baseline noise

Retrofit of an installed instrument – a drop in supply for a lamp that is still available, keeping the ignition behaviour the optical path expects

Lamp supplies are quoted with the SUPERB PCB and SUPERB PCBA boards that sequence the shutter and read the lamp current, so ignition and control are tested together.

Sourcing and Cross-Reference

The number that a retrofit hinges on is the minimum ignition voltage at the lamp, not the open circuit pulse. A Cermax lamp may break down at 7 to 10 kV on a DC basis, yet need a 13 kV pulse with a 250 ns width, and the maker recommends a target of 17 to 35 kV to cover the whole life of the lamp. A supply with a 15 kV igniter is therefore normal for a lamp that lights at 8 kV, and that margin is what the hot restrike needs.

Hot restrike is the second point. A lamp that has just been switched off needs a higher voltage for about a minute, because the gas is still hot and the density is low. A supply that ignites reliably from cold and fails a minute after a restart is not faulty; it is running out of margin. Extra trigger voltage and a clean, short lamp lead are the two fixes.

Lead length and routing matter more than the datasheet suggests. The trigger pulse is fast, and a long lead between the igniter and the lamp behaves as a capacitor that shunts part of the pulse. Keeping that run under about a metre, away from signal wiring, and using the maker lead where one is offered, solves most field ignition problems before they appear.

Supply, igniter lead and lamp connector are kitted through SUPERB Assembly against the instrument wiring drawing.

Q: Can you supply a 300 W xenon supply with a built in igniter, plus 15 kV output and an under 30 uA leakage class?

A: Yes. SUPERB Tech quotes the supply against the lamp type, the leakage class and the ignition voltage at the lamp, and states the recommended lead length.

Why Choose SUPERB for Arc Lamp Igniters and Power Supplies

  1. Igniter and supply as one unit: boost, pulse and shutoff matched by the maker, with no field matching needed
  2. Leakage class visible: the part number states the ground leakage, which is the first question in medical equipment
  3. Both lamp chemistries: xenon and mercury versions on one page, with the output current for each
  4. Installation guidance included: lead length, hot restrike and trigger margin written out beside the part numbers

Related Product Families

This page is part of the component range we supply for industrial, telecom, medical and defence builds. Continue with: Xenon and Deuterium Lamps, Lamp Housings and Shutters, Lamp Reflectors and Ellipsoidal Condensers, Wavelength Calibration Lamps, Fiber Bundles and Light Guides, PWM Controller ICs for Switching Supplies.